| ID |
Date |
Author |
Status |
Type |
Category |
Location |
Title |
|
307
|
Fri Dec 22 09:46:18 2023 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | Christmas shutdown |
this morning, I shut off all the equipements (especially the computer) except the ones powered by the UPS : laser controller, laser peltier controller, laser motors controller.
and I disconnected the cables from the power plugs of the wall (to prevent any possible overvoltage).
Jean-Noël Cayla will need to shut off the electrical shelf (the one powering the UPS) during the afternoon, I will check at this moment if the laser controller is still properly powered thanks to the UPS. |
|
308
|
Fri Dec 22 16:13:27 2023 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | Christmas shutdown |
this afternoon, I checked visually what happens when the AC power of the UPS is broken down => the laser controller has still its "emission" button ON... no AC interruption.
so, I guess the UPS switched properly the output AC line from the input AC line to the internal batteries.
| Ronic Chiche wrote: |
|
this morning, I shut off all the equipements (especially the computer) except the ones powered by the UPS : laser controller, laser peltier controller, laser motors controller.
and I disconnected the cables from the power plugs of the wall (to prevent any possible overvoltage).
Jean-Noël Cayla will need to shut off the electrical shelf (the one powering the UPS) during the afternoon, I will check at this moment if the laser controller is still properly powered thanks to the UPS.
|
|
|
309
|
Wed Jan 10 14:52:22 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | Restarting equipements after Christmas shutdown |
this morning, I plugged back the 4 power cables which supply all the equipments.
all the equipments have restarted without problem, but the shutter of the laser amplifier needs a ThomX safety system grant to let the beam propagate to the cavity...
and then continue the commissioning of the restart after Christmas shutdown. |
|
311
|
Wed Jan 10 18:43:03 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | Restarting equipements after Christmas shutdown |
the laser coupled into the amplifier is still measured at ~3mW at the input, on the Alphanov software, once the Amp stage are started at 0% of power ratio.
this afternoon, Kevin restarted the 3rd safety system of ThomX to grant the opening of the beam shutter => now the shutter can be opened.
after adjusting the phase for the PDH system, I was able to find back the laser/CFP lock.
after some minimal alignement, one gets back 50kW for 30.5% of laser amplifier ratio, as before the Christmas shutdown.
I didn't try to lock the FP cavity on the 500MHz RF reference clock as I'm not sure if it has been set to 500.25MHz (as before shutdown) or at 500.1MHz (the future RF frequency) which is very far and will require to move the FP cavity on a long rage.
=> I will ask to Vincent Chaumat to confirm the present RF frequency.
| Ronic Chiche wrote: |
|
this morning, I plugged back the 4 power cables which supply all the equipments.
all the equipments have restarted without problem, but the shutter of the laser amplifier needs a ThomX safety system grant to let the beam propagate to the cavity...
and then continue the commissioning of the restart after Christmas shutdown.
|
|
|
312
|
Thu Jan 11 13:42:16 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | Restarting equipements after Christmas shutdown |
this morning, I set the Ring frequency at 500.25MHz and after switching ON the power supply of the amplifiers for the RF signal coming from the laser,
I saw the beating frequency (between ring RF and laser harmonic frerquencies at 500.25MHz) that I canceled, playing with the Smaract motor.
Now, I need to adjust the FP-cavity length to follow the laser cavity length but the IcePap DS (controlling the FP-cavity motor) freezes too often and it is almost impossible to make the adjustment.
Kevin sent a GLPI ticket to try to solve this problem.... waiting for IT answer...
| Ronic Chiche wrote: |
|
the laser coupled into the amplifier is still measured at ~3mW at the input, on the Alphanov software, once the Amp stage are started at 0% of power ratio.
this afternoon, Kevin restarted the 3rd safety system of ThomX to grant the opening of the beam shutter => now the shutter can be opened.
after adjusting the phase for the PDH system, I was able to find back the laser/CFP lock.
after some minimal alignement, one gets back 50kW for 30.5% of laser amplifier ratio, as before the Christmas shutdown.
I didn't try to lock the FP cavity on the 500MHz RF reference clock as I'm not sure if it has been set to 500.25MHz (as before shutdown) or at 500.1MHz (the future RF frequency) which is very far and will require to move the FP cavity on a long rage.
=> I will ask to Vincent Chaumat to confirm the present RF frequency.
| Ronic Chiche wrote: |
|
this morning, I plugged back the 4 power cables which supply all the equipments.
all the equipments have restarted without problem, but the shutter of the laser amplifier needs a ThomX safety system grant to let the beam propagate to the cavity...
and then continue the commissioning of the restart after Christmas shutdown.
|
|
|
|
313
|
Tue Jan 16 12:19:35 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | Restarting equipements after Christmas shutdown |
on last friday (12/01/2024), the IcePap DS issue was solved and I was able to control the FP-cavity motors properly.
I was able to quickly lock either the laser on the FP-cavity and the FP-cavity on the RF reference as before.
then, we can close these posts on restarting the equipments after the Christmas shutdown
| Ronic Chiche wrote: |
|
this morning, I set the Ring frequency at 500.25MHz and after switching ON the power supply of the amplifiers for the RF signal coming from the laser,
I saw the beating frequency (between ring RF and laser harmonic frerquencies at 500.25MHz) that I canceled, playing with the Smaract motor.
Now, I need to adjust the FP-cavity length to follow the laser cavity length but the IcePap DS (controlling the FP-cavity motor) freezes too often and it is almost impossible to make the adjustment.
Kevin sent a GLPI ticket to try to solve this problem.... waiting for IT answer...
| Ronic Chiche wrote: |
|
the laser coupled into the amplifier is still measured at ~3mW at the input, on the Alphanov software, once the Amp stage are started at 0% of power ratio.
this afternoon, Kevin restarted the 3rd safety system of ThomX to grant the opening of the beam shutter => now the shutter can be opened.
after adjusting the phase for the PDH system, I was able to find back the laser/CFP lock.
after some minimal alignement, one gets back 50kW for 30.5% of laser amplifier ratio, as before the Christmas shutdown.
