  Status of commissioning, report also here plots are reports., Page 17 of 24  Not logged in ELOG logo
ID Date Author Status Type Category Location Titledown
  66   Thu Oct 3 14:00:13 2019 Loïc AmoudryFixedissuecablingOptical roomElectrical trouble (disjunction)

The circuit breaker on ThomX cavity drops @1.5-1.8kW when we turn on 1/2nd stage of the amplifier or 3rd stage of the amplifier.
Amplifiers are now connected to K-BOX UPS.

We are buying a clamp ammeter for further investigations.

  7   Mon Oct 22 11:03:22 2018 Ronic ChicheFixedinfovacuumOptical roomDétection de fuite

Mail de Bruno Mercier du 18/10/2018 :

Un petit CR de la journée:
Nous avons effectué une détection de fuite sur l'ensemble des cavités optiques. Une fuite très importante a été localisée au niveau de la bride de la Pompe ionique (PI) située sur la deuxième cavité optique par rapport à l'entrée de la salle blanche.
Manu a démonté la PI , la fuite provenait du fait que le joint n'était pas tout à fait centré. Après un remontage de la PI, une nouvelle détection a été faite:
détection d'une fuite sur la bride avec les sorties Sub-d M4 après resserrage c'est OK.
Détection d'une fuite sur la bride de sortie DN200 avec  hublots de la cavité 1 après resserrage c'est OK.

Demain et le début de la semaine prochaine sera consacré à la mise en place de l'étuvage, puis de la descente en pression avec pompage ionique.

bonne fin de journée,
Bruno

  112   Mon Jan 17 10:04:14 2022 Ronic ChicheFixedinfomechanicsThomX iglooDocumentation about mechanics on Atrium

https://atrium.in2p3.fr/0dded51a-1cb0-43e6-ae0d-626cd5db7078

a lot of schematics regarding the FP cavity are stored by the mechanic group on the Atrium repository.

  62   Mon Sep 30 11:20:48 2019 Loïc AmoudryFixedinfolasers and opticsOptical roomDiffusion ULE spherical mirror (S3) vs Fused Silicate injection planar mirror (P1)

Observation made with IR Beam Viewer: 

The observation of P1 and S3 simultaneously while locking shows much higher diffusion on the ULE mirror. The diffusion on the two other ULE mirrors (S2 and P4) shows similar diffusion as S3.

Attachment 1: diffusion.jpg
diffusion.jpg
  132   Mon Mar 14 18:39:40 2022 Manar AmerFixedinfomechanics | lasers and optics | detectors and electronicsThomX iglooCurrent Installed Setup

Attached is the current setup in ThomX

Attachment 1: 18-02-2022_ThomX_FPCavity_Finesse_Setup.jpg
18-02-2022_ThomX_FPCavity_Finesse_Setup.jpg
  102   Tue Oct 27 13:13:50 2020 Loïc AmoudryFixedinfolasers and opticsOptical roomCoating reflectivity

Coatings reflectivity curves and datas for HR Saphir, HR Suprasil and HR ULE.

