HOME SBOX THOMX MINICAV Utilities
THOMX orders THOMX installation THOMX commissioning THOMX control command
  Status of commissioning, report also here plots are reports., Page 9 of 24  Not logged in ELOG logo
ID Date Author Status Type Category Location Titledown
  99   Fri Feb 28 17:44:12 2020 Ronic ChicheFixedreportlasers and opticsOptical roomPointing stability with CVBG

Yesterday, with Titouan, we made some measurements on pointing stability of the laser beam after amplifier and CVBG.
the amplifier beam go through a first telescope to be small enough and colimated before going to CVBG's, then go to CVBG's and then is sent through the FP-cavity telescope to the FP-cavity itself.
the total length is about 6m to mirror M1.
the surface of the mirror M1 is imaged with several wedges to a Basler CCD.
(x is for veritcal position and y is for horizontal position)

* the 1st plot shows the pointing stability at low power of amplifier without airflow and walking around during about 5 minutes extracted from the Basler CCD video.
when walking around or with airflow the pointing stabily is much worse.

* the 2nd plot shows the pointing stability vs amplifier current.
it is comparable (a litle bit worse) to the pointing stability at low power.
one clearly see the beam expanding in vertical direction and also in horizontal direction.
the effect could come from the CVBG telescope lenses which are standard lenses and not high power lenses.
the pointing effect could come also from the same effect if the lens is not perfectly centered.

we took also some picture of the CVBG at different power with the Thermal camera but we need to get the data from camera (old software not compatible with Windows 10)

Attachment 1: Pointing_stability_without_airflow_and_no_one_walking_around.png
Pointing_stability_without_airflow_and_no_one_walking_around.png
Attachment 2: Pointing_stability_without_airflow_vs_amplifier_current.png
Pointing_stability_without_airflow_vs_amplifier_current.png
  64   Thu Oct 3 13:49:58 2019 Loïc AmoudryFixedinfomechanicsOptical roomPneumatic valves

Frederic and Manu from vacuum group came this morning to fix again the pneumatic valves which had a leak again.

  83   Tue Nov 26 16:21:27 2019 Loïc AmoudryFixedissuevacuumOptical roomPneumatic valves

The pneumatic valves are now plugged on the university air network.
The air valve is fully openned for the moment.
We still have to check if there is some minor leakage close to the pneumatic valve to improve the setup.
But it's not anymore an issue for the lock.

  208   Thu Apr 13 17:48:49 2023 Manar AmerFixedreportlasers and optics | detectors and electronicsThomX iglooPW laser Injection into cavity

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,
Attachment 1: 20230412_pulsed_cav_mode.bmp
  209   Wed Apr 19 16:39:28 2023 Manar AmerFixedreportlasers and optics | detectors and electronicsThomX iglooPW laser Injection into cavity
  • On Monday with Daniele,
    • we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
  • On Tuesday with Daniele and Kevin
    •  I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
    • removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm  , Reflection : 50 ohm  )
    • We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
  • Images:
    • 1st : showing the size of the beam just before injecting into the cavity @ 10%
    • 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
    • 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
    • 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors

Note : the voltage ranges are not the same between the images

Manar Amer wrote:

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,

 

Attachment 1: 20230418_size_of_beam_before_injection_window_at_10percetage_ampli.jpg
20230418_size_of_beam_before_injection_window_at_10percetage_ampli.jpg
Attachment 2: 20230418_coupling_at_the_start_only_alignment_improved1.jpg
20230418_coupling_at_the_start_only_alignment_improved1.jpg
Attachment 3: 20230418_zoom_on_Oscilliscope_signal_fundamental_mode.jpg
20230418_zoom_on_Oscilliscope_signal_fundamental_mode.jpg
Attachment 4: 20230418_coupling_after_improving_it_with_regulator_on_attempting_to_lock.jpg
20230418_coupling_after_improving_it_with_regulator_on_attempting_to_lock.jpg
  210   Wed Apr 19 16:41:46 2023 Manar AmerFixedreportlasers and optics | detectors and electronicsThomX iglooPW laser Injection into cavity

Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)

was done at the end due to excitation of higher order modes.

