HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports., Page 8 of 13  Not logged in ELOG logo
ID Date Author Status Type Category Location Titledown
  216   Fri Apr 12 17:18:15 2024 Xinyi LuFixedinfolasers and opticsOptical roomInstall 2-mirror cavity

Today Viktor and I completed the installation of the two-mirror cavity and managed to lock and measure the finesse.

- The finesse is 36k now (see figure 1). For the designed value of the mirror, the expected finesse is ~50k.

- The diameter of M2 transmission is 1.67 mm,1.65 mm (see figure 2).

- The installation process took a lot of time in orienting the PBS. In addition, we found that the cavity reflected beam and the window reflected beam would interfere (see figure 3). The small spot in the lower right corner is the window reflected light.

- We need to discuss whether the next step is to clean the mirrors or vacuum and move on. 

Xinyi Lu wrote:

Today, Viktor and I started installing the two-mirror cavity.
- Firstly, we cleaned the environment and the dust counter showed good cleanliness
- After opening the cavity we tried to determine the source of the strange spot with a laser detection card and found that the beam was very close to the front edge of the longitudinal D-shaped mirror. In addition there was nothing else strange.
- The setup of the two-mirror cavity is shown in Figure 1. We have to use the menhir laser of 216MHz. The mirrors used are shown in Figure 2.
- We have installed the M2 and will continue the installation tomorrow.

 

 
 

 

 

Attachment 1: finesse_2mirror.png
finesse_2mirror.png
Attachment 2: Screenshot_2024-04-12_170357.png
Screenshot_2024-04-12_170357.png
Attachment 3: Screenshot_2024-04-12_145025.png
Screenshot_2024-04-12_145025.png
  199   Wed Mar 13 19:44:41 2024 Xinyi LuFixedinfolasers and opticsOptical roomInjection at the amplifier current of 2 A

Today, Ronic and I optimized the locking at the amplifier current of 2 A and obtained ~ 21 kW inside the cavity.

- When all the iris open, the injected power is 10 W and the coupling is ~ 40%, corresponding to an effective gain of 2,100 and a full gain of 5,250. But the coupling may not be the true value because there is a large spot around the output beam.

- We have optimized the CEP, alignment, D-shaped mirrors and locking state. We optimized alignment after leaving the iris open and the inside power went from 14kW to 21kW.

- The transmission and reflection signals both have some same fluctuations, and they seem to come from the cavity. It's possible that the over-angled mirror mount could be the cause, but not sure. We will check in different power and see the stability of the signal.

- In addition, we found that the design values of the mirror incidence angles for the SBOX (3.359°, 5.900°) are different from the mirror ratings (1.146°). This may result in parameters such as reflection and transmission being different from the datasheet. It will also change the estimated maximum finesse, gain, and power inside the cavity. It might be better if the mirror parameters could be recalculated based on the actual angle of incidence.

Xinyi Lu wrote:

- Today, Ronic and I locked at the amplifier current of 2 A and obtained ~60% coupling after optimizing the CEP (see Figure 1).

- The injected power is 10 W at 2 A. We measured only 14 kW inside the cavity, which corresponds to an effective gain of 1,400 and a full gain of 2,300. The cavity finesse is 23,000 and the normal gain should be around 6,200.

- We found fluctuations in transmission, possibly because of mode degradation. Tomorrow we will use D-shape mirrors to suppress high-order modes and optimize alignment and locking.

Xinyi Lu wrote:

Today, Ronic and I installed the new telescope and locked the cavity.

- We locked at the amplifier current of 1 A and obtained 32% of coupling. (see Figure 1)

- The telescope was designed for a current of 2 A (output power ~10 W). To inject this power, we need to add some filters to devices.

- For CEP tuning, when we changed the AOM frequency while cavity locking, sometimes it caused unlock and power drops. It will be dangerous in high-power cases. So it's better to optimize the AOM frequency in low power and just tune the laser current in high power. Now the current variation range of the menhir laser is 750mA to 950mA.

Xinyi Lu wrote:

- These days, Ronic, Fatematuj and I measured the beam parameters of the output of the third-stage amplifier.

- We used 2 wedges and reflection filters to reduce the intensity on the CCD.

- We measured multiple points at pump current of 2 A (output power ~10 W). The waist diameter of the output is w_x = 792.26 um, w_y=873.90 um.

- The next step is to design the telescope and improve the coupling efficiency.

 

 

 

Attachment 1: 21kW_power_at_2A_injection.png
21kW_power_at_2A_injection.png
  26   Thu Jan 24 17:49:54 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomImpacts on mirror's surface

All the mirrors show impacts on there surface (some of them do not show deposit). Does it come from experiments or fabrication ? Are these holes or bumps ?

Attachment 1: 18_M1reflectiveface2x_compressed.jpg
18_M1reflectiveface2x_compressed.jpg
  107   Fri Jul 8 17:09:26 2022 Manar AmerFixedreportlasers and opticsOptical roomImages of ThomX mirrors

Under the microscope,

the spherical mirrors show the deposit of metal dots on the reflecting surface and not on the back of the mirrors

The mirror with fewer spots near the center was used in the setup.

 

Attachment 1: ThomX_Spherical_Front_coating.jpg
ThomX_Spherical_Front_coating.jpg
Attachment 2: ThomX_Spherical_2_Front_coating.jpg
ThomX_Spherical_2_Front_coating.jpg
Attachment 3: ThomX_Spherical_back.jpg
ThomX_Spherical_back.jpg
  108   Fri Jul 8 17:10:44 2022 Manar AmerFixedreportlasers and opticsOptical roomImages of ThomX mirrors

Mirror images in the previous log showing the metal deposit of the coating of the spherical mirrors

(not sent by mail due to large size of images)

Manar Amer wrote:

Under the microscope,

the spherical mirrors show the deposit of metal dots on the reflecting surface and not on the back of the mirrors

The mirror with fewer spots near the center was used in the setup.

 

 

  210   Wed Apr 3 21:37:07 2024 Xinyi LuFixedinfolasers and opticsOptical roomHigh power experiments (550kW)

- Today we moved the position of the D-shaped mirror at 6A. When motor1 (vertical) is 0.2mm away from the spot, the power in the cavity rises from 457kW to 483kW. Gain=8407 is similar to that at low power (Gain=8511). So the D-shaped mirror lost some of the gain in the previous experiments. At 4A and 5A we did not move the D-shaped mirror. (Figure 1)

- At 8A, we got 553 kW inside the cavity for one minute (Figure 2). The pump temperature is higher than yesterday (up to 34°C).

