    Reply   CW FP Cavit lock , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 20230406_setup.jpg20230406_transmission_S2-P4.jpg20230406_Cavity_Lock.jpg

Yesterday morning we locked the cavity on a strange mode in the cavity (we call it mode, but not sure what it is ??)

we also managed to lock on a higher order mode 02

we get an estimated 30% coupling (expected value), and note that we have not installed a telescope yet.

 

Manar Amer wrote:

Today, we prepared lenses for a telescope for the CW laser injection into the cavity, yet to be installed

Image of the transmission at P4 attached, an estimation of the beam size of the mode see is ~ 5.5 mm 

The vacuum in the 2 optical cavities is decreasing but slowly. today it is 2.6 - 2.4 * 10^-7 mbar

Manar Amer wrote:

Last week, with Ronic and Victor we aligned the cavity using CW laser (Koheras)

  • we have also installed 4 mirrors in the cavity with the best theoretical qualities (serial number and images included)
    • The Plan ULE mirror P4 thickness was not compatible with the piezoelectric mount. We had to install a metal spacer behind the mirror to be able to fix it. the metal spacers in from ThorLabs made from INOX, It is 1 inch diameter (25.4 mm), we drilled a hole in the middle of 15 mm in the Workshop in bat 200. Then cleaned it using an ultrasound bath in pure water, using a small Ultrasound bath machine in Maverics platform Vide and Surfaces next to ThomX eglo.
    •  
    • This gives some difficulties in imaging the transmission at P4. To help with the imaging, we used the plastic #D printed irises that Yann made long time ago
    • we took images of the mode inside the cavity at P4 transmission.

 

  • We also observed some oxidization on the metal bars of the controllers, more on the side of S2 and P4. we are yet to test the motion of the motors in the z-axis.
  • Friday evening we closed the windows using a new copper rings (last one we have, Ronic placed a new order).
  • The pump was turned on by Bruno and Eric yesterday afternoon. The current vacuum level is 6-7*10^-6 mbar

the pre and the turbo pimps are on so still too noisy to work, we have access all day tomorrow.

 

 

 

 

 

    Reply   CW FP Cavit lock , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 20230407_103103-2.jpg

This morning with Daniele and Viktor,

- we confirmed the "potato" mode was well centered on the injected beam and was not due to some misalignment of the injected beam compared to the cavity axis.
we look at the reflected beam during a lock. we can see the direct reflected beam superposed with the cavity mode (cf attached image).

- then we assumed the "potato" mode problem could come from the too short distance between spherical mirrors which introduces ellipticity and finally can lead to some instability.
we manage to move further spherical mirrors S2 and S3 by moving them from 1 350 000 to 800 000 steps.
=> we increased the distance by 2x 550 000 steps (with 6µm for 1000 steps) => we increased the distance between spherical mirrors by 6.6 mm
the new mode shrinked a lot and at the end was spherical at 73%.
the mode size is now about 4400 x 6000 µm but with the major and minor axis still at 45° !
 

Manar Amer wrote:

Yesterday morning we locked the cavity on a strange mode in the cavity (we call it mode, but not sure what it is ??)

we also managed to lock on a higher order mode 02

we get an estimated 30% coupling (expected value), and note that we have not installed a telescope yet.

 

Manar Amer wrote:

Today, we prepared lenses for a telescope for the CW laser injection into the cavity, yet to be installed

Image of the transmission at P4 attached, an estimation of the beam size of the mode see is ~ 5.5 mm 

The vacuum in the 2 optical cavities is decreasing but slowly. today it is 2.6 - 2.4 * 10^-7 mbar

Manar Amer wrote:

Last week, with Ronic and Victor we aligned the cavity using CW laser (Koheras)

  • we have also installed 4 mirrors in the cavity with the best theoretical qualities (serial number and images included)
    • The Plan ULE mirror P4 thickness was not compatible with the piezoelectric mount. We had to install a metal spacer behind the mirror to be able to fix it. the metal spacers in from ThorLabs made from INOX, It is 1 inch diameter (25.4 mm), we drilled a hole in the middle of 15 mm in the Workshop in bat 200. Then cleaned it using an ultrasound bath in pure water, using a small Ultrasound bath machine in Maverics platform Vide and Surfaces next to ThomX eglo.
    •  
    • This gives some difficulties in imaging the transmission at P4. To help with the imaging, we used the plastic #D printed irises that Yann made long time ago
    • we took images of the mode inside the cavity at P4 transmission.

 

  • We also observed some oxidization on the metal bars of the controllers, more on the side of S2 and P4. we are yet to test the motion of the motors in the z-axis.
  • Friday evening we closed the windows using a new copper rings (last one we have, Ronic placed a new order).
  • The pump was turned on by Bruno and Eric yesterday afternoon. The current vacuum level is 6-7*10^-6 mbar

the pre and the turbo pimps are on so still too noisy to work, we have access all day tomorrow.

