Entry   Checking the amplifier power, posted by Ronic Chiche at ThomX igloo about lasers and optics | detectors and electronics 
    Reply   Checking the amplifier power, posted by Ronic Chiche at ThomX igloo about lasers and optics | detectors and electronics 
       Reply   Checking the amplifier power, posted by Ronic Chiche at ThomX igloo about lasers and optics | detectors and electronics 
          Reply   Checking the amplifier power, posted by Ronic Chiche at ThomX igloo about lasers and optics | detectors and electronics 
Message ID: 380     Entry time: Wed Nov 20 12:42:11 2024     In reply to: 379     Reply to this: 381
 Author: Ronic Chiche 
 Status: Fixed 
 Type: info 
 Category: lasers and optics | detectors and electronics 
 Location: ThomX igloo 
 Title: Checking the amplifier power 

I put back and aligned the 1/2 waveplate only and I get 71kW max after its optimization and optimizing the CEP and alignment (without walking procedure).
this waveplate change a lot the reflected power seen on the CH2 of the scope.

to be continued this afternoon...

Ronic Chiche wrote:

we did a long term run (25 mn) with the powermeter located at the FPC position at 33% amplifier ratio.

the initial power was 18W .
then, it goes to 18.5W in 10 minutes, then 5 minutes later, it goes to 18.3W and stay stable until  the end the run.

Ronic Chiche wrote:

this morning, with Daniele, we checked the amplifier power, right after the CVBG (the power meter has to be placed on a metal plate above the large table hole), and just at the input of the FP cavity, after the 1/4 and 1/2 waveplates (the power meter has to be placed on flat beam dump+ Thorlabs beam dump + V metallic mount to be at the right height).

we compared with the power measured the 9th of september : https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/356

the 3 collumns are the measurement close to the CVBG on 9th of september / same position today / just befor the FPC today

amplifier ratio (%)                 power after compressor (W)                 (new) power @ CVBG   (new) power @ FPC      ratio FPC/CVBG power (%)
10                                               0.91                                                  0.93                                    0.83                                       89.2
20                                               8.6                                                    8.6                                      8.0                                         93.0
30                                              16.7                                                  16.9                                    15.8                                        93.5
40                                              25.5                                                  25.6                                    24.0                                        93.7
50                                              34.5                                                  34.6                                    32.5                                        93.9
60                                              42.5                                                  42.5                                    39.5                                        92.9
70                                              50.0                                                  50.0                                    45.0                                        90.0

we observed with the viewer the beam on the powermeter.
at 70% of amplifier ratio, the beam size is as big as the powermeter detector.
then, the 93% to 90% transport efficiency drop could come from this "too small" powermeter detector.

we observed also at this power ratio (70%) that the 1/4 and 1/2 waveplates were not perfectly centered and we burn a part of the platic mount at this power.


 

 