- At 7.5A and 8A, the cavity can remain stably locked, but the power fluctuation in the cavity is so large that it is difficult to optimize the alignment. This may be due to the short time the amplifier was on, the pump temperature, amplifier pointing and power fluctuations, and thermal effects in the cavity....... The amplifier operated differently at different moments.

- We measured the spectrum of the amplified laser. (Figure 3) The peak is 1032.2 nm. We will optimize the alignment and increase the power to optimize this measurement.

- Next arrangement
   Thursday: larger laser beam size
   Friday: smaller laser beam size
   Monday: finesse measurement with CW laser (Firstly check the possibility of measuring with pulsed laser)

Xinyi Lu wrote:

Yesterday, Ronic, Xing, Qili and I achieved a more stable 520kW power at 7.5A (71W injection) by optimizing the alignment and locking parameters. (Figure 1)

- The cavity can be stable locked when airflow is on. At 7.5A, the pump temperature is about 28℃. The chiller temperature didn't change, to the same 23 ℃ setting. We can try 8A later (75W injection) for a short time;

- Figure 2 demonstrates the cavity mode variation, wy/Pc ~ 1.7 mm/MW, half that of the OL paper (3.3 mm/MW). The thermal deformation of our device is much smaller.

- The experimental data are shown in Figure 3. Figure 4 shows the injection power vs circulating power.

- There are some tests that can be done at the moment. I'll update on the elog after discussing the necessity today. ^_^

Xinyi Lu wrote:

Last week, we achieved a stable intracavity average power of 500kW, limited by amplifier power. The experimental data are shown in Figure 1.

- We measured the transmitted laser with a power meter in the windows behind M2 and M4 respectively, and the results were consistent, so the measurements were credible.

- There is only one transmitted laser spot behind both M2 and M4.

- We measured 10-minute locking data at different powers (Figure 2). 480 kW data was not optimized, and we will add 500 kW locking data later.

- We compared cavity modes at different powers (Figure 3). There are fluctuations because we only saved one data at one power. More data will be collected for averaging later.

- After finishing the high-power experiments, we will measure the finesse and the transmission of the mirrors used. As well as the pulse duration, spectrum, phase noise, and repetition rate of the laser.

Xinyi Lu wrote:

Today, Ronic, Daniele, Aurélien and I measured the amplifier power and mirror transmission.

Current (A) 0 (2rd stage) 1 2 3 4 5 6 7 7.5 8
Power (W) 1 1.8 11.8 23.5 35.5 47 57.5 66.9 70.7 74.9

For transmission measurements, we used the same new mirrors as Sbox and ThomX, and installed an iris and a 2-inch mount to block the scattering laser.

The angle of incidence during the measurement was about 0.5°. We changed the angle and the measurements remained the same.

Mirror Number PL-0898 PL-10978
Nominal Value 3 ppm 115 ppm
Measured Value 1.75 ppm 113 ppm

If the mirror being used also has a transmission of 1.75 ppm, the original 270kW is actually 463kW!!! The gain is 6549 and the finesse is 28585 (70% coupling).

We will do more tests to check it.

  • Redo the experiment and check the spot behind the window at high power.
  • Move the power meter to the plane mirror M2 window. It was previously behind the curved mirror M4 window.
  • Compare locking curves, cavity mode sizes, and coupling efficiency at different powers.
  • After finishing the high-power experiments, we will measure the finesse using CW laser and the transmission of the mirrors used.
Xinyi Lu wrote:

Today, Ronic and I achieved 272kW inside the cavity at 7.5A. The coupling maintained 60%-70%.

Amp current (A) Injection power (W) Circulating power (kW) Gain
2 10 50 5000
3 22 105 4773
4 34 156 4588
5 47 210 4468
6 58(Estimated) 250 4310
7.5 76(Estimated) 272 3579

- Compared to yesterday's experiment, we moved the position of the D-shaped mirrors farther in two directions to make the higher-order modes just disappear.

- Possible reasons for higher gain: D-shaped mirrors position, high power and pump vacuum cleaned cavity mirrors so that improve the finesse.

- We didn't see the strange drops like yesterday (Figure 1). However, in the window behind the M3, we can see 3 spots correlating with the intracavity power, even though moving the D-shaped very far does not make them disappear, only weakens them. We don't know where they came from. When this round of experiments is over, we can open the cavity and observe the optical paths.

- Next steps:

  • Repeat the experiment to ensure that the gain does not drop.
  • Long-term measurement at maximum power when the amplifier temperature is safe.
  • Measure the transmittance of the cavity mirrors and the amplifier power.
  • Open the cavity and observe the optical paths and the mirror surface.
Xinyi Lu wrote:

all the injection power in the chart have not been measured recently but during the Loic thesis period.
and these old measurements stopped at 5.5A of pump current.... so, the data at "8A" is a pure estimation.

about the last measurement :
it was made at 6A/8A/8A/8A for the 4 pump diodes of the amplifier (because 1st stage has a Peltier issue and we cannot check its temperature), so the average current is 7.5A instead of 8A.
and the linear scale between pump current and amplifier power is ~ 12W/A, then the estimated amplifier power for the last measurement is 76W instead of 87W
and the estimated gain is more 2658.
for this current, the amplifier works out of its nominal limits (temperature set at 25°C but measured at 30°C !!!)  and the fans of the crate are making noise like hell.
so the last gain estimation should be treated very cautiously.

about the transmission and reflection signals behavior, one can write :
R + T + L = 1 => energy conservation for the cavity.
dR + dT + dL = 0 => dL = - (dR + dT)

if dX = Xfinal - Xinitial, dR and dT are < 0 on the last picture, then dL > 0.
it means that this picture seems to show that some losses are increasing from the beginning of the locking process.


several possibilities :
- we saw a strange D-shape effect on the large port of the cavity.
it seems that one of the D-shape mount/mirror is touching the intra-cavity beam producing some ghost effect on this large cavity port.
some cavity axis changing during the beginning of the lock could introduce some additionnal losses.
it can be easily tested by puting the D-shapes far from the beam.

- because of cavity axis changing at the beginning of the lock, the mirror losses are different.
but it is surprising that it is still going in the same direction... more losses at the end.
could be tested by slightly changing the optical axis of the cavity.