Wanted to improve the fundamental first, then increase the power.

Manar Amer wrote:
  • On Monday with Daniele,
    • we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
  • On Tuesday with Daniele and Kevin
    •  I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
    • removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm  , Reflection : 50 ohm  )
    • We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
  • Images:
    • 1st : showing the size of the beam just before injecting into the cavity @ 10%
    • 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
    • 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
    • 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors

Note : the voltage ranges are not the same between the images

Manar Amer wrote:

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,

 

 

  211   Mon May 22 16:07:12 2023 Manar AmerFixedreportlasers and optics | detectors and electronicsThomX iglooPW laser Injection into cavity

Today with Ronic and Daniele we attempted to lock the cavity, but the alphanov amplifier did not turn on because of an error in MMD3 (related to the third stage)

the new error is having the MMD3 error show on the alarm window of the LAL software, and having the Alarms on the line 3 of the alphanov software red for both EXT/CPU and Laser T Max (never been red before)

this is after a 2-week work stop, but the last time when Daniele changed the fiber connector it worked for 3 hours with no errors.

Ronic, is in contact with Guillaume from Alphanov to fix the issue !!!

Work scheduled all Wednesday, in hopes the issue is fixed by then.

Manar Amer wrote:

Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)

was done at the end due to excitation of higher order modes.

Wanted to improve the fundamental first, then increase the power.

Manar Amer wrote:
  • On Monday with Daniele,
    • we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
  • On Tuesday with Daniele and Kevin
    •  I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
    • removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm  , Reflection : 50 ohm  )
    • We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
  • Images:
    • 1st : showing the size of the beam just before injecting into the cavity @ 10%
    • 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
    • 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
    • 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors

Note : the voltage ranges are not the same between the images

Manar Amer wrote:

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,

 

 

 

Attachment 1: 2023-05-22_MMD3_error_on_alarm_window.jpg
2023-05-22_MMD3_error_on_alarm_window.jpg
Attachment 2: 2023-05-22_alphanov_software.jpg
2023-05-22_alphanov_software.jpg
  32   Mon Apr 8 11:55:40 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomP4 cleaned and finesse measurement

P4 cleaned with isopropanol.

Finesse measured afterward: ~=13 000 (clear increase)

  365   Tue Sep 17 13:04:32 2024 Ronic ChicheUnder Processinfolasers and optics | softwareThomX iglooOptimizing lock parameters for stable X-ray production

here are the (good) lock parameters used this morning.

for the RF/FPC lock, the 33MHz beating signal used to select the right bucket is 1Vpp
=> beating signal : V0 . sin(phi) with V0=0.5V.

to discrimate a 500MHz bucket, we need to get dV < V0 dphi.

dphi = 2pi / 15 = 420 mrad => dV < 0.2 V => dV < +/- 0.1V

we used dV = +/- 0.02V but perharps we can relax the constraint.

we also reduced the RF scanning speed at 0.1V/s to let the system find the right phase when the system is slowly drifting.


 

 

Attachment 1: Capture.JPG
Capture.JPG
  366   Thu Oct 3 16:38:03 2024 Ronic ChicheUnder Processinfolasers and optics | softwareThomX iglooOptimizing lock parameters for stable X-ray production

Today with Daniele and Alice, we operated to cavity.
we obtained 84kW for 33% of amplifier ratio.
we had to tune the CEP @ -565µm and the FP-cavity alignment.

then we locked properly the FP-cavity on the ring RF frequency.
we quite easily relock with the correct phase when we are loosing the lock.

the ThomX machine was running during these 2 locks, so we are ready for doing X-rays again.

