| ID |
Date |
Author |
Status |
Type |
Category |
Location |
Title |
|
242
|
Fri Sep 8 18:22:21 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | detectors and electronics | ThomX igloo | Removing high order modes |
here is the picture of the L-shape arm and the inside of the vessel.
this image corresponds to the "zero" position on each positionner.
| Ronic Chiche wrote: |
|
this morning we locked the laser and the FP-cavity with ~ 40kW => some HOM appeared "randomly" depending on the power.
we tried to play manually on the L-shape arm :
- first, we moved horizontally to put the L-shape in the beam axis
at some point, we saw the power level divided by ~2, we stopped and came back to restore the full power.
- then, we tried to move vertically.
whatever the direction, we were not able to see a clear cut of the beam.
BUT "strangely", at some point, the PZT drift followed the direction of the motion and was not really depending on the "cooling" or "heating" process when the lock stops or restart.
"strangely" again, at the endpoint, the power dropped by ~20-30% after a delock but it was impossible to restore the power when we put the arm at the initial position.
=> we had to adjust the FP-cavity alignment to restore the power !!!
=> we had the feeling that moving the arm could have misalign the cavity axis !!! :-(
=> we have to discuss with Yann and get the CAD files of the cavity to have a better understanding...
|
|
| Attachment 1: Capture.PNG
|
 |
|
245
|
Tue Sep 12 14:28:36 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | detectors and electronics | ThomX igloo | Removing high order modes |
yesterday with Daniele, we tried again to move the arm horizontally and vertically from the "zero position" to have a better understanding of the beahvior,
but it is roughly the same :
- a clear beam power reduction (=> cut of the beam) when one moves the horizontal axis
- a very small beam power reduction when one moves the vertical axis
as there are no end-position limits on the axis and as it is possible to "touch" the motors or different mecanichal pieces, we prefered to be conservative:
after several tries, one placed the arm at a position where the HOMs seem to be pretty well suppressed.
for the moment, we will consider this arm in a proper position.
| Ronic Chiche wrote: |
|
here is the picture of the L-shape arm and the inside of the vessel.
this image corresponds to the "zero" position on each positionner.
| Ronic Chiche wrote: |
|
this morning we locked the laser and the FP-cavity with ~ 40kW => some HOM appeared "randomly" depending on the power.
we tried to play manually on the L-shape arm :
- first, we moved horizontally to put the L-shape in the beam axis
at some point, we saw the power level divided by ~2, we stopped and came back to restore the full power.
- then, we tried to move vertically.
whatever the direction, we were not able to see a clear cut of the beam.
BUT "strangely", at some point, the PZT drift followed the direction of the motion and was not really depending on the "cooling" or "heating" process when the lock stops or restart.
"strangely" again, at the endpoint, the power dropped by ~20-30% after a delock but it was impossible to restore the power when we put the arm at the initial position.
=> we had to adjust the FP-cavity alignment to restore the power !!!
=> we had the feeling that moving the arm could have misalign the cavity axis !!! :-(
=> we have to discuss with Yann and get the CAD files of the cavity to have a better understanding...
|
|
|
|
8
|
Mon Oct 22 11:27:35 2018 |
Ronic Chiche | Fixed | info | mechanics | Optical room | Remontage mécanique |
Mail de Yann du 12/10/2018 :
Je vous fais le bilan des actions effectuées :
1/ La mécanique a été remontée avec changement des billes céramiques, soufflet défectueux, ect...
2/ Test et mise en place des vérins en position
3/ nettoyage des chambres et insertions de la méca
4/ connectorisation des câbles perlés et test des vérins par le soft
5/ fermeture des brides
6/ nettoyage du marbre et de la zone
On est donc prêt pour les test de vide et d'étuvage.
RAPPEL pour tous :
Des nouvelles combinaisons propres sont à disposition (on ne sort pas du flux avec la combi !)
Mettre MASQUE / CHARLOTTE / COMBI / GANTS !
Ne pas rentrer de CARTONS ou autres sources de pollution.