- "prior damage" behavior with a bump in the middle of the mirror due to thermal effect which introduces some losses at the end.
=> if it's the case, it's not a good behavior !!! :-(((
can be tested by looking at the wavefront phase in transmission.

- Non linear effect is the coatings.
but the field density seems not so much to produce this kind of effect

- A thermally induced change in the refractive index of the mirrors.
Daniele mentionned a relation between real and imaginary (related to absorption) parts of this refractive index which could explain that a reflectivity change could induce an absorption change.

Xinyi Lu wrote:

These days, Ronic and I achieved 200kW inside the cavity and 70% coupling efficiency.

- By optimizing the telescope, the coupling reached 70% with iris fully open and maintained 60%-70% coupling at high power.

- The cavity mode went from 2.2mm,2.5mm (38kW) to finally 2.3mm,2.8mm (200kW) without changing a lot.

- Gradually raising the power while optimizing alignment, CEP, and locking, we got the following stable power:

Amp current (A) Injection power (W) Circulating power (kW) Gain
2 10 38 3800
2.3 14 50 3571
3 22 70 3181
4 35 115 3285
5 48 158 3292
8 87(Estimated) 202 2322

- Next steps:

  • Explain the strange drop phenomenon that occurs at high power, where both transmission and reflection drop, as in Fig. 2.
  • Maintains a half-hour locking at 200kW. Now the temperature of the amplifier at 8A is over 40 degrees, which may be risky.

 

 

 

 

 

 

Attachment 1: record.png
record.png
Attachment 2: power_vs_time_550.jpg
power_vs_time_550.jpg
Attachment 3: 1A.JPG
1A.JPG
  209   Wed Apr 3 08:53:33 2024 Xinyi LuFixedinfolasers and opticsOptical roomHigh power experiments (520kW)

Yesterday, Ronic, Xing, Qili and I achieved a more stable 520kW power at 7.5A (71W injection) by optimizing the alignment and locking parameters. (Figure 1)

- The cavity can be stable locked when airflow is on. At 7.5A, the pump temperature is about 28℃. The chiller temperature didn't change, to the same 23 ℃ setting. We can try 8A later (75W injection) for a short time;

- Figure 2 demonstrates the cavity mode variation, wy/Pc ~ 1.7 mm/MW, half that of the OL paper (3.3 mm/MW). The thermal deformation of our device is much smaller.

- The experimental data are shown in Figure 3. Figure 4 shows the injection power vs circulating power.

- There are some tests that can be done at the moment. I'll update on the elog after discussing the necessity today. ^_^

Xinyi Lu wrote:

Last week, we achieved a stable intracavity average power of 500kW, limited by amplifier power. The experimental data are shown in Figure 1.

- We measured the transmitted laser with a power meter in the windows behind M2 and M4 respectively, and the results were consistent, so the measurements were credible.

- There is only one transmitted laser spot behind both M2 and M4.

- We measured 10-minute locking data at different powers (Figure 2). 480 kW data was not optimized, and we will add 500 kW locking data later.

- We compared cavity modes at different powers (Figure 3). There are fluctuations because we only saved one data at one power. More data will be collected for averaging later.

- After finishing the high-power experiments, we will measure the finesse and the transmission of the mirrors used. As well as the pulse duration, spectrum, phase noise, and repetition rate of the laser.

Xinyi Lu wrote:

Today, Ronic, Daniele, Aurélien and I measured the amplifier power and mirror transmission.

Current (A) 0 (2rd stage) 1 2 3 4 5 6 7 7.5 8
Power (W) 1 1.8 11.8 23.5 35.5 47 57.5 66.9 70.7 74.9

For transmission measurements, we used the same new mirrors as Sbox and ThomX, and installed an iris and a 2-inch mount to block the scattering laser.

The angle of incidence during the measurement was about 0.5°. We changed the angle and the measurements remained the same.

Mirror Number PL-0898 PL-10978
Nominal Value 3 ppm 115 ppm
Measured Value 1.75 ppm 113 ppm

If the mirror being used also has a transmission of 1.75 ppm, the original 270kW is actually 463kW!!! The gain is 6549 and the finesse is 28585 (70% coupling).

We will do more tests to check it.

  • Redo the experiment and check the spot behind the window at high power.
  • Move the power meter to the plane mirror M2 window. It was previously behind the curved mirror M4 window.
  • Compare locking curves, cavity mode sizes, and coupling efficiency at different powers.
  • After finishing the high-power experiments, we will measure the finesse using CW laser and the transmission of the mirrors used.
Xinyi Lu wrote:

Today, Ronic and I achieved 272kW inside the cavity at 7.5A. The coupling maintained 60%-70%.

Amp current (A) Injection power (W) Circulating power (kW) Gain
2 10 50 5000
3 22 105 4773
4 34 156 4588
5 47 210 4468
6 58(Estimated) 250 4310
7.5 76(Estimated) 272 3579

- Compared to yesterday's experiment, we moved the position of the D-shaped mirrors farther in two directions to make the higher-order modes just disappear.

- Possible reasons for higher gain: D-shaped mirrors position, high power and pump vacuum cleaned cavity mirrors so that improve the finesse.

- We didn't see the strange drops like yesterday (Figure 1). However, in the window behind the M3, we can see 3 spots correlating with the intracavity power, even though moving the D-shaped very far does not make them disappear, only weakens them. We don't know where they came from. When this round of experiments is over, we can open the cavity and observe the optical paths.

- Next steps:

  • Repeat the experiment to ensure that the gain does not drop.
  • Long-term measurement at maximum power when the amplifier temperature is safe.
  • Measure the transmittance of the cavity mirrors and the amplifier power.
  • Open the cavity and observe the optical paths and the mirror surface.
Xinyi Lu wrote:

all the injection power in the chart have not been measured recently but during the Loic thesis period.
and these old measurements stopped at 5.5A of pump current.... so, the data at "8A" is a pure estimation.

about the last measurement :
it was made at 6A/8A/8A/8A for the 4 pump diodes of the amplifier (because 1st stage has a Peltier issue and we cannot check its temperature), so the average current is 7.5A instead of 8A.
and the linear scale between pump current and amplifier power is ~ 12W/A, then the estimated amplifier power for the last measurement is 76W instead of 87W
and the estimated gain is more 2658.
for this current, the amplifier works out of its nominal limits (temperature set at 25°C but measured at 30°C !!!)  and the fans of the crate are making noise like hell.
so the last gain estimation should be treated very cautiously.

about the transmission and reflection signals behavior, one can write :
R + T + L = 1 => energy conservation for the cavity.
dR + dT + dL = 0 => dL = - (dR + dT)

if dX = Xfinal - Xinitial, dR and dT are < 0 on the last picture, then dL > 0.
it means that this picture seems to show that some losses are increasing from the beginning of the locking process.


several possibilities :
- we saw a strange D-shape effect on the large port of the cavity.
it seems that one of the D-shape mount/mirror is touching the intra-cavity beam producing some ghost effect on this large cavity port.
some cavity axis changing during the beginning of the lock could introduce some additionnal losses.
it can be easily tested by puting the D-shapes far from the beam.

