Entry   Electrons-Photons phase stability measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | detectors and electronics img1(1).jpg
    Reply   Electrons-Photons phase stability measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | detectors and electronics 
Message ID: 415     Entry time: Wed Mar 5 13:17:26 2025     Reply to this: 416
 Author: Ronic Chiche 
 Status: Fixed 
 Type: info 
 Category: lasers and optics | detectors and electronics 
 Location: ThomX igloo 
 Title: Electrons-Photons phase stability measurement 

we installed a measurement at the IP with the BPM for ion cleaning for electrons and we used a fast photodiode at the output of the FP-cavity and measured with a fast oscilloscope the phase jitter between the 2 signals.
in red, the BPM signal
in blue, the photodiode signal.

on the right of the picture, this is the trend of the phase difference measurement.
in yellow, this is the histogram of this phase difference.

one measured 35ps of FWHM which is equivalent to 15ps rms for a Gaussian distribution.

Attachment 1: img1(1).jpg  48 kB  | Hide | Hide all
img1(1).jpg