 

 

 

 

 

 

    Reply   FP Finesse measurement using CW, posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 20230411_first_Finesse_reading.JPG20230411_second_Finesse_reading.JPG20230411_Beam_Profile_position_differance_when_changing_FSR.JPG

This morning, with Ronic and Daniele, we locked the cavity again (The lock was relatively easy and stable during the reading of the finesse)

We took 2 reading of finesse

1st   : @ 33.2808 MHz (what the cavity length was at)           :  Finesse = 27 024

2nd :  @ 33.3382 MHz (adjusted to match the pulsed laser)  :  Finesse = 30 433

The difference between the 2 is the FSR, during the change we also did some alignment to counter the reduced transmitted power.

We took an image of the beam profile before and after changing FSR and there was ~ 70 um change in the beam position on the Horizontal axis of the beam.

 

Manar Amer wrote:

This morning with Daniele and Viktor,

- we confirmed the "potato" mode was well centered on the injected beam and was not due to some misalignment of the injected beam compared to the cavity axis.
we look at the reflected beam during a lock. we can see the direct reflected beam superposed with the cavity mode (cf attached image).

- then we assumed the "potato" mode problem could come from the too short distance between spherical mirrors which introduces ellipticity and finally can lead to some instability.
we manage to move further spherical mirrors S2 and S3 by moving them from 1 350 000 to 800 000 steps.
=> we increased the distance by 2x 550 000 steps (with 6µm for 1000 steps) => we increased the distance between spherical mirrors by 6.6 mm
the new mode shrinked a lot and at the end was spherical at 73%.
the mode size is now about 4400 x 6000 µm but with the major and minor axis still at 45° !
 

Manar Amer wrote:

Yesterday morning we locked the cavity on a strange mode in the cavity (we call it mode, but not sure what it is ??)

we also managed to lock on a higher order mode 02

we get an estimated 30% coupling (expected value), and note that we have not installed a telescope yet.

 

Manar Amer wrote:

Today, we prepared lenses for a telescope for the CW laser injection into the cavity, yet to be installed

Image of the transmission at P4 attached, an estimation of the beam size of the mode see is ~ 5.5 mm 

The vacuum in the 2 optical cavities is decreasing but slowly. today it is 2.6 - 2.4 * 10^-7 mbar

Manar Amer wrote:

Last week, with Ronic and Victor we aligned the cavity using CW laser (Koheras)

  • we have also installed 4 mirrors in the cavity with the best theoretical qualities (serial number and images included)
    • The Plan ULE mirror P4 thickness was not compatible with the piezoelectric mount. We had to install a metal spacer behind the mirror to be able to fix it. the metal spacers in from ThorLabs made from INOX, It is 1 inch diameter (25.4 mm), we drilled a hole in the middle of 15 mm in the Workshop in bat 200. Then cleaned it using an ultrasound bath in pure water, using a small Ultrasound bath machine in Maverics platform Vide and Surfaces next to ThomX eglo.
    •  
    • This gives some difficulties in imaging the transmission at P4. To help with the imaging, we used the plastic #D printed irises that Yann made long time ago
    • we took images of the mode inside the cavity at P4 transmission.

 

  • We also observed some oxidization on the metal bars of the controllers, more on the side of S2 and P4. we are yet to test the motion of the motors in the z-axis.
  • Friday evening we closed the windows using a new copper rings (last one we have, Ronic placed a new order).
  • The pump was turned on by Bruno and Eric yesterday afternoon. The current vacuum level is 6-7*10^-6 mbar

the pre and the turbo pimps are on so still too noisy to work, we have access all day tomorrow.

 

 

 

 

 

 

 

Entry   PW laser Injection into cavity , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 20230412_pulsed_cav_mode.bmp

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,
    Reply   PW laser Injection into cavity , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 20230418_size_of_beam_before_injection_window_at_10percetage_ampli.jpg20230418_coupling_at_the_start_only_alignment_improved1.jpg20230418_zoom_on_Oscilliscope_signal_fundamental_mode.jpg20230418_coupling_after_improving_it_with_regulator_on_attempting_to_lock.jpg
  • On Monday with Daniele,
    • we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
  • On Tuesday with Daniele and Kevin
    •  I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
    • removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm  , Reflection : 50 ohm  )
    • We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
  • Images:
    • 1st : showing the size of the beam just before injecting into the cavity @ 10%
    • 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
    • 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
    • 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors

Note : the voltage ranges are not the same between the images

Manar Amer wrote:

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,

 

    Reply   PW laser Injection into cavity , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 

Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)

was done at the end due to excitation of higher order modes.

Wanted to improve the fundamental first, then increase the power.