- because of cavity axis changing at the beginning of the lock, the mirror losses are different.
but it is surprising that it is still going in the same direction... more losses at the end.
could be tested by slightly changing the optical axis of the cavity.

- "prior damage" behavior with a bump in the middle of the mirror due to thermal effect which introduces some losses at the end.
=> if it's the case, it's not a good behavior !!! :-(((
can be tested by looking at the wavefront phase in transmission.

- Non linear effect is the coatings.
but the field density seems not so much to produce this kind of effect

- A thermally induced change in the refractive index of the mirrors.
Daniele mentionned a relation between real and imaginary (related to absorption) parts of this refractive index which could explain that a reflectivity change could induce an absorption change.

Xinyi Lu wrote:

These days, Ronic and I achieved 200kW inside the cavity and 70% coupling efficiency.

- By optimizing the telescope, the coupling reached 70% with iris fully open and maintained 60%-70% coupling at high power.

- The cavity mode went from 2.2mm,2.5mm (38kW) to finally 2.3mm,2.8mm (200kW) without changing a lot.

- Gradually raising the power while optimizing alignment, CEP, and locking, we got the following stable power:

Amp current (A) Injection power (W) Circulating power (kW) Gain
2 10 38 3800
2.3 14 50 3571
3 22 70 3181
4 35 115 3285
5 48 158 3292
8 87(Estimated) 202 2322

- Next steps:

  • Explain the strange drop phenomenon that occurs at high power, where both transmission and reflection drop, as in Fig. 2.
  • Maintains a half-hour locking at 200kW. Now the temperature of the amplifier at 8A is over 40 degrees, which may be risky.

 

 

 

 

 

Attachment 1: power_vs_time.png
power_vs_time.png
Attachment 2: cavitymode_vs_power_1.png
cavitymode_vs_power_1.png
Attachment 3: record.png
record.png
Attachment 4: injection_power_vs_intra_power.png
injection_power_vs_intra_power.png
  208   Tue Apr 2 08:39:17 2024 Xinyi LuFixedinfolasers and opticsOptical roomHigh power experiments (500kW)

Last week, we achieved a stable intracavity average power of 500kW, limited by amplifier power. The experimental data are shown in Figure 1.

- We measured the transmitted laser with a power meter in the windows behind M2 and M4 respectively, and the results were consistent, so the measurements were credible.

- There is only one transmitted laser spot behind both M2 and M4.

- We measured 10-minute locking data at different powers (Figure 2). 480 kW data was not optimized, and we will add 500 kW locking data later.

- We compared cavity modes at different powers (Figure 3). There are fluctuations because we only saved one data at one power. More data will be collected for averaging later.

- After finishing the high-power experiments, we will measure the finesse and the transmission of the mirrors used. As well as the pulse duration, spectrum, phase noise, and repetition rate of the laser.

Xinyi Lu wrote:

Today, Ronic, Daniele, Aurélien and I measured the amplifier power and mirror transmission.

Current (A) 0 (2rd stage) 1 2 3 4 5 6 7 7.5 8
Power (W) 1 1.8 11.8 23.5 35.5 47 57.5 66.9 70.7 74.9

For transmission measurements, we used the same new mirrors as Sbox and ThomX, and installed an iris and a 2-inch mount to block the scattering laser.

The angle of incidence during the measurement was about 0.5°. We changed the angle and the measurements remained the same.

Mirror Number PL-0898 PL-10978
Nominal Value 3 ppm 115 ppm
Measured Value 1.75 ppm 113 ppm

If the mirror being used also has a transmission of 1.75 ppm, the original 270kW is actually 463kW!!! The gain is 6549 and the finesse is 28585 (70% coupling).

We will do more tests to check it.

  • Redo the experiment and check the spot behind the window at high power.
  • Move the power meter to the plane mirror M2 window. It was previously behind the curved mirror M4 window.
  • Compare locking curves, cavity mode sizes, and coupling efficiency at different powers.
  • After finishing the high-power experiments, we will measure the finesse using CW laser and the transmission of the mirrors used.
Xinyi Lu wrote:

Today, Ronic and I achieved 272kW inside the cavity at 7.5A. The coupling maintained 60%-70%.

Amp current (A) Injection power (W) Circulating power (kW) Gain
2 10 50 5000
3 22 105 4773
4 34 156 4588
5 47 210 4468
6 58(Estimated) 250 4310
7.5 76(Estimated) 272 3579

- Compared to yesterday's experiment, we moved the position of the D-shaped mirrors farther in two directions to make the higher-order modes just disappear.

- Possible reasons for higher gain: D-shaped mirrors position, high power and pump vacuum cleaned cavity mirrors so that improve the finesse.

- We didn't see the strange drops like yesterday (Figure 1). However, in the window behind the M3, we can see 3 spots correlating with the intracavity power, even though moving the D-shaped very far does not make them disappear, only weakens them. We don't know where they came from. When this round of experiments is over, we can open the cavity and observe the optical paths.

- Next steps:

  • Repeat the experiment to ensure that the gain does not drop.
  • Long-term measurement at maximum power when the amplifier temperature is safe.
  • Measure the transmittance of the cavity mirrors and the amplifier power.
  • Open the cavity and observe the optical paths and the mirror surface.
Xinyi Lu wrote:

all the injection power in the chart have not been measured recently but during the Loic thesis period.
and these old measurements stopped at 5.5A of pump current.... so, the data at "8A" is a pure estimation.

about the last measurement :
it was made at 6A/8A/8A/8A for the 4 pump diodes of the amplifier (because 1st stage has a Peltier issue and we cannot check its temperature), so the average current is 7.5A instead of 8A.
and the linear scale between pump current and amplifier power is ~ 12W/A, then the estimated amplifier power for the last measurement is 76W instead of 87W
and the estimated gain is more 2658.
for this current, the amplifier works out of its nominal limits (temperature set at 25°C but measured at 30°C !!!)  and the fans of the crate are making noise like hell.
so the last gain estimation should be treated very cautiously.

about the transmission and reflection signals behavior, one can write :
R + T + L = 1 => energy conservation for the cavity.
dR + dT + dL = 0 => dL = - (dR + dT)

if dX = Xfinal - Xinitial, dR and dT are < 0 on the last picture, then dL > 0.
it means that this picture seems to show that some losses are increasing from the beginning of the locking process.


several possibilities :
- we saw a strange D-shape effect on the large port of the cavity.
it seems that one of the D-shape mount/mirror is touching the intra-cavity beam producing some ghost effect on this large cavity port.
some cavity axis changing during the beginning of the lock could introduce some additionnal losses.
it can be easily tested by puting the D-shapes far from the beam.