NETTOYER les outils ou objets à rentrer dans la zone
RANGER après les manips. |
|
9
|
Mon Oct 22 11:29:41 2018 |
Ronic Chiche | Fixed | info | mechanics | Optical room | Remontage des vérins |
Mail de Yann du 8/10/2018 :
Bonjour,
Tous les vérins sont maintenant OK ! Merci Ronic.
La cavité a été réglé en mode nominal (tilt à +2/-2 et Z à +5/-15 ou -5/+15).
Demain avec Christopher on remet tous dans les enceintes à vide.
Du coup, Patrick tu pourras ré-connecter les câbles céramique (à partir de demain apres midi).
On est dans les temps pour des tests de vide et d'étuvage a partir de lundi prochain.
A+
Yann. |
|
431
|
Thu Jun 26 12:37:22 2025 |
Ronic Chiche | Fixed | info | lasers and optics | ThomX igloo | Recipies for different FPC power |
For ~ 46kW power in the FP-cavity:
- Alplhanov amplifier ratio : 20%
- Axis 18 position : -0004710 steps
- PID A (for laser/FPC lock) :
input : input a
P = 0.06
I = 0.0007
D = 0.85
Sign : positive
Sampling : fast
filter : off
criterion : input d
upper th. : 10V
lower th. : 0.04V
speed : 100V/s
relock mode : none
off mode : hold
range : 50%
offset : 5V
- PID B (for RF/FPC lock) :
input : input b
P = 3
I = 0.00002
D = 2
Sign : negative
Sampling : mid
filter : off
criterion : input c
upper th. : 0.45V
lower th. : -0.45V
speed : 1V/s
relock mode : none
off mode : hold
range : 20%
offset : 5V
|
|
432
|
Tue Jul 1 12:13:18 2025 |
Ronic Chiche | Fixed | info | lasers and optics | ThomX igloo | Recipies for different FPC power |
For ~ 23kW power in the FP-cavity:
- Alplhanov amplifier ratio : 15%
- Axis 18 position : -0016014 steps
- PID A (for laser/FPC lock) :
input : input a
P = 0.1
I = 0.0016
D = 1.4
Sign : positive
Sampling : fast
filter : off
criterion : input d
upper th. : 10V
lower th. : 0.03V
speed : 100V/s
relock mode : none
off mode : hold
range : 50%
offset : 5V
- PID B (for RF/FPC lock) :
input : input b
P = 5
I = 0.0001
D = 1
Sign : negative
Sampling : mid
filter : off
criterion : input c
upper th. : 0.45V
lower th. : -0.45V
speed : 1V/s
relock mode : none
off mode : hold
range : 20%
offset : 5V
| Ronic Chiche wrote: |
|
For ~ 46kW power in the FP-cavity:
- Alplhanov amplifier ratio : 20%
- Axis 18 position : -0004710 steps
- PID A (for laser/FPC lock) :
input : input a
P = 0.06
I = 0.0007
D = 0.85
Sign : positive
Sampling : fast
filter : off
criterion : input d
upper th. : 10V
lower th. : 0.04V
speed : 100V/s
relock mode : none
off mode : hold
range : 50%
offset : 5V
- PID B (for RF/FPC lock) :
input : input b
P = 3
I = 0.00002
D = 2
Sign : negative
Sampling : mid
filter : off
criterion : input c
upper th. : 0.45V
lower th. : -0.45V
speed : 1V/s
relock mode : none
off mode : hold
range : 20%
offset : 5V
|
|
|
436
|
Wed Jul 9 14:12:54 2025 |
Ronic Chiche | Fixed | info | lasers and optics | ThomX igloo | Recipies for different FPC power |
|
For ~ 66kW power in the FP-cavity:
- Alplhanov amplifier ratio : 25%
- Axis 18 position : +0018369 steps
- PID A (for laser/FPC lock) :
input : input a
P = 0.1
I = 0.0016
D = 1.4
Sign : positive
Sampling : fast
filter : off
criterion : input d
upper th. : 10V
lower th. : 0.03V
speed : 100V/s
relock mode : none
off mode : hold
range : 50%
offset : 5V
- PID B (for RF/FPC lock) :
input : input b
P = 5
I = 0.0001
D = 1
Sign : negative
Sampling : mid
filter : off
criterion : input c
upper th. : 0.45V
lower th. : -0.45V
speed : 1V/s
relock mode : none
off mode : hold
range : 20%
offset : 5V
| Ronic Chiche wrote: |