Manar Amer wrote:
  • On Monday with Daniele,
    • we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
  • On Tuesday with Daniele and Kevin
    •  I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
    • removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm  , Reflection : 50 ohm  )
    • We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
  • Images:
    • 1st : showing the size of the beam just before injecting into the cavity @ 10%
    • 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
    • 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
    • 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors

Note : the voltage ranges are not the same between the images

Manar Amer wrote:

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,

 

 

    Reply   PW laser Injection into cavity , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 2023-05-22_MMD3_error_on_alarm_window.jpg2023-05-22_alphanov_software.jpg

Today with Ronic and Daniele we attempted to lock the cavity, but the alphanov amplifier did not turn on because of an error in MMD3 (related to the third stage)

the new error is having the MMD3 error show on the alarm window of the LAL software, and having the Alarms on the line 3 of the alphanov software red for both EXT/CPU and Laser T Max (never been red before)

this is after a 2-week work stop, but the last time when Daniele changed the fiber connector it worked for 3 hours with no errors.

Ronic, is in contact with Guillaume from Alphanov to fix the issue !!!

Work scheduled all Wednesday, in hopes the issue is fixed by then.

Manar Amer wrote:

Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)

was done at the end due to excitation of higher order modes.

Wanted to improve the fundamental first, then increase the power.

Manar Amer wrote:
  • On Monday with Daniele,
    • we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
  • On Tuesday with Daniele and Kevin
    •  I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
    • removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm  , Reflection : 50 ohm  )
    • We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
  • Images:
    • 1st : showing the size of the beam just before injecting into the cavity @ 10%
    • 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
    • 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
    • 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors

Note : the voltage ranges are not the same between the images

Manar Amer wrote:

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,

 

 

 

    Reply   Cavity Lock - Alphanov Amplifier , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics | cabling | software tek00000.png

This morning, while trying Guillaume solution of disconnecting the 24 V necessary to the second and third stage and raise the threshold of Temp preamp 1 (to only operate the ampli with first stage only)

Ronic disconnected the 24 volt connection, turned on amplifier LAL software but no MMD3 error or a big temp on Temp preamp 1 , software worked fine with no issue. 

we connected the 24 V again and the software worked with no issues.

Still not sure what happened for it to work correctly !!!!!!!!!!

  • We used the amplifier at 0% all day with no errors (to test , I turned it for 10 minutes @ 10% with no error too)
  • We changed the lase lock box with ThomX one (having Ethernet connection), it was connected to computer and works ok
  • Ronic installed a signal amplifier for the PDH error signal. (yet to be fully optimized)
  • we see a better coupling than before in the cavity.
  • The dominant mode is 00 mode, but we see many higher order modes, need to alight better.

 

 

 

 

 

Manar Amer wrote:

Today with Ronic and Daniele we attempted to lock the cavity, but the alphanov amplifier did not turn on because of an error in MMD3 (related to the third stage)

the new error is having the MMD3 error show on the alarm window of the LAL software, and having the Alarms on the line 3 of the alphanov software red for both EXT/CPU and Laser T Max (never been red before)

this is after a 2-week work stop, but the last time when Daniele changed the fiber connector it worked for 3 hours with no errors.

Ronic, is in contact with Guillaume from Alphanov to fix the issue !!!

Work scheduled all Wednesday, in hopes the issue is fixed by then.

Manar Amer wrote:

Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)

was done at the end due to excitation of higher order modes.

Wanted to improve the fundamental first, then increase the power.

Manar Amer wrote:
  • On Monday with Daniele,
    • we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
  • On Tuesday with Daniele and Kevin
    •  I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
    • removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm  , Reflection : 50 ohm  )
    • We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
  • Images:
    • 1st : showing the size of the beam just before injecting into the cavity @ 10%
    • 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
    • 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
    • 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors

Note : the voltage ranges are not the same between the images

Manar Amer wrote:

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,

 

 

 

 

    Reply   Cavity Lock - Alphanov Amplifier , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics | cabling | software 13x

Yesterday, with Ronic we locked ThomX cavity on the second resonance

Laser : 33.36 MHz , Amplifier : 0 % (worked with no errors all day)

adding image for the lock on the second resonance (locked), 00 Mode image, fit, Beam measurements

and image of the first resonance signal (not locked)

Info : when locking on the secondary resonance, we aligned better and adjusted on the oscillator CEP and increased the transmission signal 

Controllers connected to computer

through Ethernet :

  • Cavity Mirror motors
  • Lase lock
  • oscillator CEP and Frequency motors (easier to follow the drift with this controller, and we can see it moving)

by USB :

  • Amplifier controller
Manar Amer wrote:

This morning, while trying Guillaume solution of disconnecting the 24 V necessary to the second and third stage and raise the threshold of Temp preamp 1 (to only operate the ampli with first stage only)

Ronic disconnected the 24 volt connection, turned on amplifier LAL software but no MMD3 error or a big temp on Temp preamp 1 , software worked fine with no issue. 

we connected the 24 V again and the software worked with no issues.

Still not sure what happened for it to work correctly !!!!!!!!!!

  • We used the amplifier at 0% all day with no errors (to test , I turned it for 10 minutes @ 10% with no error too)
  • We changed the lase lock box with ThomX one (having Ethernet connection), it was connected to computer and works ok
  • Ronic installed a signal amplifier for the PDH error signal. (yet to be fully optimized)
  • we see a better coupling than before in the cavity.
  • The dominant mode is 00 mode, but we see many higher order modes, need to alight better.

