HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports., Page 8 of 13  Not logged in ELOG logo
Entry   Koheras CW laser lock on FP cavity, posted by Ronic Chiche at Optical room about lasers and optics | detectors and electronics 
I locked the Koheras CW on the FP cavity but the lock was pretty noisy and very difficult to acquire.

I tried to produce some modulation sidebands close to 216MHz to measure the Finesse but the power loss was very small so, the signal to extract
the Finesse would be unreadable !
Entry   OEwaves lock and Finesse measurement, posted by Ronic Chiche at Optical room about lasers and optics 
this morning, I locked the FP cavity with the OEwaves CW laser and the "Fred fiber amplifier" used at 500mA of pump current.

the lock was much more easy than with the Koheras.

I had to change the 10GHz EOM which seems damaged as the modulation depth is very low and does not allow a Finesse measurement by modulation
    Reply   "ThomX" CELIA Amplifier commissionning, posted by Ronic Chiche at Optical room about lasers and optics 
this afternoon, with Fabian, we did some measurements on the 3rd stage.

finally we used the Dichroïc mirror at AOI=45°C as first mirror and a Thorlabs BB1-E03 as 2nd mirror to send the
beam to the powermeter.
    Reply   CELIA amplifier installation + chiller, posted by Manar Amer at Optical room about lasers and optics | detectors and electronics | software 
Closing series




Manar
Amer wrote:



The software to control the CELIA amplifier is on the laptop that was
    Reply   Alignment / Table setup , posted by Manar Amer at Optical room about lasers and optics | detectors and electronics | software 
closing series




Manar
Amer wrote:



The current setup of the optical table attached.
    Reply   CVBG stretching / Fiber Injection , posted by Manar Amer at Optical room about lasers and optics | detectors and electronics | software 
closing series




Manar
Amer wrote:



Before injecting into the amplifier, the pulse needs to be stretched
Entry   Amplifier output Beam Profile, posted by Manar Amer at Optical room about lasers and optics 20220819_Setup.jpg
In preparation to measure the output beam profile from the amplifier at high power, I have placed two wedge mirrors just before the beam dump to be able
to extract the beam.

One of the wedges was taken from ThomX bunker, also the HR and AR mirrors were taken to plic room in case we might need them
Entry   Telescope / amplifier output, posted by Manar Amer at Optical room about lasers and optics 
The previous Sbox telescope was dismantled and the mechanical components cleaned.

its lenses are still in the mounts, it looks that two of them are spherical and two are cylindrical

2 are -100 mm and 2 are +150 mm, there is also a box containing fused silica lenses that could be used.
    Reply   Telescope / amplifier output, posted by Manar Amer at Optical room about lasers and optics Focal_length_divergence_method.pdfBeam_Profile_at_high_power_.jpgSBOX_CELIA_AMPLIFIER_Divergence_measurement.pdf
Beam divergence was measured using a method called  "Focal
Length Divergence
Measurement Method"
    Reply   Amplifier output Beam Profile, posted by Manar Amer at Optical room about lasers and optics 
Beam divergence after amplifier 4.46 mrad

 




Manar
    Reply   Telescope / amplifier output, posted by Manar Amer at Optical room about lasers and optics 20220823_setup.jpg
I placed a periscope to adjust the high of the beam from the amplifier output from ~ 10 cm from the table to ~ 15 cm

a dichroic mirror placed after it to reject the pump laser, all the mirrors on the path to the cavity were replaced with dielectric mirrors BB01-E03

the length of the path from the amplifier output to the cavity coupling mirror ~ 2 meters
    Reply   Telescope / amplifier output, posted by Manar Amer at Optical room about lasers and optics Capture.PNGTelescope_AMPLI_CELIA_1.xml
Here is a view of beam propagation in the optical software : GaussianBeam

the red filled shape is the model of the CELIA amplifier beam propagation with a divergence of 4.46 mrad
(the 2 black dots is the measurement of the beam size without any lens to change the beam propagation).
    Reply   2 Mirror Setup @ 216.6 MHz, posted by Manar Amer at Optical room about lasers and optics manip2miroirs.pptx2mirror_plan_shperical_beam_size.png
Adding information about the 2 mirror cavity setup (plan - spherical) that is currently installed.

From Aurélien at the start of the manipulation.

@ 0 is where the injection mirror is located
Entry   2nd stage CELIA amplifier, posted by Ronic Chiche at Optical room about lasers and optics 2nd_stage_pump_wavelength.jpg
today, we measured the 2nd stage CELIA amplifier pump wavelength : 970-990 nm
    Reply   2nd stage CELIA amplifier, posted by Ronic Chiche at Optical room about lasers and optics image_faisceau_multimode_@1030nm.PNG
today we did several test with the Dichroic shortpass mirror (Thorlabs DMSP1000) and with a 10nm optical filter around 1030nm (which works in tranmission
at AOI=0).

one used the dichroic mirror in reflection: one should cut the pump @970-990nm and we should keep only the signal @1030nm.
    Reply   Telescope / amplifier output, posted by Manar Amer at Optical room about lasers and optics Beam_Profile_Fit_using_gaussian_Beam.jpgBeam_Profile_Fit_using_gaussian_Beam.xml
removing the reading which is not correct (wrong use of software)

redone a reading similar using a lens of focal 250 mm got a FWHM-X = 0.64 mm , FWHM-Y = 0.84 mm

using the vertical to calculate the divergence, we get divergence ~ 2.17 mrad which is closer to fit obtained for the beam profile by taking
    Reply   Telescope / amplifier output, posted by Manar Amer at Optical room about lasers and optics Ampli_Celia_2.3mrad_divergence_New_Telescope.jpgBeam_waist_at_injection_image.jpg
Note the correct beam divergence is approximately ~ 2.3 mrad

M2 = 1.1 in this fit, but it is not yet optimized !!!!! could be reason for not accurate telescope
reading.
    Reply   Telescope / amplifier output, posted by Manar Amer at Optical room about lasers and optics 20220905_Telescope_lense_placement.jpg
Am adjustment on the lenses position to have a smaller waist.

+ 250 mm @ 88 cm from amplifier

-150 mm @ 111 cm from amplifier
Entry   unsuccessful telescope design :-(, posted by Ronic Chiche at Optical room about lasers and optics image_at_z_being_40mm.pngbeam_fit_at_z_being_40mm.pngpropagation_fit.png
we wanted to calculate the right telescope with 2 spherical lenses.

1) we have the FP cavity mode size which is 0.58mm at the input mirror and 0.7mm at the output mirror.

2) we planned to measure the laser beam at the output of the amplifier working at P=1W (2nd stage ON only).
    Reply   Telescope / amplifier output, posted by Manar Amer at Optical room about lasers and optics 
We increased the power of the amplifier up to 10 W to see if there is a change in the beam shape at the injection point or the transmission.

There was no change in the shape of both of them from the reading at 1 W (with only the 2nd stage on)

Only saw an increase in the transmission power, which is expected.
ELOG V3.1.4-395e101