HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports., Page 6 of 13  Not logged in ELOG logo
ID Date Authordown Status Type Category Location Title
  74   Fri Apr 8 19:27:12 2022 Manar AmerFixedissuedetectors and electronics | utilitiesOptical roomClean Airflow Dust measurements

Dust measurement done today on  top of the SBox average 10min

Manar Amer wrote:

The dust meter cap was cleaned using Alcohol, and using the filter white cap the dust count was (0 Av. 10 min)

After the dust counter was hand held in direct airflow, it counted (1581 p/m3 Av. 2 min).

I tested the count also in the SAS, and it counted twice the amount ~ 4500 p/m3

Note: discussion on the next steps to take for the airflow filtering !!!!!!!!

 

Attachment 1: 20220408_DustMeasurement.jpg
20220408_DustMeasurement.jpg
  75   Wed Apr 13 18:42:50 2022 Manar AmerFixedreportlasers and optics | utilitiesOptical roomSpincoater - Mirrror cleaning

Used the spincoater and pure water to clean dielectric mirrors which will be installed inside the SBox for directing the reflection line.

The method : place pure alchole on mirror - wipe it using white sterile tissue (in one direction) - clean with pure water in spincoater suing cotton swap .

After multible rounds, the result in the images was obtained. there is still a significant amount of dust particles (not all could be dust!!).

Not sure if the reason is the cleaning method ?? will invistegate it more !!!

Manar Amer wrote:

In preparation for cleaning the mirrors.

The spincoater has been moved with its bump into the sas of the SBox.

The pure water used will be provided by Alice, as the previous source in the workshop is not available.

 

 

Attachment 1: Reflection_one_front_after_cleaning.jpg
Reflection_one_front_after_cleaning.jpg
Attachment 2: reflection_two_front_before_cleaning.jpg
reflection_two_front_before_cleaning.jpg
Attachment 3: Reflection_two_front_after_cleaning.jpg
Reflection_two_front_after_cleaning.jpg
  76   Wed Apr 20 19:05:37 2022 Manar AmerFixedreportlasers and opticsOptical roomKBox Alignment

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

Attachment 1: CW_Koheras_1meter_ColimatorOutput_NE40A_NE10B_NE01A_Beam.bmp
Attachment 2: CW_Koheras_1meter_ColimatorOutput_NE40A_NE10B_NE01A_BeamFit.png
CW_Koheras_1meter_ColimatorOutput_NE40A_NE10B_NE01A_BeamFit.png
Attachment 3: Ma1_Horizontal_Position.png
Ma1_Horizontal_Position.png
Attachment 4: Ma1_Vertical_Position.png
Ma1_Vertical_Position.png
Attachment 5: Ma2_Horizontal_Position.png
Ma2_Horizontal_Position.png
Attachment 6: Transmission_M2_Irises_No_Mirrors_Beam.jpg
Transmission_M2_Irises_No_Mirrors_Beam.jpg
Attachment 7: Transmission_M2_Irises_No_Mirrors_BeamFit.png
Transmission_M2_Irises_No_Mirrors_BeamFit.png
  77   Mon Apr 25 17:28:29 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

Attachment 1: SBox_Setup.jpg
SBox_Setup.jpg
Attachment 2: Ma1_Ma2_Horizontal_Position.png
Ma1_Ma2_Horizontal_Position.png
Attachment 3: Ma1_Ma2_Vertical_Position.png
Ma1_Ma2_Vertical_Position.png
Attachment 4: Output_afterM2_Diaghragm_On_M1_M2_Beam.jpg
Output_afterM2_Diaghragm_On_M1_M2_Beam.jpg
Attachment 5: Output_afterM2_Diaghragm_On_M1_M2_BeamFit.png
Output_afterM2_Diaghragm_On_M1_M2_BeamFit.png
Attachment 6: Output_afterM2_Diaghragm_On_M1_M2_Beam_Outside_Irises_Installed_and_Closed.jpg
Output_afterM2_Diaghragm_On_M1_M2_Beam_Outside_Irises_Installed_and_Closed.jpg
  78   Mon Apr 25 17:58:01 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

