Text
5
Wed Oct 3 11:28:46 2018
Ronic Chiche
Fixed
report
mechanics | cabling
Optical room
test des vérins ISP 20mm et 6mm
Après avoir utilisé la chaine
: driver ISP - cable ISP DB9 - Feedthrough
enceinte DB9-DB9 - cable perlé DB9-SUB-C
6
Fri Oct 5 10:17:30 2018
Ronic Chiche
Fixed
report
mechanics | cabling
Optical room
test des vérins ISP 20mm et 6mm
12
Wed Jan 16 14:54:14 2019
Loïc Amoudry
Fixed
report
lasers and optics | utilities
Optical room
Cleaning outside IP cavity
From 7th january to 11th january.
13
Fri Feb 1 13:20:47 2019
Loïc Amoudry
Fixed
report
utilities
Optical room
End of cleaning, plastic fixed on cavity
08/02/19 - End of the cleaning process.
14
Fri Feb 1 13:39:21 2019
Loïc Amoudry
Fixed
report
lasers and optics
Optical room
Mirrors checked with binocular microscope
29/01/19 - ThomX mirrors have also been
checked with binocular microscope. They shown
the same spots as S-BOX mirrors.
15
Fri Feb 1 14:10:25 2019
Loïc Amoudry
Fixed
report
lasers and optics
Other
Mirrors sent to LMA
30/01/19 - As LMA people came, we discussed
about the spots on ThomX mirrors which should
be clean. As mentionned in "SBOX commissioning",
16
Thu Feb 14 18:20:52 2019
Loïc Amoudry
Fixed
report
mechanics | lasers and optics
Optical room
Basic alignment & alignment motors check
14/02/19: The 4 motors but Mby control
works. Mby motor home does not bring
it to its "middle" position
18
Tue Mar 19 16:35:32 2019
Loïc Amoudry
Fixed
report
lasers and optics
Optical room
Beam size and finesse
FSR initiale : 33.34 MHz (S3 = -825000
et S4 = -825000)
Beam size initiale : wx=2.05mm
20
Fri Mar 22 17:53:10 2019
Loïc Amoudry
Fixed
report
lasers and optics | vacuum
Optical room
Cavity at atm pressure
Dry air injected into the cavity.
Lock has been proceeded without
any trouble. ~~4000 finesse. Considered to
21
Tue Mar 26 18:26:32 2019
Loïc Amoudry
Fixed
report
lasers and optics
Optical room
Check of mirrors and coating directions
Direction of HR face of P1 and P4 have
been checked, there was no problem. The mirrors
are in the good direction and the coating
22
Tue Mar 26 18:32:40 2019
Loïc Amoudry
Fixed
report
mechanics | lasers and optics | cabling
Optical room
Wires in laser path
There was some ceramic wires close to the
beem path right before P4. We moved them
away (we can imagine they could move again
25
Tue Apr 2 17:59:18 2019
Loïc Amoudry
Fixed
report
lasers and optics
Optical room
Finesse after removing wires
19/04/01: We removed the wires from the
beam path in the cavity. Finesse stayed at
3400. FSR: 33.342 MHz.
26
Tue Apr 2 18:03:31 2019
Loïc Amoudry
Fixed
report
lasers and optics
Optical room
Transmission of the mirrors
We openned the cavity, removed S2 and measured
trans of P1 and same for S2.
Pin ~= 3.91 mW
28
Thu Apr 4 09:42:12 2019
Loïc Amoudry
Fixed
report
lasers and optics
Optical room
Transmission of the mirrors
After verification, transmission of P1
correspond to ~150ppm.
The 1337ppm come from a beam leakage.
29
Fri Apr 5 11:18:47 2019
Loïc Amoudry
Fixed
report
lasers and optics
Optical room
Transmission of the mirrors
P1 transmission out of the mount (means
angle of the mirror is not the real angle):
Pin = 3.2 mW
30
Fri Apr 5 11:28:39 2019
Loïc Amoudry
Fixed
report
lasers and optics
Optical room
Finesse with P1 SUPRASIL
P1 SUPRASIL has been installed instead
of P1 Sapphire. It has been cleaned 3 times
on each face with wipe+isopropanol before
31
Mon Apr 8 10:18:31 2019
Loïc Amoudry
Fixed
report
lasers and optics
Optical room
Finesse after 2days
Finesse measured this morning after P1
SUPRASIL being into the cavity for two
days. In order to check if it's getting
32
Mon Apr 8 11:55:40 2019
Loïc Amoudry
Fixed
report
lasers and optics
Optical room
P4 cleaned and finesse measurement
P4 cleaned with isopropanol.
Finesse measured afterward: ~=13
000 (clear increase)
33
Wed Apr 10 13:11:46 2019
Loïc Amoudry
Fixed
report
lasers and optics
Optical room
Mirror's cleaning and finesse
Summary of what have been cleaned
the last week:
P1 with isopropanol, finesse 11000
36
Tue May 21 17:44:46 2019
Loïc Amoudry
Fixed
report
lasers and optics
Optical room
New Suprasil mirror from LMA
14/05/19:
New Suprasil mirror received. It
will replace the S-BOX suprasil mirror on