Ronic Chiche wrote:

here are the (good) lock parameters used this morning.

for the RF/FPC lock, the 33MHz beating signal used to select the right bucket is 1Vpp
=> beating signal : V0 . sin(phi) with V0=0.5V.

to discrimate a 500MHz bucket, we need to get dV < V0 dphi.

dphi = 2pi / 15 = 420 mrad => dV < 0.2 V => dV < +/- 0.1V

we used dV = +/- 0.02V but perharps we can relax the constraint.

we also reduced the RF scanning speed at 0.1V/s to let the system find the right phase when the system is slowly drifting.


 

 

 

  368   Wed Oct 9 09:56:21 2024 Ronic ChicheUnder Processinfolasers and optics | softwareThomX iglooOptimizing lock parameters for stable X-ray production

This morning, I wanted to test quickly if the Smaract motors can be used in open loop instead of closed loop or in piezo scan.

the goal is to do very fine steps without to much vibrations (like with piezo scan mode) but with the full motor range (the piezo scan mode has a very limited range).

I can check what happens to the 33MHz beating frequency between the laser and the RF frequency without the laser amplifier or the lock of the FP-cavity.
1Hz of beating freqency variation is equivalent to 270nm of round-trip length, which is 135nm of motor displacement !

=> the full range of the piezo scan mode is difficult to estimate because the measurement sensitivity is not good enough but around 5Hz.

=> 1350nm in closed loop is equivalent to ~ 10Hz => 135nm is equivalent to 1Hz => ok

=> 1 step in open loop is equivalent to ~ 10Hz !!! => this is a way too coarse tuning !!!! => cannot be used unless one finds a way to set the motion differently in the settings parameters.
but usually, the settings parameters are used only to tune the speed, not the step size.

 

Ronic Chiche wrote:

Today with Daniele and Alice, we operated to cavity.
we obtained 84kW for 33% of amplifier ratio.
we had to tune the CEP @ -565µm and the FP-cavity alignment.

then we locked properly the FP-cavity on the ring RF frequency.
we quite easily relock with the correct phase when we are loosing the lock.

the ThomX machine was running during these 2 locks, so we are ready for doing X-rays again.

Ronic Chiche wrote:

here are the (good) lock parameters used this morning.

for the RF/FPC lock, the 33MHz beating signal used to select the right bucket is 1Vpp
=> beating signal : V0 . sin(phi) with V0=0.5V.

to discrimate a 500MHz bucket, we need to get dV < V0 dphi.

dphi = 2pi / 15 = 420 mrad => dV < 0.2 V => dV < +/- 0.1V

we used dV = +/- 0.02V but perharps we can relax the constraint.

we also reduced the RF scanning speed at 0.1V/s to let the system find the right phase when the system is slowly drifting.


 

 

 

 

  369   Wed Oct 9 18:39:34 2024 Ronic ChicheUnder Processinfolasers and optics | softwareThomX iglooOptimizing lock parameters for stable X-ray production

This afternoon, we scanned the optical table vertically with the hexapod in asynchronous mode to find its optimum position, looking at the X-ray production.
then, we searched for the correct bucket and phase in the bucket thanks to the Kevin script on the 500MHz and the 33MHz phases.

we got relatively easily some stable X-rays.
on the gain "0" on the current amplifier of the X-rays photodiode, we got 15 000 pA (380k Xrays / pA => 5.7.10^9 Xrays).
but we saw, when the cathode charge was fluctuating, that we could be saturated above 20 000 pA !

the power in the cavity was ~84kW after CEP and alignment tuning.

the command to launch the X-rays measurement window is:
taurustrend -r 100 /XLI/OP/TMD.01/I1

Ronic Chiche wrote:

This morning, I wanted to test quickly if the Smaract motors can be used in open loop instead of closed loop or in piezo scan.

the goal is to do very fine steps without to much vibrations (like with piezo scan mode) but with the full motor range (the piezo scan mode has a very limited range).

I can check what happens to the 33MHz beating frequency between the laser and the RF frequency without the laser amplifier or the lock of the FP-cavity.
1Hz of beating freqency variation is equivalent to 270nm of round-trip length, which is 135nm of motor displacement !

=> the full range of the piezo scan mode is difficult to estimate because the measurement sensitivity is not good enough but around 5Hz.