Attachment 1: Ma1_Ma2_Horizontal_Position_Optimized.png
Ma1_Ma2_Horizontal_Position_Optimized.png
Attachment 2: Ma1_Ma2_Vertical_Position_Optimized.png
Ma1_Ma2_Vertical_Position_Optimized.png
Attachment 3: After_M1_diaphram_1m_Saturated_Beam_Optimized.jpg
After_M1_diaphram_1m_Saturated_Beam_Optimized.jpg
Attachment 4: After_M2_M1_diaphram_Saturated_Beam_Optimized.jpg
After_M2_M1_diaphram_Saturated_Beam_Optimized.jpg
Attachment 5: After_M2_M1_diaphram_Beam_Optimized.jpg
After_M2_M1_diaphram_Beam_Optimized.jpg
  79   Mon Apr 25 18:02:09 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

  80   Tue Apr 26 20:08:10 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

Attachment 1: Pinhole_One.jpg
Pinhole_One.jpg
Attachment 2: Pinhole_Two.jpg
Pinhole_Two.jpg
Attachment 3: After_M1_diaphram_Saturated_Beam_Optimized.jpg
After_M1_diaphram_Saturated_Beam_Optimized.jpg
Attachment 4: After_M2_diaphram_Saturated_Beam_Optimized.jpg
After_M2_diaphram_Saturated_Beam_Optimized.jpg
Attachment 5: both_diaphrams_M1_M2.jpg
both_diaphrams_M1_M2.jpg
Attachment 6: both_diaphrams_M1_M2_Optimized.jpg
both_diaphrams_M1_M2_Optimized.jpg
Attachment 7: After_M2_diaphram_M1_Mirror_Installed_SBox_Saturated_Beam_Optimized.jpg
After_M2_diaphram_M1_Mirror_Installed_SBox_Saturated_Beam_Optimized.jpg
Attachment 8: Ma1_Ma2_Horizontal_Position_Optimized_M1_Mirror_Installes_SBox.png
Ma1_Ma2_Horizontal_Position_Optimized_M1_Mirror_Installes_SBox.png
Attachment 9: Ma1_Ma2_Vertical_Position_Optimized_M1_Mirror_Installes_SBox.png
Ma1_Ma2_Vertical_Position_Optimized_M1_Mirror_Installes_SBox.png
  81   Wed Apr 27 22:13:37 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

Finished the alignment of the SBox with M1 of SBox placed, to be changed to gamma factory mirror later

Observed what seems to be transmission through M2 mirror, to be confirmed. image attached shows it

a PowerPoint attached showing the details of today's work.

 

 

Manar Amer wrote:
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

 

Attachment 1: all_mirrorsplaced_after_m2_window.jpg
all_mirrorsplaced_after_m2_window.jpg
Attachment 2: 20222704_SBox_alignment.pptx
  82   Wed Apr 27 22:30:14 2022 Manar AmerFixedreportlasers and optics | utilitiesOptical roomSpincoater - Mirrror cleaning

Images of the SBox mirrors cleaned at the same time when the cavity was aligned.

Directly from the spincoater to the SBox directly.

PowerPoint attached shows details

Manar Amer wrote:

Used the spincoater and pure water to clean dielectric mirrors which will be installed inside the SBox for directing the reflection line.

The method : place pure alchole on mirror - wipe it using white sterile tissue (in one direction) - clean with pure water in spincoater suing cotton swap .

After multible rounds, the result in the images was obtained. there is still a significant amount of dust particles (not all could be dust!!).

Not sure if the reason is the cleaning method ?? will invistegate it more !!!

Manar Amer wrote:

In preparation for cleaning the mirrors.