- because of cavity axis changing at the beginning of the lock, the mirror losses are different.
but it is surprising that it is still going in the same direction... more losses at the end.
could be tested by slightly changing the optical axis of the cavity.

- "prior damage" behavior with a bump in the middle of the mirror due to thermal effect which introduces some losses at the end.
=> if it's the case, it's not a good behavior !!! :-(((
can be tested by looking at the wavefront phase in transmission.

- Non linear effect is the coatings.
but the field density seems not so much to produce this kind of effect

- A thermally induced change in the refractive index of the mirrors.
Daniele mentionned a relation between real and imaginary (related to absorption) parts of this refractive index which could explain that a reflectivity change could induce an absorption change.

Xinyi Lu wrote:

These days, Ronic and I achieved 200kW inside the cavity and 70% coupling efficiency.

- By optimizing the telescope, the coupling reached 70% with iris fully open and maintained 60%-70% coupling at high power.

- The cavity mode went from 2.2mm,2.5mm (38kW) to finally 2.3mm,2.8mm (200kW) without changing a lot.

- Gradually raising the power while optimizing alignment, CEP, and locking, we got the following stable power:

Amp current (A) Injection power (W) Circulating power (kW) Gain
2 10 38 3800
2.3 14 50 3571
3 22 70 3181
4 35 115 3285
5 48 158 3292
8 87(Estimated) 202 2322

- Next steps:

  • Explain the strange drop phenomenon that occurs at high power, where both transmission and reflection drop, as in Fig. 2.
  • Maintains a half-hour locking at 200kW. Now the temperature of the amplifier at 8A is over 40 degrees, which may be risky.

 

 

 

 

 
 

 

 

Attachment 1: record.png
record.png
Attachment 2: power_vs_time.png
power_vs_time.png
Attachment 3: cavitymode_vs_power.png
cavitymode_vs_power.png
  205   Wed Mar 27 22:37:02 2024 Xinyi LuFixedinfolasers and opticsOptical roomHigh power experiments (272kW)

Today, Ronic and I achieved 272kW inside the cavity at 7.5A. The coupling maintained 60%-70%.

Amp current (A) Injection power (W) Circulating power (kW) Gain
2 10 50 5000
3 22 105 4773
4 34 156 4588
5 47 210 4468
6 58(Estimated) 250 4310
7.5 76(Estimated) 272 3579

- Compared to yesterday's experiment, we moved the position of the D-shaped mirrors farther in two directions to make the higher-order modes just disappear.

- Possible reasons for higher gain: D-shaped mirrors position, high power and pump vacuum cleaned cavity mirrors so that improve the finesse.

- We didn't see the strange drops like yesterday (Figure 1). However, in the window behind the M3, we can see 3 spots correlating with the intracavity power, even though moving the D-shaped very far does not make them disappear, only weakens them. We don't know where they came from. When this round of experiments is over, we can open the cavity and observe the optical paths.

- Next steps:

  • Repeat the experiment to ensure that the gain does not drop.
  • Long-term measurement at maximum power when the amplifier temperature is safe.
  • Measure the transmittance of the cavity mirrors and the amplifier power.
  • Open the cavity and observe the optical paths and the mirror surface.
Xinyi Lu wrote:

all the injection power in the chart have not been measured recently but during the Loic thesis period.
and these old measurements stopped at 5.5A of pump current.... so, the data at "8A" is a pure estimation.

about the last measurement :
it was made at 6A/8A/8A/8A for the 4 pump diodes of the amplifier (because 1st stage has a Peltier issue and we cannot check its temperature), so the average current is 7.5A instead of 8A.
and the linear scale between pump current and amplifier power is ~ 12W/A, then the estimated amplifier power for the last measurement is 76W instead of 87W
and the estimated gain is more 2658.
for this current, the amplifier works out of its nominal limits (temperature set at 25°C but measured at 30°C !!!)  and the fans of the crate are making noise like hell.
so the last gain estimation should be treated very cautiously.

about the transmission and reflection signals behavior, one can write :
R + T + L = 1 => energy conservation for the cavity.
dR + dT + dL = 0 => dL = - (dR + dT)

if dX = Xfinal - Xinitial, dR and dT are < 0 on the last picture, then dL > 0.
it means that this picture seems to show that some losses are increasing from the beginning of the locking process.


several possibilities :
- we saw a strange D-shape effect on the large port of the cavity.
it seems that one of the D-shape mount/mirror is touching the intra-cavity beam producing some ghost effect on this large cavity port.
some cavity axis changing during the beginning of the lock could introduce some additionnal losses.
it can be easily tested by puting the D-shapes far from the beam.

- because of cavity axis changing at the beginning of the lock, the mirror losses are different.
but it is surprising that it is still going in the same direction... more losses at the end.
could be tested by slightly changing the optical axis of the cavity.

- "prior damage" behavior with a bump in the middle of the mirror due to thermal effect which introduces some losses at the end.
=> if it's the case, it's not a good behavior !!! :-(((
can be tested by looking at the wavefront phase in transmission.

- Non linear effect is the coatings.
but the field density seems not so much to produce this kind of effect

- A thermally induced change in the refractive index of the mirrors.
Daniele mentionned a relation between real and imaginary (related to absorption) parts of this refractive index which could explain that a reflectivity change could induce an absorption change.

Xinyi Lu wrote:

These days, Ronic and I achieved 200kW inside the cavity and 70% coupling efficiency.

- By optimizing the telescope, the coupling reached 70% with iris fully open and maintained 60%-70% coupling at high power.

- The cavity mode went from 2.2mm,2.5mm (38kW) to finally 2.3mm,2.8mm (200kW) without changing a lot.