=> 1350nm in closed loop is equivalent to ~ 10Hz => 135nm is equivalent to 1Hz => ok

=> 1 step in open loop is equivalent to ~ 10Hz !!! => this is a way too coarse tuning !!!! => cannot be used unless one finds a way to set the motion differently in the settings parameters.
but usually, the settings parameters are used only to tune the speed, not the step size.

 

Ronic Chiche wrote:

Today with Daniele and Alice, we operated to cavity.
we obtained 84kW for 33% of amplifier ratio.
we had to tune the CEP @ -565µm and the FP-cavity alignment.

then we locked properly the FP-cavity on the ring RF frequency.
we quite easily relock with the correct phase when we are loosing the lock.

the ThomX machine was running during these 2 locks, so we are ready for doing X-rays again.

Ronic Chiche wrote:

here are the (good) lock parameters used this morning.

for the RF/FPC lock, the 33MHz beating signal used to select the right bucket is 1Vpp
=> beating signal : V0 . sin(phi) with V0=0.5V.

to discrimate a 500MHz bucket, we need to get dV < V0 dphi.

dphi = 2pi / 15 = 420 mrad => dV < 0.2 V => dV < +/- 0.1V

we used dV = +/- 0.02V but perharps we can relax the constraint.

we also reduced the RF scanning speed at 0.1V/s to let the system find the right phase when the system is slowly drifting.


 

 

 

 

 

  374   Tue Oct 22 18:58:56 2024 Ronic ChicheUnder Processinfolasers and optics | softwareThomX iglooOptimizing lock parameters for stable X-ray production

Today, after setting the locking parameters, I got 80kW in the FPC for 33% amplifier ratio after CEP tuning (Smaract CH2 ~ -423.5µm) and alignment.

maybe the alignment has to be improved by some walking procedure.

 

Ronic Chiche wrote:

This afternoon, we scanned the optical table vertically with the hexapod in asynchronous mode to find its optimum position, looking at the X-ray production.
then, we searched for the correct bucket and phase in the bucket thanks to the Kevin script on the 500MHz and the 33MHz phases.

we got relatively easily some stable X-rays.
on the gain "0" on the current amplifier of the X-rays photodiode, we got 15 000 pA (380k Xrays / pA => 5.7.10^9 Xrays).
but we saw, when the cathode charge was fluctuating, that we could be saturated above 20 000 pA !

the power in the cavity was ~84kW after CEP and alignment tuning.

the command to launch the X-rays measurement window is:
taurustrend -r 100 /XLI/OP/TMD.01/I1

Ronic Chiche wrote:

This morning, I wanted to test quickly if the Smaract motors can be used in open loop instead of closed loop or in piezo scan.

the goal is to do very fine steps without to much vibrations (like with piezo scan mode) but with the full motor range (the piezo scan mode has a very limited range).

I can check what happens to the 33MHz beating frequency between the laser and the RF frequency without the laser amplifier or the lock of the FP-cavity.
1Hz of beating freqency variation is equivalent to 270nm of round-trip length, which is 135nm of motor displacement !

=> the full range of the piezo scan mode is difficult to estimate because the measurement sensitivity is not good enough but around 5Hz.

=> 1350nm in closed loop is equivalent to ~ 10Hz => 135nm is equivalent to 1Hz => ok

=> 1 step in open loop is equivalent to ~ 10Hz !!! => this is a way too coarse tuning !!!! => cannot be used unless one finds a way to set the motion differently in the settings parameters.
but usually, the settings parameters are used only to tune the speed, not the step size.

 

Ronic Chiche wrote:

Today with Daniele and Alice, we operated to cavity.
we obtained 84kW for 33% of amplifier ratio.
we had to tune the CEP @ -565µm and the FP-cavity alignment.

then we locked properly the FP-cavity on the ring RF frequency.
we quite easily relock with the correct phase when we are loosing the lock.

the ThomX machine was running during these 2 locks, so we are ready for doing X-rays again.

