ID Date Author Status Type Category Location Title Text Attachments
  4   Mon Oct 1 15:19:48 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomPhase measurement with HASOwith just intensity, the coupling is 98%,
and with phase, x direction coupling is 95%,
y direction 97%, so the telescope is good.
  
  3   Mon Oct 1 14:27:35 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomPhase measurement with HASO* Loic has to fix the number of files (3)
regarding the number of measurements (2)

* splitted Intensity and phase
 8x 
  2   Mon Oct 1 11:30:39 2018 Loïc AmoudryFixedreportlasers and opticsOptical roomPhase measurement with HASO2 measurement.

First one locked, in transmission
of M2 with 2nd stage 0A.Total 89 cm from
 180928_M2_0A.has1801001_input_0A.has180928_M2_trans_0A.txt 
  1   Wed Sep 26 18:12:44 2018 Loïc AmoudryFixedinfolasers and opticsOptical roomCavity polarization states (Koheras), Finesse of 2 polarization statesMeasurments of Finesse with the 2 polarization
states, let's call them H (higher) and
L (lower): 24500 for the H and 23500 for
 Finesse_higher.isfFinesse_lower.isfThe2polarizations_states_direct_trans.isfThe2polarizations_states_direct_trans.png 