Attachment 1: HR_Saphir_Théorique.PNG
HR_Saphir_Théorique.PNG
Attachment 2: HR_Suprasil_Théorique.PNG
HR_Suprasil_Théorique.PNG
Attachment 3: HR_ULE_Théorique.PNG
HR_ULE_Théorique.PNG
Attachment 4: Table_HR_Saphir.TXT
400.0	85.1750
401.0	87.3579
402.0	89.6588
403.0	91.2610
404.0	91.5817
405.0	90.5819
406.0	88.8006
407.0	87.0591
408.0	86.0823
409.0	86.2646
410.0	87.5842
411.0	89.5943
412.0	91.5095
413.0	92.4844
414.0	92.0541
415.0	90.4258
416.0	88.3396
417.0	86.6428
418.0	85.9386
419.0	86.4444
420.0	87.9765
421.0	89.9869
422.0	91.6931
423.0	92.3704
424.0	91.7171
425.0	90.0207
426.0	87.9708
427.0	86.2994
428.0	85.5165
429.0	85.8139
430.0	87.0628
431.0	88.8460
432.0	90.5433
433.0	91.5207
434.0	91.3963
435.0	90.2238
436.0	88.4481
437.0	86.6722
438.0	85.4180
439.0	84.9957
440.0	85.4699
441.0	86.6720
442.0	88.2333
443.0	89.6609
444.0	90.4796
445.0	90.4096
446.0	89.4823
447.0	88.0125
448.0	86.4504
449.0	85.2159
450.0	84.5929
451.0	84.6908
452.0	85.4459
453.0	86.6395
454.0	87.9385
455.0	88.9689
456.0	89.4245
457.0	89.1716
458.0	88.2943
459.0	87.0511
460.0	85.7733
461.0	84.7630
462.0	84.2266
463.0	84.2490
464.0	84.7926
465.0	85.7091
466.0	86.7665
467.0	87.6948
468.0	88.2511
469.0	88.2900
470.0	87.8065
471.0	86.9318
472.0	85.8842
473.0	84.9012
474.0	84.1822
475.0	83.8540
476.0	83.9574
477.0	84.4483
478.0	85.2074
479.0	86.0603
480.0	86.8101
481.0	87.2800
482.0	87.3580
483.0	87.0263
484.0	86.3627
485.0	85.5151
486.0	84.6592
487.0	83.9585
488.0	83.5339
489.0	83.4489
490.0	83.7039
491.0	84.2394
492.0	84.9441
493.0	85.6722
494.0	86.2679
495.0	86.5977
496.0	86.5806
497.0	86.2082
498.0	85.5457
499.0	84.7139
500.0	83.8602
501.0	83.1284
502.0	82.6350
503.0	82.4549
504.0	82.6140
505.0	83.0868
506.0	83.7970
507.0	84.6233
508.0	85.4110
509.0	85.9927
510.0	86.2173
511.0	85.9810
512.0	85.2518
513.0	84.0769
514.0	82.5717
515.0	80.8932
516.0	79.2103
517.0	77.6790
518.0	76.4268
519.0	75.5474
520.0	75.1010
521.0	75.1173
522.0	75.5997
523.0	76.5265
524.0	77.8522
525.0	79.5075
526.0	81.3991
527.0	83.4129
528.0	85.4198
529.0	87.2874
530.0	88.8939
531.0	90.1440
532.0	90.9816
533.0	91.3958
534.0	91.4184
535.0	91.1151
536.0	90.5717
537.0	89.8801
538.0	89.1269
539.0	88.3862
540.0	87.7151
541.0	87.1536
542.0	86.7252
543.0	86.4394
544.0	86.2944
545.0	86.2791
546.0	86.3760
547.0	86.5627
548.0	86.8142
549.0	87.1038
550.0	87.4051
551.0	87.6927
552.0	87.9441
553.0	88.1400
554.0	88.2660
555.0	88.3132
556.0	88.2786
557.0	88.1654
558.0	87.9828
559.0	87.7453
560.0	87.4718
561.0	87.1836
562.0	86.9035
563.0	86.6537
564.0	86.4544
565.0	86.3222
566.0	86.2689
567.0	86.3007
568.0	86.4177
569.0	86.6133
570.0	86.8746
571.0	87.1832
572.0	87.5159
573.0	87.8467
574.0	88.1484
575.0	88.3956
576.0	88.5666
577.0	88.6460
578.0	88.6262
579.0	88.5085
580.0	88.3032
581.0	88.0282
582.0	87.7077
583.0	87.3695
584.0	87.0428
585.0	86.7555
586.0	86.5322
587.0	86.3927
588.0	86.3503
589.0	86.4112
590.0	86.5743
591.0	86.8309
592.0	87.1653
593.0	87.5555
594.0	87.9748
595.0	88.3932
596.0	88.7800
597.0	89.1058
598.0	89.3457
599.0	89.4812
600.0	89.5022
601.0	89.4082
602.0	89.2079
603.0	88.9187
604.0	88.5647
605.0	88.1744
606.0	87.7786
607.0	87.4076
608.0	87.0890
609.0	86.8466
610.0	86.6983
611.0	86.6559
612.0	86.7246
613.0	86.9024
614.0	87.1806
615.0	87.5441
616.0	87.9723
617.0	88.4395