I didn't try to lock the FP cavity on the 500MHz RF reference clock as I'm not sure if it has been set to 500.25MHz (as before shutdown) or at 500.1MHz (the future RF frequency) which is very far and will require to move the FP cavity on a long rage.
=> I will ask to Vincent Chaumat to confirm the present RF frequency.
| Ronic Chiche wrote: |
|
this morning, I plugged back the 4 power cables which supply all the equipments.
all the equipments have restarted without problem, but the shutter of the laser amplifier needs a ThomX safety system grant to let the beam propagate to the cavity...
and then continue the commissioning of the restart after Christmas shutdown.
|
|
|
|
|
348
|
Tue Aug 27 09:59:18 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | restarting CFP system after the long summer shutdown (RF section conditionning) and Alphanov amplifier issue |
temperature in the casemate ~ 19.5°C stable
this morning, I measured the power directly at the output of the fiber coupler (through the output fiber), after the strecher : ~5mW
then, I connected the EOM used for the PDH technic : ~2.5mW
then, I connected the fiber of the EOM to the Alphanov amplifier input fiber.
LAL Alphnov software :
PD_IN = 3.289mW
PD_PULSE = 33.372MHz
=> nothing to do ! :-)))
I switched ON the preamplifier (3rd stage @ 0%) => PD_Preamp2 = 152.44mW
|
| Attachment 1: Capture.JPG
|
 |
| Attachment 2: Capture2.JPG
|
 |
|
349
|
Tue Aug 27 10:19:08 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | restarting CFP system after the long summer shutdown (RF section conditionning) and Alphanov amplifier issue |
I refilled the chiller and the stabilized temperature is @ 25°C
we started the 3rd stage of the Alphanov amplifier at 30% => PD_OUT is fluctuating around 1-1.4W => it's not related to the real output power !
the photodiode connected to the scope CH2 is around 210mV @ 50ohms
we clearly see some small peak on the transmission signal (scope CH1) => we need to optimize the laser cavity length and the CEP
| Ronic Chiche wrote: |
|
temperature in the casemate ~ 19.5°C stable
this morning, I measured the power directly at the output of the fiber coupler (through the output fiber), after the strecher : ~5mW
then, I connected the EOM used for the PDH technic : ~2.5mW
then, I connected the fiber of the EOM to the Alphanov amplifier input fiber.
LAL Alphnov software :
PD_IN = 3.289mW
PD_PULSE = 33.372MHz
=> nothing to do ! :-)))
I switched ON the preamplifier (3rd stage @ 0%) => PD_Preamp2 = 152.44mW
|
|
| Attachment 1: Capture.JPG
|
 |
|
350
|
Tue Aug 27 11:33:24 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | restarting CFP system after the long summer shutdown (RF section conditionning) and Alphanov amplifier issue |
we had to move the laser cavity motor to find the resonances and change a lot the CEP.
the CEP motor is at -524.1µm
we got ~75kW in the FP cavity without alignment but with CEP optimization.
the coupling (the cyan color) is ~ 62% !
=> we need to do some alignment
| Ronic Chiche wrote: |
|
I refilled the chiller and the stabilized temperature is @ 25°C
we started the 3rd stage of the Alphanov amplifier at 30% => PD_OUT is fluctuating around 1-1.4W => it's not related to the real output power !
the photodiode connected to the scope CH2 is around 210mV @ 50ohms
we clearly see some small peak on the transmission signal (scope CH1) => we need to optimize the laser cavity length and the CEP
| Ronic Chiche wrote: |
|
temperature in the casemate ~ 19.5°C stable
this morning, I measured the power directly at the output of the fiber coupler (through the output fiber), after the strecher : ~5mW
then, I connected the EOM used for the PDH technic : ~2.5mW
then, I connected the fiber of the EOM to the Alphanov amplifier input fiber.
LAL Alphnov software :
PD_IN = 3.289mW
PD_PULSE = 33.372MHz
=> nothing to do ! :-)))
I switched ON the preamplifier (3rd stage @ 0%) => PD_Preamp2 = 152.44mW
|
|
|
| Attachment 1: Sans_titre.png
|
 |
|
351
|
Tue Aug 27 12:31:19 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | restarting CFP system after the long summer shutdown (RF section conditionning) and Alphanov amplifier issue |
after doing some alignment (walking procedure on the Y axis), and CEP optimization, I got 83kW for 33% on the laser amplifier with a coupling of ~60%
=> we have to check the input power after the amplifier.
I adjusted the FP and laser cavity motors to have a small 33MHz detuning with the RF reference (~10 Hz of beating @ 33MHz)
pb : I don't see the 500MHz beating signal => to be checked
during the motion of the FP-cavity motors, I observed a systematic delock when moving MOT.06 on P1z and no delock with MOT.03 on P4z when I move by 10 steps.
Kevin checked that both IcePap controller have the same configuration and then, the problem is maybe coming from the rust observed on the mechanic.
| Ronic Chiche wrote: |
|
we had to move the laser cavity motor to find the resonances and change a lot the CEP.
the CEP motor is at -524.1µm
we got ~75kW in the FP cavity without alignment but with CEP optimization.
the coupling (the cyan color) is ~ 62% !
=> we need to do some alignment
| Ronic Chiche wrote: |
|
I refilled the chiller and the stabilized temperature is @ 25°C
we started the 3rd stage of the Alphanov amplifier at 30% => PD_OUT is fluctuating around 1-1.4W => it's not related to the real output power !
the photodiode connected to the scope CH2 is around 210mV @ 50ohms
we clearly see some small peak on the transmission signal (scope CH1) => we need to optimize the laser cavity length and the CEP
| Ronic Chiche wrote: |
|
temperature in the casemate ~ 19.5°C stable
this morning, I measured the power directly at the output of the fiber coupler (through the output fiber), after the strecher : ~5mW
then, I connected the EOM used for the PDH technic : ~2.5mW
then, I connected the fiber of the EOM to the Alphanov amplifier input fiber.
