HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports., Page 6 of 13  Not logged in ELOG logo
ID Date Author Status Type Category Location Titledown
  99   Tue Jun 21 23:52:28 2022 Manar AmerFixedreportmechanics | lasers and optics | detectors and electronicsOptical roomMenhir Pulsed Laser

The SBox Length (FSR), was changed to match the repetition rate of the Menhir pulsed laser Frep = 216.6625 MHz

It was changed by removing the 2 D-shaped mirrors placed between M1 and M2, distance between M1-M2 ~= 169 mm 

it was still needed to be adjusted with M1 motor to reach the correct FSR.

The pulsed laser is placed on the able with injection into a fiber

The Finesse of the cavity has progressed from ~2500  to ~ 18 000

 

** Note: PowerPoint attached shows the progress over the last two weeks on the laser and 

Attachment 1: MENHIR_injection_into_fiber.jpg
MENHIR_injection_into_fiber.jpg
Attachment 2: SBox_M.jpg
SBox_M.jpg
Attachment 3: 20220620_SBox_MENHIR.pptx
  227   Fri Nov 29 16:50:04 2024 Alice RenauxUnder Processinfolasers and opticsOptical roomMenhir 216MHz laser

The Menhir 216MHz laser has been put back in place on the cavity table. Its output power is measured to be 160mW with an attenuator. Its spectrum is available in "spectre.xlsx" and "spectre.png". The main wavelength is a bit shorter (1028.75nm) and the spectrum a bit narrower (4.73nm) than expected.

Attachment 1: spectre.xlsx
Attachment 2: spectre.png
spectre.png
  226   Fri Nov 29 12:06:05 2024 Alice RenauxFixedinfolasers and opticsOptical roomMenhir 160MHz laser

The Menhir 160MHz has been put back in place on the CELIA amplifier setup. Its output power is measured to be 150-160mW with an attenuator as expected. Its spectrum is available in "spectre_avant_cvbg.xlsx" and "spectre_avant_cvbg.png".

The pulses are stretched by means of a CVBG. Their spectrum is available in "spectre_apres_cvbg.xlsx" and "spectre_apres_cvbg.png".

The laser is coupled into an optical fiber with an output power of 11.5mW for a 32mW input.

 

Attachment 1: spectre_avant_cvbg.xlsx
Attachment 2: spectre_apres_cvbg.xlsx
Attachment 3: spectre_avant_cvbg.png
spectre_avant_cvbg.png
Attachment 4: spectre_apres_cvbg.png
spectre_apres_cvbg.png
  28   Fri Feb 1 14:04:27 2019 Loïc AmoudryFixedinfolasers and opticsOtherMeeting with LMA people, AFM

30/01/19 - The following powerpoint shows the results discussed with LMA people.

What came out from the discussion is:

- Their cleaning method uses demineralized water drop on a spinner. It is probably the explanation of the circular traces on the mirror's surfaces but we still don't know what is this deposit (XPS is running out on 2 of this mirrors at this time).

-  According to their point of view, the spots could come from the coating deposition technique and are "normal". No real explaination, should not come from the substrat which is ultra-polished but can come from some clustering in the coating.

We gave them 2 of the mirrors so they can check if it is possible to clean them. They'll also do a measurement of the mirror's topology.

Attachment 1: POINT_SUR_LES_MIROIRS.pptx
  19   Wed Oct 31 14:03:22 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomM4 transmission measurement

The kept transmission for M4 is 2.17 ppm. Which is the value measured after alignement (center of the mirror).

These measurements have been taken before alignement.

I (A) Input power (mW) Transmitted power after window (mW) Transmission (ppm)
0 74.3 195.2e-6 2.63
0 73.5 212.8e-6 2.89
0 52.4 151e-6 2.88
6 28 400 72e-3 2.54

High dependence against the position. Few measurement several mm away from the center each one at different position give 194, 193, 208, 200 ppm.

  20   Wed Oct 31 14:04:50 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomM4 transmission measurement
Loïc Amoudry wrote:

The kept transmission for M4 is 2.17 ppm. Which is the value measured after alignement (center of the mirror).

These measurements have been taken before alignement.

I (A) Input power (mW) Transmitted power after window (mW) Transmission (ppm)
0 74.3 195.2e-6 2.63
0 73.5 212.8e-6 2.89
0 52.4 151e-6 2.88
6 28 400 72e-3 2.54

High dependence against the position. Few measurement several mm away from the center each one at different position give 2.86, 2.68, 2.93, 3.01, 3.06, 3.35 ppm.

 

  172   Wed Sep 28 18:43:51 2022 Manar AmerFixedreportlasers and opticsOptical roomM2 ThomX spherical Cleaning

Today, M2 Spherical-3 from ThomX that was installed inside the SBox was removed, there was one big dust on the surface of the mirror, mirror was cleaned

using pure ethanol and pure water with spincoater. (images taken with arrow far from us)

M1 from Gamma factory, fixed with the addition of the ring with 3 screws.

