  Status of commissioning, report also here plots are reports., Page 20 of 24  Not logged in ELOG logo
    Reply   CW FP Cavit lock , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 20230406_setup.jpg20230406_transmission_S2-P4.jpg20230406_Cavity_Lock.jpg
Yesterday morning we locked the cavity on a strange mode in the cavity (we call it mode, but not sure what it is ??)

we also managed to lock on a higher order mode 02

we get an estimated 30% coupling (expected value), and note that we have not installed a telescope yet.
    Reply   CW FP Cavit lock , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 20230407_103103-2.jpg
This morning with Daniele and Viktor,

- we confirmed the "potato" mode was well centered on the injected beam and was not due to some misalignment of the injected beam compared
to the cavity axis.
    Reply   FP Finesse measurement using CW, posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 20230411_first_Finesse_reading.JPG20230411_second_Finesse_reading.JPG20230411_Beam_Profile_position_differance_when_changing_FSR.JPG
This morning, with Ronic and Daniele, we locked the cavity again (The lock was relatively easy and stable during the reading of the finesse)

We took 2 reading of finesse

1st   : @ 33.2808 MHz (what the cavity length was at)           :  Finesse =
Entry   PW laser Injection into cavity , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 20230412_pulsed_cav_mode.bmp
Yesterday,


we installed the second high power reflecting mirrors at the output of the amplifier.
We aligned and installed a telescope
that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
Amplifier
    Reply   PW laser Injection into cavity , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 20230418_size_of_beam_before_injection_window_at_10percetage_ampli.jpg20230418_coupling_at_the_start_only_alignment_improved1.jpg20230418_zoom_on_Oscilliscope_signal_fundamental_mode.jpg20230418_coupling_after_improving_it_with_regulator_on_attempting_to_lock.jpg

On Monday with Daniele,

we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly
and injected into the cavity aligned and improved on the fundamental mode.


On Tuesday with Daniele and Kevin

 I
added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
removed the external resistors
    Reply   PW laser Injection into cavity , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 
Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)

was done at the end due to excitation of higher order modes.

Wanted to improve the fundamental first, then increase the power.
    Reply   PW laser Injection into cavity , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 2023-05-22_MMD3_error_on_alarm_window.jpg2023-05-22_alphanov_software.jpg
Today with Ronic and Daniele we attempted to lock the cavity, but the alphanov amplifier did not turn on because of an error in MMD3 (related to the
third stage)

the new error is having the MMD3 error show on the alarm window of the LAL software, and having the Alarms on the line 3 of the alphanov software
    Reply   Cavity Lock - Alphanov Amplifier , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics | cabling | software tek00000.png
This morning, while trying Guillaume solution of disconnecting the 24 V necessary to the second and third stage and raise the threshold of Temp preamp
1 (to only operate the ampli with first stage only)

Ronic disconnected the 24 volt connection, turned on amplifier LAL software but no MMD3 error or a big temp on Temp preamp 1 , software worked
    Reply   Cavity Lock - Alphanov Amplifier , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics | cabling | software 13x
Yesterday, with Ronic we locked ThomX cavity on the second resonance 

Laser : 33.36 MHz , Amplifier : 0 % (worked with no errors all day)

adding image for the lock on the second resonance (locked), 00 Mode image, fit, Beam measurements
    Reply   Cavity Lock - Alphanov Amplifier , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics | cabling | software 20230605_tek00001_33uW.png20230605_tek00002_82uW.png
Today, I installed a power meter at the transmission point of the 1st spherical mirror , transmission 3 ppm (direct from cavity window beam profiler,
no filter)

We locked the cavity at 2 different resonances of the fundamental mode, the lock in both cases was stable for a round 1 minute.
    Reply   Cavity Lock - Alphanov Amplifier , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics | cabling | software tek00001.png20230607_183626_Frep_CEP_motors_smartAct.jpg20230607_183641_transmitted_power.jpgtek00002.png
Today with Ronic,

starting with major events that happened (Water circulation + ThomX valves )  


Early in the morning a company worked on the water network, it seemed it was not restarted properly so there was no water circulation
    Reply   Cavity Lock - Alphanov Amplifier , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics | cabling | software tek00001_noise_reduced_first_resonance.pngtek00000_cavity_lock_with_noise_from_odcillator_controller.png
In the morning, we locked on the first resonance. With an increase on the amplifier power @ 20-25% and 30%, which reduced the noise

we manage to get a coupling when improving the CEP up to 55 - 60 % coupling : put there is still a drift on the CEP

The controller for the oscillator frep and CEP produces a lot of noise, even if it is disconnected. we need to switch it off to remove its effect
    Reply   Cavity Lock - Alphanov Amplifier , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics | cabling | software 
The computer is possible to connect from the control room,

we are able to run remotely the amplifier, cavity motors, oscillator motors, lase lock (note: there is an issue with the keyboard, we are not
able to use it with remote access!!!!!!!!)
Entry   Onefive Origami 33MHz specs, posted by Loïc Amoudry at Optical room about lasers and optics 6x
Power during ~20h, power after turn on, spectrum image and excel datas, polarisation and pulse length.

Powermeter measurments were done with a /10 filter.. 3 mW means 30 mW

Important: Onefive does not automatically locks after being turned on.
Entry   End of vacuum test and cavity ready for cleaning, posted by Loïc Amoudry at Optical room about vacuum IMG_20181221_1110152.jpgIMG_20181221_1119082.jpgIMG_20181221_1113122.jpgIMG_20181221_1124492.jpg
Vacuum tests finished on 19 december 2018.

Spent last two days to discuss the importance of removing aluminium and baking wires before to insert mirrors for cleanliness issues. Finally
we took pictures to support our demand.
Entry   Cleaning outside IP cavity, posted by Loïc Amoudry at Optical room about lasers and optics | utilities Avant_nettoyage.jpgApres_une_semaine_nettoyage_lingettes_et_brosseadents.jpgSans_UV_traces_scotch_nettoyees_apres_baking.jpg
From 7th january to 11th january.

 
Entry   End of cleaning, plastic fixed on cavity, posted by Loïc Amoudry at Optical room about utilities covered_by_plastic.pngcovered_by_plastic2.png
08/02/19 - End of the cleaning process.
Entry   Mirrors checked with binocular microscope, posted by Loïc Amoudry at Optical room about lasers and optics 41_THOMX_SphericalZoom1x.jpg49_THOMX_M1ReflectiveZoom4x.jpg
29/01/19 - ThomX mirrors have also been checked with binocular microscope. They shown the same spots as S-BOX mirrors. 

1st image shows reflective face of a spherical mirror.

2nd image shows reflective face of M1 (sapphire).
Entry   Mirrors sent to LMA, posted by Loïc Amoudry at Other about lasers and optics 
30/01/19 - As LMA people came, we discussed about the spots on ThomX mirrors which should be clean. As mentionned in "SBOX commissioning",
these spots are normal.

An other question appeared during the conversation about the storage time and location of these mirrors. They have been stored for more than
Entry   Basic alignment & alignment motors check, posted by Loïc Amoudry at Optical room about mechanics | lasers and optics 
14/02/19: The 4 motors but Mby control works. Mby motor home does not bring it to its "middle" position but to the end of its range.
Issue circumvented using the higher rights software.

The beam is now going through the cavity windows (from P1 to S2).