LAL Alphnov software :
PD_IN = 3.289mW
PD_PULSE = 33.372MHz
=> nothing to do ! :-)))
I switched ON the preamplifier (3rd stage @ 0%) => PD_Preamp2 = 152.44mW
|
|
|
|
| Attachment 1: Capture.JPG
|
 |
|
352
|
Fri Aug 30 10:50:45 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | restarting CFP system after the long summer shutdown (RF section conditionning) and Alphanov amplifier issue |
this morning with Daniele, we are measuring the amplifier power just before the FP-cavity.
Casemate temperature @19.5°C
@0% : 205mW
@10%: 755mW
@20%: 7.44W
@30%: 14.3W
@33%:16.3W
@40% : 20.3W
@50% : 24W
@60% : 26W
=> there is a power issue !!!
the normal casemate temperature should be in between 21 and 22°C.
cf this post (https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/133) to get previous amplifier power.
=> it seems we need to redo the compressor CVBG alignment !
first, we will check the amplifier power before the compressor.
| Ronic Chiche wrote: |
|
after doing some alignment (walking procedure on the Y axis), and CEP optimization, I got 83kW for 33% on the laser amplifier with a coupling of ~60%
=> we have to check the input power after the amplifier.
I adjusted the FP and laser cavity motors to have a small 33MHz detuning with the RF reference (~10 Hz of beating @ 33MHz)
pb : I don't see the 500MHz beating signal => to be checked
during the motion of the FP-cavity motors, I observed a systematic delock when moving MOT.06 on P1z and no delock with MOT.03 on P4z when I move by 10 steps.
Kevin checked that both IcePap controller have the same configuration and then, the problem is maybe coming from the rust observed on the mechanic.
| Ronic Chiche wrote: |
|
we had to move the laser cavity motor to find the resonances and change a lot the CEP.
the CEP motor is at -524.1µm
we got ~75kW in the FP cavity without alignment but with CEP optimization.
the coupling (the cyan color) is ~ 62% !
=> we need to do some alignment
| Ronic Chiche wrote: |
|
I refilled the chiller and the stabilized temperature is @ 25°C
we started the 3rd stage of the Alphanov amplifier at 30% => PD_OUT is fluctuating around 1-1.4W => it's not related to the real output power !
the photodiode connected to the scope CH2 is around 210mV @ 50ohms
we clearly see some small peak on the transmission signal (scope CH1) => we need to optimize the laser cavity length and the CEP
| Ronic Chiche wrote: |
|
temperature in the casemate ~ 19.5°C stable
this morning, I measured the power directly at the output of the fiber coupler (through the output fiber), after the strecher : ~5mW
then, I connected the EOM used for the PDH technic : ~2.5mW
then, I connected the fiber of the EOM to the Alphanov amplifier input fiber.
LAL Alphnov software :
PD_IN = 3.289mW
PD_PULSE = 33.372MHz
=> nothing to do ! :-)))
I switched ON the preamplifier (3rd stage @ 0%) => PD_Preamp2 = 152.44mW
|
|
|
|
|
|
353
|
Fri Aug 30 12:05:38 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | restarting CFP system after the long summer shutdown (RF section conditionning) and Alphanov amplifier issue |
After the summer power shutdown, I restarted all the equipements and got quickly ~80kW for 33% amplifier ratio with CEP optimization and without Alignment optimization.
CEP motor position : -506µm
| Ronic Chiche wrote: |
|
this morning with Daniele, we are measuring the amplifier power just before the FP-cavity.
Casemate temperature @19.5°C
@0% : 205mW
@10%: 755mW
@20%: 7.44W
@30%: 14.3W
@33%:16.3W
@40% : 20.3W
@50% : 24W
@60% : 26W
=> there is a power issue !!!
the normal casemate temperature should be in between 21 and 22°C.
cf this post (https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/133) to get previous amplifier power.
=> it seems we need to redo the compressor CVBG alignment !
first, we will check the amplifier power before the compressor.
| Ronic Chiche wrote: |
|
after doing some alignment (walking procedure on the Y axis), and CEP optimization, I got 83kW for 33% on the laser amplifier with a coupling of ~60%
=> we have to check the input power after the amplifier.
I adjusted the FP and laser cavity motors to have a small 33MHz detuning with the RF reference (~10 Hz of beating @ 33MHz)
pb : I don't see the 500MHz beating signal => to be checked
during the motion of the FP-cavity motors, I observed a systematic delock when moving MOT.06 on P1z and no delock with MOT.03 on P4z when I move by 10 steps.
Kevin checked that both IcePap controller have the same configuration and then, the problem is maybe coming from the rust observed on the mechanic.
| Ronic Chiche wrote: |
|
we had to move the laser cavity motor to find the resonances and change a lot the CEP.
the CEP motor is at -524.1µm
we got ~75kW in the FP cavity without alignment but with CEP optimization.
the coupling (the cyan color) is ~ 62% !
=> we need to do some alignment
| Ronic Chiche wrote: |
|
I refilled the chiller and the stabilized temperature is @ 25°C
we started the 3rd stage of the Alphanov amplifier at 30% => PD_OUT is fluctuating around 1-1.4W => it's not related to the real output power !
the photodiode connected to the scope CH2 is around 210mV @ 50ohms
we clearly see some small peak on the transmission signal (scope CH1) => we need to optimize the laser cavity length and the CEP
| Ronic Chiche wrote: |
|
temperature in the casemate ~ 19.5°C stable
this morning, I measured the power directly at the output of the fiber coupler (through the output fiber), after the strecher : ~5mW
then, I connected the EOM used for the PDH technic : ~2.5mW
then, I connected the fiber of the EOM to the Alphanov amplifier input fiber.
