    Reply   optical table cleaning, posted by Ronic Chiche at Optical room about utilities 
this afternoon, we checked the dust meter which is at 0 for 1 and 5µm dust particles but ~ 2000-3000 for 0.3µm particles.

we opened the two top panels to let the air flow clean the inside of the vessel.

we observed some other minor oxydised regions (than the one taken in picture) on the external parts of the inox panels but at first sight, nothing
    Reply   optical table cleaning, posted by Ronic Chiche at Optical room about utilities 
now, it seems the table is clean enough (dust meter counts 0 particles) to install a 2 mirrors FP-cavity.




Ronic
Chiche wrote:



this afternoon, we checked the dust meter which is at 0 for 1 and
Entry   Damaged mirrors test, posted by Alice Renaux at Optical room about lasers and optics 
Today with Dorian we tested three mirrors with which we previously haven't been able to obtain any resonance or optical beating :

- C16111/11 : The mirror looks normal under a microscope, apart from a few inclusions and maybe a small scratch towards the edge. We tested it
with a 161186 M1 mirror and we weren't able to obtain any optical beating on the cavity's mirrors or resonance.
Entry   Installation of the avalanche photodiode, posted by Ronic Chiche at Optical room about lasers and optics | detectors and electronics 
After installing the 2nd EOM, we had some trouble to be able to lock again.
One possible reason was the very low signal level in transmission, which is important to trigger the locking system (and stop it).

See the Alice post for details, but we were able to measure only once the Finesse of the cavity at around 2600.