Ronic Chiche wrote:

here are the (good) lock parameters used this morning.

for the RF/FPC lock, the 33MHz beating signal used to select the right bucket is 1Vpp
=> beating signal : V0 . sin(phi) with V0=0.5V.

to discrimate a 500MHz bucket, we need to get dV < V0 dphi.

dphi = 2pi / 15 = 420 mrad => dV < 0.2 V => dV < +/- 0.1V

we used dV = +/- 0.02V but perharps we can relax the constraint.

we also reduced the RF scanning speed at 0.1V/s to let the system find the right phase when the system is slowly drifting.


 

 

 

 

 

 

  375   Wed Oct 23 12:23:10 2024 Ronic ChicheUnder Processinfolasers and optics | softwareThomX iglooOptimizing lock parameters for stable X-ray production

this morning, I did some walking procedure and I got 82kW in the FPC with 33% amplifier ratio (after CEP optimization too).
but when I move the FP cavity motors to adjust the frequency, I cannot keep this power and it is reduced.
could it be we get a stronger correlation between axis than before in the mechanics has more rust than before ?

I'm able to keep the power quite easily with a stable phase related to the 33MHz RF generator,
so, we are ready to produce X-rays again.

we could do also a measurement of the amplifier power vs ratio, as maybe it reduced a bit which could explain why we have this small power drop compare to before.

Ronic Chiche wrote:

Today, after setting the locking parameters, I got 80kW in the FPC for 33% amplifier ratio after CEP tuning (Smaract CH2 ~ -423.5µm) and alignment.

maybe the alignment has to be improved by some walking procedure.

 

Ronic Chiche wrote:

This afternoon, we scanned the optical table vertically with the hexapod in asynchronous mode to find its optimum position, looking at the X-ray production.
then, we searched for the correct bucket and phase in the bucket thanks to the Kevin script on the 500MHz and the 33MHz phases.

we got relatively easily some stable X-rays.
on the gain "0" on the current amplifier of the X-rays photodiode, we got 15 000 pA (380k Xrays / pA => 5.7.10^9 Xrays).
but we saw, when the cathode charge was fluctuating, that we could be saturated above 20 000 pA !

the power in the cavity was ~84kW after CEP and alignment tuning.

the command to launch the X-rays measurement window is:
taurustrend -r 100 /XLI/OP/TMD.01/I1

Ronic Chiche wrote:

This morning, I wanted to test quickly if the Smaract motors can be used in open loop instead of closed loop or in piezo scan.

the goal is to do very fine steps without to much vibrations (like with piezo scan mode) but with the full motor range (the piezo scan mode has a very limited range).

I can check what happens to the 33MHz beating frequency between the laser and the RF frequency without the laser amplifier or the lock of the FP-cavity.
1Hz of beating freqency variation is equivalent to 270nm of round-trip length, which is 135nm of motor displacement !

=> the full range of the piezo scan mode is difficult to estimate because the measurement sensitivity is not good enough but around 5Hz.

=> 1350nm in closed loop is equivalent to ~ 10Hz => 135nm is equivalent to 1Hz => ok

=> 1 step in open loop is equivalent to ~ 10Hz !!! => this is a way too coarse tuning !!!! => cannot be used unless one finds a way to set the motion differently in the settings parameters.
but usually, the settings parameters are used only to tune the speed, not the step size.

 

Ronic Chiche wrote:

Today with Daniele and Alice, we operated to cavity.
we obtained 84kW for 33% of amplifier ratio.
we had to tune the CEP @ -565µm and the FP-cavity alignment.

then we locked properly the FP-cavity on the ring RF frequency.
we quite easily relock with the correct phase when we are loosing the lock.

the ThomX machine was running during these 2 locks, so we are ready for doing X-rays again.

Ronic Chiche wrote:

here are the (good) lock parameters used this morning.

for the RF/FPC lock, the 33MHz beating signal used to select the right bucket is 1Vpp
=> beating signal : V0 . sin(phi) with V0=0.5V.

to discrimate a 500MHz bucket, we need to get dV < V0 dphi.

dphi = 2pi / 15 = 420 mrad => dV < 0.2 V => dV < +/- 0.1V

we used dV = +/- 0.02V but perharps we can relax the constraint.

we also reduced the RF scanning speed at 0.1V/s to let the system find the right phase when the system is slowly drifting.