LAL Alphnov software :
PD_IN = 3.289mW
PD_PULSE = 33.372MHz
=> nothing to do ! :-)))
I switched ON the preamplifier (3rd stage @ 0%) => PD_Preamp2 = 152.44mW
|
|
|
|
|
|
|
354
|
Fri Aug 30 17:59:01 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | restarting CFP system after the long summer shutdown (RF section conditionning) and Alphanov amplifier issue |
amplifier power measurement before the compressor CVBG.
we used a mirror in between the amplifier output and the compressor + a 2-mirror periscope to match the powermeter height.
amp ratio (%) power (W)
0 0.7
10 1.65
20 13
30 25
40 37.7
50 50.5
60 62
70 74
80 86
90 99
100 110
we had an alarm @100% of amplifier ratio => the reason is not clear... maybe it's coming from a discrepency between some internal measurement and the expected value.
we checked the power after the alarm => still OK
conclusion, the amplifier seems to be OK... we need to redo the alignment of the compressor CVBG.
| Ronic Chiche wrote: |
|
After the summer power shutdown, I restarted all the equipements and got quickly ~80kW for 33% amplifier ratio with CEP optimization and without Alignment optimization.
CEP motor position : -506µm
| Ronic Chiche wrote: |
|
this morning with Daniele, we are measuring the amplifier power just before the FP-cavity.
Casemate temperature @19.5°C
@0% : 205mW
@10%: 755mW
@20%: 7.44W
@30%: 14.3W
@33%:16.3W
@40% : 20.3W
@50% : 24W
@60% : 26W
=> there is a power issue !!!
the normal casemate temperature should be in between 21 and 22°C.
cf this post (https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/133) to get previous amplifier power.
=> it seems we need to redo the compressor CVBG alignment !
first, we will check the amplifier power before the compressor.
| Ronic Chiche wrote: |
|
after doing some alignment (walking procedure on the Y axis), and CEP optimization, I got 83kW for 33% on the laser amplifier with a coupling of ~60%
=> we have to check the input power after the amplifier.
I adjusted the FP and laser cavity motors to have a small 33MHz detuning with the RF reference (~10 Hz of beating @ 33MHz)
pb : I don't see the 500MHz beating signal => to be checked
during the motion of the FP-cavity motors, I observed a systematic delock when moving MOT.06 on P1z and no delock with MOT.03 on P4z when I move by 10 steps.
Kevin checked that both IcePap controller have the same configuration and then, the problem is maybe coming from the rust observed on the mechanic.
| Ronic Chiche wrote: |
|
we had to move the laser cavity motor to find the resonances and change a lot the CEP.
the CEP motor is at -524.1µm
we got ~75kW in the FP cavity without alignment but with CEP optimization.
the coupling (the cyan color) is ~ 62% !
=> we need to do some alignment
| Ronic Chiche wrote: |
|
I refilled the chiller and the stabilized temperature is @ 25°C
we started the 3rd stage of the Alphanov amplifier at 30% => PD_OUT is fluctuating around 1-1.4W => it's not related to the real output power !
the photodiode connected to the scope CH2 is around 210mV @ 50ohms
we clearly see some small peak on the transmission signal (scope CH1) => we need to optimize the laser cavity length and the CEP
| Ronic Chiche wrote: |
|
temperature in the casemate ~ 19.5°C stable
this morning, I measured the power directly at the output of the fiber coupler (through the output fiber), after the strecher : ~5mW
then, I connected the EOM used for the PDH technic : ~2.5mW
then, I connected the fiber of the EOM to the Alphanov amplifier input fiber.
LAL Alphnov software :
PD_IN = 3.289mW
PD_PULSE = 33.372MHz
=> nothing to do ! :-)))
I switched ON the preamplifier (3rd stage @ 0%) => PD_Preamp2 = 152.44mW
|
|
|
|
|
|
|
|
358
|
Fri Sep 6 12:10:51 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | restarting CFP system after the long summer shutdown (RF section conditionning) and Alphanov amplifier issue |
after redoing the alignment of the CVBG, we got back the initial power in the FPC (87kW)
now, I checked the RF locking signals on the 2nd oscilloscope.
I get a strange ratio between the 33MHz beating and the 500MHz beating signals.
beat @ 33MHz ~ 40kHz !
beat @ 500MHz ~ 700Hz !
maybe the 33MHz generator frequency is not matching the 500MHz RF frequency ?
| Ronic Chiche wrote: |
|
amplifier power measurement before the compressor CVBG.
we used a mirror in between the amplifier output and the compressor + a 2-mirror periscope to match the powermeter height.
amp ratio (%) power (W)
0 0.7
10 1.65
20 13
30 25
40 37.7
50 50.5
60 62
70 74
80 86
90 99
100 110
we had an alarm @100% of amplifier ratio => the reason is not clear... maybe it's coming from a discrepency between some internal measurement and the expected value.
we checked the power after the alarm => still OK
conclusion, the amplifier seems to be OK... we need to redo the alignment of the compressor CVBG.
| Ronic Chiche wrote: |
|
After the summer power shutdown, I restarted all the equipements and got quickly ~80kW for 33% amplifier ratio with CEP optimization and without Alignment optimization.
CEP motor position : -506µm
| Ronic Chiche wrote: |
|
this morning with Daniele, we are measuring the amplifier power just before the FP-cavity.
Casemate temperature @19.5°C
@0% : 205mW
@10%: 755mW
@20%: 7.44W
@30%: 14.3W
@33%:16.3W
@40% : 20.3W
@50% : 24W
@60% : 26W
=> there is a power issue !!!
the normal casemate temperature should be in between 21 and 22°C.
cf this post (https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/133) to get previous amplifier power.
=> it seems we need to redo the compressor CVBG alignment !
first, we will check the amplifier power before the compressor.
| Ronic Chiche wrote: |
|
after doing some alignment (walking procedure on the Y axis), and CEP optimization, I got 83kW for 33% on the laser amplifier with a coupling of ~60%
=> we have to check the input power after the amplifier.
I adjusted the FP and laser cavity motors to have a small 33MHz detuning with the RF reference (~10 Hz of beating @ 33MHz)
pb : I don't see the 500MHz beating signal => to be checked
during the motion of the FP-cavity motors, I observed a systematic delock when moving MOT.06 on P1z and no delock with MOT.03 on P4z when I move by 10 steps.
