HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports., Page 12 of 13  Not logged in ELOG logo
ID Date Authordown Status Type Category Location Title
  51   Fri Jan 24 17:58:45 2020 Loïc AmoudryFixedinfolasers and opticsOptical roomCFBG removed from ThomX CELIA amplifier

The CFBG has been removed successfully and the amplifier is closed now.

The soldering machine had a calibration trouble and was sent to the company for review during holidays.

The output power has been checked and is the same as before (tested until 30W ouput power).

  52   Wed Jan 29 19:00:02 2020 Loïc AmoudryFixedreportlasers and opticsOptical roomGlow discharge cleaner on SBOX mirrors

The glow discharge cleaner has been tested on the SBOX mirrors.

I've put them 1 by 1. Each run was 15min long at 15mA. The mirrors HR face was always away from the electrode and ~35° angle with the support. These values have been choosen thanks to the reference:

We have learned that Air can be compared to Azote and Azote to Oxygen in glow discharge. The main difference is H2O in air which make the glow discharge less stable as say several papers.

 

Attachment 1: glow.PNG
glow.PNG
  53   Wed Jan 29 19:04:38 2020 Loïc AmoudryFixedreportlasers and opticsOptical roomSBOX mirrors cleaned

Today, SBOX mirrors have been cleaned with spin coater on HR face and isoprop on back face. They show similar spots as before.

They also have been installed in the SBOX. The M1 mount has been displaced so the beam doesn't go in its center (spot).

  54   Wed Feb 5 14:47:16 2020 Loïc AmoudryFixedreportlasers and opticsOptical roomSBOX mirrors cleaned

Yesterday, after installation and alignment of mirrors and cavity, the Finesse has been measured with cavity at air pressure, with Koheras and modulation technique.
the measurement has been done 7 times with quite different fits for the Finesse : 21.5k, 21.5k, 23.7k, 21.3k, 22.3k, 22k, 21.5k

But, as the cavity is at air pressure, the lock is not very stable.
we will pump the cavity and make the measurement again.

 

Loïc Amoudry wrote:

Today, SBOX mirrors have been cleaned with spin coater on HR face and isoprop on back face. They show similar spots as before.

They also have been installed in the SBOX. The M1 mount has been displaced so the beam doesn't go in its center (spot).

 

  55   Wed Feb 5 16:46:01 2020 Loïc AmoudryFixedreportlasers and opticsOptical roomSBOX mirrors cleaned

Today we put the cavity to vacuum (~ 0.1 mBar) and we measured again the Finesse with Koheras and modulation technique.
the measurement has been done 3 times with Finesse of 23.2k, 23.8k, 23k

Loïc Amoudry wrote:

Yesterday, after installation and alignment of mirrors and cavity, the Finesse has been measured with cavity at air pressure, with Koheras and modulation technique.
the measurement has been done 5 times with quite different fits for the Finesse : 21.5k, 21.5k, 23.7k, 21.3k, 22.3k.

But, as the cavity is at air pressure, the lock is not very stable.
we will pump the cavity and make the measurement again.

 

Loïc Amoudry wrote:

Today, SBOX mirrors have been cleaned with spin coater on HR face and isoprop on back face. They show similar spots as before.

They also have been installed in the SBOX. The M1 mount has been displaced so the beam doesn't go in its center (spot).

 

 

  57   Thu Feb 6 16:58:57 2020 Loïc AmoudryFixedreportlasers and opticsOptical roomPower up to 48kW crater effect again

As we said, after glow discharge and spin coater we recovered a low power finesse ~23000 (aligning away from the M1 crater).

Today, we have made some power up to 48kW (3A on 3rd stage). At this power, there was no important effect on the transmission which can be compared with the coupling (see image "3A_136mWtrans"). We waited ~20min without aligning and the power dropped slowly (misalignment) to 39kW. 

Then we increased the amplifier power to 4A and the power went up to 44kW. At this moment, the strange transmission behavior that we observed before mirror crater, appeared again (see image "4A_126mWtrans"). We also observed a mode deformation with the camera, see image "mode_strange". This shape was not depend of the camera/filter rotation angle and still appear with another camera.

Loïc Amoudry wrote:

The mirrors went in the cavity the 28th of november (We did several power up to 30kW stored and only one to 40kW then the power went down to 2kW during the run).

Microscope study shows that mirrors get some dust during the handling [travel from microscope to SBOX --> installation --> in SBOX for +1month and power up --> travel to microscope].

Almost all of these dusts can be removed with cleaning.