 

 

 

 

 

 

 

  376   Fri Nov 8 15:23:11 2024 Ronic ChicheUnder Processinfolasers and optics | softwareThomX iglooOptimizing lock parameters for stable X-ray production

this morning, I did some walking procedure and CEP alignment to get ~80kW in the FPC with 33% amplifier ratio.

we have to check if this power drop comes from:

- a laser amplifier power drop
- or related to some cavity axis shift which could change the cavity gain due to L-shapes

Ronic Chiche wrote:

this morning, I did some walking procedure and I got 82kW in the FPC with 33% amplifier ratio (after CEP optimization too).
but when I move the FP cavity motors to adjust the frequency, I cannot keep this power and it is reduced.
could it be we get a stronger correlation between axis than before in the mechanics has more rust than before ?

I'm able to keep the power quite easily with a stable phase related to the 33MHz RF generator,
so, we are ready to produce X-rays again.

we could do also a measurement of the amplifier power vs ratio, as maybe it reduced a bit which could explain why we have this small power drop compare to before.

Ronic Chiche wrote:

Today, after setting the locking parameters, I got 80kW in the FPC for 33% amplifier ratio after CEP tuning (Smaract CH2 ~ -423.5µm) and alignment.

maybe the alignment has to be improved by some walking procedure.

 

Ronic Chiche wrote:

This afternoon, we scanned the optical table vertically with the hexapod in asynchronous mode to find its optimum position, looking at the X-ray production.
then, we searched for the correct bucket and phase in the bucket thanks to the Kevin script on the 500MHz and the 33MHz phases.

we got relatively easily some stable X-rays.
on the gain "0" on the current amplifier of the X-rays photodiode, we got 15 000 pA (380k Xrays / pA => 5.7.10^9 Xrays).
but we saw, when the cathode charge was fluctuating, that we could be saturated above 20 000 pA !

the power in the cavity was ~84kW after CEP and alignment tuning.

the command to launch the X-rays measurement window is:
taurustrend -r 100 /XLI/OP/TMD.01/I1

Ronic Chiche wrote:

This morning, I wanted to test quickly if the Smaract motors can be used in open loop instead of closed loop or in piezo scan.

the goal is to do very fine steps without to much vibrations (like with piezo scan mode) but with the full motor range (the piezo scan mode has a very limited range).

I can check what happens to the 33MHz beating frequency between the laser and the RF frequency without the laser amplifier or the lock of the FP-cavity.
1Hz of beating freqency variation is equivalent to 270nm of round-trip length, which is 135nm of motor displacement !

=> the full range of the piezo scan mode is difficult to estimate because the measurement sensitivity is not good enough but around 5Hz.

=> 1350nm in closed loop is equivalent to ~ 10Hz => 135nm is equivalent to 1Hz => ok

=> 1 step in open loop is equivalent to ~ 10Hz !!! => this is a way too coarse tuning !!!! => cannot be used unless one finds a way to set the motion differently in the settings parameters.
but usually, the settings parameters are used only to tune the speed, not the step size.

 

Ronic Chiche wrote:

Today with Daniele and Alice, we operated to cavity.
we obtained 84kW for 33% of amplifier ratio.
we had to tune the CEP @ -565µm and the FP-cavity alignment.

then we locked properly the FP-cavity on the ring RF frequency.
we quite easily relock with the correct phase when we are loosing the lock.

the ThomX machine was running during these 2 locks, so we are ready for doing X-rays again.

Ronic Chiche wrote:

here are the (good) lock parameters used this morning.

for the RF/FPC lock, the 33MHz beating signal used to select the right bucket is 1Vpp
=> beating signal : V0 . sin(phi) with V0=0.5V.

to discrimate a 500MHz bucket, we need to get dV < V0 dphi.

dphi = 2pi / 15 = 420 mrad => dV < 0.2 V => dV < +/- 0.1V

we used dV = +/- 0.02V but perharps we can relax the constraint.

we also reduced the RF scanning speed at 0.1V/s to let the system find the right phase when the system is slowly drifting.