 

 

 

 

 

Manar Amer wrote:

Today with Ronic and Daniele we attempted to lock the cavity, but the alphanov amplifier did not turn on because of an error in MMD3 (related to the third stage)

the new error is having the MMD3 error show on the alarm window of the LAL software, and having the Alarms on the line 3 of the alphanov software red for both EXT/CPU and Laser T Max (never been red before)

this is after a 2-week work stop, but the last time when Daniele changed the fiber connector it worked for 3 hours with no errors.

Ronic, is in contact with Guillaume from Alphanov to fix the issue !!!

Work scheduled all Wednesday, in hopes the issue is fixed by then.

Manar Amer wrote:

Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)

was done at the end due to excitation of higher order modes.

Wanted to improve the fundamental first, then increase the power.

Manar Amer wrote:
  • On Monday with Daniele,
    • we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
  • On Tuesday with Daniele and Kevin
    •  I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
    • removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm  , Reflection : 50 ohm  )
    • We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
  • Images:
    • 1st : showing the size of the beam just before injecting into the cavity @ 10%
    • 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
    • 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
    • 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors

Note : the voltage ranges are not the same between the images

Manar Amer wrote:

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,

 

 

 

 

 

    Reply   Cavity Lock - Alphanov Amplifier , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics | cabling | software 20230605_tek00001_33uW.png20230605_tek00002_82uW.png

Today, I installed a power meter at the transmission point of the 1st spherical mirror , transmission 3 ppm (direct from cavity window beam profiler, no filter)

We locked the cavity at 2 different resonances of the fundamental mode, the lock in both cases was stable for a round 1 minute.

at different transmitted power of 33.85 uW ( 11 W average inside cavity) and then locked again at 82.65 uW (27.5 W average inside cavity) transmitted power 

Note : the coupling is almost zero for both of the resonances locked !!!!

Oscillator : 33.33 MHz ( 33.356 MHz , frequency read on LAL software)

power injected into amplifier (after injecting into fiber and an EOM) : 3.886 mW (on LAL software)

Amplifier power : 0% (injected power into cavity ~ 300 mW)

Both images of oscilloscope have same voltage scales, only difference timescale and the color code is : 

yellow : transmission

blue    : reflection

green : piezo voltage

red     : error signal

 

 

Manar Amer wrote:

Yesterday, with Ronic we locked ThomX cavity on the second resonance

Laser : 33.36 MHz , Amplifier : 0 % (worked with no errors all day)

adding image for the lock on the second resonance (locked), 00 Mode image, fit, Beam measurements

and image of the first resonance signal (not locked)

Info : when locking on the secondary resonance, we aligned better and adjusted on the oscillator CEP and increased the transmission signal 

Controllers connected to computer

through Ethernet :

  • Cavity Mirror motors
  • Lase lock
  • oscillator CEP and Frequency motors (easier to follow the drift with this controller, and we can see it moving)

by USB :

  • Amplifier controller
Manar Amer wrote:

This morning, while trying Guillaume solution of disconnecting the 24 V necessary to the second and third stage and raise the threshold of Temp preamp 1 (to only operate the ampli with first stage only)

Ronic disconnected the 24 volt connection, turned on amplifier LAL software but no MMD3 error or a big temp on Temp preamp 1 , software worked fine with no issue. 

we connected the 24 V again and the software worked with no issues.

Still not sure what happened for it to work correctly !!!!!!!!!!

  • We used the amplifier at 0% all day with no errors (to test , I turned it for 10 minutes @ 10% with no error too)
  • We changed the lase lock box with ThomX one (having Ethernet connection), it was connected to computer and works ok
  • Ronic installed a signal amplifier for the PDH error signal. (yet to be fully optimized)
  • we see a better coupling than before in the cavity.
  • The dominant mode is 00 mode, but we see many higher order modes, need to alight better.

 

 

 

 

 

Manar Amer wrote:

Today with Ronic and Daniele we attempted to lock the cavity, but the alphanov amplifier did not turn on because of an error in MMD3 (related to the third stage)

the new error is having the MMD3 error show on the alarm window of the LAL software, and having the Alarms on the line 3 of the alphanov software red for both EXT/CPU and Laser T Max (never been red before)

this is after a 2-week work stop, but the last time when Daniele changed the fiber connector it worked for 3 hours with no errors.

Ronic, is in contact with Guillaume from Alphanov to fix the issue !!!

Work scheduled all Wednesday, in hopes the issue is fixed by then.

Manar Amer wrote:

Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)

was done at the end due to excitation of higher order modes.

Wanted to improve the fundamental first, then increase the power.

