    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 

Yesterday, Guillaume Machinet and Vincent Clet from Alphanov came to fixe some issues with the laser amplifier:

- 1) they fixed the "rep. rate" measurement issue.
previously, the voltage range in which it was possible to detect the correct rep rate (33MHz) was very narrow ~ 50mV.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software WhatsApp_Image_2023-02-27_at_6.07.00_PM.jpeg
This morning with Victor we took a reading of the beam profile using a 400 mm lens at 70 % and 100 % amplification.

Will process the data and add after

Note : after an hour and half of amplifier on at 70% we increased the power to 100%, and after ~ 10 minutes the amplifier turned off and showed
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 20230308_124313.jpg
this morning with Manar,
we checked the Alphanov amplifier issue : "The laser has been stopped because its power is too far from the nominal power".

the photodiode voltage measuring the output power has drifted again and then, the estimated output power was wrong, triggering an alarm issue.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software Beam_wavefront_30cm_no_Lens.jpg
We took reading of the beam profile of the amplifier without a lens @ 30 cm from the 2nd wedge for amplification percentages

20-30-40-50-60-70-80-90-100 % another additional 3 readings will be taken at different locations for all amplification percentages.

 
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 
due to several "case temperature" issues, I change the "Max case temperature" in the "general config" of the "Alphanov
Software" from 50°C to 55°C.
=> validated by Alphanov.
Entry   Alphanov amplifier beam profile Data , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 8x
Adding the schematic of the measurements.,

the 4° Beam Deviation, AR Coating ( B coating)

The 3 data points taken of the beam profile are fitted in Gaussian Beam with taking the beam waist at the output of the compressor box.
    Reply   Alphanov amplifier beam profile Data , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 
This Morning with Victor the amplifier was turned on @ 20% amplification (output ~ 9 W ) for approximately 3 hours and no error appeared.

I took data of the beam profile at different points without a lens after the 2 wedges 

@ 0% (only pump beam), output ~ 300 mW (taken from previous data) , output after 2 wedges 2.2 uW  
Entry   Alphanov amplifier Beam Pulse Duration, posted by Manar Amer at ThomX igloo about lasers and optics pulse-check_compressed.pdf
The pulse duration from the amplifier compressor output will be measured using PulseCheck autocorelator

a first alignment of the device was done, continuation in the afternoon.

YouTube link attached of the general operation of pulse check < 5 min :  https://www.youtube.com/watch?v=J1pNHYySSYg
    Reply   Alphanov amplifier Beam Pulse Duration, posted by Manar Amer at ThomX igloo about lasers and optics 20220401_Alphanov_Pulse_Duration.jpg20220408_AMplifier_133MHz_Pulse_duration.jpg20220408_AMplifier_133MHz_Pulse_duration_better_alignment.jpg
Measurement of the pulse was done at the end of the afternoon, Pulse Duration FWHM = 12.9 ps

After a long process of adjusting the alignment and angle of the crystal inside the device

The power input on the PulseCheck device should be above the sensitivity, in our case we had 1 mW input.
    Reply   Alphanov amplifier Beam Pulse Duration, posted by Manar Amer at ThomX igloo about lasers and optics 
Pulse Duration FWHM = 12.9 ps




Manar
Amer wrote:



Measurement of the pulse was done at the end of the afternoon, Pulse
Entry   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 8x
Today: Manar, Ronic and Aurlien started the study of the beam profile of the alphanov amplifier at high power.

The setup shown in the image shows how the power is reduced by using Anti-reflective mirrors,

pick up 1 : Anti-reflective coating on both sides , pick up 2 : Anti-reflective coating on one side and High
    Reply   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 
 




Manar
Amer wrote:



Today: Manar, Ronic and Aurlien started the study of the beam profile
    Reply   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 6x
Adding the fitted beam in 1D in Horizontal and vertical axis.

for 10%   to  60%




Manar
    Reply   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 
Today a second reading of the beam profile was done , with Victor ,Ronic, Aurlien and Manar

an additional set-up was made using two 4 deg wedges instead of High reflecting (HR) and Anti-reflecting (AR) mirrors.

Two reading were taken:
    Reply   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 9x
Readings of the beam shape and horizontal (upper plot) and vertical (lower plot) axis fit at the direct output of the amplifier.

The beam shape relatively shaped like an egg.




Manar
    Reply   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 20220401_Alphanov_Beam_setup_after_compressor.jpg20220401_Amplifier_AfterCompressor_50perc_133MHz.bmp20220401_Amplifier_AfterCompressor_60perc_133MHz.bmp20220401_Amplifier_AfterCompressor_50perc_133MHz_BeamFit.png20220401_Amplifier_AfterCompressor_60perc_133MHz_BeamFit.png
Readings of the beam shape and fit at the output of the compressor

differance from previous reading is that here we used wedges instead of HR mirrors.

Similar to previous readings, the beam shape deforms and the Gaussian fit is bad after 50% amplification
    Reply   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 6x
Adding the readings  20% , 30% ,  40%  for output of amplifier after the compressor

 




Manar
    Reply   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 12x
This Morning, another reading was taken for the output of the amplifier after the compressor using 2 wedge mirrors.

In addition, a start of the calibration of a PulseCheck "Autocorrelator" to measure the pulse duration. 

note, the pulsed laser used is OneFive 133.33 MHz
    Reply   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 20x
Victor, Ronic, Manar

Today  was dedicated to the study of the CVBG, and it's effect on the beam profile

we opened the box containing the 2 CVBGs : The 2CVBG are glued from the bottom on a copper surface .
    Reply   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 8x
For better comparison ,

the measurements done previously with 2 CVBGs was done at ~ 30 cm and the one done on Friday was for 1 CVBG is at 1 m

so we did additional measurement for 2 CVBGs at 1 m, for 20 , 50 and 70 % amplification 