  441   Thu Jul 10 18:02:54 2025 Ronic ChicheFixedinfodetectors and electronicsThomX igloolaser PZT amplifier

this afternoon, I added the Gain = 2.8 amplifier at the output of the Laselock to drive the laser PZT.
the power supply of the board is disymetric to address the correct maximal dynamic range of the amplifier (~ 30V) and voltage drop (~ 2V) related to the power supply voltage.
so, the power supply is set to -2V / 32V which gives roughly 0-28V of dynamic range at the ouput for 0-10V at the input.

despite the additionnal noise added by the amplifier, we are able to lock easily, and reach ~ 93kW in the FPC for 33% amplifier ratio.
obviously, as the dynamic range of this PZT has been multiplied by ~3, it is much more comfortable to operate the motors.
but the overall stability seems a bit degraded... to be checked, as finding the right PID + fast loop gain is not easy.

=> to be done tomorow.

  449   Wed Aug 27 12:01:41 2025 Ronic ChicheFixedinfodetectors and electronicsThomX iglooPower shut down on August 29th

This morning, all the equipments have been switched off due to a power shut down tomorrow morning.
Only the Onefive oscillator is still ON, connected to the Uninterruptable Power Supply (UPS).

  450   Thu Aug 28 18:01:46 2025 Ronic ChicheFixedinfodetectors and electronicsThomX iglooPower shut down on August 29th

after alignment, CEP adjustment and optimizing the PDH phase, I got back ~92kW in the FPC for 33% amplifier ratio.

Ronic Chiche wrote:

This morning, all the equipments have been switched off due to a power shut down tomorrow morning.
Only the Onefive oscillator is still ON, connected to the Uninterruptable Power Supply (UPS).

 

  66   Thu Oct 3 14:00:13 2019 Loïc AmoudryFixedissuecablingOptical roomElectrical trouble (disjunction)

The circuit breaker on ThomX cavity drops @1.5-1.8kW when we turn on 1/2nd stage of the amplifier or 3rd stage of the amplifier.
Amplifiers are now connected to K-BOX UPS.

We are buying a clamp ammeter for further investigations.