HOME SBOX THOMX MINICAV Utilities
THOMX orders THOMX installation THOMX commissioning THOMX control command
  Status of commissioning, report also here plots are reports.  Not logged in ELOG logo
Message ID: 150     Entry time: Tue Apr 26 11:15:23 2022
 Author: Ronic Chiche 
 Status: Fixed 
 Type: issue 
 Category: lasers and optics 
 Location: ThomX igloo 
 Title: Alphanov amplifier compressor's CVBG thermal images 

last Friday with Manar, we took some thermal images of the Alpnavov compressor box.

I post a small document to summarize some points :

- the 2 mirrors used to make the CVBG injection are polarized thin-film mirrors... thus sensitive to the beam polarization.
there is also a half-wave plate on the input path...
are we sure that the beam polarization is properly set with the Onefive laser injecting the amplifier?
(Alphanov used their own seeder).

- the CVBG temperature seems normal without too much difference between profiles and values.

- the 2 thin-film mirrors seem quite hot (45°) compared to the rest of the setup.
a basic calculation of power thermally radiated gives an absorbed power of around 55mW which is equivalent to about 500ppm of absorption.
is this estimation correct? could it be the confirmation of some misalignment that traps some power inside the mirror substrate?
(document corrected to take into account the room temperature)

Attachment 1: Thermal_images.pdf  373 kB  Uploaded Wed Apr 27 11:12:16 2022
ELOG V3.1.4-395e101