618.0	88.9172
619.0	89.3752
620.0	89.7838
621.0	90.1164
622.0	90.3511
623.0	90.4728
624.0	90.4747
625.0	90.3586
626.0	90.1347
627.0	89.8206
628.0	89.4397
629.0	89.0191
630.0	88.5878
631.0	88.1743
632.0	87.8049
633.0	87.5025
634.0	87.2858
635.0	87.1678
636.0	87.1564
637.0	87.2536
638.0	87.4557
639.0	87.7534
640.0	88.1325
641.0	88.5740
642.0	89.0555
643.0	89.5514
644.0	90.0351
645.0	90.4801
646.0	90.8614
647.0	91.1580
648.0	91.3536
649.0	91.4382
650.0	91.4087
651.0	91.2692
652.0	91.0305
653.0	90.7091
654.0	90.3257
655.0	89.9040
656.0	89.4689
657.0	89.0450
658.0	88.6552
659.0	88.3198
660.0	88.0558
661.0	87.8762
662.0	87.7897
663.0	87.8008
664.0	87.9093
665.0	88.1109
666.0	88.3971
667.0	88.7558
668.0	89.1712
669.0	89.6251
670.0	90.0968
671.0	90.5648
672.0	91.0073
673.0	91.4034
674.0	91.7346
675.0	91.9854
676.0	92.1448
677.0	92.2065
678.0	92.1694
679.0	92.0378
680.0	91.8207
681.0	91.5313
682.0	91.1858
683.0	90.8027
684.0	90.4016
685.0	90.0018
686.0	89.6219
687.0	89.2786
688.0	88.9866
689.0	88.7576
690.0	88.6007
691.0	88.5214
692.0	88.5225
693.0	88.6035
694.0	88.7608
695.0	88.9881
696.0	89.2763
697.0	89.6142
698.0	89.9884
699.0	90.3843
... 702 more lines ...
Attachment 5: Table_HR_Suprasil.TXT
400.0	88.7768
401.0	86.2695
402.0	86.9869
403.0	90.4219
404.0	93.9311
405.0	94.4202
406.0	91.5736
407.0	88.2471
408.0	87.3107
409.0	89.5856
410.0	93.4985
411.0	95.7352
412.0	94.1050
413.0	90.1872
414.0	87.3945
415.0	87.8016
416.0	91.1922
417.0	95.1117
418.0	96.1514
419.0	93.2754
420.0	89.0618
421.0	86.7369
422.0	87.7741
423.0	91.5348
424.0	95.4121
425.0	96.1862
426.0	93.1187
427.0	88.7806
428.0	86.2246
429.0	86.8956
430.0	90.3858
431.0	94.5243
432.0	96.2184
433.0	94.0233
434.0	89.7026
435.0	86.2514
436.0	85.5664
437.0	87.9247
438.0	92.0610
439.0	95.3794
440.0	95.4098
441.0	92.1048
442.0	87.8382
443.0	85.1256
444.0	85.2208
445.0	88.0207
446.0	92.1272
447.0	95.1008
448.0	94.8653
449.0	91.5769
450.0	87.4159
451.0	84.6395
452.0	84.4255
453.0	86.8006
454.0	90.7043
455.0	94.0931
456.0	94.8538
457.0	92.4633
458.0	88.4538
459.0	85.0028
460.0	83.5861
461.0	84.6760
462.0	87.8282
463.0	91.6865
464.0	94.2531
465.0	93.9852
466.0	91.0438
467.0	87.1133
468.0	84.0546
469.0	83.0053
470.0	84.2668
471.0	87.3829
472.0	91.1293
473.0	93.7318
474.0	93.7554
475.0	91.1805
476.0	87.3920
477.0	84.1027
478.0	82.4884
479.0	83.0065
480.0	85.4731
481.0	89.0633
482.0	92.3464
483.0	93.7547
484.0	92.5348
485.0	89.3177
486.0	85.6011
487.0	82.7975
488.0	81.7479
489.0	82.7037
490.0	85.3914
491.0	88.9875
492.0	92.1563
493.0	93.4820
494.0	92.2910
495.0	89.1389
496.0	85.3795
497.0	82.3416
498.0	80.8731
499.0	81.3072
500.0	83.5345
501.0	86.9877
502.0	90.5950
503.0	92.9602
504.0	92.9836
505.0	90.5767
506.0	86.7214
507.0	82.8027
508.0	79.9454
509.0	78.7948
510.0	79.5788
511.0	82.1721
512.0	86.0363
513.0	90.0979
514.0	92.8237
515.0	92.8080
516.0	89.6601
517.0	84.2984
518.0	78.2762
519.0	72.9235
520.0	69.0109
521.0	66.8539
522.0	66.5189
523.0	67.9560
524.0	71.0307
525.0	75.4775
526.0	80.8144
527.0	86.2988
528.0	91.0341
529.0	94.2687
530.0	95.7257
531.0	95.6946
532.0	94.8141
533.0	93.7444
534.0	92.9478
535.0	92.6210
536.0	92.7218
537.0	93.0387
538.0	93.2833
539.0	93.1983
540.0	92.6557
541.0	91.7049
542.0	90.5482
543.0	89.4625
544.0	88.7103
545.0	88.4738
546.0	88.8188