|
For ~ 23kW power in the FP-cavity:
- Alplhanov amplifier ratio : 15%
- Axis 18 position : -0016014 steps
- PID A (for laser/FPC lock) :
input : input a
P = 0.09
I = 0.0016
D = 1.35
Sign : positive
Sampling : fast
filter : off
criterion : input d
upper th. : 10V
lower th. : 0.03V
speed : 100V/s
relock mode : none
off mode : hold
range : 50%
offset : 5V
- PID B (for RF/FPC lock) :
input : input b
P = 5
I = 0.0001
D = 1
Sign : negative
Sampling : mid
filter : off
criterion : input c
upper th. : 0.45V
lower th. : -0.45V
speed : 1V/s
relock mode : none
off mode : hold
range : 20%
offset : 5V
| Ronic Chiche wrote: |
|
For ~ 46kW power in the FP-cavity:
- Alplhanov amplifier ratio : 20%
- Axis 18 position : -0004710 steps
- PID A (for laser/FPC lock) :
input : input a
P = 0.06
I = 0.0007
D = 0.85
Sign : positive
Sampling : fast
filter : off
criterion : input d
upper th. : 10V
lower th. : 0.04V
speed : 100V/s
relock mode : none
off mode : hold
range : 50%
offset : 5V
- PID B (for RF/FPC lock) :
input : input b
P = 3
I = 0.00002
D = 2
Sign : negative
Sampling : mid
filter : off
criterion : input c
upper th. : 0.45V
lower th. : -0.45V
speed : 1V/s
relock mode : none
off mode : hold
range : 20%
offset : 5V
|
|
|
|
56
|
Fri Sep 6 14:48:14 2019 |
Loïc Amoudry | Fixed | report | mechanics | vacuum | Optical room | Primary vacuum to check dust |
The cavity has been cleaned then put under primary vacuum (primary pump without vacuum flow control). Dusts presence was checked at air and no significant changes have been observed.
1ST image before cleaning
2ND image after cleaning
No changes after coming back from primary vacuum |
| Attachment 1: Image3.jpg
|
 |
| Attachment 2: Image6.jpg
|
 |
|
93
|
Wed Feb 19 17:21:28 2020 |
Loïc Amoudry | Fixed | report | lasers and optics | Optical room | Powerup with 2.5ps pulses |
A power up has been performed on ThomX until 65kW intracavity power. We didn't see any modes but stored power was really unstable since ~60kW (see image tek00004.png).
Measurement report is shown in the table below. Gain seems to decrease against intracavity power.
As a conclusion, pulse length compression does not seems to bring any trouble in the PDH loop. But at relatively high intracavity power, power start to be really unstable even if we did not see any mode.
We can notice that we tried to align, change CEP but it had no important impact on the quality of the lock. See further experiments with Ronic and D-shape. |
| Attachment 1: report.PNG
|
 |
| Attachment 2: tek00004.png
|
 |
| Attachment 3: 4A.PNG
|
 |
|
453
|
Thu Sep 18 18:29:04 2025 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | vacuum | detectors and electronics | cabling | software | utilities | ThomX igloo | Power shut down on September 23th |
to anticipate the power shut down on next tuesday 23/09,
this afternoon, I switched off all the equipments except the oscillator connected on the UPS. |
|
454
|
Fri Sep 26 11:08:21 2025 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | vacuum | detectors and electronics | cabling | software | utilities | ThomX igloo | Power shut down on September 23th |
this morning, I restarted all the equipments of the FPC.
everything was fine after being able to get back the correct PDH phase between modulation and demodulation...I got 88kW with 33% amplifier ratio after basic alignment and CEP tuning.
the Pico1 server used for the X-ray photodiode TMD.01 doesn't want to start...