The mode was immediately seen after, did not have to align. After locking the cavity, we do not see the degeneracy resonance we saw yesterday.

But we still see the spot on the bottom left of the mode in transmission. The integration time for both centers maximums were 0.34 for mode and 200 for spot.

After optimizing the polarization and the CEP, we managed to get a coupling of ~ 25%

 

Manar Amer wrote:

In the morning, Vacuum broken and rotated M1 ThomX 90 degrees clockwise, locked the cavity in air and we observe a degeneracy close to the fundamental mode.

In the afternoon, I cleaned M1 from Gamma factory using pure ethanol and pure water with the spincoater

then placed it as the coupling mirror, aligned and locked in air. we observed similar degeneracy to before next to the fundamental mode.

During the process, M2 spherical from ThomX installed in the cavity was not changed. There could be damage on it, will investigate tomorrow.

 

Manar Amer wrote:

On Wednesday 21st , I opened the cavity did an additional 2 mm shift of the injection mirror and put it under vacuum again.

Locked the cavity, and observed the transmitted beam.

The second spot is still visible on the beam profiler , the distance difference between the 2 spots is ~ 5.2 mm (the same as before )

no difference in distance, decreases the likelihood that it is from the damage (to be investigated more)

in addition, we have locked at the reflection from the cavity to confirm the spot next to the beam.

We took two images when the laser was locked with the cavity and when it was not.

We clearly see that the spot is indeed related to the mode of the cavity. And probably the damaged spot.

(Difference is size on the reflection image is due to the distance is larger than the transmission + the spherical mirror effect is not there)

 

Manar Amer wrote:

To compare between the 2 images of the cavity mode:

  • the mode by itself has an integration time of 0.06 ms, position (x, y) = (1142.969, -53.932) um on the beam profiler 
  • the mode saturated with the spot next to it almost at max intensity has an integration time of 50 ms, position (-3700, -2000) um

comparing the positions of both spots, they have difference (4842.969, 2053.932) um

------ > total difference on the beam profiler ~ 5.3 mm , the distance from the spherical mirror to the beam profiler is ~ 40 cm

 

Manar Amer wrote:

Yesterday , we locked the cavity and we see a sign of a high finesse on the transmission signal, but no measurement of Finesse was done.

we have a coupling of ~ 45%, which is a loss of 20% from the previous coupling of 60%

an estimate done by Ronic MATLAB simulation for the coupling drop where we have 200 pp additional losses and gain of 2.6 k we should get a transmission of 1.1 mW for injected power of ~ 300 mW

which is consistent with the power measured after a 50% beam splitter on transmission we got 0.51 mW (total would be 1.02mW)

in addition, there is a beam that is next to the mode of the cavity , confirmed it was not a reflection from the beam splitter or the optics.

it could be that we are still close to the damaged spot ?

Manar Amer wrote:

ThomX injection mirror has been cleaned and placed again inside the optical cavity.

This time to avoid the damaged spot I have displaced the mirror mount horizontally to have a distance between center of the beam and the spot ~ 2.5 - 3 mm.

The alignment was affected slightly but recovered by adjusting the mirror mount nobs, (00 mode observed in air)

The cavity was closed is being pumped with vacuum.

To be done: adjust the cavity length and find the resonance, improve the outer alignment, lock the cavity

Manar Amer wrote:

a better image of the damaged spot, image taken with the arrow for the reflective surface facing the other direction (image shows position)

The image of M1 for ThomX reflective surface was taken at min zoom (full image scale 13 mm) and max zoom (full image scale 2 mm) on microscope

The spot appears to be not close to the center of the mirror, at max zoom in the center we do not see the spot it is just out of the image 

the last image has the mirror position adjusted to center the damaged spot for a better image of it.

Manar Amer wrote:

After discussing, we have decided against shifting the mirror to avoid the time lost.

We changed the injection mirror to a different mirror from Mighty Laser set, Transmission of mirror 80 ppm. (no visible damage at the center of the mirror, only a small scratch on the back)

mirror cleaned using pure ethanol and pure water with spin coater, also the spherical mirror was cleaned again.

 

Manar Amer wrote:

Tomorrow will try to shift the injection mirror to avoid hitting the damaged spot.

 

Manar Amer wrote:

Following the storage of ~ 50 kW inside the cavity and a sudden drop in transmitted power from the cavity 

damage to the mirror surface was suspected.

We broke vacuum and took images of the surface of the 2 mirrors in the cavity, the spherical and the planar mirror 

image 1 , spherical reflective surface (no visible damage with the UV light, and no visible damage under the microscope)

image 2 , planar coupler mirror reflective surface (no visible damage under UV light, but under the microscope there is a damaged spot close to the center)

image 4 is the planar surface reflective surface at zoom 8 on the microscope.