Kevin checked that both IcePap controller have the same configuration and then, the problem is maybe coming from the rust observed on the mechanic.
| Ronic Chiche wrote: |
|
we had to move the laser cavity motor to find the resonances and change a lot the CEP.
the CEP motor is at -524.1µm
we got ~75kW in the FP cavity without alignment but with CEP optimization.
the coupling (the cyan color) is ~ 62% !
=> we need to do some alignment
| Ronic Chiche wrote: |
|
I refilled the chiller and the stabilized temperature is @ 25°C
we started the 3rd stage of the Alphanov amplifier at 30% => PD_OUT is fluctuating around 1-1.4W => it's not related to the real output power !
the photodiode connected to the scope CH2 is around 210mV @ 50ohms
we clearly see some small peak on the transmission signal (scope CH1) => we need to optimize the laser cavity length and the CEP
| Ronic Chiche wrote: |
|
temperature in the casemate ~ 19.5°C stable
this morning, I measured the power directly at the output of the fiber coupler (through the output fiber), after the strecher : ~5mW
then, I connected the EOM used for the PDH technic : ~2.5mW
then, I connected the fiber of the EOM to the Alphanov amplifier input fiber.
LAL Alphnov software :
PD_IN = 3.289mW
PD_PULSE = 33.372MHz
=> nothing to do ! :-)))
I switched ON the preamplifier (3rd stage @ 0%) => PD_Preamp2 = 152.44mW
|
|
|
|
|
|
|
|
|
359
|
Fri Sep 6 15:20:20 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | restarting CFP system after the long summer shutdown (RF section conditionning) and Alphanov amplifier issue |
now, the Ring RF frequency is set to 500.067MHz.
so the CFP/laser frequency should be 500.067MHz/15 = 33.3378MHz.
the frequency was set to 33.378MHz !!! => it explains the frequency shift only on the 33MHz beating => I corrected the frequency on the generator and informed Nicolas Delerue.
=> now, I can try to lock the CFP to the RF frequency.
| Ronic Chiche wrote: |
|
after redoing the alignment of the CVBG, we got back the initial power in the FPC (87kW)
now, I checked the RF locking signals on the 2nd oscilloscope.
I get a strange ratio between the 33MHz beating and the 500MHz beating signals.
beat @ 33MHz ~ 40kHz !
beat @ 500MHz ~ 700Hz !
maybe the 33MHz generator frequency is not matching the 500MHz RF frequency ?
| Ronic Chiche wrote: |
|
amplifier power measurement before the compressor CVBG.
we used a mirror in between the amplifier output and the compressor + a 2-mirror periscope to match the powermeter height.
amp ratio (%) power (W)
0 0.7
10 1.65
20 13
30 25
40 37.7
50 50.5
60 62
70 74
80 86
90 99
100 110
we had an alarm @100% of amplifier ratio => the reason is not clear... maybe it's coming from a discrepency between some internal measurement and the expected value.
we checked the power after the alarm => still OK
conclusion, the amplifier seems to be OK... we need to redo the alignment of the compressor CVBG.
| Ronic Chiche wrote: |
|
After the summer power shutdown, I restarted all the equipements and got quickly ~80kW for 33% amplifier ratio with CEP optimization and without Alignment optimization.
CEP motor position : -506µm
| Ronic Chiche wrote: |
|
this morning with Daniele, we are measuring the amplifier power just before the FP-cavity.
Casemate temperature @19.5°C
@0% : 205mW
@10%: 755mW
@20%: 7.44W
@30%: 14.3W
@33%:16.3W
@40% : 20.3W
@50% : 24W
@60% : 26W
=> there is a power issue !!!
the normal casemate temperature should be in between 21 and 22°C.
cf this post (https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/133) to get previous amplifier power.
=> it seems we need to redo the compressor CVBG alignment !
first, we will check the amplifier power before the compressor.
| Ronic Chiche wrote: |
|
after doing some alignment (walking procedure on the Y axis), and CEP optimization, I got 83kW for 33% on the laser amplifier with a coupling of ~60%
=> we have to check the input power after the amplifier.
I adjusted the FP and laser cavity motors to have a small 33MHz detuning with the RF reference (~10 Hz of beating @ 33MHz)
pb : I don't see the 500MHz beating signal => to be checked
during the motion of the FP-cavity motors, I observed a systematic delock when moving MOT.06 on P1z and no delock with MOT.03 on P4z when I move by 10 steps.
Kevin checked that both IcePap controller have the same configuration and then, the problem is maybe coming from the rust observed on the mechanic.
| Ronic Chiche wrote: |
|
we had to move the laser cavity motor to find the resonances and change a lot the CEP.
the CEP motor is at -524.1µm
we got ~75kW in the FP cavity without alignment but with CEP optimization.
the coupling (the cyan color) is ~ 62% !
=> we need to do some alignment
| Ronic Chiche wrote: |
|
I refilled the chiller and the stabilized temperature is @ 25°C
we started the 3rd stage of the Alphanov amplifier at 30% => PD_OUT is fluctuating around 1-1.4W => it's not related to the real output power !
the photodiode connected to the scope CH2 is around 210mV @ 50ohms
we clearly see some small peak on the transmission signal (scope CH1) => we need to optimize the laser cavity length and the CEP
| Ronic Chiche wrote: |
|
temperature in the casemate ~ 19.5°C stable
this morning, I measured the power directly at the output of the fiber coupler (through the output fiber), after the strecher : ~5mW
then, I connected the EOM used for the PDH technic : ~2.5mW
then, I connected the fiber of the EOM to the Alphanov amplifier input fiber.
LAL Alphnov software :
PD_IN = 3.289mW
PD_PULSE = 33.372MHz
=> nothing to do ! :-)))
I switched ON the preamplifier (3rd stage @ 0%) => PD_Preamp2 = 152.44mW
|
|
|
|
|
|
|
|
|
|
360
|
Fri Sep 6 17:21:51 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | restarting CFP system after the long summer shutdown (RF section conditionning) and Alphanov amplifier issue |
I changed a little bit the locking parameters (see picture) for the RF/FPC loop and it locked rapidely.