There is only one important difference between 28th of november and today, a large spot on M1

 

Attachment 1: 3A_136mWtrans.png
3A_136mWtrans.png
Attachment 2: 4A_126mWtrans.png
4A_126mWtrans.png
Attachment 3: mode_strange.png
mode_strange.png
  60   Wed Dec 16 12:36:37 2020 Loïc AmoudryFixedreportlasers and opticsOptical roomMirror's cleaning and microscope imaging

On 10th of december 2020 we cleaned the SBOX mirrors and took microscope images (the name of the images indicates what they are).

There are 7 mirror, the initial M1 (spot in the center), M2 (spot on the edge), M3 and M4 which made the 200-400kW and the M2, M3 and M4 SPARE. The difference we make between M3 and M4 SPARE is the number on the box (11 or 13).

We used 3 different cleaning methods : 1st, one spin coater on HR, 2nd one, tissu wipe on AR (wipe with the optical tissu and isoprop) or 3rd one, mirror wiped on tissue (put isoprop on tissu and press AR face of the mirror doing "8" shape 3 times).

The second method is far les efficient as a cleaning method. The image "M3_M4_spare_11_after_cleaning_back.tif " shows the traces let by it and removed by the 3rd method on image "M3_M4_spare_11_after_cleaning_back_second_time_on_tissu.tif".

We can also notice that the spin coater let some trace on the HR face, round shaped, see Image "M3_M4_spare_13_after_cleaning_back.tif". We can propose to use the third method with Acetone on HR face before using spin coater to remove oil or organic particles.

It also lets a trace on the AR face, this is why we clean the AR face with the 2nd method after cleaning it with the spin coater.

 

Note : The position of the mirrors in the microscope is always the same here. Meaning mirrors are directed so that the arrow (which shows the HR face and is placed on the side of the mirror) is placed on the top of the images.

Attachment 1: M1_before_cleaning.tif
Attachment 2: M1_after_cleaning.tif
Attachment 3: M2_after_cleaning.tif
Attachment 4: M3_after_cleaning.tif
Attachment 5: M4_after_cleaning.tif
Attachment 6: M2_spare_before_cleaning.tif
Attachment 7: M2_spare_after_cleaning.tif
Attachment 8: M2_spare_after_cleaning_back.tif
Attachment 9: M3_M4_spare_11_before_cleaning.tif
Attachment 10: M3_M4_spare_11_after_cleaning_back.tif
Attachment 11: M3_M4_spare_11_after_cleaning_back_second_time_on_tissu.tif
Attachment 12: M3_M4_spare_13_before_cleaning.tif
Attachment 13: M3_M4_spare_13_after_cleaning.tif
Attachment 14: M3_M4_spare_13_after_cleaning_back.tif
  58   Wed Jun 19 18:54:00 2019 Huan WangFixedinfolasers and opticsOptical roomamplifier

Onefive output power is 24mW now, and 2.41mW(after EOM) injected into fiber,

a injection power monitor added, 99% (2.06mW)  injected into amplifier, 1% (16.1uW) monitored with photodiode DET36A/M, which gives ~500mV DC signal on oscilloscope with 1Mohm impedanc;

 

First stage amplifier works good, monitoring phtodiode gives more than 200mV DC signal with 50ohm impedance on oscilloscope (as attached photo);

 

Second stage, the old monitoring photodiode is broken, a new monitoring photodiode is connected, which we don't have reference data for it,

on the optical output port of the monitoring signal, it's written 150mW, but at where we measured 40mW.

Attachment 1: 1st_stage_amplifier_monitoring_photodiode_signal.jpg
1st_stage_amplifier_monitoring_photodiode_signal.jpg
  59   Mon Jun 24 16:15:08 2019 Huan WangFixedinfolasers and opticsOptical roomamplifier

I just had a phone call with Jérome and he told me 2 things :

* Be carefull ! the MightyLaser amplifier is not designed to work with 33MHz laser : the streching level is not sufficient !
One could worsen the phase noise by self-modulation due to peak power or even distroy the amplifier !
One should use it only at low power !!!

* He thinks we should more or less find back the same DC levels than before even with lower seeding power and lower repetition rate.
He thinks we should look at the optical spectrum to check if we don't have some ASE in the 1st stage and 2nd stage signal !
We can send him plots or call him to discuss these points.

