HOME SBOX THOMX MINICAV Utilities
THOMX orders THOMX installation THOMX commissioning THOMX control command
  Status of commissioning, report also here plots are reports., Page 15 of 24  Not logged in ELOG logo
ID Date Author Status Type Categorydown Location Title
  43   Tue Jun 18 20:44:34 2019 Huan WangFixedinfolasers and opticsOptical roomtelescope for Koheras

telescope of Koheras is removed, information record here:

- distance between collimator edge and -50 lens mount edge is 14.7cm,

- distance between another edge of -50 lens mount and +250 lens mount edge is 14.3cm,

- distance between another edge of +250 lens mount and edge of bread board is 25cm.

Attachment 1: 2019.06.18_telescope_of_Koheras.jpg
2019.06.18_telescope_of_Koheras.jpg
  47   Thu Jul 4 18:28:55 2019 Loïc AmoudryFixedissuelasers and opticsOptical roomOnefive 33MHz unmodelocked

Oscillator randomly unmodelocked on 3rd july afternoon (during oscillator power measurement, maybe a reflection from the powermeter power filter ?)

Modelocked this afternoon by putting Onefive motor and phase at reference on the PTC software. 

  48   Thu Jul 4 18:30:39 2019 Loïc AmoudryFixedreportlasers and opticsOptical room2nd stage output power validated

Onefive output power : 28mW
Power before fiber injection : 7.8mW
After EOM and 99% coupling fiber : 1.8mW
After EOM and 1% coupling fiber : 14 uW (~1%), ~7.5V @ 1MOhm on DET36
Frep : 33.326 MHz
Output 1st stage + collimator & dichroic mirror: 36uW
Output 2nd stage @6A (~260mV @50Ohm on DET36): 300mW --> after 20-30min

Output 2nd stage is the same value as Onefive 133.33MHz

Attachment 1: 190704_2ndstage6A_spectrum.PNG
190704_2ndstage6A_spectrum.PNG
  50   Fri Aug 30 09:44:56 2019 Ronic ChicheFixedreportlasers and opticsOptical roomFinesse measurement after mirror cleaning using the spin coater

Last Finesse measurement in air : ~ 17 000 the 20th of may (see logbook)

28th and 29th of august, we cleanned, with Loïc, mirrors S3 then P1 successively using only ultra pure water with the spin coater.

we measured the finesse after each cleaning by modulation technique and in both cases, F ~ 12 500 !!!

maybe the mirror surfaces have been covered by some hydrocarbure during the summer ?
or by an other contaminant which is difficult to clean with only water ?
we decided to clean the mirrors first with an isopropanol solution and then with ultra pure water and the spin coater.

  51   Fri Aug 30 17:46:15 2019 Ronic ChicheFixedreportlasers and opticsOptical roomFinesse measurement after mirror cleaning using the spin coater

today I cleaned S3 with isopropanol and then with pure water on spin coater.
Finesse is only 10 700.(coupling 50% ? => to be confirmed...)

maybe we have to redo the procedure with a different way of cleanig (pushing less on the mirror ... ?)

Ronic Chiche wrote:

Last Finesse measurement in air : ~ 17 000 the 20th of may (see logbook)

from 28th to 30th of august, we cleanned, with Loïc, mirrors S3 then P1 successively using only ultra pure water with the spin coater.

we measured the finesse after each cleaning by modulation technique and in both cases, the finesse did not change !

maybe the mirror surfaces have been covered by some hydrocarbure during the summer ?
or by an other contaminant which is difficult to clean with only water ?
we decided to clean the mirrors first with an isopropanol solution and then with ultra pure water and the spin coater.

 

  54   Tue Sep 3 18:04:03 2019 Ronic ChicheFixedreportlasers and opticsOptical roomFinesse measurement after mirror cleaning using the spin coater

Following finesse measurements (start @10700): 
Clean P1 with SC (water) + isoprop (wipe) finesse @15000
Clean S2 with SC (water) + isoprop (wipe) finesse @13000
Clean P4 with SC (water) + isoprop (wipe) finesse @11000

We have to take into account that "cleaning" the mirror change the alignment axis in the cavity so the finesse can change because of that. It would explain the erratic nature of these results which seem to not show any effective cleaning.