I tried with Astor. the server start to run (red => blue => green colors)
but when I start the Jive panel, it gets down again (red color)... to be solved
| Ronic Chiche wrote: |
|
to anticipate the power shut down on next tuesday 23/09,
this afternoon, I switched off all the equipments except the oscillator connected on the UPS.
|
|
|
455
|
Fri Sep 26 11:13:35 2025 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | vacuum | detectors and electronics | cabling | software | utilities | ThomX igloo | Power shut down on September 23th |
the 33MHz/500MHz RF frequencies from the Ring seems different from the previous state (maybe working at an energy of 50MeV instead of 70MeV which was the previous energy ?).
so, I let the FPC at the present position and wait to discuss if it is necessary or not to move the FPC motors for the correct energy...
| Ronic Chiche wrote: |
|
this morning, I restarted all the equipments of the FPC.
everything was fine after being able to get back the correct PDH phase between modulation and demodulation...I got 88kW with 33% amplifier ratio after basic alignment and CEP tuning.
the Pico1 server used for the X-ray photodiode TMD.01 doesn't want to start...
I tried with Astor. the server start to run (red => blue => green colors)
but when I start the Jive panel, it gets down again (red color)... to be solved
| Ronic Chiche wrote: |
|
to anticipate the power shut down on next tuesday 23/09,
this afternoon, I switched off all the equipments except the oscillator connected on the UPS.
|
|
|
|
456
|
Mon Sep 29 10:22:32 2025 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | vacuum | detectors and electronics | cabling | software | utilities | ThomX igloo | Power shut down on September 23th |
This morning I did a phase alignment on the PDH signal generator (Rigol DG4102)
this procedure ensure that both phases of the sine wave signals are aligned before finding the correct phase for maximizing the error signal.
so, from now on, in case of power shut down, one just need to :
1) do a phase alignment
2) put the correct phase number on the 2 channels according to the picture (if not automatically done by the generator itself).
340° on channel 1
270° on channel 2
1) and 2) can be done in any order.
| Ronic Chiche wrote: |
|
the 33MHz/500MHz RF frequencies from the Ring seems different from the previous state (maybe working at an energy of 50MeV instead of 70MeV which was the previous energy ?).
so, I let the FPC at the present position and wait to discuss if it is necessary or not to move the FPC motors for the correct energy...
| Ronic Chiche wrote: |
|
this morning, I restarted all the equipments of the FPC.
everything was fine after being able to get back the correct PDH phase between modulation and demodulation...I got 88kW with 33% amplifier ratio after basic alignment and CEP tuning.
the Pico1 server used for the X-ray photodiode TMD.01 doesn't want to start...
I tried with Astor. the server start to run (red => blue => green colors)
but when I start the Jive panel, it gets down again (red color)... to be solved
| Ronic Chiche wrote: |
|
to anticipate the power shut down on next tuesday 23/09,
this afternoon, I switched off all the equipments except the oscillator connected on the UPS.
|
|
|
|
| Attachment 1: img1.jpg
|
 |
|
199
|
Mon Mar 20 09:33:54 2023 |
Ronic Chiche | Fixed | info | utilities | ThomX igloo | Power shut down on Saturday the 18th |
All the equipments have been switched off on Friday 17th due to a power shut down on Saturday 18th :
CW and Onefive oscillators, computers, laser amplifier, cavity mirror motors, chiller, etc... |
|
201
|
Mon Mar 20 17:39:28 2023 |
Ronic Chiche | Fixed | info | utilities | ThomX igloo | Power shut down on Saturday the 18th |
Today, I restarted all the equipements :
Chiller : ok, temperature converged quickly to the set temperature : 25°C
Cavity mirrors motors controller : I had to configure the general network as 192.168.xxx.xxx in order to operate the software designed by Didier Jehanno
Computer : ok, now, it IP address is 192.168.1.1
Laser amplifier controller : I had a LOT of issues in restarting the LAL software. I had to restart several times the amplifier controller in order to access it by software. I quickly check the power increase to 50% after it has been correctly seeded
CW oscillator : ok, Koheras restarted without any issue.