 

 

 

 

 

 

 

 

  377   Tue Nov 19 17:37:55 2024 Ronic ChicheUnder Processinfolasers and optics | softwareThomX iglooOptimizing lock parameters for stable X-ray production

the RF frequency is now 500.1003MHz which is equivalent (if divided by 15) to 33.34002MHz.

the frequency has to be changed by ~2kHz @33MHz <=> ~550µm /2 for one motor.

I changed the laser and the FP cavities frequency.

for the laser frequency, the smaract motor CH1 is at 1.500627m.

for the FPC frequency, the plane mirror motors are at MOT.03= - 123 130 steps and MOT.06

BE CARREFUL, when the offset frequency is large, as we measure it on a scope with a beat frequency, if the scope window is too large,
one gets some stromboscope effect and one measures a lower frequency depending on the number of points in the window.

we got ~81kW for 33% amplifier ratio

Ronic Chiche wrote:

this morning, I did some walking procedure and CEP alignment to get ~80kW in the FPC with 33% amplifier ratio.

we have to check if this power drop comes from:

- a laser amplifier power drop
- or related to some cavity axis shift which could change the cavity gain due to L-shapes

Ronic Chiche wrote:

this morning, I did some walking procedure and I got 82kW in the FPC with 33% amplifier ratio (after CEP optimization too).
but when I move the FP cavity motors to adjust the frequency, I cannot keep this power and it is reduced.
could it be we get a stronger correlation between axis than before in the mechanics has more rust than before ?

I'm able to keep the power quite easily with a stable phase related to the 33MHz RF generator,
so, we are ready to produce X-rays again.

we could do also a measurement of the amplifier power vs ratio, as maybe it reduced a bit which could explain why we have this small power drop compare to before.

Ronic Chiche wrote:

Today, after setting the locking parameters, I got 80kW in the FPC for 33% amplifier ratio after CEP tuning (Smaract CH2 ~ -423.5µm) and alignment.

maybe the alignment has to be improved by some walking procedure.

 

Ronic Chiche wrote:

This afternoon, we scanned the optical table vertically with the hexapod in asynchronous mode to find its optimum position, looking at the X-ray production.
then, we searched for the correct bucket and phase in the bucket thanks to the Kevin script on the 500MHz and the 33MHz phases.

we got relatively easily some stable X-rays.
on the gain "0" on the current amplifier of the X-rays photodiode, we got 15 000 pA (380k Xrays / pA => 5.7.10^9 Xrays).
but we saw, when the cathode charge was fluctuating, that we could be saturated above 20 000 pA !

the power in the cavity was ~84kW after CEP and alignment tuning.

the command to launch the X-rays measurement window is:
taurustrend -r 100 /XLI/OP/TMD.01/I1

Ronic Chiche wrote:

This morning, I wanted to test quickly if the Smaract motors can be used in open loop instead of closed loop or in piezo scan.

the goal is to do very fine steps without to much vibrations (like with piezo scan mode) but with the full motor range (the piezo scan mode has a very limited range).

I can check what happens to the 33MHz beating frequency between the laser and the RF frequency without the laser amplifier or the lock of the FP-cavity.
1Hz of beating freqency variation is equivalent to 270nm of round-trip length, which is 135nm of motor displacement !

=> the full range of the piezo scan mode is difficult to estimate because the measurement sensitivity is not good enough but around 5Hz.

=> 1350nm in closed loop is equivalent to ~ 10Hz => 135nm is equivalent to 1Hz => ok

=> 1 step in open loop is equivalent to ~ 10Hz !!! => this is a way too coarse tuning !!!! => cannot be used unless one finds a way to set the motion differently in the settings parameters.
but usually, the settings parameters are used only to tune the speed, not the step size.