I asked Vincent to connect on the same scope the synchro signals to check if the FPC lock was OK => he connected 33MHz signals coming from the laser and from the 33MHz RF generator.
=> the signals are not locked even when the FPC seems to be locked to the RF.
=> it can be normal because he forgot to connect also the synchro trigger signal which gives the moment of synchronization of the machine... to be finished on Monday.
| Ronic Chiche wrote: |
|
now, the Ring RF frequency is set to 500.067MHz.
so the CFP/laser frequency should be 500.067MHz/15 = 33.3378MHz.
the frequency was set to 33.378MHz !!! => it explains the frequency shift only on the 33MHz beating => I corrected the frequency on the generator and informed Nicolas Delerue.
=> now, I can try to lock the CFP to the RF frequency.
| Ronic Chiche wrote: |
|
after redoing the alignment of the CVBG, we got back the initial power in the FPC (87kW)
now, I checked the RF locking signals on the 2nd oscilloscope.
I get a strange ratio between the 33MHz beating and the 500MHz beating signals.
beat @ 33MHz ~ 40kHz !
beat @ 500MHz ~ 700Hz !
maybe the 33MHz generator frequency is not matching the 500MHz RF frequency ?
| Ronic Chiche wrote: |
|
amplifier power measurement before the compressor CVBG.
we used a mirror in between the amplifier output and the compressor + a 2-mirror periscope to match the powermeter height.
amp ratio (%) power (W)
0 0.7
10 1.65
20 13
30 25
40 37.7
50 50.5
60 62
70 74
80 86
90 99
100 110
we had an alarm @100% of amplifier ratio => the reason is not clear... maybe it's coming from a discrepency between some internal measurement and the expected value.
we checked the power after the alarm => still OK
conclusion, the amplifier seems to be OK... we need to redo the alignment of the compressor CVBG.
| Ronic Chiche wrote: |
|
After the summer power shutdown, I restarted all the equipements and got quickly ~80kW for 33% amplifier ratio with CEP optimization and without Alignment optimization.
CEP motor position : -506µm
| Ronic Chiche wrote: |
|
this morning with Daniele, we are measuring the amplifier power just before the FP-cavity.
Casemate temperature @19.5°C
@0% : 205mW
@10%: 755mW
@20%: 7.44W
@30%: 14.3W
@33%:16.3W
@40% : 20.3W
@50% : 24W
@60% : 26W
=> there is a power issue !!!
the normal casemate temperature should be in between 21 and 22°C.
cf this post (https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/133) to get previous amplifier power.
=> it seems we need to redo the compressor CVBG alignment !
first, we will check the amplifier power before the compressor.
| Ronic Chiche wrote: |
|
after doing some alignment (walking procedure on the Y axis), and CEP optimization, I got 83kW for 33% on the laser amplifier with a coupling of ~60%
=> we have to check the input power after the amplifier.
I adjusted the FP and laser cavity motors to have a small 33MHz detuning with the RF reference (~10 Hz of beating @ 33MHz)
pb : I don't see the 500MHz beating signal => to be checked
during the motion of the FP-cavity motors, I observed a systematic delock when moving MOT.06 on P1z and no delock with MOT.03 on P4z when I move by 10 steps.
Kevin checked that both IcePap controller have the same configuration and then, the problem is maybe coming from the rust observed on the mechanic.
| Ronic Chiche wrote: |
|
we had to move the laser cavity motor to find the resonances and change a lot the CEP.
the CEP motor is at -524.1µm
we got ~75kW in the FP cavity without alignment but with CEP optimization.
the coupling (the cyan color) is ~ 62% !
=> we need to do some alignment
| Ronic Chiche wrote: |
|
I refilled the chiller and the stabilized temperature is @ 25°C
we started the 3rd stage of the Alphanov amplifier at 30% => PD_OUT is fluctuating around 1-1.4W => it's not related to the real output power !
the photodiode connected to the scope CH2 is around 210mV @ 50ohms
we clearly see some small peak on the transmission signal (scope CH1) => we need to optimize the laser cavity length and the CEP
| Ronic Chiche wrote: |
|
temperature in the casemate ~ 19.5°C stable
this morning, I measured the power directly at the output of the fiber coupler (through the output fiber), after the strecher : ~5mW
then, I connected the EOM used for the PDH technic : ~2.5mW
then, I connected the fiber of the EOM to the Alphanov amplifier input fiber.
LAL Alphnov software :
PD_IN = 3.289mW
PD_PULSE = 33.372MHz
=> nothing to do ! :-)))
I switched ON the preamplifier (3rd stage @ 0%) => PD_Preamp2 = 152.44mW
|
|
|
|
|
|
|
|
|
|
| Attachment 1: 20240906_161926.jpg
|
 |
|
362
|
Mon Sep 9 11:00:43 2024 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | restarting CFP system after the long summer shutdown (RF section conditionning) and Alphanov amplifier issue |
this morning, with Vincent :
- we added the 10Hz trigger to the CH2 on the remote scope (192.168.229.21) to check the synchronization at the right timing.
CH1 : 33MHz RING
CH2 : LINAC/RING synchro signal
CH3 : 500MHz RING
CH4 : 33MHz FPC
we successfully lock both laser on FPC (87kW for 33% amp ratio) and FPC on 500MHz RF.
we saw the result on the scope.
suprisingly, the FPC/RF lock seems much robust than before.
the only problem is the MOT.06 which make the lock being lost at almost every move.
one reason could be the rust on the mechanics... we can try to change the position of both plan mirror motors to work in a proper region.
before, it was working well at -900 000 steps on MOT.06, now it is -780 00.
it's a quite long travel...
end of this posts thread
| Ronic Chiche wrote: |
|
I changed a little bit the locking parameters (see picture) for the RF/FPC loop and it locked rapidely.