547.0	89.6831
548.0	90.8803
549.0	92.1261
550.0	93.0952
551.0	93.5083
552.0	93.2254
553.0	92.2962
554.0	90.9411
555.0	89.4748
556.0	88.2148
557.0	87.4127
558.0	87.2185
559.0	87.6694
560.0	88.6885
561.0	90.0902
562.0	91.5989
563.0	92.8906
564.0	93.6662
565.0	93.7361
566.0	93.0822
567.0	91.8625
568.0	90.3549
569.0	88.8746
570.0	87.7015
571.0	87.0363
572.0	86.9846
573.0	87.5537
574.0	88.6554
575.0	90.1112
576.0	91.6670
577.0	93.0282
578.0	93.9181
579.0	94.1489
580.0	93.6768
581.0	92.6139
582.0	91.1902
583.0	89.6868
584.0	88.3714
585.0	87.4543
586.0	87.0688
587.0	87.2669
588.0	88.0202
589.0	89.2232
590.0	90.6988
591.0	92.2131
592.0	93.5070
593.0	94.3454
594.0	94.5721
595.0	94.1526
596.0	93.1809
597.0	91.8511
598.0	90.4048
599.0	89.0794
600.0	88.0701
601.0	87.5117
602.0	87.4726
603.0	87.9558
604.0	88.9008
605.0	90.1865
606.0	91.6388
607.0	93.0464
608.0	94.1900
609.0	94.8829
610.0	95.0123
611.0	94.5667
612.0	93.6365
613.0	92.3892
614.0	91.0295
615.0	89.7592
616.0	88.7484
617.0	88.1199
618.0	87.9438
619.0	88.2373
620.0	88.9666
621.0	90.0492
622.0	91.3571
623.0	92.7263
624.0	93.9737
625.0	94.9227
626.0	95.4356
627.0	95.4408
628.0	94.9490
629.0	94.0483
630.0	92.8835
631.0	91.6250
632.0	90.4403
633.0	89.4723
634.0	88.8271
635.0	88.5697
636.0	88.7235
637.0	89.2712
638.0	90.1569
639.0	91.2880
640.0	92.5411
641.0	93.7712
642.0	94.8279
643.0	95.5772
644.0	95.9234
645.0	95.8275
646.0	95.3135
647.0	94.4627
648.0	93.3961
649.0	92.2526
650.0	91.1682
651.0	90.2604
652.0	89.6190
653.0	89.3025
654.0	89.3368
655.0	89.7164
656.0	90.4055
657.0	91.3394
658.0	92.4284
659.0	93.5629
660.0	94.6232
661.0	95.4926
662.0	96.0735
663.0	96.3022
664.0	96.1591
665.0	95.6715
666.0	94.9076
667.0	93.9639
668.0	92.9501
669.0	91.9742
670.0	91.1319
671.0	90.4986
672.0	90.1268
673.0	90.0442
674.0	90.2546
675.0	90.7383
676.0	91.4535
677.0	92.3386
678.0	93.3151
679.0	94.2936
680.0	95.1811
681.0	95.8911
682.0	96.3540
683.0	96.5266
684.0	96.3981
685.0	95.9906
686.0	95.3556
687.0	94.5651
688.0	93.7017
689.0	92.8487
690.0	92.0821
691.0	91.4649
692.0	91.0437
693.0	90.8475
694.0	90.8873
695.0	91.1568
696.0	91.6327
697.0	92.2769
698.0	93.0377
699.0	93.8535
... 702 more lines ...
Attachment 6: Table_HR_ULE.TXT
400.0	88.8169
401.0	86.3032
402.0	86.9425
403.0	90.2811
404.0	93.7560
405.0	94.3189
406.0	91.5829
407.0	88.2983
408.0	87.3068
409.0	89.4657
410.0	93.2829
411.0	95.5426
412.0	94.0486
413.0	90.2468
414.0	87.4569
415.0	87.7596
416.0	91.0002
417.0	94.8351
418.0	95.9534
419.0	93.2567
420.0	89.1596
421.0	86.8145
422.0	87.7168
423.0	91.3002
424.0	95.0839
425.0	95.9528
426.0	93.0943
427.0	88.9008
428.0	86.3412
429.0	86.8771
430.0	90.1658
431.0	94.1580
432.0	95.8952
433.0	93.9149
434.0	89.8013
435.0	86.4179
436.0	85.6475
437.0	87.8104
438.0	91.7310
439.0	94.9586
440.0	95.1187
441.0	92.0672
442.0	87.9911
443.0	85.3139
444.0	85.2940
445.0	87.8758
446.0	91.7558
447.0	94.6419
448.0	94.5451
449.0	91.5227
450.0	87.5732
451.0	84.8583
452.0	84.5518
453.0	86.7172
454.0	90.3711
455.0	93.6048
456.0	94.4271
457.0	92.2831
458.0	88.5372
459.0	85.2320
460.0	83.8052
461.0	84.7450
462.0	87.6525
463.0	91.2651
464.0	93.7253
465.0	93.5764
466.0	90.9056
467.0	87.2331
468.0	84.3094
469.0	83.2445
470.0	84.3542
471.0	87.2224
472.0	90.7103
473.0	93.1764
474.0	93.2819
475.0	90.9621
476.0	87.4531