Manar Amer wrote:
  • On Monday with Daniele,
    • we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
  • On Tuesday with Daniele and Kevin
    •  I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
    • removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm  , Reflection : 50 ohm  )
    • We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
  • Images:
    • 1st : showing the size of the beam just before injecting into the cavity @ 10%
    • 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
    • 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
    • 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors

Note : the voltage ranges are not the same between the images

Manar Amer wrote:

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,

 

 

 

 

 

 

    Reply   Cavity Lock - Alphanov Amplifier , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics | cabling | software tek00001.png20230607_183626_Frep_CEP_motors_smartAct.jpg20230607_183641_transmitted_power.jpgtek00002.png

Today with Ronic,

starting with major events that happened (Water circulation + ThomX valves )  

  • Early in the morning a company worked on the water network, it seemed it was not restarted properly so there was no water circulation in the main ThomX pipes.
    • There was an error on the ThomX amplifier chiller , which was noticed at the end of the morning work , where we couldn't find the cavity resonance , could be due to the temperature increase of the amplifer.
    • Solved :  the issue was solved by restarting the water network and the amplifier chiller was restarted an no error found and temp. Stabilized around 25°
  • Around 12h15 ThomX there was a power cut off for less than a second (micro cutt), which cased all the valves to close, the valve air compressor did not restart after the power cut off, we restarted it with Daniele in the evening with the help of Marie and the valves open around 5-6 bar , you will see then all the controllers green and the valves will open.

Results of the day :

  • All the power supplies and function generators are under the table, nothing on the table (reduced noise on the signal)
  • EOM of Oscillator Off (for now, might turn on if needed)
  • Amplifier on 0% , output ~ 300 mW
  • Locked on the first resonance of the cavity,  lase lock parameters to be optimized better
  • Transmitted power increased to ~ 267.2 uW
  • Average power inside cavity = 89. 06 W , peak power = 0.2 kW (pulse width = 12.6 ps , frep = 33.36 MHz )
  • current effective cavity gain > 300 

Oscilloscope :

  •  error signal, not shown as it was too noisy to have in the image
  • blue : reflection (low bypass filter ), yellow : transmission (resistance of 100 k ohm added), green : piezo signal
  • 1st image : showing lock and scan regions.
  • 2nd image: time zoom on the locked signal 

oscillators controller / smartAct:

  • Frequency and CEP control , the parameters in the photo attached are the best for the moment to see the smooth change of the resonance peaks when operating the motors, and we can stay FSR range and compensate the variation easily. 
  • even though we stayed at closed loop (for easier adjustment), we still managed to lock , will try to switch off later to see if reduces a lot of noise. 

 

 

Manar Amer wrote:

Today, I installed a power meter at the transmission point of the 1st spherical mirror , transmission 3 ppm (direct from cavity window beam profiler, no filter)

We locked the cavity at 2 different resonances of the fundamental mode, the lock in both cases was stable for a round 1 minute.

at different transmitted power of 33.85 uW ( 11 W average inside cavity) and then locked again at 82.65 uW (27.5 W average inside cavity) transmitted power 

Note : the coupling is almost zero for both of the resonances locked !!!!

Oscillator : 33.33 MHz ( 33.356 MHz , frequency read on LAL software)

power injected into amplifier (after injecting into fiber and an EOM) : 3.886 mW (on LAL software)

Amplifier power : 0% (injected power into cavity ~ 300 mW)

Both images of oscilloscope have same voltage scales, only difference timescale and the color code is : 

yellow : transmission

blue    : reflection

green : piezo voltage

red     : error signal

 

 

Manar Amer wrote:

Yesterday, with Ronic we locked ThomX cavity on the second resonance

Laser : 33.36 MHz , Amplifier : 0 % (worked with no errors all day)

adding image for the lock on the second resonance (locked), 00 Mode image, fit, Beam measurements

and image of the first resonance signal (not locked)

Info : when locking on the secondary resonance, we aligned better and adjusted on the oscillator CEP and increased the transmission signal 

Controllers connected to computer

through Ethernet :

  • Cavity Mirror motors
  • Lase lock
  • oscillator CEP and Frequency motors (easier to follow the drift with this controller, and we can see it moving)

by USB :

  • Amplifier controller
Manar Amer wrote:

This morning, while trying Guillaume solution of disconnecting the 24 V necessary to the second and third stage and raise the threshold of Temp preamp 1 (to only operate the ampli with first stage only)

Ronic disconnected the 24 volt connection, turned on amplifier LAL software but no MMD3 error or a big temp on Temp preamp 1 , software worked fine with no issue. 

we connected the 24 V again and the software worked with no issues.

Still not sure what happened for it to work correctly !!!!!!!!!!

  • We used the amplifier at 0% all day with no errors (to test , I turned it for 10 minutes @ 10% with no error too)
  • We changed the lase lock box with ThomX one (having Ethernet connection), it was connected to computer and works ok
  • Ronic installed a signal amplifier for the PDH error signal. (yet to be fully optimized)
  • we see a better coupling than before in the cavity.
  • The dominant mode is 00 mode, but we see many higher order modes, need to alight better.

