    Reply   Alphanov amplifier issue, posted by Ronic Chiche at ThomX igloo about lasers and optics Sécurité_et_CC_ampli_Alphanov.pdf
Today, I did a Teams meeting with Guillaume.
He didn't see any problem with the controller.
normally, the controller will be shipped to the lab tomorrow.
    Reply   Alphanov amplifier issue, posted by Ronic Chiche at ThomX igloo about lasers and optics 
Yesterday, we received the controller back from Alphanov
and I tested it immediately in the optical room.
if the software is properly talking to the controller, everything works normally.
    Reply   Alphanov amplifier issue, posted by Ronic Chiche at ThomX igloo about lasers and optics 
FYI : to access special menus on the Alphanov software,

login : "Administrator"
pwd : no password
    Reply   Alphanov amplifier issue, posted by Ronic Chiche at ThomX igloo about lasers and optics 
the present situation is :

Guillaume asked me to disconnect the "Laser head" cable from the pump diode module, installed on the optical table.
on the pump diode module, the connector has several small pins and 3 larges pins : A1, A2, A3.
    Reply   Alphanov amplifier issue, posted by Ronic Chiche at ThomX igloo about lasers and optics 
Guillaume and Julien Bour from Alphanov will come on wednesday 24/07 to have a first look and check the amplifier.

to prepare their intervention, I tested this morning the OneFive oscillator : it's still working at 33MHz.
and we have ~5mW in the LAL software without the EOM.
    Reply   Alphanov amplifier issue, posted by Ronic Chiche at ThomX igloo about lasers and optics connecteur_laser_head_sur_module_pompes_ampli.jpgconnecteur_laser_head_sur_chassis_ampli.jpg
picture of the connectors used to carry signals between the amplifier pump module and the rack under the table.
DB13W3 standard has been used to carry high current on the 3 dedicated pins:

https://en.wikipedia.org/wiki/DB13W3
Entry   Alphanov amplifier compressor's CVBG thermal images, posted by Ronic Chiche at ThomX igloo about lasers and optics Thermal_images.pdf
last Friday with Manar, we took some thermal images of the Alpnavov compressor box.

I post a small document to summarize some points :

- the 2 mirrors used to make the CVBG injection are polarized thin-film mirrors... thus sensitive to the beam polarization.
Entry   Alphanov amplifier compressor's CVBG thermal images, posted by Ronic Chiche at ThomX igloo about lasers and optics Thermal_images.pdf
last Friday with Manar, we took some thermal images of the Alpnavov compressor box.

I post a small document to summarize some points :

- the 2 mirrors used to make the CVBG injection are polarized thin-film mirrors... thus sensitive to the beam polarization.
Entry   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics 
today with Manar, we brought everything to make the profile measurement.

the chiller was in error because of the water level : we filled it in.
! warning ! there is no alarm signal, only a message on the screen.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics Tutoriel_PD_PULS.pdf2023_Alphanov_Beam_Profile_setup.jpg
yesterday morning with Manar,

1) we fixed the threshold level of the laser input signal which makes the measurement of the repetition rate.
(if the rep rate is not measured properly, the amplifier safety interlock stops immediately the amplification).
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics 
this morning with Manar,

- we checked first the chiller problem.
the fluctations of +/-2° around 25°C are still there.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software error_signal_Temperature_case_amplifier.jpg
Today with Ronic 

For the Temperature fluctuation :


we checked the valves of the tubes bumping water into the chiller from ThomX main water supply, they were
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics 
this morning with Manar,

- we fixed the chiller problem (+/-2°C fluctuation around the set temperature value = 25°C).
we opened and closed several times the water valve of the primary circuit to remove air bubles => no effect on the temperature variations.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 
Ronic on Friday operated the Amplifier from the Alphanov software, switched it off then turned on the LAL software,

and it worked for 40 min without issues or error appearing,

 
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 
the laser amplifier worked for some days but it ended to a "watchdog" alarm which switches off the amplifier... investigation is ongoing




Ronic
Chiche wrote:



Ronic on Friday operated the Amplifier from the Alphanov software,
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software scope.jpg
we have had this new issue (watchdog alarm switching off the amplifier) for several days.

Alphanov mentionned it could come from the input oscillator stability (modelock loss for example).

Today,
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 20230208_error_watchdog.jpeg
With Daniele after, we took a readings of the beam with a lens of 750 mm.

From the observed data, we can state that the beam is not completely Gaussian M2 > 1, due to the variation of the ellipticity of the beam
along the focused path 
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 
This morning,

- with the "Alphanov" software => Central panel => "system" button => Watchdog "Period Max" was at 100ns
and has been set to 200ns.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 
This morning with Daniele, we did a second measurement of the beam profile using a different lens of focal 400 mm.

at a power amplification of 20%

the amplifier stayed on for around one hour and a half with no errors. 
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software Analyse_Données_20_fevrier.pptx
On Monday 20th of February, Daniele and Viktor took some beam profile measurement data @ 50% of power amplification.

data and analysis can be found in the attached file.




Ronic