Huan Wang wrote:

Onefive output power is 24mW now, and 2.41mW(after EOM) injected into fiber,

a injection power monitor added, 99% (2.06mW)  injected into amplifier, 1% (16.1uW) monitored with photodiode DET36A/M, which gives ~500mV DC signal on oscilloscope with 1Mohm impedanc;

 

First stage amplifier works good, monitoring phtodiode gives more than 200mV DC signal with 50ohm impedance on oscilloscope (as attached photo);

 

Second stage, the old monitoring photodiode is broken, a new monitoring photodiode is connected, which we don't have reference data for it,

on the optical output port of the monitoring signal, it's written 150mW, but at where we measured 40mW.

 

  121   Thu Jul 28 13:52:23 2022 Aurélien MartensFixedreportutilitiesOptical roomchiller for CELIA amplifier

The chiller of CELIA amplifier has been moved from the corridor to the changing room in front of PLIC. 

It was not functionning well (quick drop of flow below 0.9l/min). After several cleanings of the filter, shorting the water circuit a stable 2.1l/min flow was obtained. Then the water circuit was elongated to go close to the amplifier but not into it. Allowing to get about 1.3l/min stable. Now the amplifier is conencted, the filter has been cleaned again twice and the flow seem stable with 1.4l/min. New filters have been ordered and must arrive quickly. Changing it is necessary I think. 

  122   Mon Aug 8 11:11:28 2022 Aurélien MartensFixedreportutilitiesOptical roomchiller for CELIA amplifier

today, temperature of the chiller has been set to 23°C.

measured temperature in the chiller reaches 23°C after approximatively 10 minutes

Aurélien Martens wrote:

The chiller of CELIA amplifier has been moved from the corridor to the changing room in front of PLIC. 

It was not functionning well (quick drop of flow below 0.9l/min). After several cleanings of the filter, shorting the water circuit a stable 2.1l/min flow was obtained. Then the water circuit was elongated to go close to the amplifier but not into it. Allowing to get about 1.3l/min stable. Now the amplifier is conencted, the filter has been cleaned again twice and the flow seem stable with 1.4l/min. New filters have been ordered and must arrive quickly. Changing it is necessary I think. 

 

  226   Fri Nov 29 12:06:05 2024 Alice RenauxFixedinfolasers and opticsOptical roomMenhir 160MHz laser

The Menhir 160MHz has been put back in place on the CELIA amplifier setup. Its output power is measured to be 150-160mW with an attenuator as expected. Its spectrum is available in "spectre_avant_cvbg.xlsx" and "spectre_avant_cvbg.png".

The pulses are stretched by means of a CVBG. Their spectrum is available in "spectre_apres_cvbg.xlsx" and "spectre_apres_cvbg.png".

The laser is coupled into an optical fiber with an output power of 11.5mW for a 32mW input.

 

Attachment 1: spectre_avant_cvbg.xlsx
Attachment 2: spectre_apres_cvbg.xlsx
Attachment 3: spectre_avant_cvbg.png
spectre_avant_cvbg.png
Attachment 4: spectre_apres_cvbg.png
spectre_apres_cvbg.png
  227   Fri Nov 29 16:50:04 2024 Alice RenauxUnder Processinfolasers and opticsOptical roomMenhir 216MHz laser

The Menhir 216MHz laser has been put back in place on the cavity table. Its output power is measured to be 160mW with an attenuator. Its spectrum is available in "spectre.xlsx" and "spectre.png". The main wavelength is a bit shorter (1028.75nm) and the spectrum a bit narrower (4.73nm) than expected.

Attachment 1: spectre.xlsx
Attachment 2: spectre.png
spectre.png
  233   Fri Mar 21 18:27:36 2025 Alice RenauxUnder Processreportmechanics | lasers and opticsOptical room2-mirror cavity alignment

At the beginning of the week, I installed a half-wave plate, a beamsplitter cube and a quarter-wave plate, as well as the end-of-cavity mirror, which I unsuccessfully tried to align.

I removed the end-of-cavity mirror and started the laser injection over again with the two motorized mirrors placed in front of the vacuum chamber (but left Ronic's setup as it was).

Today, I aligned the laser by putting an iris on the cavity mirrors' mounts. I tested the repeatability of such a setup, both by removing and putting back the same iris and by using a different iris. The result is in the attached pictures. Is it good enough ?

The mirrors' motors positions are :

- 1 : 3.349670 mm ;

- 3 : 2.060280 mm ;

- 4 : 3.484560 mm ;

- 5 : 3.269710 mm.

Attachment 1: 2-mirror_cavity_alignment.zip
  234   Tue May 27 18:12:22 2025 Alice RenauxUnder Processreportlasers and opticsOptical room2-mirror cavity alignment

Yesterday and today, I replaced the OEWaves CW laser with the NKT CW laser. Its screen does not display anything, so it has to be operated through the GraphiK software.