- Gradually raising the power while optimizing alignment, CEP, and locking, we got the following stable power:

Amp current (A) Injection power (W) Circulating power (kW) Gain
2 10 38 3800
2.3 14 50 3571
3 22 70 3181
4 35 115 3285
5 48 158 3292
8 87(Estimated) 202 2322

- Next steps:

  • Explain the strange drop phenomenon that occurs at high power, where both transmission and reflection drop, as in Fig. 2.
  • Maintains a half-hour locking at 200kW. Now the temperature of the amplifier at 8A is over 40 degrees, which may be risky.

 

 

 
 

 

 

Attachment 1: Screenshot_2024-03-27_7_270kW_7.5A.png
Screenshot_2024-03-27_7_270kW_7.5A.png
Attachment 2: 210_kW_power_at_5A_injection.png
210_kW_power_at_5A_injection.png
Attachment 3: 272.3kW_at_7.5A.jpg
272.3kW_at_7.5A.jpg
  203   Tue Mar 26 19:36:48 2024 Xinyi LuFixedinfolasers and opticsOptical roomHigh power experiments (200kW)

These days, Ronic and I achieved 200kW inside the cavity and 70% coupling efficiency.

- By optimizing the telescope, the coupling reached 70% with iris fully open and maintained 60%-70% coupling at high power.

- The cavity mode went from 2.2mm,2.5mm (38kW) to finally 2.3mm,2.8mm (200kW) without changing a lot.

- Gradually raising the power while optimizing alignment, CEP, and locking, we got the following stable power:

Amp current (A) Injection power (W) Circulating power (kW) Gain
2 10 38 3800
2.3 14 50 3571
3 22 70 3181
4 35 115 3285
5 48 158 3292
8 87(Estimated) 202 2322

- Next steps:

  • Explain the strange drop phenomenon that occurs at high power, where both transmission and reflection drop, as in Fig. 2.
  • Maintains a half-hour locking at 200kW. Now the temperature of the amplifier at 8A is over 40 degrees, which may be risky.
Attachment 1: 202kW_power_at_8A_injection.png
202kW_power_at_8A_injection.png
Attachment 2: Strange_drops.jpg
Strange_drops.jpg
  204   Wed Mar 27 09:47:37 2024 Xinyi LuFixedinfolasers and opticsOptical roomHigh power experiments (200kW)

all the injection power in the chart have not been measured recently but during the Loic thesis period.
and these old measurements stopped at 5.5A of pump current.... so, the data at "8A" is a pure estimation.

about the last measurement :
it was made at 6A/8A/8A/8A for the 4 pump diodes of the amplifier (because 1st stage has a Peltier issue and we cannot check its temperature), so the average current is 7.5A instead of 8A.
and the linear scale between pump current and amplifier power is ~ 12W/A, then the estimated amplifier power for the last measurement is 76W instead of 87W
and the estimated gain is more 2658.
for this current, the amplifier works out of its nominal limits (temperature set at 25°C but measured at 30°C !!!)  and the fans of the crate are making noise like hell.
so the last gain estimation should be treated very cautiously.

about the transmission and reflection signals behavior, one can write :
R + T + L = 1 => energy conservation for the cavity.
dR + dT + dL = 0 => dL = - (dR + dT)

if dX = Xfinal - Xinitial, dR and dT are < 0 on the last picture, then dL > 0.
it means that this picture seems to show that some losses are increasing from the beginning of the locking process.


several possibilities :
- we saw a strange D-shape effect on the large port of the cavity.
it seems that one of the D-shape mount/mirror is touching the intra-cavity beam producing some ghost effect on this large cavity port.
some cavity axis changing during the beginning of the lock could introduce some additionnal losses.
it can be easily tested by puting the D-shapes far from the beam.

- because of cavity axis changing at the beginning of the lock, the mirror losses are different.
but it is surprising that it is still going in the same direction... more losses at the end.
could be tested by slightly changing the optical axis of the cavity.

- "prior damage" behavior with a bump in the middle of the mirror due to thermal effect which introduces some losses at the end.
=> if it's the case, it's not a good behavior !!! :-(((
can be tested by looking at the wavefront phase in transmission.

- Non linear effect is the coatings.
but the field density seems not so much to produce this kind of effect

- A thermally induced change in the refractive index of the mirrors.
Daniele mentionned a relation between real and imaginary (related to absorption) parts of this refractive index which could explain that a reflectivity change could induce an absorption change.

Xinyi Lu wrote:

These days, Ronic and I achieved 200kW inside the cavity and 70% coupling efficiency.

- By optimizing the telescope, the coupling reached 70% with iris fully open and maintained 60%-70% coupling at high power.

- The cavity mode went from 2.2mm,2.5mm (38kW) to finally 2.3mm,2.8mm (200kW) without changing a lot.

- Gradually raising the power while optimizing alignment, CEP, and locking, we got the following stable power:

Amp current (A) Injection power (W) Circulating power (kW) Gain
2 10 38 3800
2.3 14 50 3571
3 22 70 3181
4 35 115 3285
5 48 158 3292
8 87(Estimated) 202 2322

- Next steps:

  • Explain the strange drop phenomenon that occurs at high power, where both transmission and reflection drop, as in Fig. 2.
  • Maintains a half-hour locking at 200kW. Now the temperature of the amplifier at 8A is over 40 degrees, which may be risky.

 

  52   Wed Jan 29 19:00:02 2020 Loïc AmoudryFixedreportlasers and opticsOptical roomGlow discharge cleaner on SBOX mirrors

The glow discharge cleaner has been tested on the SBOX mirrors.

I've put them 1 by 1. Each run was 15min long at 15mA. The mirrors HR face was always away from the electrode and ~35° angle with the support. These values have been choosen thanks to the reference:

We have learned that Air can be compared to Azote and Azote to Oxygen in glow discharge. The main difference is H2O in air which make the glow discharge less stable as say several papers.

 

Attachment 1: glow.PNG
glow.PNG
  91   Tue May 24 22:11:40 2022 Manar AmerFixedreportlasers and opticsOptical roomFundamental Mode TEM00

The current setup for the SBox has a beam viewer and 2 photodiodes placed at reflection and transmission.

The CW laser piezoelectric crystal was driven using the lase lock system to observe the 00 mode.

Mode Observed and has no reflection (from what was previously seen)

higher order modes are resonating with 00 mode , example 11 mode (shown in image attached)

the coupling is yet to be seen and measured.