 

 

 

 

 

 

 

 

 

 

Attachment 1: MS_directfromcavity_reduceSize.jpg
MS_directfromcavity_reduceSize.jpg
Attachment 2: MS_afteralcholeswipe_reducedSize.jpg
MS_afteralcholeswipe_reducedSize.jpg
Attachment 3: Screenshot_2022-09-28_0_172610.png
Screenshot_2022-09-28_0_172610.png
Attachment 4: 00mode0point34ms_transmission.jpg
00mode0point34ms_transmission.jpg
Attachment 5: 00mode_200ms.jpg
00mode_200ms.jpg
Attachment 6: reflection_lock.jpg
reflection_lock.jpg
  17   Wed Oct 31 13:48:58 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomM1 transmission measurement

Center of M1:

I (A) Input power (mW) Transmitted power (mW) Transmission (ppm)
0 78 0.01477 189
4 16500 3.2 194
  21   Wed Oct 31 14:06:17 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomM1 transmission measurement

At different positions, transmission give 194, 193, 208, 200 ppm?

Loïc Amoudry wrote:

Center of M1:

I (A) Input power (mW) Transmitted power (mW) Transmission (ppm)
0 78 0.01477 189
4 16500 3.2 194

 

  168   Tue Sep 20 12:50:40 2022 Manar AmerFixedreportlasers and opticsOptical roomM1 ThomX used while shifted from damaged spot

Yesterday , we locked the cavity and we see a sign of a high finesse on the transmission signal, but no measurement of Finesse was done.

we have a coupling of ~ 45%, which is a loss of 20% from the previous coupling of 60%

an estimate done by Ronic MATLAB simulation for the coupling drop where we have 200 pp additional losses and gain of 2.6 k we should get a transmission of 1.1 mW for injected power of ~ 300 mW

which is consistent with the power measured after a 50% beam splitter on transmission we got 0.51 mW (total would be 1.02mW)

in addition, there is a beam that is next to the mode of the cavity , confirmed it was not a reflection from the beam splitter or the optics.

it could be that we are still close to the damaged spot ?

Manar Amer wrote:

ThomX injection mirror has been cleaned and placed again inside the optical cavity.

This time to avoid the damaged spot I have displaced the mirror mount horizontally to have a distance between center of the beam and the spot ~ 2.5 - 3 mm.

The alignment was affected slightly but recovered by adjusting the mirror mount nobs, (00 mode observed in air)

The cavity was closed is being pumped with vacuum.

To be done: adjust the cavity length and find the resonance, improve the outer alignment, lock the cavity

Manar Amer wrote:

a better image of the damaged spot, image taken with the arrow for the reflective surface facing the other direction (image shows position)

The image of M1 for ThomX reflective surface was taken at min zoom (full image scale 13 mm) and max zoom (full image scale 2 mm) on microscope

The spot appears to be not close to the center of the mirror, at max zoom in the center we do not see the spot it is just out of the image 

the last image has the mirror position adjusted to center the damaged spot for a better image of it.

Manar Amer wrote:

After discussing, we have decided against shifting the mirror to avoid the time lost.

We changed the injection mirror to a different mirror from Mighty Laser set, Transmission of mirror 80 ppm. (no visible damage at the center of the mirror, only a small scratch on the back)

mirror cleaned using pure ethanol and pure water with spin coater, also the spherical mirror was cleaned again.

 

Manar Amer wrote:

Tomorrow will try to shift the injection mirror to avoid hitting the damaged spot.

 

Manar Amer wrote:

Following the storage of ~ 50 kW inside the cavity and a sudden drop in transmitted power from the cavity 

damage to the mirror surface was suspected.

We broke vacuum and took images of the surface of the 2 mirrors in the cavity, the spherical and the planar mirror 

image 1 , spherical reflective surface (no visible damage with the UV light, and no visible damage under the microscope)

image 2 , planar coupler mirror reflective surface (no visible damage under UV light, but under the microscope there is a damaged spot close to the center)

image 4 is the planar surface reflective surface at zoom 8 on the microscope.

 

 

 

 

 

 

Attachment 1: Screenshot_2022-09-19_2_172943.png
Screenshot_2022-09-19_2_172943.png
Attachment 2: Screenshot_2022-09-19_1_171834.png
Screenshot_2022-09-19_1_171834.png
Attachment 3: 00mode_LOCK.jpg
00mode_LOCK.jpg
Attachment 4: 00mode_LOCK_saturaed.jpg
00mode_LOCK_saturaed.jpg
  169   Wed Sep 21 12:10:07 2022 Manar AmerFixedreportlasers and opticsOptical roomM1 ThomX used while shifted from damaged spot

To compare between the 2 images of the cavity mode:

  • the mode by itself has an integration time of 0.06 ms, position (x, y) = (1142.969, -53.932) um on the beam profiler 
  • the mode saturated with the spot next to it almost at max intensity has an integration time of 50 ms, position (-3700, -2000) um

comparing the positions of both spots, they have difference (4842.969, 2053.932) um