The spincoater has been moved with its bump into the sas of the SBox.

The pure water used will be provided by Alice, as the previous source in the workshop is not available.

 

 

 

Attachment 1: Images_SBox_Mirrors.pptx
  83   Mon May 9 18:36:48 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

Last week In FP cavity we had all SBox mirrors including the damaged M1 mirror.

The beating was observed with just a slight adjustment on M4 alignment.

In addition, it was also observed while driving the piezo of the CW laser.

In the attached image of the beam at transmission point at M2 mirror, the shape of the beam is deformed.

one of the factors could be related to the damaged spot on M1 mirror. The deformation position can also be seen in the higher order modes.

after changing the injection mirror to Gamma factory M1, this deformation should disappear.

Manar Amer wrote:

Finished the alignment of the SBox with M1 of SBox placed, to be changed to gamma factory mirror later

Observed what seems to be transmission through M2 mirror, to be confirmed. image attached shows it

a PowerPoint attached showing the details of today's work.

 

 

Manar Amer wrote:
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

 

 

Attachment 1: Beating_TM00_SBox_damaged_M1.jpg
Beating_TM00_SBox_damaged_M1.jpg
Attachment 2: Beating_HigherOrder_SBox_damaged_M1.jpg
Beating_HigherOrder_SBox_damaged_M1.jpg
  84   Mon May 9 19:05:35 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

Today; replaced the M1 SBox mirror with M1 from gamma factory , following the previous alignment.

Removed all mirrors and cleaned them one by one.

Started with no M1, placed M2 did its alignment through M3 pinhole, then placed M3 aligned it through M4 pinhole, then placed M4 aligned it through M1 pinhole.

Placed M1 Gamma factory at the end

Note: mirror is new was not used before, took image on microscope there was no dust in the center, but there was little on the edges (cleaned it using spin-coater)

beating of fundamental mode observed (at M2 transmission) immediately with very higher order modes, slight adjustment on M4 suppressed them.

The beam is not circular and still has some deformation, M1 was changed, so it could be a damaged point at one of the other mirrors (to be investigated tomorrow)!!

No significant damage seen on the other mirrors.

 

 

Manar Amer wrote:

Last week In FP cavity we had all SBox mirrors including the damaged M1 mirror.

The beating was observed with just a slight adjustment on M4 alignment.

In addition, it was also observed while driving the piezo of the CW laser.

In the attached image of the beam at transmission point at M2 mirror, the shape of the beam is deformed.

one of the factors could be related to the damaged spot on M1 mirror. The deformation position can also be seen in the higher order modes.

after changing the injection mirror to Gamma factory M1, this deformation should disappear.

Manar Amer wrote:

Finished the alignment of the SBox with M1 of SBox placed, to be changed to gamma factory mirror later

Observed what seems to be transmission through M2 mirror, to be confirmed. image attached shows it

a PowerPoint attached showing the details of today's work.

 

 

Manar Amer wrote:
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

 

 

 

Attachment 1: M2_alignment_M3_pinhole.jpg
M2_alignment_M3_pinhole.jpg
Attachment 2: M3_alignment_M4_pinhole.jpg
M3_alignment_M4_pinhole.jpg
Attachment 3: M4_alignment_M1_pinhole.jpg
M4_alignment_M1_pinhole.jpg
Attachment 4: Beating_TM00_CW_piezo_drive.jpg
Beating_TM00_CW_piezo_drive.jpg
  85   Mon May 9 19:27:35 2022 Manar AmerFixedreportlasers and optics | utilitiesOptical roomSpincoater - Mirrror cleaning

PowerPoint shows the detailed images of all the mirrors before and after cleaning.

Images of SBox mirrors and M1 from Gamma factory attached only for the front after cleaning

the lines showing are on the back of the mirrors, they come from the spinning mount the mirrors are fixed on in the spin-coater.