 

 

 

 

 

Manar Amer wrote:

Today with Ronic and Daniele we attempted to lock the cavity, but the alphanov amplifier did not turn on because of an error in MMD3 (related to the third stage)

the new error is having the MMD3 error show on the alarm window of the LAL software, and having the Alarms on the line 3 of the alphanov software red for both EXT/CPU and Laser T Max (never been red before)

this is after a 2-week work stop, but the last time when Daniele changed the fiber connector it worked for 3 hours with no errors.

Ronic, is in contact with Guillaume from Alphanov to fix the issue !!!

Work scheduled all Wednesday, in hopes the issue is fixed by then.

Manar Amer wrote:

Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)

was done at the end due to excitation of higher order modes.

Wanted to improve the fundamental first, then increase the power.

Manar Amer wrote:
  • On Monday with Daniele,
    • we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
  • On Tuesday with Daniele and Kevin
    •  I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
    • removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm  , Reflection : 50 ohm  )
    • We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
  • Images:
    • 1st : showing the size of the beam just before injecting into the cavity @ 10%
    • 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
    • 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
    • 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors

Note : the voltage ranges are not the same between the images

Manar Amer wrote:

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,

 

 

 

 

 

 

 

    Reply   Cavity Lock - Alphanov Amplifier , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics | cabling | software tek00001_noise_reduced_first_resonance.pngtek00000_cavity_lock_with_noise_from_odcillator_controller.png

In the morning, we locked on the first resonance. With an increase on the amplifier power @ 20-25% and 30%, which reduced the noise

we manage to get a coupling when improving the CEP up to 55 - 60 % coupling : put there is still a drift on the CEP

The controller for the oscillator frep and CEP produces a lot of noise, even if it is disconnected. we need to switch it off to remove its effect (Kevin will order a new one)

The image attached is of the oscilloscope lock on the first resonance @30% amplifier power ~ 10 W injected into cavity,

transmitted power ~ 36 mW (cavity average power 12 kW)

cavity effective gain > 1200

 

in the afternoon, we installed and aligned a gentec power meter we can monitor from the computer. (will need to buy a permeant one for ThomX)  

there was an issue with the laselock USB connection, yet to be solved.

and the control of the computer.

Manar Amer wrote:

Today with Ronic,

starting with major events that happened (Water circulation + ThomX valves )  

  • Early in the morning a company worked on the water network, it seemed it was not restarted properly so there was no water circulation in the main ThomX pipes.
    • There was an error on the ThomX amplifier chiller , which was noticed at the end of the morning work , where we couldn't find the cavity resonance , could be due to the temperature increase of the amplifer.
    • Solved :  the issue was solved by restarting the water network and the amplifier chiller was restarted an no error found and temp. Stabilized around 25°
  • Around 12h15 ThomX there was a power cut off for less than a second (micro cutt), which cased all the valves to close, the valve air compressor did not restart after the power cut off, we restarted it with Daniele in the evening with the help of Marie and the valves open around 5-6 bar , you will see then all the controllers green and the valves will open.

Results of the day :

  • All the power supplies and function generators are under the table, nothing on the table (reduced noise on the signal)
  • EOM of Oscillator Off (for now, might turn on if needed)
  • Amplifier on 0% , output ~ 300 mW
  • Locked on the first resonance of the cavity,  lase lock parameters to be optimized better
  • Transmitted power increased to ~ 267.2 uW
  • Average power inside cavity = 89. 06 W , peak power = 0.2 kW (pulse width = 12.6 ps , frep = 33.36 MHz )
  • current effective cavity gain > 300 

Oscilloscope :

  •  error signal, not shown as it was too noisy to have in the image
  • blue : reflection (low bypass filter ), yellow : transmission (resistance of 100 k ohm added), green : piezo signal
  • 1st image : showing lock and scan regions.
  • 2nd image: time zoom on the locked signal 

oscillators controller / smartAct:

  • Frequency and CEP control , the parameters in the photo attached are the best for the moment to see the smooth change of the resonance peaks when operating the motors, and we can stay FSR range and compensate the variation easily. 
  • even though we stayed at closed loop (for easier adjustment), we still managed to lock , will try to switch off later to see if reduces a lot of noise. 

 

 

Manar Amer wrote:

Today, I installed a power meter at the transmission point of the 1st spherical mirror , transmission 3 ppm (direct from cavity window beam profiler, no filter)

We locked the cavity at 2 different resonances of the fundamental mode, the lock in both cases was stable for a round 1 minute.

at different transmitted power of 33.85 uW ( 11 W average inside cavity) and then locked again at 82.65 uW (27.5 W average inside cavity) transmitted power 

Note : the coupling is almost zero for both of the resonances locked !!!!