477.0	84.3465
478.0	82.7737
479.0	83.1942
480.0	85.4476
481.0	88.7648
482.0	91.8222
483.0	93.1775
484.0	92.1183
485.0	89.1833
486.0	85.7296
487.0	83.0799
488.0	82.0489
489.0	82.8935
490.0	85.3625
491.0	88.6854
492.0	91.6271
493.0	92.8886
494.0	91.8394
495.0	88.9570
496.0	85.4671
497.0	82.6097
498.0	81.2008
499.0	81.5726
500.0	83.6272
501.0	86.8278
502.0	90.1722
503.0	92.3738
504.0	92.4226
505.0	90.2205
506.0	86.6455
507.0	82.9720
508.0	80.2650
509.0	79.1543
510.0	79.8697
511.0	82.2930
512.0	85.9080
513.0	89.7006
514.0	92.2446
515.0	92.2375
516.0	89.2957
517.0	84.2338
518.0	78.4829
519.0	73.3093
520.0	69.4765
521.0	67.3161
522.0	66.9105
523.0	68.2200
524.0	71.1171
525.0	75.3494
526.0	80.4646
527.0	85.7709
528.0	90.4283
529.0	93.7121
530.0	95.3162
531.0	95.4591
532.0	94.7081
533.0	93.6840
534.0	92.8442
535.0	92.4054
536.0	92.3607
537.0	92.5419
538.0	92.7036
539.0	92.6187
540.0	92.1647
541.0	91.3689
542.0	90.3932
543.0	89.4708
544.0	88.8299
545.0	88.6321
546.0	88.9374
547.0	89.6905
548.0	90.7259
549.0	91.7930
550.0	92.6095
551.0	92.9374
552.0	92.6603
553.0	91.8249
554.0	90.6244
555.0	89.3340
556.0	88.2335
557.0	87.5460
558.0	87.4044
559.0	87.8387
560.0	88.7740
561.0	90.0371
562.0	91.3761
563.0	92.5019
564.0	93.1542
565.0	93.1743
566.0	92.5554
567.0	91.4435
568.0	90.0877
569.0	88.7697
570.0	87.7392
571.0	87.1759
572.0	87.1720
573.0	87.7289
574.0	88.7592
575.0	90.0929
576.0	91.4943
577.0	92.6967
578.0	93.4565
579.0	93.6150
580.0	93.1441
581.0	92.1523
582.0	90.8506
583.0	89.4931
584.0	88.3207
585.0	87.5226
586.0	87.2173
587.0	87.4477
588.0	88.1815
589.0	89.3144
590.0	90.6766
591.0	92.0500
592.0	93.1986
593.0	93.9145
594.0	94.0664
595.0	93.6340
596.0	92.7113
597.0	91.4791
598.0	90.1586
599.0	88.9650
600.0	88.0747
601.0	87.6085
602.0	87.6251
603.0	88.1216
604.0	89.0353
605.0	90.2475
606.0	91.5915
607.0	92.8700
608.0	93.8837
609.0	94.4682
610.0	94.5301
611.0	94.0689
612.0	93.1755
613.0	92.0080
614.0	90.7552
615.0	89.6018
616.0	88.7022
617.0	88.1666
618.0	88.0562
619.0	88.3820
620.0	89.1067
621.0	90.1473
622.0	91.3790
623.0	92.6457
624.0	93.7769
625.0	94.6129
626.0	95.0332
627.0	94.9812
628.0	94.4756
629.0	93.6045
630.0	92.5054
631.0	91.3368
632.0	90.2529
633.0	89.3841
634.0	88.8265
635.0	88.6369
636.0	88.8330
637.0	89.3938
638.0	90.2616
639.0	91.3449
640.0	92.5238
641.0	93.6609
642.0	94.6170
643.0	95.2712
644.0	95.5411
645.0	95.3981
646.0	94.8721
647.0	94.0438
648.0	93.0292
649.0	91.9585
650.0	90.9577
651.0	90.1350
652.0	89.5720
653.0	89.3205
654.0	89.4015
655.0	89.8058
656.0	90.4954
657.0	91.4052
658.0	92.4467
659.0	93.5142
660.0	94.4943
661.0	95.2793
662.0	95.7814
663.0	95.9466
664.0	95.7630
665.0	95.2619
666.0	94.5118
667.0	93.6059
668.0	92.6479
669.0	91.7390
670.0	90.9675
671.0	90.4027
672.0	90.0913
673.0	90.0569
674.0	90.2996
675.0	90.7974
676.0	91.5074
677.0	92.3678
678.0	93.3021
679.0	94.2238
680.0	95.0453
681.0	95.6864
682.0	96.0849
683.0	96.2044
684.0	96.0396
685.0	95.6160
686.0	94.9856
687.0	94.2187
688.0	93.3942
689.0	92.5910
690.0	91.8798
691.0	91.3188
692.0	90.9503
693.0	90.7997
694.0	90.8750
695.0	91.1675
696.0	91.6526
697.0	92.2913
698.0	93.0324
699.0	93.8155
... 702 more lines ...
  12   Wed Jan 16 14:54:14 2019 Loïc AmoudryFixedreportlasers and optics | utilitiesOptical roomCleaning outside IP cavity