Naming code :

               M#BB - M# mirror Back Before cleaning

               M#BA - M# mirror Back After cleaning

               M#FB - M# mirror Front Before cleaning

               M#FA - M# mirror Front After cleaning

               G : Gamma factory

 

Manar Amer wrote:

Images of the SBox mirrors cleaned at the same time when the cavity was aligned.

Directly from the spincoater to the SBox directly.

PowerPoint attached shows details

Manar Amer wrote:

Used the spincoater and pure water to clean dielectric mirrors which will be installed inside the SBox for directing the reflection line.

The method : place pure alchole on mirror - wipe it using white sterile tissue (in one direction) - clean with pure water in spincoater suing cotton swap .

After multible rounds, the result in the images was obtained. there is still a significant amount of dust particles (not all could be dust!!).

Not sure if the reason is the cleaning method ?? will invistegate it more !!!

Manar Amer wrote:

In preparation for cleaning the mirrors.

The spincoater has been moved with its bump into the sas of the SBox.

The pure water used will be provided by Alice, as the previous source in the workshop is not available.

 

 

 

 

Attachment 1: 20220509_Images_SBox_Mirrors.pptx
Attachment 2: M1GFA.jpg
M1GFA.jpg
Attachment 3: M2FA.jpg
M2FA.jpg
Attachment 4: M3FA.jpg
M3FA.jpg
Attachment 5: M4FA.jpg
M4FA.jpg
  86   Wed May 11 09:19:48 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

Yesterday, to find the source of why the mode is deformed, I tried to rotate the mirrors to see if there is a change.

I started with M2 and rotated it by ~ 90 degrees, there is a change in the shape of the beam.(image attached)

Will see if changing M2 to the spare M2 from SBox can fix the issue.

At the end of the day, alignment was lost.

will be redone today while changing M2 mirror.

Manar Amer wrote:

Today; replaced the M1 SBox mirror with M1 from gamma factory , following the previous alignment.

Removed all mirrors and cleaned them one by one.

Started with no M1, placed M2 did its alignment through M3 pinhole, then placed M3 aligned it through M4 pinhole, then placed M4 aligned it through M1 pinhole.

Placed M1 Gamma factory at the end

Note: mirror is new was not used before, took image on microscope there was no dust in the center, but there was little on the edges (cleaned it using spin-coater)

beating of fundamental mode observed (at M2 transmission) immediately with very higher order modes, slight adjustment on M4 suppressed them.

The beam is not circular and still has some deformation, M1 was changed, so it could be a damaged point at one of the other mirrors (to be investigated tomorrow)!!

No significant damage seen on the other mirrors.

 

 

Manar Amer wrote:

Last week In FP cavity we had all SBox mirrors including the damaged M1 mirror.

The beating was observed with just a slight adjustment on M4 alignment.

In addition, it was also observed while driving the piezo of the CW laser.

In the attached image of the beam at transmission point at M2 mirror, the shape of the beam is deformed.

one of the factors could be related to the damaged spot on M1 mirror. The deformation position can also be seen in the higher order modes.

after changing the injection mirror to Gamma factory M1, this deformation should disappear.

Manar Amer wrote:

Finished the alignment of the SBox with M1 of SBox placed, to be changed to gamma factory mirror later

Observed what seems to be transmission through M2 mirror, to be confirmed. image attached shows it

a PowerPoint attached showing the details of today's work.

 

 

Manar Amer wrote:
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

 

 

 

 

Attachment 1: Beating00_after_rotatingM2_2.jpg
Beating00_after_rotatingM2_2.jpg
  87   Wed May 11 20:51:28 2022 Manar AmerFixedreportlasers and optics | utilitiesOptical roomSpincoater - Mirrror cleaning

Following the observation that when rotating M2 plan mirror of SBox the deformation position of the mode changes.

M2 was replaced with M2 spare of SBox

images of the mirror front side attached before and after cleaning, no spots at the center, but there seems to be 2 spots far from the center of the mirror.