Ronic Chiche wrote:

today I cleaned S3 with isopropanol and then with pure water on spin coater.
Finesse is only 10 700.(coupling 50% ? => to be confirmed...)

maybe we have to redo the procedure with a different way of cleanig (pushing less on the mirror ... ?)

Ronic Chiche wrote:

Last Finesse measurement in air : ~ 17 000 the 20th of may (see logbook)

from 28th to 30th of august, we cleanned, with Loïc, mirrors S3 then P1 successively using only ultra pure water with the spin coater.

we measured the finesse after each cleaning by modulation technique and in both cases, the finesse did not change !

maybe the mirror surfaces have been covered by some hydrocarbure during the summer ?
or by an other contaminant which is difficult to clean with only water ?
we decided to clean the mirrors first with an isopropanol solution and then with ultra pure water and the spin coater.

 

 

  55   Tue Sep 3 18:06:29 2019 Loïc AmoudryFixedinfolasers and opticsOptical roomMirrors sent to LMA

Mirrors have been sent today to LMA for cleaning and maybe annealing (recuit). 3 ULE of ThomX and the fused silicate for injection (trans @140ppm ).

  57   Fri Sep 6 14:59:00 2019 Loïc AmoudryFixedinfolasers and opticsOptical roomLMA cleaning technique

Christophe MICHEL told us more precise informations about spin coater cleaning.

They always clean with pur water and cotton tipped sticks.
First a spray of water, then they put the wet cotton tipped sitck onto the mirror from the center to the edge ~x3 while continuing to spread water. They finish by putting more water without the cotton tipped stick.

WARNING: WATER HAS TO BE CHANGED EVERY DAY 

They said they only use clean water but in case of grease/oil onto the mirror they would use first acetone then clean water.

  60   Fri Sep 27 10:43:17 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomNew Suprasil mirror from LMA

19/09/2019 : email from Chrisophe Michel about AR coating of LMA mirrors after reading that Ta2O5 layers can be damaged in vacuum if not protected by a SiO2 layer :

"oui tous nos coating standards HR se terminent par une couche de protection SiO2 Lambda/2.
Dans certains cas (spécifications du projet) l'épaisseur de la couche de silice de sortie peut être ajustée. 
Par exemple pour matcher une transmission très précise.
Pour les coating AR la couche de sortie est en silice mais pas lambda/2.

Pour ThomX je te confirme que la couche de sortie est une lambda/2 de silice. La formule de l'empilement est : (HB)19 H2B avec H:Ta2O5-TiO2 et B: Silice pour les substrats en ULE et (HB)13 H2B pour le substrat en saphir.
La plupart de nos miroirs sont montés en cavité sous vide (notamment ceux de VIRGO) et ce phénomène n'a pas été observé (hors présence de plasma)."
 

20/09/2019 : new email from Chrisophe Michel after he sent back the mirrors (LMA made a cleaning and thermal annealing) :

"le colis avec les 4 optiques, nettoyées 2 faces après traitement thermique et caractérisation, a été envoyé cet après midi par DHL (tracking number 4995967550).

Dans le colis j'ai ajouté le miroir en saphir C16115-11.

Les mesures donnent des absorptions entre 0.12 et 0.5 ppm (nous avons réussit à faire fonctionner de nouveau le banc mais il est mode dégradé avec une incertitude plus grande sur les mesures mais l'ordre de grandeur est bon et correspond aux valeurs mesurées en 2016 qui était de 0.24 ppm).

