    Reply   Cleaning, Motors Controllers mini-rack, hexapod controller, chiller and vacuum, posted by Ronic Chiche at ThomX igloo about detectors and electronics | utilities 
This afternoon, we continued to clean the housing with Manar and Daniele.
we did a dust measurement after cleaning with the two ceil doors opened and the dust meter in the middle of the table, and it counted ZERO!
it doesn't mean the housing is perfectly clean but it is clean enough to pay attention now to cleanness.
Entry   Chiller of the new laser amplifier , posted by Ronic Chiche at Optical room about utilities 
We observed 2 issues for the chiller of the new CELIA laser amplifier:


setting the water temperature to 17°C, the chiller water turns to ice close to the water filling port !!! => temperature too low
?
even when the water is fluid, one has a flow sensor error ! => pump charge too strong ? => reduce the water pipe length.


We
    Reply   Chiller of the new laser amplifier , posted by Ronic Chiche at Optical room about utilities 
Today, the chiller worked at 17°C during 3 hours without error, no freezing nor flow sensor error.




Ronic
Chiche wrote:



We observed 2 issues for the chiller of the new CELIA laser amplifier:
Entry   Chiller installation + interlock, posted by Ronic Chiche at ThomX igloo about mechanics 
- this morning with Aurélien, we filled the water tank of the chiller and we opened the water valves (last pipes on water arrival/departure panel).

there is no leakage on the external water circuit but there is a small one on one of the Swagelock adapters of the water circuit connected to
the Alphanov Amplifier.
    Reply   Chiller installation + interlock, posted by Ronic Chiche at ThomX igloo about mechanics IMG_20220309_093541.pngChiller_Sub-d15_pinout.jpg
- this morning the chiller leakage has been fixed by adding some Teflon on the faulty Swagelock adapter.
a long leakage test has to be done to be sure there is no leakage at all => can be done on Friday during our tests.

- here is a picture of the 2 Alphanov amplifier controller interlocks.
    Reply   Chiller installation + interlock, posted by Ronic Chiche at ThomX igloo about mechanics 
This morning, I started the chiller and opened the valves on the external water circuit => no water leakage during the morning.

now, the chiller will stay ON all the time => we will check the installation at the end of the afternoon.




Ronic
Entry   Alphanov amplifier chiller instruction manual , posted by Manar Amer at ThomX igloo about utilities HEC-OM-U015-B.pdf
Adding the instructions manual for the Alphanov Amplifier SMC Chiller

Installation and Maintenance Manual
Entry   Airflow power breakdown, posted by Ronic Chiche at ThomX igloo about utilities 
since around 2 weeks, the airflow above the optical table housing has a power breakdown when one tries to switch it on.

the power breaker is in the electrical bay, on the left of the south entrance of ThomX.
the name of the power breaker switch related to the airflow is "TABLEAU ANNEAU 4".
Entry   Airflow installation, posted by Ronic Chiche at ThomX igloo about mechanics 
On the 27th of May, the Airflow has been dismounted from the optical room to be mounted around the optical cavity by the AirC2 company.

here is email from Yann :

Le flux est rentré dans la casemate.
    Reply   Airflow installation, posted by Ronic Chiche at ThomX igloo about mechanics 
The AC plug and the power cord of the airflow were quite hot because they have been dimensioned for 16A current and the total current is above.
AirC2 company should replace them with 32A equipment (6mm2 power cord + 32A Hypra AC plug) in the next days.
 