After comparing with old images of M2 SBox spare, they do not appear to be damages, but dust !! (old image attached, taken on 10/12/2020, was never opened after)

Manar Amer wrote:

PowerPoint shows the detailed images of all the mirrors before and after cleaning.

Images of SBox mirrors and M1 from Gamma factory attached only for the front after cleaning

the lines showing are on the back of the mirrors, they come from the spinning mount the mirrors are fixed on in the spin-coater.

Naming code :

               M#BB - M# mirror Back Before cleaning

               M#BA - M# mirror Back After cleaning

               M#FB - M# mirror Front Before cleaning

               M#FA - M# mirror Front After cleaning

               G : Gamma factory

 

Manar Amer wrote:

Images of the SBox mirrors cleaned at the same time when the cavity was aligned.

Directly from the spincoater to the SBox directly.

PowerPoint attached shows details

Manar Amer wrote:

Used the spincoater and pure water to clean dielectric mirrors which will be installed inside the SBox for directing the reflection line.

The method : place pure alchole on mirror - wipe it using white sterile tissue (in one direction) - clean with pure water in spincoater suing cotton swap .

After multible rounds, the result in the images was obtained. there is still a significant amount of dust particles (not all could be dust!!).

Not sure if the reason is the cleaning method ?? will invistegate it more !!!

Manar Amer wrote:

In preparation for cleaning the mirrors.

The spincoater has been moved with its bump into the sas of the SBox.

The pure water used will be provided by Alice, as the previous source in the workshop is not available.

 

 

 

 

 

Attachment 1: M2_spare_fb_2.jpg
M2_spare_fb_2.jpg
Attachment 2: M2_spare_fa_2.jpg
M2_spare_fa_2.jpg
Attachment 3: 20201210_M2_spare_after_cleaning.tif
  88   Wed May 11 21:10:27 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

M2 Sbox changed to spare. Alignment of M2 was redone. 

With the help of Ronic , We observe fundamental mode of the cavity with higher order modes. A frequency scan on the piezo was done to find the beating frequency , the shape is elliptical with no clear deformation.

The beam seems to be hitting something, which causes a reflection seen on the beam profiler.

We started a slow scan to improvement of the outside alignment.

At the end of the day, I can only get the TM01 and TM10 modes,

alignment to be continued tomorrow + the scan offset

Note: observation on TM01 and TM10 modes, we clearly see that the beam position is different from the mode center position !!

 

Manar Amer wrote:

Yesterday, to find the source of why the mode is deformed, I tried to rotate the mirrors to see if there is a change.

I started with M2 and rotated it by ~ 90 degrees, there is a change in the shape of the beam.(image attached)

Will see if changing M2 to the spare M2 from SBox can fix the issue.

At the end of the day, alignment was lost.

will be redone today while changing M2 mirror.

Manar Amer wrote:

Today; replaced the M1 SBox mirror with M1 from gamma factory , following the previous alignment.

Removed all mirrors and cleaned them one by one.

Started with no M1, placed M2 did its alignment through M3 pinhole, then placed M3 aligned it through M4 pinhole, then placed M4 aligned it through M1 pinhole.

Placed M1 Gamma factory at the end

Note: mirror is new was not used before, took image on microscope there was no dust in the center, but there was little on the edges (cleaned it using spin-coater)

beating of fundamental mode observed (at M2 transmission) immediately with very higher order modes, slight adjustment on M4 suppressed them.

The beam is not circular and still has some deformation, M1 was changed, so it could be a damaged point at one of the other mirrors (to be investigated tomorrow)!!

No significant damage seen on the other mirrors.

 

 

Manar Amer wrote:

Last week In FP cavity we had all SBox mirrors including the damaged M1 mirror.

The beating was observed with just a slight adjustment on M4 alignment.

In addition, it was also observed while driving the piezo of the CW laser.