------ > total difference on the beam profiler ~ 5.3 mm , the distance from the spherical mirror to the beam profiler is ~ 40 cm

 

Manar Amer wrote:

Yesterday , we locked the cavity and we see a sign of a high finesse on the transmission signal, but no measurement of Finesse was done.

we have a coupling of ~ 45%, which is a loss of 20% from the previous coupling of 60%

an estimate done by Ronic MATLAB simulation for the coupling drop where we have 200 pp additional losses and gain of 2.6 k we should get a transmission of 1.1 mW for injected power of ~ 300 mW

which is consistent with the power measured after a 50% beam splitter on transmission we got 0.51 mW (total would be 1.02mW)

in addition, there is a beam that is next to the mode of the cavity , confirmed it was not a reflection from the beam splitter or the optics.

it could be that we are still close to the damaged spot ?

Manar Amer wrote:

ThomX injection mirror has been cleaned and placed again inside the optical cavity.

This time to avoid the damaged spot I have displaced the mirror mount horizontally to have a distance between center of the beam and the spot ~ 2.5 - 3 mm.

The alignment was affected slightly but recovered by adjusting the mirror mount nobs, (00 mode observed in air)

The cavity was closed is being pumped with vacuum.

To be done: adjust the cavity length and find the resonance, improve the outer alignment, lock the cavity

Manar Amer wrote:

a better image of the damaged spot, image taken with the arrow for the reflective surface facing the other direction (image shows position)

The image of M1 for ThomX reflective surface was taken at min zoom (full image scale 13 mm) and max zoom (full image scale 2 mm) on microscope

The spot appears to be not close to the center of the mirror, at max zoom in the center we do not see the spot it is just out of the image 

the last image has the mirror position adjusted to center the damaged spot for a better image of it.

Manar Amer wrote:

After discussing, we have decided against shifting the mirror to avoid the time lost.

We changed the injection mirror to a different mirror from Mighty Laser set, Transmission of mirror 80 ppm. (no visible damage at the center of the mirror, only a small scratch on the back)

mirror cleaned using pure ethanol and pure water with spin coater, also the spherical mirror was cleaned again.

 

Manar Amer wrote:

Tomorrow will try to shift the injection mirror to avoid hitting the damaged spot.

 

Manar Amer wrote:

Following the storage of ~ 50 kW inside the cavity and a sudden drop in transmitted power from the cavity 

damage to the mirror surface was suspected.

We broke vacuum and took images of the surface of the 2 mirrors in the cavity, the spherical and the planar mirror 

image 1 , spherical reflective surface (no visible damage with the UV light, and no visible damage under the microscope)

image 2 , planar coupler mirror reflective surface (no visible damage under UV light, but under the microscope there is a damaged spot close to the center)

image 4 is the planar surface reflective surface at zoom 8 on the microscope.

 

 

 

 

 

 

 

  170   Mon Sep 26 16:15:06 2022 Manar AmerFixedreportlasers and opticsOptical roomM1 ThomX used while shifted from damaged spot

On Wednesday 21st , I opened the cavity did an additional 2 mm shift of the injection mirror and put it under vacuum again.

Locked the cavity, and observed the transmitted beam.

The second spot is still visible on the beam profiler , the distance difference between the 2 spots is ~ 5.2 mm (the same as before )

no difference in distance, decreases the likelihood that it is from the damage (to be investigated more)

in addition, we have locked at the reflection from the cavity to confirm the spot next to the beam.

We took two images when the laser was locked with the cavity and when it was not.

We clearly see that the spot is indeed related to the mode of the cavity. And probably the damaged spot.

(Difference is size on the reflection image is due to the distance is larger than the transmission + the spherical mirror effect is not there)

 

Manar Amer wrote:

To compare between the 2 images of the cavity mode:

  • the mode by itself has an integration time of 0.06 ms, position (x, y) = (1142.969, -53.932) um on the beam profiler 
  • the mode saturated with the spot next to it almost at max intensity has an integration time of 50 ms, position (-3700, -2000) um

comparing the positions of both spots, they have difference (4842.969, 2053.932) um

------ > total difference on the beam profiler ~ 5.3 mm , the distance from the spherical mirror to the beam profiler is ~ 40 cm

 

Manar Amer wrote:

Yesterday , we locked the cavity and we see a sign of a high finesse on the transmission signal, but no measurement of Finesse was done.

we have a coupling of ~ 45%, which is a loss of 20% from the previous coupling of 60%

an estimate done by Ronic MATLAB simulation for the coupling drop where we have 200 pp additional losses and gain of 2.6 k we should get a transmission of 1.1 mW for injected power of ~ 300 mW

which is consistent with the power measured after a 50% beam splitter on transmission we got 0.51 mW (total would be 1.02mW)

in addition, there is a beam that is next to the mode of the cavity , confirmed it was not a reflection from the beam splitter or the optics.

it could be that we are still close to the damaged spot ?

Manar Amer wrote:

ThomX injection mirror has been cleaned and placed again inside the optical cavity.