I then re-aligned the cavity with a new adjustment adaptation tool between the mirror mounts and the irises.

The motor positions are :

- 1 : 3.354420 mm

- 3 : 2.128850 mm

- 4 : 3.468480 mm

- 5 : 3.157300 mm

I then connected the LaseLock module to scan the NKT laser wavelength on a roughly 0-10 V range at a 2Hz rate, so that it could match the cavity's resonance frequency.

Without optimizing the injection, I monitored the transmitted power with a photodiode paired with an amplifier. The pictures are available through this link : https://box.in2p3.fr/s/TGgwkKgYik7MyqW, and an example picture is attached. Their timestamp is in their filenames, and it seems that the transmission varies quite a lot on a 10 s scale, and these variations seem to be periodic on a 1 minute scale. The peaks seem weirdly wide, almost up to 100-200 MHz (≈FSR).

Attachment 1: IMG20250527160111.jpg
IMG20250527160111.jpg
  235   Wed May 28 10:55:15 2025 Alice RenauxUnder Processreportlasers and opticsOptical room2-mirror cavity alignment

The NKT laser PZT sensitivity is ~0.09pm/V of wavelength variation, which is equivalent to ~ 25MHz/V
So, the full range of the scan is roughly 250MHz (more than a FSR) for 10V.

It seems impossible to get such large resonances unless the Finesse is very low => let's try to change M1 by a spare GammaFactory plan or 10m ROC mirror.

Alice Renaux wrote:

Yesterday and today, I replaced the OEWaves CW laser with the NKT CW laser. Its screen does not display anything, so it has to be operated through the GraphiK software.

I then re-aligned the cavity with a new adjustment adaptation tool between the mirror mounts and the irises.

The motor positions are :

- 1 : 3.354420 mm

- 3 : 2.128850 mm

- 4 : 3.468480 mm

- 5 : 3.157300 mm

I then connected the LaseLock module to scan the NKT laser wavelength on a roughly 0-10 V range at a 2Hz rate, so that it could match the cavity's resonance frequency.

Without optimizing the injection, I monitored the transmitted power with a photodiode paired with an amplifier. The pictures are available through this link : https://box.in2p3.fr/s/TGgwkKgYik7MyqW, and an example picture is attached. Their timestamp is in their filenames, and it seems that the transmission varies quite a lot on a 10 s scale, and these variations seem to be periodic on a 1 minute scale. The peaks seem weirdly wide, almost up to 100-200 MHz (≈FSR).

 

Attachment 1: Sans_titre.jpg
Sans_titre.jpg
  236   Wed May 28 18:27:27 2025 Alice RenauxUnder Processreportlasers and opticsOptical room2-mirror cavity full setup

Today with Ronic, we changed the FP cavity's mirrors to Layertek (Gamma Factory) mirrors :

- M1 : n°161186 --> fused silica, unknown absorption, unknown diffusion, T1=2500ppm transmission planar mirror ;
- M2 : n°161182 --> fused silica, unknown absorption, unknown diffusion, T2=10ppm transmission, 5m ROC mirror.

RTL (round-trip losses) ~ 2500 + 10 ppm (we forget the unknown parameters for absorption and diffusion).
The maximal finesse we could expect is thus F=π([(1-T1)(1-T2)]^(1/4))/(1-[(1-T1)(1-2)]^(1/2)) ~  2*pi /RTL ≈ 2500 assuming no absorption and no scattering.

The FP cavity's FSR is ≈ 216 MHz given its length.

We managed to see some optical beating on the FP cavity's mirrors and to reach the fundamental transverse mode of the FP cavity by adjusting the injection mirrors, but when scanning the laser's wavelength, some higher-order modes appear and the fundamental mode is reached when the voltage applied to the piezoelectric actuator of the laser's cavity is ≈0V. The actuator is not meant to work with negative voltages, so we translated one of the FP cavity's mirrors so that the fundamental transverse mode's resonance frequency is in the middle of the voltage range.

We also removed the D-shaped mirrors, as they are only useful when working with high power.

  237   Fri Jun 13 20:10:00 2025 Alice RenauxFixedreportlasers and opticsOptical roomDamaged mirrors test

Today with Dorian we tested three mirrors with which we previously haven't been able to obtain any resonance or optical beating :

- C16111/11 : The mirror looks normal under a microscope, apart from a few inclusions and maybe a small scratch towards the edge. We tested it with a 161186 M1 mirror and we weren't able to obtain any optical beating on the cavity's mirrors or resonance.