In the attached image of the beam at transmission point at M2 mirror, the shape of the beam is deformed.

one of the factors could be related to the damaged spot on M1 mirror. The deformation position can also be seen in the higher order modes.

after changing the injection mirror to Gamma factory M1, this deformation should disappear.

Manar Amer wrote:

Finished the alignment of the SBox with M1 of SBox placed, to be changed to gamma factory mirror later

Observed what seems to be transmission through M2 mirror, to be confirmed. image attached shows it

a PowerPoint attached showing the details of today's work.

 

 

Manar Amer wrote:
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

 

 

 

 

 

Attachment 1: TM00_mode.jpg
TM00_mode.jpg
Attachment 2: TM01_mode.jpg
TM01_mode.jpg
Attachment 3: TM10_mode.jpg
TM10_mode.jpg
  89   Tue May 24 21:51:01 2022 Manar AmerFixedreportlasers and optics | utilitiesOptical roomSpincoater - Mirrror cleaning

While repeating the alignment, the mirrors were cleaned using the spin-coater and placed again.

The current mirrors placed in the SBox are

M1 plane injection from Gamma factory

M2 plane from SBox spare

M3/M4 spherical from SBox main set

no visible damage seen on the mirrors with the microscope

no images of the surfaces were taken due to time constraint.

Manar Amer wrote:

Following the observation that when rotating M2 plan mirror of SBox the deformation position of the mode changes.

M2 was replaced with M2 spare of SBox

images of the mirror front side attached before and after cleaning, no spots at the center, but there seems to be 2 spots far from the center of the mirror.

After comparing with old images of M2 SBox spare, they do not appear to be damages, but dust !! (old image attached, taken on 10/12/2020, was never opened after)

Manar Amer wrote:

PowerPoint shows the detailed images of all the mirrors before and after cleaning.

Images of SBox mirrors and M1 from Gamma factory attached only for the front after cleaning

the lines showing are on the back of the mirrors, they come from the spinning mount the mirrors are fixed on in the spin-coater.

Naming code :

               M#BB - M# mirror Back Before cleaning

               M#BA - M# mirror Back After cleaning

               M#FB - M# mirror Front Before cleaning

               M#FA - M# mirror Front After cleaning

               G : Gamma factory

 

Manar Amer wrote:

Images of the SBox mirrors cleaned at the same time when the cavity was aligned.

Directly from the spincoater to the SBox directly.

PowerPoint attached shows details

Manar Amer wrote:

Used the spincoater and pure water to clean dielectric mirrors which will be installed inside the SBox for directing the reflection line.

The method : place pure alchole on mirror - wipe it using white sterile tissue (in one direction) - clean with pure water in spincoater suing cotton swap .

After multible rounds, the result in the images was obtained. there is still a significant amount of dust particles (not all could be dust!!).

Not sure if the reason is the cleaning method ?? will invistegate it more !!!

Manar Amer wrote:

In preparation for cleaning the mirrors.

The spincoater has been moved with its bump into the sas of the SBox.

The pure water used will be provided by Alice, as the previous source in the workshop is not available.

 

 

 

 

 

 

  90   Tue May 24 22:05:48 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

The external and internal alignment was redone.

External Alignment : the collimator position moved so in order to have a good start I placed 2 metallic mirrors and did a rough retrace of it's line through the cavity box.

It helped as I did not have to do a lot of adjustment on Ma1 and Ma2 position (new position attached)

 

Internal Alignment: I replaced the 2 pinholes inside the cavity box with another set where the holes are more symmetric, it shows in the alignment.

Images attached of the beam diffraction pattern after each mirror pinhole.

The injection mirror was placed at the end of the alignment.

Alignment Done !!!!!!!!!!!!!

 

Manar Amer wrote:

M2 Sbox changed to spare. Alignment of M2 was redone. 

With the help of Ronic , We observe fundamental mode of the cavity with higher order modes. A frequency scan on the piezo was done to find the beating frequency , the shape is elliptical with no clear deformation.