This time to avoid the damaged spot I have displaced the mirror mount horizontally to have a distance between center of the beam and the spot ~ 2.5 - 3 mm.

The alignment was affected slightly but recovered by adjusting the mirror mount nobs, (00 mode observed in air)

The cavity was closed is being pumped with vacuum.

To be done: adjust the cavity length and find the resonance, improve the outer alignment, lock the cavity

Manar Amer wrote:

a better image of the damaged spot, image taken with the arrow for the reflective surface facing the other direction (image shows position)

The image of M1 for ThomX reflective surface was taken at min zoom (full image scale 13 mm) and max zoom (full image scale 2 mm) on microscope

The spot appears to be not close to the center of the mirror, at max zoom in the center we do not see the spot it is just out of the image 

the last image has the mirror position adjusted to center the damaged spot for a better image of it.

Manar Amer wrote:

After discussing, we have decided against shifting the mirror to avoid the time lost.

We changed the injection mirror to a different mirror from Mighty Laser set, Transmission of mirror 80 ppm. (no visible damage at the center of the mirror, only a small scratch on the back)

mirror cleaned using pure ethanol and pure water with spin coater, also the spherical mirror was cleaned again.

 

Manar Amer wrote:

Tomorrow will try to shift the injection mirror to avoid hitting the damaged spot.

 

Manar Amer wrote:

Following the storage of ~ 50 kW inside the cavity and a sudden drop in transmitted power from the cavity 

damage to the mirror surface was suspected.

We broke vacuum and took images of the surface of the 2 mirrors in the cavity, the spherical and the planar mirror 

image 1 , spherical reflective surface (no visible damage with the UV light, and no visible damage under the microscope)

image 2 , planar coupler mirror reflective surface (no visible damage under UV light, but under the microscope there is a damaged spot close to the center)

image 4 is the planar surface reflective surface at zoom 8 on the microscope.

 

 

 

 

 

 

 

 

Attachment 1: 00mode_damaged_0_26ms.jpg
00mode_damaged_0_26ms.jpg
Attachment 2: 00mode_damaged_saturted_50ms.jpg
00mode_damaged_saturted_50ms.jpg
Attachment 3: reflection_noLock.jpg
reflection_noLock.jpg
Attachment 4: reflection_Lock_0_15ms.jpg
reflection_Lock_0_15ms.jpg
  171   Tue Sep 27 19:30:49 2022 Manar AmerFixedreportlasers and opticsOptical roomM1 Gamma Factory

In the morning, Vacuum broken and rotated M1 ThomX 90 degrees clockwise, locked the cavity in air and we observe a degeneracy close to the fundamental mode.

In the afternoon, I cleaned M1 from Gamma factory using pure ethanol and pure water with the spincoater

then placed it as the coupling mirror, aligned and locked in air. we observed similar degeneracy to before next to the fundamental mode.

During the process, M2 spherical from ThomX installed in the cavity was not changed. There could be damage on it, will investigate tomorrow.

 

Manar Amer wrote:

On Wednesday 21st , I opened the cavity did an additional 2 mm shift of the injection mirror and put it under vacuum again.

Locked the cavity, and observed the transmitted beam.

The second spot is still visible on the beam profiler , the distance difference between the 2 spots is ~ 5.2 mm (the same as before )

no difference in distance, decreases the likelihood that it is from the damage (to be investigated more)

in addition, we have locked at the reflection from the cavity to confirm the spot next to the beam.

We took two images when the laser was locked with the cavity and when it was not.

We clearly see that the spot is indeed related to the mode of the cavity. And probably the damaged spot.

(Difference is size on the reflection image is due to the distance is larger than the transmission + the spherical mirror effect is not there)

 

Manar Amer wrote:

To compare between the 2 images of the cavity mode:

  • the mode by itself has an integration time of 0.06 ms, position (x, y) = (1142.969, -53.932) um on the beam profiler 
  • the mode saturated with the spot next to it almost at max intensity has an integration time of 50 ms, position (-3700, -2000) um

comparing the positions of both spots, they have difference (4842.969, 2053.932) um

------ > total difference on the beam profiler ~ 5.3 mm , the distance from the spherical mirror to the beam profiler is ~ 40 cm

 

Manar Amer wrote:

Yesterday , we locked the cavity and we see a sign of a high finesse on the transmission signal, but no measurement of Finesse was done.

we have a coupling of ~ 45%, which is a loss of 20% from the previous coupling of 60%

an estimate done by Ronic MATLAB simulation for the coupling drop where we have 200 pp additional losses and gain of 2.6 k we should get a transmission of 1.1 mW for injected power of ~ 300 mW

which is consistent with the power measured after a 50% beam splitter on transmission we got 0.51 mW (total would be 1.02mW)

in addition, there is a beam that is next to the mode of the cavity , confirmed it was not a reflection from the beam splitter or the optics.

it could be that we are still close to the damaged spot ?