Pour les pertes par diffusion voila ce que Laurent à mesurer (les valeurs sont identiques à celles mesurées en 2016 sauf pour le C16116/11):

- C16116/11: 9 ppm sur diamètre 10mm (6 ppm mesurées en 2016); 4.5 ppm sur diamètre 6 mm; T=1.5 ppm @ 1064 & 1031 nm (simulée)

- C16116/12: 8.5 ppm sur diamètre 10mm; 4.5 ppm sur diamètre 6 mm; T=1.5 ppm @ 1064 & 1031 nm (simulée)

- C16116/13: 14 ppm sur diamètre 10mm; 10 ppm sur diamètre 6 mm; T=1.5 ppm @ 1064 & 1031 nm (simulée)

- C16115/11 (saphir): 9 ppm sur diamètre 10mm; 4 ppm sur diamètre 6 mm; T=148 ppm @ 1064 nm soit 120 ppm simulée @ 1031 nm"

Loïc Amoudry wrote:

14/05/19:

New Suprasil mirror received. It will replace the S-BOX suprasil mirror on ThomX which was already replacing the sapphire mirror.

First finesse measurement: 14000

Features sent by Christophe Michel:

"- Transmission mesurée (1064 nm)= 172 ppm ; Transmission estimée (1031 nm)= 140 ppm
- Réflexion (1064 nm)= 1025 ppm; doit être inférieure à 1031 nm
L'AR n'est pas bon alors que l'on obtenu 170 ppm pour le run de calage!
J'espère que vous pourrez néanmoins l'utiliser.
- Diffusion moyenne sur diamètre 12 mm: 2.6 ppm"

 

 

  61   Fri Sep 27 16:17:36 2019 Ronic ChicheFixedinfolasers and opticsOptical roomFinesse measurement

After mirrors were going back from LMA for cleaning and thermal anealing, they have been installed in FP cavity for Finesse measurement.

after getting strange results when the cavity is closed (very low Finesse...maybe too much ethanol inside cavity procucing some lineic losses), we get 12k-12k5 Finesse with primary vacuum (~4 mbar) and a clean lock (very flat transmission signal and quite noiseless reflection signal)
we let the cavity pumped all the WE to see if we can measure a difference or not at lower pressure.

  62   Mon Sep 30 11:20:48 2019 Loïc AmoudryFixedinfolasers and opticsOptical roomDiffusion ULE spherical mirror (S3) vs Fused Silicate injection planar mirror (P1)

Observation made with IR Beam Viewer: 

The observation of P1 and S3 simultaneously while locking shows much higher diffusion on the ULE mirror. The diffusion on the two other ULE mirrors (S2 and P4) shows similar diffusion as S3.

Attachment 1: diffusion.jpg
diffusion.jpg
  63   Mon Sep 30 11:30:32 2019 Ronic ChicheFixedinfolasers and opticsOptical roomFinesse measurement

To complete these observations:

Finesse has been measured on thursday, right after cleaning the cavity with ethanol @6000.
Finesse has been measured on friday, after opening the cavity for few tens of minutes @12000-12500. The cavity was openned while locking and the transmission was continuously and ~linearly increasing.
Finesse has been measured today after ~50h pumping with primary vacuum @17000. It's back to the highest value we've ever measured on this cavity.

Parameters @17000 finesse:

Input power: 7.85 mW (Koheras + AOM + 2*EOM)
Coupling: 70-75% => Pin ~ 5.9 mW
Transmission through S2 and P4: 50 µW (transmission estimated to 1.5 ppm) => Pcav ~ 33 W => G ~ 5600
It is compatible with a simulation with nominal parameters for mirrors and adding 47ppm by mirror of losses (absorption and scattering) which indicates 17k of Finesse and 4100 of power gain.
in this simulation, the coupling should be 94% instead of the 70-75% measured. It is possible that the losses (47 ppm) are not the same on all mirrors.

Ronic Chiche wrote:

After mirrors were going back from LMA for cleaning and thermal anealing, they have been installed in FP cavity for Finesse measurement.

after getting strange results when the cavity is closed (very low Finesse...maybe too much ethanol inside cavity procucing some lineic losses), we get 12k-12k5 Finesse with primary vacuum (~4 mbar) and a clean lock (very flat transmission signal and quite noiseless reflection signal)
we let the cavity pumped all the WE to see if we can measure a difference or not at lower pressure.