- 161185 (1) : The mirror looks normal with the naked eye. We tested it with a 161182 M2 mirror and we were able to notice some optical beating on the cavity's mirrors as well as small resonance peaks and a higher-order transverse mode.

- 161185 (2) : The mirror shows some damage on the substrate side (not the coated side). We tested it with a 161186 M2 mirror and we were able to notice some optical beating on the cavity's mirrors as well as huge resonance peaks and a Gaussian transverse mode.

The updated recap file is available here, as well as a few pictures.

  239   Thu Aug 28 15:29:00 2025 Alice RenauxUnder Processreportlasers and opticsOptical room2-mirror cavity locking

In June, we encountered some problems regarding the transmission and error signals (see images here), looking as if the laser was switched off before the cavity was filled.

 

Aurélien, Ronic and I discussed this on 08/27, resulting in a list of tests to perform. We :

- checked the mirrors' thickness (maybe there was some mechanical stress if they were too thick). The mirrors' thickness is 6.35mm (1/4 inch) and is consistent with the mounts size ;

- checked the mirrors' mounts' screws' tightening (maybe the mirrors were either moving if the screws were not tight enough or some mechanical stress if they were too tight). We tightened the mirrors' mounts' upper screws.

Next thing we did with Ronic was check if the error signal depended on the modulation/demodulation relative phase, which was not the case, but it should have.

Ronic added a quarter waveplate before the half waveplate in the injection system.

 

Today, on 08/29, the succession of the higher-order transverse modes when scanning the seeder laser's piezoelectric actuator's voltage to scan the seeder laser's optical frequency seemed a bit strange, so we checked the resonance frequencies of several occurrences of the same transverse mode :

- (0, 0) : @ -0.4V and 6.0V ;

- (1, 0) and (0, 1) : @0.5V and 6.9V.

In both cases, several occurrences were separated by about 6.4V, which corresponds to the voltage difference to scan a full FSR. The spacing between (0, 0) and (1, 0)/(0, 1) is then about 0.14*FSR.

I checked this by writing a small piece of Python code to calculate the cavity's fundamental transverse mode and its Rayleigh length, which is displayed in Figure 1, and then by calculating the resonance frequencies wrt the (0, 0) resonance frequency for the (1, 0), (0, 1) and (1, 1) transverse modes, which is displayed in Figure 2, with the following formula: $\nu_{p, n, m}=(p+\frac{(n+\frac{1}{2})\arctan(\frac{2L_{\text{cav}}}{z_{\text{R}}})+(m+\frac{1}{2})\arctan(\frac{2L_{\text{cav}}}{z_{\text{R}}})}{2\pi})\times\text{FSR}$ for a $p$ longitudinal and ($n$, $m$) transverse mode. Here, $p=1$. The spacing betwen (0, 0) and (1, 0)/(0, 1) is about 0.11*FSR, making the previous observation consistent with the calculation. Everything seems normal.

Ronic also increased the EOM modulation voltage, increasing the modulation depth for the generation of the error signal (from 100mV RMS to 300mV RMS), making the error signal depend on the modulation/demodulation relative phase, as it should. He managed to lock the laser onto the cavity for about 1s at a time.

 

Next steps are to optimize the PID parameters and to add a low-pass filter/AOM to cut the higher frequencies off and improve the feedback loop.

Attachment 1: Figure_1.png
Figure_1.png
Attachment 2: Figure_2.png
Figure_2.png
  240   Wed Sep 3 18:10:34 2025 Alice RenauxUnder Processreportlasers and opticsOptical roomSuccessful lock (finally !)

Today, with Ronic, we managed to get a successful lock with the NKT laser by setting up a new PDH box from scratch (photodiode + amplifer + mixer).

Me measured the injected power (5.4mW) and the transmitted power (33µW) after a wedge (92% transmission) and a 7ppm transmission mirror, so the intracavity power was 5.1W. We have a 950 enhancement factor for a 3100 finesse cavity, so a nominal enhancement factor of 2500. The lock was very stable, as shown in the attached picture (yellow signal is the transmitted power, orange signal is the error signal and green signal is the voltage sent from the Laselock module to the pizeoelectric actuator of the NKT laser cavity).

Then, we added a second EOM in order to perform a finesse measurement, but we weren't able to inject more than 3mW at full laser power in the cavity or to lock the laser onto the cavity.

Attachment 1: IMG_20250903_113758.jpg
IMG_20250903_113758.jpg
ELOG V3.1.4-395e101