Manar Amer wrote:

ThomX injection mirror has been cleaned and placed again inside the optical cavity.

This time to avoid the damaged spot I have displaced the mirror mount horizontally to have a distance between center of the beam and the spot ~ 2.5 - 3 mm.

The alignment was affected slightly but recovered by adjusting the mirror mount nobs, (00 mode observed in air)

The cavity was closed is being pumped with vacuum.

To be done: adjust the cavity length and find the resonance, improve the outer alignment, lock the cavity

Manar Amer wrote:

a better image of the damaged spot, image taken with the arrow for the reflective surface facing the other direction (image shows position)

The image of M1 for ThomX reflective surface was taken at min zoom (full image scale 13 mm) and max zoom (full image scale 2 mm) on microscope

The spot appears to be not close to the center of the mirror, at max zoom in the center we do not see the spot it is just out of the image 

the last image has the mirror position adjusted to center the damaged spot for a better image of it.

Manar Amer wrote:

After discussing, we have decided against shifting the mirror to avoid the time lost.

We changed the injection mirror to a different mirror from Mighty Laser set, Transmission of mirror 80 ppm. (no visible damage at the center of the mirror, only a small scratch on the back)

mirror cleaned using pure ethanol and pure water with spin coater, also the spherical mirror was cleaned again.

 

Manar Amer wrote:

Tomorrow will try to shift the injection mirror to avoid hitting the damaged spot.

 

Manar Amer wrote:

Following the storage of ~ 50 kW inside the cavity and a sudden drop in transmitted power from the cavity 

damage to the mirror surface was suspected.

We broke vacuum and took images of the surface of the 2 mirrors in the cavity, the spherical and the planar mirror 

image 1 , spherical reflective surface (no visible damage with the UV light, and no visible damage under the microscope)

image 2 , planar coupler mirror reflective surface (no visible damage under UV light, but under the microscope there is a damaged spot close to the center)

image 4 is the planar surface reflective surface at zoom 8 on the microscope.

 

 

 

 

 

 

 

 

 

Attachment 1: 00mode_ThomX_Mirror_degeneracySpot_1point04ms.jpg
00mode_ThomX_Mirror_degeneracySpot_1point04ms.jpg
Attachment 2: 00mode_ThomX_Mirror_degeneracySpot_saturateat50.jpg
00mode_ThomX_Mirror_degeneracySpot_saturateat50.jpg
Attachment 3: Coupling_mirror_.jpeg
Coupling_mirror_.jpeg
  146   Wed Sep 7 18:58:03 2022 Manar AmerFixedreportlasers and opticsOptical roomLocking Amplifier with Cavity

Today we managed to observe the fundamental mode and stabilize the scan on it until we improved the alignment enough.

We see some coupling, but it is very week < 5% , We improved the alignment and the polarization, but there is still no explanation to why it is very low.

The mode shape is circular with radius = 0.89 mm at transmission point, ~ 40 cm from circular mirror.

One EOM was installed along the injection into the amplifier, we saw a drop in the power measured by the first stage monitor from ~ 8.2 mW to 7.2 mW

we improved the injection up to 7.6 mW, but it still fluctuates a lot. We need to be careful about it.

The error signal looks clean, but it is very week which is due to the weak coupling.

 

 
Attachment 1: 20220907_00mode.jpg
20220907_00mode.jpg
Attachment 2: 20220907_00mode_diameter.jpg
20220907_00mode_diameter.jpg
Attachment 3: Screenshot_2022-09-07_0_174654.png
Screenshot_2022-09-07_0_174654.png
Attachment 4: Screenshot_2022-09-07_2_182520.png
Screenshot_2022-09-07_2_182520.png
  147   Thu Sep 8 11:16:40 2022 Manar AmerFixedreportlasers and opticsOptical roomLocking Amplifier with Cavity

Today, we played on the CEP but using the I-tune input of the Menhir Laser (+/- 5V maximal range).

unfortunately, one only saw a very weak improvement of the transmission by 10-20%... and the coupling improvement is almost zero.

the best improvement was for the maximal I-tune range (+5V) which maybe means that we could improve more the effect if were able to get a full range of 2pi for the CEP (instead of the present pi/2 range).

Manar Amer wrote:

Today we managed to observe the fundamental mode and stabilize the scan on it until we improved the alignment enough.

We see some coupling, but it is very week < 5% , We improved the alignment and the polarization, but there is still no explanation to why it is very low.

The mode shape is circular with radius = 0.89 mm at transmission point, ~ 40 cm from circular mirror.

One EOM was installed along the injection into the amplifier, we saw a drop in the power measured by the first stage monitor from ~ 8.2 mW to 7.2 mW

we improved the injection up to 7.6 mW, but it still fluctuates a lot. We need to be careful about it.

The error signal looks clean, but it is very week which is due to the weak coupling.