 

  67   Thu Oct 3 18:00:10 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomThomX amplifier M2

The ThomX M2 has been measured between 1.5-1.6 for x and y

Attachment 1: Fit_matlab_M2.PNG
Fit_matlab_M2.PNG
  68   Wed Oct 9 10:08:24 2019 Loïc AmoudryFixedinfolasers and opticsOptical roomImprovement holder amplifier fiber

We've added a clamp in the 3rd stage to prevent the pump fiber from moving and tilting (it was tilting to the left).

Attachment 1: 1.png
1.png
Attachment 2: 2.png
2.png
  69   Wed Oct 9 10:10:01 2019 Loïc AmoudryFixedinfolasers and opticsOptical roomFree Spectral Range (FSR)

FSR is 33.347 MHz @1e-8 mbar

 

  70   Wed Oct 9 10:16:24 2019 Loïc AmoudryFixedinfolasers and opticsOptical roomFiber injection and onefive output power

Fiber injection is finished with the new ThomX stretching CVBG in double pass (+50% injection power compared to K-BOX CVBG in simple pass).

- Output Onefive: 27 mW
- After Onefive + lambda/2 + isolator: 25.8 mW
- After Onefive + lambda/2 + isolator + lambda/2 + PBS + CVBG*2 + PBS: 9.6 mW
- After injection into fiber and EOM (+all preceding): 3.2 mW
- Input 99% of the amplifier: 2.86 mW

> 660 mW on 1% photodiode DET100 @1Mohm

  71   Wed Oct 9 11:41:11 2019 Loïc AmoudryFixedinfolasers and opticsOptical roomOnefive spectrum and CVBG spectrum

The Onefive spectrum has shifted a little bit to lower lambda. It's still enough for the experiment (3mW injection of ampli intead of 1mW needed).

Attachment 1: 191008_spectrumbeforecvbg.PNG
191008_spectrumbeforecvbg.PNG
Attachment 2: 191008_spectrumaftercvbg.PNG
191008_spectrumaftercvbg.PNG
Attachment 3: 191008_beamshapeaftercvbgthomx.PNG
191008_beamshapeaftercvbgthomx.PNG
  76   Tue Oct 22 09:10:45 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomFirst mode degeneracy on ThomX

Yesterday, we've observed the first degeneracy on ThomX cavity at ~115 kW stored. The shape of the stored power on the oscilloscope is really close to what has been published in the D-shape mirror's paper ("tek0000" transmission in yellow, piezoelectric activator in green, error signal in blue).
This mode could be a 8.0 or 9.0. 
We have also observed a change of the angle of the mode (the pictures "degenere", "degenere2" and "degenere3" were taken by order). The only changing parameters are the alignement and the length of the cavity. 
The last thing we noticed is the shape of the mode which could not be an hermite-gauss mode (Bessel mode ?).

Attachment 1: degenere.PNG
degenere.PNG
Attachment 2: degenere2.PNG
degenere2.PNG
Attachment 3: degenere3.PNG
degenere3.PNG
Attachment 4: tek0000.png
tek0000.png
Attachment 5: 100kWThomX.png
100kWThomX.png
  77   Wed Nov 6 16:11:12 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomSpherical mirrors at closest position. FSR, finesse and beam size.

The spherical mirrors are at the closest position to get a FSR ~=33.3435 MHz. Planar mirrors are at the largest position.

The finesse has been measured about 17 000. 

Beam diameter (2*w) is ~= 4.6 mm x and y.

Attachment 1: data_finesse.isf
Attachment 2: fit_finesse_17000.png
fit_finesse_17000.png
Attachment 3: beamsize.PNG
beamsize.PNG
Attachment 4: fit.PNG
fit.PNG
  78   Wed Nov 13 14:44:56 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomAmplifier output power

The ouput power of the ThomX amplifier has been measured as a function of the 3rd stage pumps current (0A is 1st and 2nd stage turned on).

We measured the total power, the total power without the "donuts" pump signal dumped by an iris and the transmission of a dichroic mirror (for ref see image).

Attachment 1: dichroic.PNG
dichroic.PNG
Attachment 2: courbe.PNG
courbe.PNG
Attachment 3: power.xlsx
ELOG V3.1.4-395e101