Entry   Finesse measurement, posted by Ronic Chiche at Optical room about lasers and optics 
    Reply   Finesse measurement, posted by Ronic Chiche at Optical room about lasers and optics 
Message ID: 61     Entry time: Fri Sep 27 16:17:36 2019     Reply to this: 63
 Author: Ronic Chiche 
 Status: Fixed 
 Type: info 
 Category: lasers and optics 
 Location: Optical room 
 Title: Finesse measurement 

After mirrors were going back from LMA for cleaning and thermal anealing, they have been installed in FP cavity for Finesse measurement.

after getting strange results when the cavity is closed (very low Finesse...maybe too much ethanol inside cavity procucing some lineic losses), we get 12k-12k5 Finesse with primary vacuum (~4 mbar) and a clean lock (very flat transmission signal and quite noiseless reflection signal)
we let the cavity pumped all the WE to see if we can measure a difference or not at lower pressure.