 

Ronic Chiche wrote:

Today with Daniele and Alice, we operated to cavity.
we obtained 84kW for 33% of amplifier ratio.
we had to tune the CEP @ -565µm and the FP-cavity alignment.

then we locked properly the FP-cavity on the ring RF frequency.
we quite easily relock with the correct phase when we are loosing the lock.

the ThomX machine was running during these 2 locks, so we are ready for doing X-rays again.

Ronic Chiche wrote:

here are the (good) lock parameters used this morning.

for the RF/FPC lock, the 33MHz beating signal used to select the right bucket is 1Vpp
=> beating signal : V0 . sin(phi) with V0=0.5V.

to discrimate a 500MHz bucket, we need to get dV < V0 dphi.

dphi = 2pi / 15 = 420 mrad => dV < 0.2 V => dV < +/- 0.1V

we used dV = +/- 0.02V but perharps we can relax the constraint.

we also reduced the RF scanning speed at 0.1V/s to let the system find the right phase when the system is slowly drifting.


 

 

 

 

 

 

 

 

 

  97   Wed Feb 26 16:48:02 2020 Ronic ChicheFixedinfolasers and opticsOptical roomOptical spectrum before and after CVBG

Measurement of the optical spectrum with Avantes OSA before (1st picture) and after CVBG (2nd picture) with 2nd stage on the laser amplifier.
The 1st CVBG stretches the beam horizontally due to the default incident angle and the fact that different wavelength are reflected in the CVBG with different depth.
as a result, the optical spectrum is varying along the transversal axis of the beam.
The 2nd CVBG is injected with the opposiste incident angle and should compensate the stretch effect to get back a circular beam.
spatially, the beam is quite circular but one can still see a dependance between position in the beam and optical spectrum.

Then, it is difficult to show the "right" optical spectrum after CVBG (one could use a diffuser for that) but it is clear that the spectral width is quite the same (~ 2nm due to the CFBG at the input of the amplifier) before and after CVBG.

Attachment 1: OSA_after_amplifier_CFBG_and_before_CVBG.PNG
OSA_after_amplifier_CFBG_and_before_CVBG.PNG
Attachment 2: OSA_after_amplifier_CFBG_2_CVBG.PNG
OSA_after_amplifier_CFBG_2_CVBG.PNG
  95   Wed Feb 26 10:38:37 2020 Ronic ChicheFixedinfolasers and opticsOptical roomOptical spectrum analyzer resolution

The Koheras @0.5mW is directly connected with fibers FC/APC -> FC/PC to the Avantes optical spectrum analyzer.

The FWHM wavelength, measured with the Avantes software, is 0.126nm

Attachment 1: Capture.PNG
Capture.PNG
  96   Wed Feb 26 11:05:09 2020 Ronic ChicheFixedinfolasers and opticsOptical roomOptical spectrum analyzer resolution

Same measurement with Koheras @100mW and using fiber coupling lenses between Koheras and Avantes OSA.

The FWHM wavelength, measured with the Avantes software, is 0.116nm

Ronic Chiche wrote:

The Koheras @0.5mW is directly connected with fibers FC/APC -> FC/PC to the Avantes optical spectrum analyzer.

The FWHM wavelength, measured with the Avantes software, is 0.126nm

 

Attachment 1: Capture2.PNG
Capture2.PNG
  71   Wed Oct 9 11:41:11 2019 Loïc AmoudryFixedinfolasers and opticsOptical roomOnefive spectrum and CVBG spectrum

The Onefive spectrum has shifted a little bit to lower lambda. It's still enough for the experiment (3mW injection of ampli intead of 1mW needed).

Attachment 1: 191008_spectrumbeforecvbg.PNG
191008_spectrumbeforecvbg.PNG
Attachment 2: 191008_spectrumaftercvbg.PNG
191008_spectrumaftercvbg.PNG
Attachment 3: 191008_beamshapeaftercvbgthomx.PNG
191008_beamshapeaftercvbgthomx.PNG
ELOG V3.1.4-395e101