Attachment 1: 20220524_SBox_Setup.png
20220524_SBox_Setup.png
Attachment 2: 00mode_CW_piezo_drive.bmp
Attachment 3: 00mode_11mode.bmp
  92   Wed May 25 17:34:29 2022 Manar AmerFixedreportlasers and opticsOptical roomFundamental Mode TEM00

Adding images from the transmission after the spherical mirrors M3 and M4 window for a resonating mode inside the cavity.

There appears to be some misalignment from the center of the mirror ???

 

Manar Amer wrote:

The current setup for the SBox has a beam viewer and 2 photodiodes placed at reflection and transmission.

The CW laser piezoelectric crystal was driven using the lase lock system to observe the 00 mode.

Mode Observed and has no reflection (from what was previously seen)

higher order modes are resonating with 00 mode , example 11 mode (shown in image attached)

the coupling is yet to be seen and measured.

 

Attachment 1: beamviwer_transmission_at_M3.bmp
Attachment 2: beamviwer_transmission_at_M4.bmp
  94   Wed Jun 8 10:58:15 2022 Manar AmerFixedreportlasers and opticsOptical roomFundamental Mode TEM00

Images of the Transmitted beam at M2 mirror at different days. 

The first one was from last week, the second from yesterday.

There is misalignment (drift due to temperature) from the position where the beam was initially. ( reference taken between centers of mode and direct injected beam) 

it is most clear in the vertical position.

Manar Amer wrote:

Adding images from the transmission after the spherical mirrors M3 and M4 window for a resonating mode inside the cavity.

There appears to be some misalignment from the center of the mirror ???

 

Manar Amer wrote:

The current setup for the SBox has a beam viewer and 2 photodiodes placed at reflection and transmission.

The CW laser piezoelectric crystal was driven using the lase lock system to observe the 00 mode.

Mode Observed and has no reflection (from what was previously seen)

higher order modes are resonating with 00 mode , example 11 mode (shown in image attached)

the coupling is yet to be seen and measured.

 

 

Attachment 1: 20220530_00_Mode.bmp
Attachment 2: 00_Mode_Beam_Transmitted_Position_drift_due_to_temperature.bmp
  102   Wed Jul 6 18:35:38 2022 Manar AmerFixedreportlasers and opticsOptical roomFundamental Mode TEM00

Closing series of 00 mode for FSR @ 133.33 MHz

 

Manar Amer wrote:

Images of the Transmitted beam at M2 mirror at different days. 

The first one was from last week, the second from yesterday.

There is misalignment (drift due to temperature) from the position where the beam was initially. ( reference taken between centers of mode and direct injected beam) 

it is most clear in the vertical position.

Manar Amer wrote:

Adding images from the transmission after the spherical mirrors M3 and M4 window for a resonating mode inside the cavity.

There appears to be some misalignment from the center of the mirror ???

 

Manar Amer wrote:

The current setup for the SBox has a beam viewer and 2 photodiodes placed at reflection and transmission.

The CW laser piezoelectric crystal was driven using the lase lock system to observe the 00 mode.

Mode Observed and has no reflection (from what was previously seen)

higher order modes are resonating with 00 mode , example 11 mode (shown in image attached)

the coupling is yet to be seen and measured.

 

 

 

  22   Mon Nov 5 13:11:08 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomFirst run with spare mirrors
I (A) Trans (mW) Coupling (%) Pin (W) Gain
0 4 70 0.37 4982
2 58 80 5.2 5140
4 180 75 16.1 5152
6 270 63 27.3 4558
8 340 60 36.7 4269

 

Attachment 1: Plot_power_and_gain_vs_all.JPG
Plot_power_and_gain_vs_all.JPG
Attachment 2: powerup.xlsx
  12   Wed Oct 31 11:36:30 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomFinesse vs power by difference between main and second resonance

Measurements show that ratio decrease versus power. BUT, the second resonance measurement induce lower power in the cavity so the ratio is not directly true.

Also, simulation of the main/second resonance power by Pierre's simulation has shown ratio ~50, ~47.6 and 43.5 respectively for 0A, 2A and 4A.

I (A) Main resonance (mW) Second resonance (mW) Ratio
0 8.07 0.416 19.4
2 121 6.77 17.9
4 324 20.2 16

 

Attachment 1: CrossSecondaryResonance.m
clear all; close all;
tic
addpath(genpath('C:\Users\amoudry\desktop\Fichiers Labo\Fichiers Pierre\Simulation\Personal codes\Various'))
[TAS,~,r] = GetCavity('SBOX_ULE');
[F,~] = Get_info(TAS(1:4),TAS(5:8),TAS(9:12));
lambda = 1030e-9;
c = 299792458;
FSR = 133.33e6;
w0 = 2*pi*c/lambda;
tau = 1e-12; % FWHM duration
a = 4*log(2)/tau^2;
E0 = 1;%(pi/2/a)^(1/2); % Energy to normalize gaussian spectrum (Input beam power = 1)
DeltaPhiCE = 0; % CEP
N = 1e5;

dk = 0:(N-1);
Aa = (r.^dk-r.^(2.*N-dk));
Bb = E0*TAS(1)./(1-r.^2);
Cc = (1-r.^(2.*N));

Nn = 5e2; % Increase Nn <-> increase resolution
dtt = -Nn:Nn;
dtt = dtt*lambda/c/(0.1*Nn); %1e6
Ecn = zeros(numel(dtt),1);
for ii = 1:numel(dtt)
    for ll = 0:3
%         ll = 0;
        dt = dtt(ii)+ll*lambda/c;
        Phid = DeltaPhiCE + w0.*dt;
        temp_vect = Aa.*cos(dk.*Phid).*exp(-a.*dk.^2.*dt.^2./2);
        Ecn(ii,ll+1) = Bb.*(2*sum(temp_vect)-Cc);
        disp([num2str(ii)]);
    end
end
toc
%% Time plots
% figure
% semilogy(dtt/lambda*c,Ecn/max(Ecn),'LineWidth',2)
% hold on
% semilogy(dtt/lambda*c,Ecn2/max(Ecn),'LineWidth',2)
% hold on
% semilogy(dtt/lambda*c,Ecn3/max(Ecn),'LineWidth',2)
% hold on
% semilogy(dtt/lambda*c,Ecn4/max(Ecn),'LineWidth',2)
% set(gca,'FontSize',15)
% xlabel('\DeltaT (\lambda/c)')
% ylabel('Energy (A.U.)')
% grid on
% legend('\DeltaT = 0','\DeltaT = \lambda/c','\DeltaT = 2\lambda/c','\DeltaT = 3\lambda/c')
% axis square

% figure
% semilogy(dtt/lambda*c,Ecn/max(Ecn),'LineWidth',2)
% grid on
% set(gca,'FontSize',25)
% % set(gca,'YLim',[1e-9 1e0])
% xlabel('\DeltaT (\lambda_0/c)')
% ylabel('log(Energie (u.a.))')