 

 

 

  148   Thu Sep 8 13:39:03 2022 Manar AmerFixedreportlasers and opticsOptical roomLocking Amplifier with Cavity

We played on the CEP using the USB command "id0=xxxxx" of the Menhir.

we put id0=48650 and we improved A LOT the transmission and coupling (~ 60%)

here is an image of the first attempts to lock... but the locking is quite difficult.
yellow : transmission
green : reflection
amplifier is still with 2nd stage only (Pout=~1W)

we are adding an AOM in the loop...

Manar Amer wrote:

Today, we played on the CEP but using the I-tune input of the Menhir Laser (+/- 5V maximal range).

unfortunately, one only saw a very weak improvement of the transmission by 10-20%... and the coupling improvement is almost zero.

the best improvement was for the maximal I-tune range (+5V) which maybe means that we could improve more the effect if were able to get a full range of 2pi for the CEP (instead of the present pi/2 range).

Manar Amer wrote:

Today we managed to observe the fundamental mode and stabilize the scan on it until we improved the alignment enough.

We see some coupling, but it is very week < 5% , We improved the alignment and the polarization, but there is still no explanation to why it is very low.

The mode shape is circular with radius = 0.89 mm at transmission point, ~ 40 cm from circular mirror.

One EOM was installed along the injection into the amplifier, we saw a drop in the power measured by the first stage monitor from ~ 8.2 mW to 7.2 mW

we improved the injection up to 7.6 mW, but it still fluctuates a lot. We need to be careful about it.

The error signal looks clean, but it is very week which is due to the weak coupling.

 

 

 

 

Attachment 1: Screenshot_2022-09-08_0_133245.png
Screenshot_2022-09-08_0_133245.png
  149   Thu Sep 8 14:58:51 2022 Manar AmerFixedreportlasers and opticsOptical roomLocking Amplifier with Cavity

We managed to lock the cavity by adding the AOM , but the lock is still difficult to stabilize.

There is some high frequency compensated by the AOM at ~ 170 kHz (yet to understand from where it comes from)

Manar Amer wrote:

We played on the CEP using the USB command "id0=xxxxx" of the Menhir.

we put id0=48650 and we improved A LOT the transmission and coupling (~ 60%)

here is an image of the first attempts to lock... but the locking is quite difficult.
yellow : transmission
green : reflection
amplifier is still with 2nd stage only (Pout=~1W)

we are adding an AOM in the loop...

Manar Amer wrote:

Today, we played on the CEP but using the I-tune input of the Menhir Laser (+/- 5V maximal range).

unfortunately, one only saw a very weak improvement of the transmission by 10-20%... and the coupling improvement is almost zero.

the best improvement was for the maximal I-tune range (+5V) which maybe means that we could improve more the effect if were able to get a full range of 2pi for the CEP (instead of the present pi/2 range).

Manar Amer wrote:

Today we managed to observe the fundamental mode and stabilize the scan on it until we improved the alignment enough.

We see some coupling, but it is very week < 5% , We improved the alignment and the polarization, but there is still no explanation to why it is very low.

The mode shape is circular with radius = 0.89 mm at transmission point, ~ 40 cm from circular mirror.

One EOM was installed along the injection into the amplifier, we saw a drop in the power measured by the first stage monitor from ~ 8.2 mW to 7.2 mW

we improved the injection up to 7.6 mW, but it still fluctuates a lot. We need to be careful about it.

The error signal looks clean, but it is very week which is due to the weak coupling.

 

 

 

 

 

Attachment 1: Screenshot_2022-09-08_2_144338.png
Screenshot_2022-09-08_2_144338.png
  150   Thu Sep 8 18:28:51 2022 Manar AmerFixedreportlasers and opticsOptical roomLocking Amplifier with Cavity

We wanted to increase the power of the amplifier to measure the transmission output at M2,

but locking the cavity again was very, very difficult.

We will try again tomorrow.

Manar Amer wrote:

We managed to lock the cavity by adding the AOM , but the lock is still difficult to stabilize.

There is some high frequency compensated by the AOM at ~ 170 kHz (yet to understand from where it comes from)

Manar Amer wrote:

We played on the CEP using the USB command "id0=xxxxx" of the Menhir.

we put id0=48650 and we improved A LOT the transmission and coupling (~ 60%)

here is an image of the first attempts to lock... but the locking is quite difficult.
yellow : transmission
green : reflection
amplifier is still with 2nd stage only (Pout=~1W)

we are adding an AOM in the loop...

Manar Amer wrote:

Today, we played on the CEP but using the I-tune input of the Menhir Laser (+/- 5V maximal range).

unfortunately, one only saw a very weak improvement of the transmission by 10-20%... and the coupling improvement is almost zero.

the best improvement was for the maximal I-tune range (+5V) which maybe means that we could improve more the effect if were able to get a full range of 2pi for the CEP (instead of the present pi/2 range).

Manar Amer wrote:

Today we managed to observe the fundamental mode and stabilize the scan on it until we improved the alignment enough.

We see some coupling, but it is very week < 5% , We improved the alignment and the polarization, but there is still no explanation to why it is very low.