Oscillator : 33.33 MHz ( 33.356 MHz , frequency read on LAL software)

power injected into amplifier (after injecting into fiber and an EOM) : 3.886 mW (on LAL software)

Amplifier power : 0% (injected power into cavity ~ 300 mW)

Both images of oscilloscope have same voltage scales, only difference timescale and the color code is : 

yellow : transmission

blue    : reflection

green : piezo voltage

red     : error signal

 

 

Manar Amer wrote:

Yesterday, with Ronic we locked ThomX cavity on the second resonance

Laser : 33.36 MHz , Amplifier : 0 % (worked with no errors all day)

adding image for the lock on the second resonance (locked), 00 Mode image, fit, Beam measurements

and image of the first resonance signal (not locked)

Info : when locking on the secondary resonance, we aligned better and adjusted on the oscillator CEP and increased the transmission signal 

Controllers connected to computer

through Ethernet :

  • Cavity Mirror motors
  • Lase lock
  • oscillator CEP and Frequency motors (easier to follow the drift with this controller, and we can see it moving)

by USB :

  • Amplifier controller
Manar Amer wrote:

This morning, while trying Guillaume solution of disconnecting the 24 V necessary to the second and third stage and raise the threshold of Temp preamp 1 (to only operate the ampli with first stage only)

Ronic disconnected the 24 volt connection, turned on amplifier LAL software but no MMD3 error or a big temp on Temp preamp 1 , software worked fine with no issue. 

we connected the 24 V again and the software worked with no issues.

Still not sure what happened for it to work correctly !!!!!!!!!!

  • We used the amplifier at 0% all day with no errors (to test , I turned it for 10 minutes @ 10% with no error too)
  • We changed the lase lock box with ThomX one (having Ethernet connection), it was connected to computer and works ok
  • Ronic installed a signal amplifier for the PDH error signal. (yet to be fully optimized)
  • we see a better coupling than before in the cavity.
  • The dominant mode is 00 mode, but we see many higher order modes, need to alight better.

 

 

 

 

 

Manar Amer wrote:

Today with Ronic and Daniele we attempted to lock the cavity, but the alphanov amplifier did not turn on because of an error in MMD3 (related to the third stage)

the new error is having the MMD3 error show on the alarm window of the LAL software, and having the Alarms on the line 3 of the alphanov software red for both EXT/CPU and Laser T Max (never been red before)

this is after a 2-week work stop, but the last time when Daniele changed the fiber connector it worked for 3 hours with no errors.

Ronic, is in contact with Guillaume from Alphanov to fix the issue !!!

Work scheduled all Wednesday, in hopes the issue is fixed by then.

Manar Amer wrote:

Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)

was done at the end due to excitation of higher order modes.

Wanted to improve the fundamental first, then increase the power.

Manar Amer wrote:
  • On Monday with Daniele,
    • we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
  • On Tuesday with Daniele and Kevin
    •  I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
    • removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm  , Reflection : 50 ohm  )
    • We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
  • Images:
    • 1st : showing the size of the beam just before injecting into the cavity @ 10%
    • 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
    • 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
    • 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors

Note : the voltage ranges are not the same between the images

Manar Amer wrote:

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,

 

 

 

 

 

 

 

 

    Reply   Cavity Lock - Alphanov Amplifier , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics | cabling | software 

The computer is possible to connect from the control room,

we are able to run remotely the amplifier, cavity motors, oscillator motors, lase lock (note: there is an issue with the keyboard, we are not able to use it with remote access!!!!!!!!)

Note on Amplifier: On Friday, Daniele and Ronic faced again the temperature issue that came up before. it is due to the fluctuation of the diode temperature.

It was fixed today by Ronic and Daniele. it seems there are three diodes for monitoring the temperature, only one was connected, and it had issues, with some directions from alphanov. 

Ronic just removed the defective diode and soldered (connected) a different one. The amplifier should work without this issue. 

now we start searching for X-rays after locking the cavity remotely  

Manar Amer wrote:

In the morning, we locked on the first resonance. With an increase on the amplifier power @ 20-25% and 30%, which reduced the noise

we manage to get a coupling when improving the CEP up to 55 - 60 % coupling : put there is still a drift on the CEP

The controller for the oscillator frep and CEP produces a lot of noise, even if it is disconnected. we need to switch it off to remove its effect (Kevin will order a new one)

The image attached is of the oscilloscope lock on the first resonance @30% amplifier power ~ 10 W injected into cavity,

transmitted power ~ 36 mW (cavity average power 12 kW)

cavity effective gain > 1200

 

in the afternoon, we installed and aligned a gentec power meter we can monitor from the computer. (will need to buy a permeant one for ThomX)  

there was an issue with the laselock USB connection, yet to be solved.

and the control of the computer.

Manar Amer wrote:

Today with Ronic,

starting with major events that happened (Water circulation + ThomX valves )  

  • Early in the morning a company worked on the water network, it seemed it was not restarted properly so there was no water circulation in the main ThomX pipes.
    • There was an error on the ThomX amplifier chiller , which was noticed at the end of the morning work , where we couldn't find the cavity resonance , could be due to the temperature increase of the amplifer.
    • Solved :  the issue was solved by restarting the water network and the amplifier chiller was restarted an no error found and temp. Stabilized around 25°
  • Around 12h15 ThomX there was a power cut off for less than a second (micro cutt), which cased all the valves to close, the valve air compressor did not restart after the power cut off, we restarted it with Daniele in the evening with the help of Marie and the valves open around 5-6 bar , you will see then all the controllers green and the valves will open.

