    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software scope.jpg
we have had this new issue (watchdog alarm switching off the amplifier) for several days.

Alphanov mentionned it could come from the input oscillator stability (modelock loss for example).

Today,
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 20230208_error_watchdog.jpeg
With Daniele after, we took a readings of the beam with a lens of 750 mm.

From the observed data, we can state that the beam is not completely Gaussian M2 > 1, due to the variation of the ellipticity of the beam
along the focused path 
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 
This morning,

- with the "Alphanov" software => Central panel => "system" button => Watchdog "Period Max" was at 100ns
and has been set to 200ns.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 
This morning with Daniele, we did a second measurement of the beam profile using a different lens of focal 400 mm.

at a power amplification of 20%

the amplifier stayed on for around one hour and a half with no errors. 
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software Analyse_DonnĂ©es_20_fevrier.pptx
On Monday 20th of February, Daniele and Viktor took some beam profile measurement data @ 50% of power amplification.

data and analysis can be found in the attached file.




Ronic
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 

Yesterday, Guillaume Machinet and Vincent Clet from Alphanov came to fixe some issues with the laser amplifier:

- 1) they fixed the "rep. rate" measurement issue.
previously, the voltage range in which it was possible to detect the correct rep rate (33MHz) was very narrow ~ 50mV.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software WhatsApp_Image_2023-02-27_at_6.07.00_PM.jpeg
This morning with Victor we took a reading of the beam profile using a 400 mm lens at 70 % and 100 % amplification.

Will process the data and add after

Note : after an hour and half of amplifier on at 70% we increased the power to 100%, and after ~ 10 minutes the amplifier turned off and showed
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 20230308_124313.jpg
this morning with Manar,
we checked the Alphanov amplifier issue : "The laser has been stopped because its power is too far from the nominal power".

the photodiode voltage measuring the output power has drifted again and then, the estimated output power was wrong, triggering an alarm issue.
Entry   CW laser installation and cleaning, posted by Ronic Chiche at ThomX igloo about lasers and optics 20230308_121713.jpg
This morning with Manar, in order to prepare the installation of the news FP cavity mirrors,

- we installed the Koheras below the optical table
- we put a fiber prolongator at the end of the Koheras fiber and we put a mounted collimator at the end of the fiber.
    Reply   CW laser installation and cleaning, posted by Ronic Chiche at ThomX igloo about lasers and optics 
This morning with Ronic , in preparation for the mirror installation, we removed additional mechanics and cables from the table. 

We organized and cleaned the permanently installed cables on the table away from the area of the cavity opening. 

We opened the two top vents and left the shutters open to circulate clean air , (note the shutters surrounding the laser and amplifier are half
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software Beam_wavefront_30cm_no_Lens.jpg
We took reading of the beam profile of the amplifier without a lens @ 30 cm from the 2nd wedge for amplification percentages

20-30-40-50-60-70-80-90-100 % another additional 3 readings will be taken at different locations for all amplification percentages.

 
Entry   Power shut down on Saturday the 18th, posted by Ronic Chiche at ThomX igloo about utilities 
All the equipments have been switched off on Friday 17th due to a power shut down on Saturday 18th :
CW and Onefive oscillators, computers, laser amplifier, cavity mirror motors, chiller, etc...
    Reply   33MHz laser SN2439 back to Orsay, posted by Ronic Chiche at Optical room about lasers and optics MCS_Network_Interface_Configuration_Guide-1.pdf20230320_154643[1].jpg
As the general local network for the FP-cavity devices is 192.168.xxx.xxx, I changed the IP adress of the smaract laser motors (Frep and CEP) controller.

For that, I used the embedded web server as described in the manual in attached file.

the new adresses are :
    Reply   Power shut down on Saturday the 18th, posted by Ronic Chiche at ThomX igloo about utilities 
Today, I restarted all the equipements :

Chiller                                                
: ok, temperature converged quickly to the set temperature : 25°C
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 
due to several "case temperature" issues, I change the "Max case temperature" in the "general config" of the "Alphanov
Software" from 50°C to 55°C.
=> validated by Alphanov.
Entry   33MHz oscillator + Alphanov amplifier lock, posted by Ronic Chiche at ThomX igloo about lasers and optics 120705_Displacement_of_cylindrical_piezo_ceramics.xls
the last tries to lock the 33MHz + amplifier to the 30k Finesse FP-cavity were unsuccessful.

during a laser Frep scan using the Laselock, one observes that the main cavity resonance is not able to stay inside the PZT scan range from one
scan to another (500ms-1s period)
    Reply   33MHz oscillator + Alphanov amplifier lock, posted by Ronic Chiche at ThomX igloo about lasers and optics 
finding the right modulation/demodulation PDH phase is very difficult on the main resonance because the we get non stationnary signals with a lot of
oscillations.
changing the phase, in this condition, does not really change the error signal.
    Reply   33MHz oscillator + Alphanov amplifier lock, posted by Ronic Chiche at ThomX igloo about lasers and optics M250_Video_Amplifier.pdfM250_manual.pdf
today with Daniele, we locked easily (but with a noisy lock) on the secundary resonance and we tried to lock on the main resonance (with very low coupling
~10% which mean a CEP ~Pi)
the lock was possible but was very noisy.
    Reply   33MHz oscillator + Alphanov amplifier lock, posted by Ronic Chiche at ThomX igloo about lasers and optics 
After removing the 2 generators from the optical table, the lock is much more stable and now, it is possible to lock on the main resonance with a poor
CEP but with quite good stability.
the coupling is still very low ~ 5% for that CEP but if one improves it (CEP ->0), using the laser double-wedge motor, one clearly sees an improvement
    Reply   33MHz oscillator + Alphanov amplifier lock, posted by Ronic Chiche at ThomX igloo about lasers and optics PDH_error_signal_with_Smaract_motors_ON-OFF.jpgLock_at_amp_@30%.jpg10kW_in_remote.jpg
this morning, I tested the laser+amplifier @ 30% lock on the FP cavity with and without Smaract motors.

I recorded the PDH error signal during a lock:
- blue   : with Smaract motors controller powered ON but motors are stopped