I asked Vincent to connect on the same scope the synchro signals to check if the FPC lock was OK => he connected 33MHz signals coming from the laser and from the 33MHz RF generator.
=> the signals are not locked even when the FPC seems to be locked to the RF.
=> it can be normal because he forgot to connect also the synchro trigger signal which gives the moment of synchronization of the machine... to be finished on Monday.
| Ronic Chiche wrote: |
|
now, the Ring RF frequency is set to 500.067MHz.
so the CFP/laser frequency should be 500.067MHz/15 = 33.3378MHz.
the frequency was set to 33.378MHz !!! => it explains the frequency shift only on the 33MHz beating => I corrected the frequency on the generator and informed Nicolas Delerue.
=> now, I can try to lock the CFP to the RF frequency.
| Ronic Chiche wrote: |
|
after redoing the alignment of the CVBG, we got back the initial power in the FPC (87kW)
now, I checked the RF locking signals on the 2nd oscilloscope.
I get a strange ratio between the 33MHz beating and the 500MHz beating signals.
beat @ 33MHz ~ 40kHz !
beat @ 500MHz ~ 700Hz !
maybe the 33MHz generator frequency is not matching the 500MHz RF frequency ?
| Ronic Chiche wrote: |
|
amplifier power measurement before the compressor CVBG.
we used a mirror in between the amplifier output and the compressor + a 2-mirror periscope to match the powermeter height.
amp ratio (%) power (W)
0 0.7
10 1.65
20 13
30 25
40 37.7
50 50.5
60 62
70 74
80 86
90 99
100 110
we had an alarm @100% of amplifier ratio => the reason is not clear... maybe it's coming from a discrepency between some internal measurement and the expected value.
we checked the power after the alarm => still OK
conclusion, the amplifier seems to be OK... we need to redo the alignment of the compressor CVBG.
| Ronic Chiche wrote: |
|
After the summer power shutdown, I restarted all the equipements and got quickly ~80kW for 33% amplifier ratio with CEP optimization and without Alignment optimization.
CEP motor position : -506µm
| Ronic Chiche wrote: |
|
this morning with Daniele, we are measuring the amplifier power just before the FP-cavity.
Casemate temperature @19.5°C
@0% : 205mW
@10%: 755mW
@20%: 7.44W
@30%: 14.3W
@33%:16.3W
@40% : 20.3W
@50% : 24W
@60% : 26W
=> there is a power issue !!!
the normal casemate temperature should be in between 21 and 22°C.
cf this post (https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/133) to get previous amplifier power.
=> it seems we need to redo the compressor CVBG alignment !
first, we will check the amplifier power before the compressor.
| Ronic Chiche wrote: |
|
after doing some alignment (walking procedure on the Y axis), and CEP optimization, I got 83kW for 33% on the laser amplifier with a coupling of ~60%
=> we have to check the input power after the amplifier.
I adjusted the FP and laser cavity motors to have a small 33MHz detuning with the RF reference (~10 Hz of beating @ 33MHz)
pb : I don't see the 500MHz beating signal => to be checked
during the motion of the FP-cavity motors, I observed a systematic delock when moving MOT.06 on P1z and no delock with MOT.03 on P4z when I move by 10 steps.
Kevin checked that both IcePap controller have the same configuration and then, the problem is maybe coming from the rust observed on the mechanic.
| Ronic Chiche wrote: |
|
we had to move the laser cavity motor to find the resonances and change a lot the CEP.
the CEP motor is at -524.1µm
we got ~75kW in the FP cavity without alignment but with CEP optimization.
the coupling (the cyan color) is ~ 62% !
=> we need to do some alignment
| Ronic Chiche wrote: |
|
I refilled the chiller and the stabilized temperature is @ 25°C
we started the 3rd stage of the Alphanov amplifier at 30% => PD_OUT is fluctuating around 1-1.4W => it's not related to the real output power !
the photodiode connected to the scope CH2 is around 210mV @ 50ohms
we clearly see some small peak on the transmission signal (scope CH1) => we need to optimize the laser cavity length and the CEP
| Ronic Chiche wrote: |
|
temperature in the casemate ~ 19.5°C stable
this morning, I measured the power directly at the output of the fiber coupler (through the output fiber), after the strecher : ~5mW
then, I connected the EOM used for the PDH technic : ~2.5mW
then, I connected the fiber of the EOM to the Alphanov amplifier input fiber.
LAL Alphnov software :
PD_IN = 3.289mW
PD_PULSE = 33.372MHz
=> nothing to do ! :-)))
I switched ON the preamplifier (3rd stage @ 0%) => PD_Preamp2 = 152.44mW
|
|
|
|
|
|
|
|
|
|
|
|
426
|
Wed Jun 11 11:35:22 2025 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | software | ThomX igloo | 33MHz phase adjustment |
the 33MHz beating signal (phase) is used to start and stop automatically the lock on the 500MHz beating signal.
this 33MHz beating phase has a fixed range (typically +/-0.5V), so it is important to center this beating phase in the middle of its own range when the 500MHz signal is locked
=> tune the 33MHz phase in order to get ~ 0V on 33MHz beating signal when the 500MHz locking is ON.
this can be done by using the python script "Write_Phase_Rigol_33MHz located in the path /tmp_mnt/data/shared/commissioning_scripts/common
|
|
212
|
Tue May 30 18:32:37 2023 |
Manar Amer | Fixed | report | lasers and optics | detectors and electronics | cabling | software | ThomX igloo | Cavity Lock - Alphanov Amplifier |
This morning, while trying Guillaume solution of disconnecting the 24 V necessary to the second and third stage and raise the threshold of Temp preamp 1 (to only operate the ampli with first stage only)
Ronic disconnected the 24 volt connection, turned on amplifier LAL software but no MMD3 error or a big temp on Temp preamp 1 , software worked fine with no issue.
we connected the 24 V again and the software worked with no issues.
Still not sure what happened for it to work correctly !!!!!!!!!!