The beam seems to be hitting something, which causes a reflection seen on the beam profiler.

We started a slow scan to improvement of the outside alignment.

At the end of the day, I can only get the TM01 and TM10 modes,

alignment to be continued tomorrow + the scan offset

Note: observation on TM01 and TM10 modes, we clearly see that the beam position is different from the mode center position !!

 

Manar Amer wrote:

Yesterday, to find the source of why the mode is deformed, I tried to rotate the mirrors to see if there is a change.

I started with M2 and rotated it by ~ 90 degrees, there is a change in the shape of the beam.(image attached)

Will see if changing M2 to the spare M2 from SBox can fix the issue.

At the end of the day, alignment was lost.

will be redone today while changing M2 mirror.

Manar Amer wrote:

Today; replaced the M1 SBox mirror with M1 from gamma factory , following the previous alignment.

Removed all mirrors and cleaned them one by one.

Started with no M1, placed M2 did its alignment through M3 pinhole, then placed M3 aligned it through M4 pinhole, then placed M4 aligned it through M1 pinhole.

Placed M1 Gamma factory at the end

Note: mirror is new was not used before, took image on microscope there was no dust in the center, but there was little on the edges (cleaned it using spin-coater)

beating of fundamental mode observed (at M2 transmission) immediately with very higher order modes, slight adjustment on M4 suppressed them.

The beam is not circular and still has some deformation, M1 was changed, so it could be a damaged point at one of the other mirrors (to be investigated tomorrow)!!

No significant damage seen on the other mirrors.

 

 

Manar Amer wrote:

Last week In FP cavity we had all SBox mirrors including the damaged M1 mirror.

The beating was observed with just a slight adjustment on M4 alignment.

In addition, it was also observed while driving the piezo of the CW laser.

In the attached image of the beam at transmission point at M2 mirror, the shape of the beam is deformed.

one of the factors could be related to the damaged spot on M1 mirror. The deformation position can also be seen in the higher order modes.

after changing the injection mirror to Gamma factory M1, this deformation should disappear.

Manar Amer wrote:

Finished the alignment of the SBox with M1 of SBox placed, to be changed to gamma factory mirror later

Observed what seems to be transmission through M2 mirror, to be confirmed. image attached shows it

a PowerPoint attached showing the details of today's work.

 

 

Manar Amer wrote:
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

 

 

 

 

 

 

Attachment 1: Ma1Ma2Horizontal.jpg
Ma1Ma2Horizontal.jpg
Attachment 2: Ma1Ma2Vertical.jpg
Ma1Ma2Vertical.jpg
Attachment 3: M1_pinhole.bmp
Attachment 4: M2_pinhole.bmp
Attachment 5: M1_M2_pinhole.bmp
Attachment 6: M3_pinhole.bmp
Attachment 7: M4_pinhole.bmp
  91   Tue May 24 22:11:40 2022 Manar AmerFixedreportlasers and opticsOptical roomFundamental Mode TEM00

The current setup for the SBox has a beam viewer and 2 photodiodes placed at reflection and transmission.

The CW laser piezoelectric crystal was driven using the lase lock system to observe the 00 mode.

Mode Observed and has no reflection (from what was previously seen)

higher order modes are resonating with 00 mode , example 11 mode (shown in image attached)

the coupling is yet to be seen and measured.

Attachment 1: 20220524_SBox_Setup.png
20220524_SBox_Setup.png
Attachment 2: 00mode_CW_piezo_drive.bmp
Attachment 3: 00mode_11mode.bmp
  92   Wed May 25 17:34:29 2022 Manar AmerFixedreportlasers and opticsOptical roomFundamental Mode TEM00

Adding images from the transmission after the spherical mirrors M3 and M4 window for a resonating mode inside the cavity.

There appears to be some misalignment from the center of the mirror ???