%% Frequency
nu0 = w0/2/pi;
frep = (1/FSR-nu0/FSR*dtt).^(-1); % Infinity in dtt = 1/nu0

fprintf('\nFinesse : %g\n\n',F);
% Get linewidth
figure
for jj = 1:4
%     Find the 2 minimas of Ecn_half. Take the corresponding frep and
%     substract them
    Ecn_half = abs(Ecn(:,jj)-max(Ecn(:,jj))/2);
    Ecn_half2 = sort(Ecn_half);
    [row1,~] = find(Ecn_half==Ecn_half2(1),1);
    [row2,~] = find(Ecn_half==Ecn_half2(2),2);
    if numel(row2)>1  % Sometimes row can be a vector
        row2 = row2(2);
    end
    dnu = abs(frep(row2)-frep(row1));
    fprintf('RES %g\nMax gain : %g. Linewidth : %g kHz\n\n',jj-1,max(Ecn(:,jj)),dnu/1e3);
    
%     plot((frep-FSR)/FSR,Ecn(:,jj)/max(Ecn(:,1)),'LineWidth',2)
%     hold on
    semilogy((frep-FSR)/FSR,Ecn(:,jj)/max(Ecn(:,1)),'LineWidth',2)
    xlim([-0.05 0.05])
    hold on
end
set(gca,'FontSize',15)
xlabel('(f_r_e_p-FSR)/FSR')
% ylabel('Energy (A.U.)')
ylabel('log(Energie (u.a.))')
grid on
legend('\DeltaT = 0','\DeltaT = \lambda/c','\DeltaT = 2\lambda/c','\DeltaT = 3\lambda/c')
axis square
% axis([-0.01 0.01 10^-6 1])
Attachment 2: GetCavity.m
function [TAS,r,r_prod] = GetCavity(cav_name,varargin)
% Return T and r coefficient of a given cavity
% TAS vector contains the 4 T coeffs, then 4 A coeffs, then 4 S coeffs

if strcmp(cav_name,'SBOX_ULE')==1
    TAS(1) = 180e-6;   % T
    TAS(2) = 2e-6;
    TAS(3) = 2e-6;
    TAS(4) = 2e-6;
    TAS(5) = 1.15e-6;  % A
    TAS(6) = 1.27e-6;
    TAS(7) = 1.2e-6;
    TAS(8) = 1e-6;
    TAS(9) =  7e-6;     % S
    TAS(10) = 4.5e-6;
    TAS(11) = 3.6e-6;
    TAS(12) = 9e-6;
%     TAS(1) = 180e-6;   % T
%     TAS(2) = 3.2e-6;
%     TAS(3) = 2.8e-6;
%     TAS(4) = 2.85e-6;
%     TAS(5) = 30e-6;  % A
%     TAS(6) = 30e-6;
%     TAS(7) = 30e-6;
%     TAS(8) = 30e-6;
%     TAS(9) = 20e-6;     % S
%     TAS(10) = 20e-6;
%     TAS(11) = 20e-6;
%     TAS(12) = 20e-6;
    
elseif strcmp(cav_name,'ThomX')==1
    TAS(1) = 120e-6;   % T
    TAS(2) = 1.5e-6;
    TAS(3) = 1.5e-6;
    TAS(4) = 1.5e-6;
    TAS(5) = 0.4e-6;  % A
    TAS(6) = 0.24e-6;
    TAS(7) = 0.24e-6;
    TAS(8) = 0.27e-6;
    TAS(9) = 4e-6;     % S
    TAS(10) = 4.5e-6;
    TAS(11) = 10e-6;
    TAS(12) = 4.5e-6;
    
elseif strcmp(cav_name,'MIGHTY_low')==1
    TAS(1) = 1060e-6;
    TAS(2) = 330e-6;
    TAS(3) = 330e-6;
    TAS(4) = 330e-6;
    TAS(5:12) = 0;

elseif strcmp(cav_name,'Fab_cav')==1
    TAS(1) = 100e-6;
    TAS(2) = 10e-6;
    TAS(3) = 10e-6;
    TAS(4) = 10e-6;
    TAS(5:12) = 0;
end

switch nargin
    case 2
        TAS = repmat(TAS,numel(varargin{1}),1);
        TAS(:,1) = varargin{1};
    case 3
        TAS = repmat(TAS,numel(varargin{1}),1);
        TAS(:,1) = varargin{1};
        TAS(:,2) = varargin{2};
end
    
% Field reflection coeffs
rr = @(TAS) (1-sum(TAS,2)).^(1/2);
for ii = 1:4
    r(:,ii) = rr(TAS(:,ii:4:12));
end
r_prod = prod(r,2);
end
  15   Wed Oct 31 13:43:28 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomFinesse vs power by difference between main and second resonance

Measurement on 24/10/18

Loïc Amoudry wrote:

Measurements show that ratio decrease versus power. BUT, the second resonance measurement induce lower power in the cavity so the ratio is not directly true.

Also, simulation of the main/second resonance power by Pierre's simulation has shown ratio ~50, ~47.6 and 43.5 respectively for 0A, 2A and 4A.

I (A) Main resonance (mW) Second resonance (mW) Ratio
0 8.07 0.416 19.4
2 121 6.77 17.9
4 324 20.2 16

 

 

  16   Wed Oct 31 13:44:22 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomFinesse vs power by difference between main and second resonance

Measurement on 25/10/2018

Loïc Amoudry wrote:

Measurements show that ratio decrease versus power. BUT, the second resonance measurement induce lower power in the cavity so the ratio is not directly true.

Also, simulation of the main/second resonance power by Pierre's simulation has shown ratio ~50, ~47.6 and 43.5 respectively for 0A, 2A and 4A.

I (A) Main resonance (mW) Second resonance (mW) Ratio
0 8.07 0.416 19.4
2 121 6.77 17.9
4 324 20.2 16

 

 

ELOG V3.1.4-395e101