OneFive oscillator : the power ON was ok but it did not modelock. I tried several gently knocks on the case but without success (one clearly see the effect on the scope, the oscillator "tries" to modelock).
fortunately, after about 1h, the oscillator modelocked alone (check with Buddy fiber photodiode + scope).... the power is the same as before. the Alphanov software measures 5.4mW (it was 5.5 mW the 2/1/2023)
Smaract controller for OneFive : ok, I changed its IP address to 192.168.1.10. I tested the PC control in moving forth and back the "CEP" motor by 1mm... the OneFive was still modelocked.
| Ronic Chiche wrote: |
|
All the equipments have been switched off on Friday 17th due to a power shut down on Saturday 18th :
CW and Onefive oscillators, computers, laser amplifier, cavity mirror motors, chiller, etc...
|
|
|
449
|
Wed Aug 27 12:01:41 2025 |
Ronic Chiche | Fixed | info | detectors and electronics | ThomX igloo | Power shut down on August 29th |
This morning, all the equipments have been switched off due to a power shut down tomorrow morning.
Only the Onefive oscillator is still ON, connected to the Uninterruptable Power Supply (UPS). |
|
450
|
Thu Aug 28 18:01:46 2025 |
Ronic Chiche | Fixed | info | detectors and electronics | ThomX igloo | Power shut down on August 29th |
after alignment, CEP adjustment and optimizing the PDH phase, I got back ~92kW in the FPC for 33% amplifier ratio.
| Ronic Chiche wrote: |
|
This morning, all the equipments have been switched off due to a power shut down tomorrow morning.
Only the Onefive oscillator is still ON, connected to the Uninterruptable Power Supply (UPS).
|
|
|
427
|
Wed Jun 11 18:59:34 2025 |
Ronic Chiche | Under Process | issue | mechanics | lasers and optics | vacuum | detectors and electronics | ThomX igloo | Possible causes for the noise in the FP-cavity system |
list of possible causes of the high frequency noise which makes the system loosing the lock => to be checked.
(if you have some new idea, I will edit the post)
1- Laser
1.1- vibrations on the laser box coming from the optical table / housing
1.2- vibrations coming from the Smaract translation stages (Frep or CEP) inside the laser box
1.3- pressure noise coming from the weather (the laser cavity is sealed) on the laser box
2- Amplifier
2.1- vibrations coming from the fans in the controler crate
2.2- vibrations coming from the chiller (water cooling cavitation)
3- Beam propagation
3.1- vibration noise coming from the housing / the table
3.2- vibration noise coming from the motors of the injection mirrors
4- Fabry-Perot cavity
4.1- pressure noise on the cavity vessels coming from the weather/air cooling
4.2- vibrations coming from the rust on the mechanics
4.3- vibrations of the mirror mounts due to temperature related to cavity power
4.4- vibrations coming from the electron ring mechanics
4.5- "vacuum" index variation coming from local ionization due to the electron beam
4.6- vibrations coming from the far position (related to the middle range) of some mounts and due to the spring of the translations stage.
5- Feedback system / Electronics / CEM
5.1- CEM votage noise on the laser or FPC PZT's
|
|
429
|
Wed Jun 25 09:57:56 2025 |
Ronic Chiche | Under Process | issue | mechanics | lasers and optics | vacuum | detectors and electronics | ThomX igloo | Possible causes for the noise in the FP-cavity system |
4.3 - Vibrations of the mirror mounts due to temperature related to cavity power (the electrons machine is OFF)
1) to test this possible issue, we plan to make a run at low power in the FP-cavity, around 10kW, instead of ~90kW as usual.
for that, I will need to change the diffuser position on axis 18 to get the same error signal for the locking.
the initial axis 18 position is +11304
I used the Alphanov amplifier at its standard value : 33%
first, I did a step at 26kW in the FPC by tuning the CEP
at this power, I saw exactly the same kind of behavior with a sudden large high frequency noise which prevent the lock to run properly and which is impossible to compensate.