From 7th january to 11th january.

 

Attachment 1: Avant_nettoyage.jpg
Avant_nettoyage.jpg
Attachment 2: Apres_une_semaine_nettoyage_lingettes_et_brosseadents.jpg
Apres_une_semaine_nettoyage_lingettes_et_brosseadents.jpg
Attachment 3: Sans_UV_traces_scotch_nettoyees_apres_baking.jpg
Sans_UV_traces_scotch_nettoyees_apres_baking.jpg
  307   Fri Dec 22 09:46:18 2023 Ronic ChicheFixedinfolasers and optics | detectors and electronics | softwareThomX iglooChristmas shutdown

this morning, I shut off all the equipements (especially the computer) except the ones powered by the UPS : laser controller, laser peltier controller, laser motors controller.
and I disconnected the cables from the power plugs of the wall (to prevent any possible overvoltage).

Jean-Noël Cayla will need to shut off the electrical shelf (the one powering the UPS) during the afternoon, I will check at this moment if the laser controller is still properly powered thanks to the UPS.

  308   Fri Dec 22 16:13:27 2023 Ronic ChicheFixedinfolasers and optics | detectors and electronics | softwareThomX iglooChristmas shutdown

this afternoon, I checked visually what happens when the AC power of the UPS is broken down => the laser controller has still its "emission" button ON... no AC interruption.
so, I guess the UPS switched properly the output AC line from the input AC line to the internal batteries.

Ronic Chiche wrote:

this morning, I shut off all the equipements (especially the computer) except the ones powered by the UPS : laser controller, laser peltier controller, laser motors controller.
and I disconnected the cables from the power plugs of the wall (to prevent any possible overvoltage).

Jean-Noël Cayla will need to shut off the electrical shelf (the one powering the UPS) during the afternoon, I will check at this moment if the laser controller is still properly powered thanks to the UPS.

 

  378   Wed Nov 20 11:28:01 2024 Ronic ChicheFixedinfolasers and optics | detectors and electronicsThomX iglooChecking the amplifier power

this morning, with Daniele, we checked the amplifier power, right after the CVBG (the power meter has to be placed on a metal plate above the large table hole), and just at the input of the FP cavity, after the 1/4 and 1/2 waveplates (the power meter has to be placed on flat beam dump+ Thorlabs beam dump + V metallic mount to be at the right height).

we compared with the power measured the 9th of september : https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/356

the 3 collumns are the measurement close to the CVBG on 9th of september / same position today / just befor the FPC today

amplifier ratio (%)                 power after compressor (W)                 (new) power @ CVBG   (new) power @ FPC      ratio FPC/CVBG power (%)
10                                               0.91                                                  0.93                                    0.83                                       89.2
20                                               8.6                                                    8.6                                      8.0                                         93.0
30                                              16.7                                                  16.9                                    15.8                                        93.5
40                                              25.5                                                  25.6                                    24.0                                        93.7
50                                              34.5                                                  34.6                                    32.5                                        93.9
60                                              42.5                                                  42.5                                    39.5                                        92.9
70                                              50.0                                                  50.0                                    45.0                                        90.0

we observed with the viewer the beam on the powermeter.
at 70% of amplifier ratio, the beam size is as big as the powermeter detector.
then, the 93% to 90% transport efficiency drop could come from this "too small" powermeter detector.

we observed also at this power ratio (70%) that the 1/4 and 1/2 waveplates were not perfectly centered and we burn a part of the platic mount at this power.


  379   Wed Nov 20 12:00:47 2024 Ronic ChicheFixedinfolasers and optics | detectors and electronicsThomX iglooChecking the amplifier power

we did a long term run (25 mn) with the powermeter located at the FPC position at 33% amplifier ratio.

the initial power was 18W .
then, it goes to 18.5W in 10 minutes, then 5 minutes later, it goes to 18.3W and stay stable until  the end the run.

Ronic Chiche wrote:

this morning, with Daniele, we checked the amplifier power, right after the CVBG (the power meter has to be placed on a metal plate above the large table hole), and just at the input of the FP cavity, after the 1/4 and 1/2 waveplates (the power meter has to be placed on flat beam dump+ Thorlabs beam dump + V metallic mount to be at the right height).

we compared with the power measured the 9th of september : https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/356

the 3 collumns are the measurement close to the CVBG on 9th of september / same position today / just befor the FPC today

amplifier ratio (%)                 power after compressor (W)                 (new) power @ CVBG   (new) power @ FPC      ratio FPC/CVBG power (%)
10                                               0.91                                                  0.93                                    0.83                                       89.2
20                                               8.6                                                    8.6                                      8.0                                         93.0
30                                              16.7                                                  16.9                                    15.8                                        93.5
40                                              25.5                                                  25.6                                    24.0                                        93.7
50                                              34.5                                                  34.6                                    32.5                                        93.9
60                                              42.5                                                  42.5                                    39.5                                        92.9
70                                              50.0                                                  50.0                                    45.0                                        90.0

we observed with the viewer the beam on the powermeter.
at 70% of amplifier ratio, the beam size is as big as the powermeter detector.
then, the 93% to 90% transport efficiency drop could come from this "too small" powermeter detector.

we observed also at this power ratio (70%) that the 1/4 and 1/2 waveplates were not perfectly centered and we burn a part of the platic mount at this power.