The mode shape is circular with radius = 0.89 mm at transmission point, ~ 40 cm from circular mirror.

One EOM was installed along the injection into the amplifier, we saw a drop in the power measured by the first stage monitor from ~ 8.2 mW to 7.2 mW

we improved the injection up to 7.6 mW, but it still fluctuates a lot. We need to be careful about it.

The error signal looks clean, but it is very week which is due to the weak coupling.

 

 

 

 

 

 

  151   Fri Sep 9 11:50:17 2022 Manar AmerFixedreportlasers and opticsOptical roomLocking Amplifier with Cavity

We locked the cavity, and it is stable using the Transmission,

the high frequency that we thought could have a reason for instability, is due to the high power on the photo diode of the PDH box which can cause non-linearity effects in the signal.

We also closed the fan of the third stage and there was no significant change on the error signal and the piezo+AOM compensation signal.  

        

Manar Amer wrote:

We managed to lock the cavity by adding the AOM , but the lock is still difficult to stabilize.

There is some high frequency compensated by the AOM at ~ 170 kHz (yet to understand from where it comes from)

Manar Amer wrote:

We played on the CEP using the USB command "id0=xxxxx" of the Menhir.

we put id0=48650 and we improved A LOT the transmission and coupling (~ 60%)

here is an image of the first attempts to lock... but the locking is quite difficult.
yellow : transmission
green : reflection
amplifier is still with 2nd stage only (Pout=~1W)

we are adding an AOM in the loop...

Manar Amer wrote:

Today, we played on the CEP but using the I-tune input of the Menhir Laser (+/- 5V maximal range).

unfortunately, one only saw a very weak improvement of the transmission by 10-20%... and the coupling improvement is almost zero.

the best improvement was for the maximal I-tune range (+5V) which maybe means that we could improve more the effect if were able to get a full range of 2pi for the CEP (instead of the present pi/2 range).

Manar Amer wrote:

Today we managed to observe the fundamental mode and stabilize the scan on it until we improved the alignment enough.

We see some coupling, but it is very week < 5% , We improved the alignment and the polarization, but there is still no explanation to why it is very low.

The mode shape is circular with radius = 0.89 mm at transmission point, ~ 40 cm from circular mirror.

One EOM was installed along the injection into the amplifier, we saw a drop in the power measured by the first stage monitor from ~ 8.2 mW to 7.2 mW

we improved the injection up to 7.6 mW, but it still fluctuates a lot. We need to be careful about it.

The error signal looks clean, but it is very week which is due to the weak coupling.

 

 

 

 

 

 

  152   Fri Sep 9 12:45:19 2022 Manar AmerFixedreportlasers and opticsOptical roomLocking Amplifier with Cavity

image of the lock with Coupling of ~ 60%

 

Manar Amer wrote:

We locked the cavity, and it is stable using the Transmission,

the high frequency that we thought could have a reason for instability, is due to the high power on the photo diode of the PDH box which can cause non-linearity effects in the signal.

We also closed the fan of the third stage and there was no significant change on the error signal and the piezo+AOM compensation signal.  

        

Manar Amer wrote:

We managed to lock the cavity by adding the AOM , but the lock is still difficult to stabilize.

There is some high frequency compensated by the AOM at ~ 170 kHz (yet to understand from where it comes from)

Manar Amer wrote:

We played on the CEP using the USB command "id0=xxxxx" of the Menhir.

we put id0=48650 and we improved A LOT the transmission and coupling (~ 60%)

here is an image of the first attempts to lock... but the locking is quite difficult.
yellow : transmission
green : reflection
amplifier is still with 2nd stage only (Pout=~1W)

we are adding an AOM in the loop...

Manar Amer wrote:

Today, we played on the CEP but using the I-tune input of the Menhir Laser (+/- 5V maximal range).

unfortunately, one only saw a very weak improvement of the transmission by 10-20%... and the coupling improvement is almost zero.

the best improvement was for the maximal I-tune range (+5V) which maybe means that we could improve more the effect if were able to get a full range of 2pi for the CEP (instead of the present pi/2 range).

Manar Amer wrote:

Today we managed to observe the fundamental mode and stabilize the scan on it until we improved the alignment enough.

We see some coupling, but it is very week < 5% , We improved the alignment and the polarization, but there is still no explanation to why it is very low.

The mode shape is circular with radius = 0.89 mm at transmission point, ~ 40 cm from circular mirror.

One EOM was installed along the injection into the amplifier, we saw a drop in the power measured by the first stage monitor from ~ 8.2 mW to 7.2 mW

we improved the injection up to 7.6 mW, but it still fluctuates a lot. We need to be careful about it.

The error signal looks clean, but it is very week which is due to the weak coupling.

 

 

 

 

 

 

 

Attachment 1: Screenshot_2022-09-09_0_120737.png
Screenshot_2022-09-09_0_120737.png
ELOG V3.1.4-395e101