=> one just have to wait... until it becomes more stable.
then, I went to 13kW in the FPC by tuning the CEP.
- I did the same alignement optimisation than for high power to remove possible coupling to high order modes in the error signal
- I tuned the power in the FPC by changing the CEP => the equivalent LW of the cavity will be different => I got ~ 13kW
- I changed the diffuser position on axis 18 to -23079
- I tuned the PID parameters :
initial => final
P = 0.055 => 0.1
I = 0.0005 => 0.002
D = 0.6 => 1
I started to record a long trend of the FPC power => see attached picture
I still observed the ~20Hz noise and sometime some high frequency noise, but not at a level which prevents the locking system to work.
CONCLUSION :
at this power and with a larger LW (CEP is not optimized), the FPC seems much more stable without any lock loss during 1h !
2) then I will do a test by decreasing the Alphanov amplifier ratio to work at a lower input power but with an optimized CEP.
Now, I optimized the CEP and alignment and reduce the Alphanov amplifier ratio to 15% => 24kW power in the FPC.
on CH2, I got 60mV (when I got 250mV when I'm at 97kW in the FPC) => ratio ~ 4.2 => 97kW / 4,2 ~ 23 kW
I checked that the PID parameters are not over-valuated which can produce a power reduction and the CEP is optimum as well as the alignment.
Axis 18 position : -13188
PID parameters (P = 0.1 / I = 0.002 / D = 1.5)
I started to record a long trend of the FPC power => cf 2nd trend with gradually increasing of the power because of the drift of the CEP + feedback adjustments.
CONCLUSION :
at this power and with nominal LW (CEP is optimized), the FPC seems much more stable without any lock loss during 1h !
3) I found a stable position for FPC around 46kW
I used the recepie for 46kW => after some time, I observed a lot of high frequency noise which induces some lock loss.
| Ronic Chiche wrote: |
|
list of possible causes of the high frequency noise which makes the system loosing the lock => to be checked.
(if you have some new idea, I will edit the post)
1- Laser
1.1- vibrations on the laser box coming from the optical table / housing
1.2- vibrations coming from the Smaract translation stages (Frep or CEP) inside the laser box
1.3- pressure noise coming from the weather (the laser cavity is sealed) on the laser box
2- Amplifier
2.1- vibrations coming from the fans in the controler crate
2.2- vibrations coming from the chiller (water cooling cavitation)
3- Beam propagation
3.1- vibration noise coming from the housing / the table
3.2- vibration noise coming from the motors of the injection mirrors
4- Fabry-Perot cavity
4.1- pressure noise on the cavity vessels coming from the weather/air cooling
4.2- vibrations coming from the rust on the mechanics
4.3- vibrations of the mirror mounts due to temperature related to cavity power
4.4- vibrations coming from the electron ring mechanics
4.5- "vacuum" index variation coming from local ionization due to the electron beam
5- Feedback system / Electronics / CEM
5.1- CEM votage noise on the laser or FPC PZT's
|
|
| Attachment 1: long_run_wihtout_lock_loss.png
|
 |
| Attachment 2: long_run_wihtout_lock_loss_with_optimum_CEP.png
|
 |
|
42
|
Wed Jun 12 17:49:20 2019 |
Loïc Amoudry | Fixed | info | lasers and optics | Optical room | Polarization |
Polariztion measured with Input vertically/horizontally polarized (~100% DOLP)
- Transmission (P4) locked on cavity horizontal polarization is ~50% linearly polarized and ~50% unpolarized ?
- Transmission (P4) locked on cavity vertical polarization is ~50% linearly polarized and ~50% unpolarized ? |
| Attachment 1: Polar_horiz_trans_P4.PNG
|
 |
| Attachment 2: Polar_vert_trans_P4.PNG
|
 |
| Attachment 3: Polar_vert_input.PNG
|
 |