 

  380   Wed Nov 20 12:42:11 2024 Ronic ChicheFixedinfolasers and optics | detectors and electronicsThomX iglooChecking the amplifier power

I put back and aligned the 1/2 waveplate only and I get 71kW max after its optimization and optimizing the CEP and alignment (without walking procedure).
this waveplate change a lot the reflected power seen on the CH2 of the scope.

to be continued this afternoon...

Ronic Chiche wrote:

we did a long term run (25 mn) with the powermeter located at the FPC position at 33% amplifier ratio.

the initial power was 18W .
then, it goes to 18.5W in 10 minutes, then 5 minutes later, it goes to 18.3W and stay stable until  the end the run.

Ronic Chiche wrote:

this morning, with Daniele, we checked the amplifier power, right after the CVBG (the power meter has to be placed on a metal plate above the large table hole), and just at the input of the FP cavity, after the 1/4 and 1/2 waveplates (the power meter has to be placed on flat beam dump+ Thorlabs beam dump + V metallic mount to be at the right height).

we compared with the power measured the 9th of september : https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/356

the 3 collumns are the measurement close to the CVBG on 9th of september / same position today / just befor the FPC today

amplifier ratio (%)                 power after compressor (W)                 (new) power @ CVBG   (new) power @ FPC      ratio FPC/CVBG power (%)
10                                               0.91                                                  0.93                                    0.83                                       89.2
20                                               8.6                                                    8.6                                      8.0                                         93.0
30                                              16.7                                                  16.9                                    15.8                                        93.5
40                                              25.5                                                  25.6                                    24.0                                        93.7
50                                              34.5                                                  34.6                                    32.5                                        93.9
60                                              42.5                                                  42.5                                    39.5                                        92.9
70                                              50.0                                                  50.0                                    45.0                                        90.0

we observed with the viewer the beam on the powermeter.
at 70% of amplifier ratio, the beam size is as big as the powermeter detector.
then, the 93% to 90% transport efficiency drop could come from this "too small" powermeter detector.

we observed also at this power ratio (70%) that the 1/4 and 1/2 waveplates were not perfectly centered and we burn a part of the platic mount at this power.


 

 

  381   Wed Nov 20 18:55:37 2024 Ronic ChicheFixedinfolasers and optics | detectors and electronicsThomX iglooChecking the amplifier power

this afternoon, we put back the 1/4 and 1/2 waveplates and after a long alignment/angle tuning/CEP optimization process, we hardly got 80-81kW in the cavity for 33% amplifier ratio. the transmission photodiode is then ~ 7.5 divisions (500mV/division) on the scope.

BUT during the 1/2 and 1/4 waveplate angle tuning process, we clearly saw:
- a reflection signal level which is very sensitive to the input polarization
- a coupling which can be almost zero with ~70kW inside the cavity
- the more important: ~9.5 divisions (500mV/division) on the scope with only 70-75kW in the powermeter
        => which device should we believe ?

the problem is all these photodiodes (reflection, error signal, transmission) and powermeter are aligned with mirrors which seem to be sensitive to the polarization state of the beam... and for the "output" beam side of the cavity, we cannot remove the mirrors because of the mechanics of the cavity which prevent a direct view of the ports. for the "input" beam side of the cavity, the output port is used for the cavity beam size measurement with a Basler camera.

we have to think to how to solve this issue !

Ronic Chiche wrote:

I put back and aligned the 1/2 waveplate only and I get 71kW max after its optimization and optimizing the CEP and alignment (without walking procedure).
this waveplate change a lot the reflected power seen on the CH2 of the scope.

to be continued this afternoon...

Ronic Chiche wrote:

we did a long term run (25 mn) with the powermeter located at the FPC position at 33% amplifier ratio.

the initial power was 18W .
then, it goes to 18.5W in 10 minutes, then 5 minutes later, it goes to 18.3W and stay stable until  the end the run.

Ronic Chiche wrote:

this morning, with Daniele, we checked the amplifier power, right after the CVBG (the power meter has to be placed on a metal plate above the large table hole), and just at the input of the FP cavity, after the 1/4 and 1/2 waveplates (the power meter has to be placed on flat beam dump+ Thorlabs beam dump + V metallic mount to be at the right height).

we compared with the power measured the 9th of september : https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/356

the 3 collumns are the measurement close to the CVBG on 9th of september / same position today / just befor the FPC today

amplifier ratio (%)                 power after compressor (W)                 (new) power @ CVBG   (new) power @ FPC      ratio FPC/CVBG power (%)
10                                               0.91                                                  0.93                                    0.83                                       89.2
20                                               8.6                                                    8.6                                      8.0                                         93.0
30                                              16.7                                                  16.9                                    15.8                                        93.5
40                                              25.5                                                  25.6                                    24.0                                        93.7
50                                              34.5                                                  34.6                                    32.5                                        93.9
60                                              42.5                                                  42.5                                    39.5                                        92.9
70                                              50.0                                                  50.0                                    45.0                                        90.0

we observed with the viewer the beam on the powermeter.
at 70% of amplifier ratio, the beam size is as big as the powermeter detector.
then, the 93% to 90% transport efficiency drop could come from this "too small" powermeter detector.

we observed also at this power ratio (70%) that the 1/4 and 1/2 waveplates were not perfectly centered and we burn a part of the platic mount at this power.


