Message ID: 16     Entry time: Wed Oct 31 13:44:22 2018     In reply to: 12
 Author: Loïc Amoudry 
 Status: Fixed 
 Type: report 
 Category: lasers and optics 
 Location: Optical room 
 Title: Finesse vs power by difference between main and second resonance 

Measurement on 25/10/2018

Loïc Amoudry wrote:

Measurements show that ratio decrease versus power. BUT, the second resonance measurement induce lower power in the cavity so the ratio is not directly true.

Also, simulation of the main/second resonance power by Pierre's simulation has shown ratio ~50, ~47.6 and 43.5 respectively for 0A, 2A and 4A.

I (A) Main resonance (mW) Second resonance (mW) Ratio
0 8.07 0.416 19.4
2 121 6.77 17.9
4 324 20.2 16

 

 