- We used the amplifier at 0% all day with no errors (to test , I turned it for 10 minutes @ 10% with no error too)
- We changed the lase lock box with ThomX one (having Ethernet connection), it was connected to computer and works ok
- Ronic installed a signal amplifier for the PDH error signal. (yet to be fully optimized)
- we see a better coupling than before in the cavity.
- The dominant mode is 00 mode, but we see many higher order modes, need to alight better.
| Manar Amer wrote: |
|
Today with Ronic and Daniele we attempted to lock the cavity, but the alphanov amplifier did not turn on because of an error in MMD3 (related to the third stage)
the new error is having the MMD3 error show on the alarm window of the LAL software, and having the Alarms on the line 3 of the alphanov software red for both EXT/CPU and Laser T Max (never been red before)
this is after a 2-week work stop, but the last time when Daniele changed the fiber connector it worked for 3 hours with no errors.
Ronic, is in contact with Guillaume from Alphanov to fix the issue !!!
Work scheduled all Wednesday, in hopes the issue is fixed by then.
| Manar Amer wrote: |
|
Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)
was done at the end due to excitation of higher order modes.
Wanted to improve the fundamental first, then increase the power.
| Manar Amer wrote: |
- On Monday with Daniele,
- we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
- On Tuesday with Daniele and Kevin
- I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
- removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm , Reflection : 50 ohm )
- We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
- Images:
- 1st : showing the size of the beam just before injecting into the cavity @ 10%
- 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
- 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
- 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors
Note : the voltage ranges are not the same between the images
| Manar Amer wrote: |
|
Yesterday,
- we installed the second high power reflecting mirrors at the output of the amplifier.
- We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of ~ 4.5 mm.
- Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
- Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
- Changed the FSR of the cavity to match it.
- Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same
- We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
- a reason could be the CEP of the oscillator ,
- bad matching of the beam waist position,
- noise from the amplifier (as we were operating it at very low power)
- additional alignment needed,
|
|
|
|
|
| Attachment 1: tek00000.png
|
 |
|
215
|
Fri Jun 2 10:54:25 2023 |
Manar Amer | Fixed | report | lasers and optics | detectors and electronics | cabling | software | ThomX igloo | Cavity Lock - Alphanov Amplifier |
Yesterday, with Ronic we locked ThomX cavity on the second resonance
Laser : 33.36 MHz , Amplifier : 0 % (worked with no errors all day)
adding image for the lock on the second resonance (locked), 00 Mode image, fit, Beam measurements
and image of the first resonance signal (not locked)
Info : when locking on the secondary resonance, we aligned better and adjusted on the oscillator CEP and increased the transmission signal
Controllers connected to computer
through Ethernet :
- Cavity Mirror motors
- Lase lock
- oscillator CEP and Frequency motors (easier to follow the drift with this controller, and we can see it moving)
by USB :
| Manar Amer wrote: |
|
This morning, while trying Guillaume solution of disconnecting the 24 V necessary to the second and third stage and raise the threshold of Temp preamp 1 (to only operate the ampli with first stage only)
Ronic disconnected the 24 volt connection, turned on amplifier LAL software but no MMD3 error or a big temp on Temp preamp 1 , software worked fine with no issue.
we connected the 24 V again and the software worked with no issues.
Still not sure what happened for it to work correctly !!!!!!!!!!
- We used the amplifier at 0% all day with no errors (to test , I turned it for 10 minutes @ 10% with no error too)
- We changed the lase lock box with ThomX one (having Ethernet connection), it was connected to computer and works ok
- Ronic installed a signal amplifier for the PDH error signal. (yet to be fully optimized)
- we see a better coupling than before in the cavity.
- The dominant mode is 00 mode, but we see many higher order modes, need to alight better.
| Manar Amer wrote: |
|
Today with Ronic and Daniele we attempted to lock the cavity, but the alphanov amplifier did not turn on because of an error in MMD3 (related to the third stage)
the new error is having the MMD3 error show on the alarm window of the LAL software, and having the Alarms on the line 3 of the alphanov software red for both EXT/CPU and Laser T Max (never been red before)
this is after a 2-week work stop, but the last time when Daniele changed the fiber connector it worked for 3 hours with no errors.
Ronic, is in contact with Guillaume from Alphanov to fix the issue !!!
Work scheduled all Wednesday, in hopes the issue is fixed by then.
| Manar Amer wrote: |
|
Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)
was done at the end due to excitation of higher order modes.
Wanted to improve the fundamental first, then increase the power.
| Manar Amer wrote: |
- On Monday with Daniele,
- we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
- On Tuesday with Daniele and Kevin
- I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
- removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm , Reflection : 50 ohm )
- We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
- Images:
- 1st : showing the size of the beam just before injecting into the cavity @ 10%
- 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
- 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
- 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors
Note : the voltage ranges are not the same between the images
| Manar Amer wrote: |
|
Yesterday,
- we installed the second high power reflecting mirrors at the output of the amplifier.
- We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of ~ 4.5 mm.
- Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
- Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
- Changed the FSR of the cavity to match it.
- Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same
- We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
- a reason could be the CEP of the oscillator ,
- bad matching of the beam waist position,
- noise from the amplifier (as we were operating it at very low power)
- additional alignment needed,
|
|
|
|
|
|
| Attachment 1: tek00003_secondary_resonance_fundamental_mode_lock.png
|
 |
| Attachment 2: tek0002CH1.isf
|
| Attachment 3: tek0002CH2.isf
|
| Attachment 4: tek0002CH3.isf
|
| Attachment 5: tek0002CH4.isf
|
| Attachment 6: 20230602_00mode_second_resonance.jpg
|
 |
| Attachment 7: 20230602_00mode_second_resonance_fit.jpg
|
 |
| Attachment 8: 20230602_00mode_second_resonance_measurements.jpg
|
 |
| Attachment 9: tek00002_first_resonance_fundamental_mode_signal.png
|
 |
| Attachment 10: tek0001CH1.isf
|
| Attachment 11: tek0001CH2.isf
|
| Attachment 12: tek0001CH3.isf
|
| Attachment 13: tek0001CH4.isf
|