 

Manar Amer wrote:

The current setup for the SBox has a beam viewer and 2 photodiodes placed at reflection and transmission.

The CW laser piezoelectric crystal was driven using the lase lock system to observe the 00 mode.

Mode Observed and has no reflection (from what was previously seen)

higher order modes are resonating with 00 mode , example 11 mode (shown in image attached)

the coupling is yet to be seen and measured.

 

Attachment 1: beamviwer_transmission_at_M3.bmp
Attachment 2: beamviwer_transmission_at_M4.bmp
  93   Wed Jun 8 10:47:45 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

Closing Series for Alignment.

Beating observed, beam is in the locking stages

 

Manar Amer wrote:

The external and internal alignment was redone.

External Alignment : the collimator position moved so in order to have a good start I placed 2 metallic mirrors and did a rough retrace of it's line through the cavity box.

It helped as I did not have to do a lot of adjustment on Ma1 and Ma2 position (new position attached)

 

Internal Alignment: I replaced the 2 pinholes inside the cavity box with another set where the holes are more symmetric, it shows in the alignment.

Images attached of the beam diffraction pattern after each mirror pinhole.

The injection mirror was placed at the end of the alignment.

Alignment Done !!!!!!!!!!!!!

 

Manar Amer wrote:

M2 Sbox changed to spare. Alignment of M2 was redone. 

With the help of Ronic , We observe fundamental mode of the cavity with higher order modes. A frequency scan on the piezo was done to find the beating frequency , the shape is elliptical with no clear deformation.

The beam seems to be hitting something, which causes a reflection seen on the beam profiler.

We started a slow scan to improvement of the outside alignment.

At the end of the day, I can only get the TM01 and TM10 modes,

alignment to be continued tomorrow + the scan offset

Note: observation on TM01 and TM10 modes, we clearly see that the beam position is different from the mode center position !!

 

Manar Amer wrote:

Yesterday, to find the source of why the mode is deformed, I tried to rotate the mirrors to see if there is a change.

I started with M2 and rotated it by ~ 90 degrees, there is a change in the shape of the beam.(image attached)

Will see if changing M2 to the spare M2 from SBox can fix the issue.

At the end of the day, alignment was lost.

will be redone today while changing M2 mirror.

Manar Amer wrote:

Today; replaced the M1 SBox mirror with M1 from gamma factory , following the previous alignment.

Removed all mirrors and cleaned them one by one.

Started with no M1, placed M2 did its alignment through M3 pinhole, then placed M3 aligned it through M4 pinhole, then placed M4 aligned it through M1 pinhole.

Placed M1 Gamma factory at the end

Note: mirror is new was not used before, took image on microscope there was no dust in the center, but there was little on the edges (cleaned it using spin-coater)

beating of fundamental mode observed (at M2 transmission) immediately with very higher order modes, slight adjustment on M4 suppressed them.

The beam is not circular and still has some deformation, M1 was changed, so it could be a damaged point at one of the other mirrors (to be investigated tomorrow)!!

No significant damage seen on the other mirrors.

 

 

Manar Amer wrote:

Last week In FP cavity we had all SBox mirrors including the damaged M1 mirror.

The beating was observed with just a slight adjustment on M4 alignment.

In addition, it was also observed while driving the piezo of the CW laser.

In the attached image of the beam at transmission point at M2 mirror, the shape of the beam is deformed.

one of the factors could be related to the damaged spot on M1 mirror. The deformation position can also be seen in the higher order modes.

after changing the injection mirror to Gamma factory M1, this deformation should disappear.

Manar Amer wrote:

Finished the alignment of the SBox with M1 of SBox placed, to be changed to gamma factory mirror later

Observed what seems to be transmission through M2 mirror, to be confirmed. image attached shows it

a PowerPoint attached showing the details of today's work.

 

 

Manar Amer wrote:
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

 

 

 

 

 

 

 

ELOG V3.1.4-395e101