 

 

 

  21   Tue Mar 26 18:26:32 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomCheck of mirrors and coating directions

Direction of HR face of P1 and P4 have been checked, there was no problem. The mirrors are in the good direction and the coating is on the good face.

Once P1 and P4 outside of the cavity, we checked the coating position by sending the HeNe laser with an angle. There was two reflections and the most intense was the one closer to the normal to the mirror.

On P4 we noticed a spot which seems to be in the coating (does not move while sending nitrogen).

Attachment 1: ThomX_P4.jpg
ThomX_P4.jpg
  367   Tue Oct 8 11:59:27 2024 Fatematuj JohoraUnder Processreportlasers and opticsOptical roomCharacterizing Focus Tunable Lens

Yesterday Ronic and I worked on the high laser power measurements with the Vertical set-up of the Focus Tunable Lens. We went upto 38W@4 dpt at the begining and then also changed the focal power at 38W.

Observation:

1. We observed nice rings @38W 4dpt, maybe it hits some optics we have to check.

2. We also observed flactuations of power for two adjacent index of the beam profiler when we increase the laser power, which we need to understand.

3. We could see that the vertical deformation is not present for this set-up, we could remove the effect of gravity by this set-up.

Attachment 1: Updates_of_Focus_Tunable_lenses.pdf
  65   Thu Oct 3 13:53:44 2019 Loïc AmoudryFixedreportvacuumOptical roomCavity under vacuum

The cavity is right now at 6.5e-8 mbar (turbo pumps on for 1 night and ionic pumps for 3h), primary pumps and turbo pumps are cut, ionic pump are on and the pressure level is still going down. 
Therefore there aren't any big leak in the cavity and we can expect a final vacuum level close to e-9 mbar.

  58   Thu Sep 12 16:19:26 2019 Loïc AmoudryFixedreportmechanics | lasers and opticsOptical roomCavity dust source check

Several experiments have been done in order to check the potential source of dust in the ThomX cavity. For each step, we look at a clean surface, put the cavity under vacuum, go back at ambient pressure and check back the "clean surface".

1st: Primary pump on
2nd: Primary pump + Turbo pump on
3rd: Primary pump + Turbo pump + Ion pump on
4th: Primary pump + Turbo pump + Ion pump on + 10 "open/close" displacement of the pneumatic valve
5th: Primary pump + Turbo pump + Ion pump on + 1 full range displacement of P1 z motor
6th : Primary pump + Turbo pump + Ion pump on + RGA

None of these experiences have shown good correlation with dust generation. 
The RGA has shown reasonable values.
Conclusion is that dust was already in the cavity and was only displaced by all combined motors, pneumatic valves, cavity opening and closing movements.

  129   Fri Mar 11 12:59:35 2022 Manar AmerFixedreportlasers and opticsThomX iglooCavity beam Image

Updated beam image after change on the mode axis.

Taken after P4.

Attachment 1: Capture.PNG
Capture.PNG
Attachment 2: Capture2.PNG
Capture2.PNG
  20   Fri Mar 22 17:53:10 2019 Loïc AmoudryFixedreportlasers and optics | vacuumOptical roomCavity at atm pressure

Dry air injected into the cavity.

Lock has been proceeded without any trouble. ~~4000 finesse. Considered to be the same than before.

x and y motors axis put back to the last "cavity at air" position. z position kept at maximum distance between spherical mirrors.

Attachment 1: finessealair.isf
  121   Thu Feb 3 14:11:50 2022 Manar AmerFixedreportlasers and optics | detectors and electronicsThomX iglooCavity Lock and Finess measurment

This morning, me and Ronic managed to obtain the FSR and image of the finesse which is to be analyzed.

Adding:

  • Beam Image at P4 propagation, no lens is added.
  • Image of the Finesse on Oscilloscope

more information to be included later

Will continue in the afternoon .

Attachment 1: Capture3.PNG
Capture3.PNG
Attachment 2: Capture4.PNG
Capture4.PNG
Attachment 3: tek0011.png
tek0011.png
Attachment 4: tek0011CH1.isf
Attachment 5: tek0011CH3.isf
Attachment 6: tek0011CH4.isf