Results of the day :

  • All the power supplies and function generators are under the table, nothing on the table (reduced noise on the signal)
  • EOM of Oscillator Off (for now, might turn on if needed)
  • Amplifier on 0% , output ~ 300 mW
  • Locked on the first resonance of the cavity,  lase lock parameters to be optimized better
  • Transmitted power increased to ~ 267.2 uW
  • Average power inside cavity = 89. 06 W , peak power = 0.2 kW (pulse width = 12.6 ps , frep = 33.36 MHz )
  • current effective cavity gain > 300 

Oscilloscope :

  •  error signal, not shown as it was too noisy to have in the image
  • blue : reflection (low bypass filter ), yellow : transmission (resistance of 100 k ohm added), green : piezo signal
  • 1st image : showing lock and scan regions.
  • 2nd image: time zoom on the locked signal 

oscillators controller / smartAct:

  • Frequency and CEP control , the parameters in the photo attached are the best for the moment to see the smooth change of the resonance peaks when operating the motors, and we can stay FSR range and compensate the variation easily. 
  • even though we stayed at closed loop (for easier adjustment), we still managed to lock , will try to switch off later to see if reduces a lot of noise. 

 

 

Manar Amer wrote:

Today, I installed a power meter at the transmission point of the 1st spherical mirror , transmission 3 ppm (direct from cavity window beam profiler, no filter)

We locked the cavity at 2 different resonances of the fundamental mode, the lock in both cases was stable for a round 1 minute.

at different transmitted power of 33.85 uW ( 11 W average inside cavity) and then locked again at 82.65 uW (27.5 W average inside cavity) transmitted power 

Note : the coupling is almost zero for both of the resonances locked !!!!

Oscillator : 33.33 MHz ( 33.356 MHz , frequency read on LAL software)

power injected into amplifier (after injecting into fiber and an EOM) : 3.886 mW (on LAL software)

Amplifier power : 0% (injected power into cavity ~ 300 mW)

Both images of oscilloscope have same voltage scales, only difference timescale and the color code is : 

yellow : transmission

blue    : reflection

green : piezo voltage

red     : error signal

 

 

Manar Amer wrote:

Yesterday, with Ronic we locked ThomX cavity on the second resonance

Laser : 33.36 MHz , Amplifier : 0 % (worked with no errors all day)

adding image for the lock on the second resonance (locked), 00 Mode image, fit, Beam measurements

and image of the first resonance signal (not locked)

Info : when locking on the secondary resonance, we aligned better and adjusted on the oscillator CEP and increased the transmission signal 

Controllers connected to computer

through Ethernet :

  • Cavity Mirror motors
  • Lase lock
  • oscillator CEP and Frequency motors (easier to follow the drift with this controller, and we can see it moving)

by USB :

  • Amplifier controller
Manar Amer wrote:

This morning, while trying Guillaume solution of disconnecting the 24 V necessary to the second and third stage and raise the threshold of Temp preamp 1 (to only operate the ampli with first stage only)

Ronic disconnected the 24 volt connection, turned on amplifier LAL software but no MMD3 error or a big temp on Temp preamp 1 , software worked fine with no issue. 

we connected the 24 V again and the software worked with no issues.

Still not sure what happened for it to work correctly !!!!!!!!!!

  • We used the amplifier at 0% all day with no errors (to test , I turned it for 10 minutes @ 10% with no error too)
  • We changed the lase lock box with ThomX one (having Ethernet connection), it was connected to computer and works ok
  • Ronic installed a signal amplifier for the PDH error signal. (yet to be fully optimized)
  • we see a better coupling than before in the cavity.
  • The dominant mode is 00 mode, but we see many higher order modes, need to alight better.

 

 

 

 

 

Manar Amer wrote:

Today with Ronic and Daniele we attempted to lock the cavity, but the alphanov amplifier did not turn on because of an error in MMD3 (related to the third stage)

the new error is having the MMD3 error show on the alarm window of the LAL software, and having the Alarms on the line 3 of the alphanov software red for both EXT/CPU and Laser T Max (never been red before)

this is after a 2-week work stop, but the last time when Daniele changed the fiber connector it worked for 3 hours with no errors.

Ronic, is in contact with Guillaume from Alphanov to fix the issue !!!

Work scheduled all Wednesday, in hopes the issue is fixed by then.

Manar Amer wrote:

Added Note: the coupling we get is at low power, only preamplifier is on (~ 300 mW)

was done at the end due to excitation of higher order modes.

Wanted to improve the fundamental first, then increase the power.

Manar Amer wrote:
  • On Monday with Daniele,
    • we did full characterization of amplifier beam @ 10% amplification, adjusted the telescope accordingly and injected into the cavity aligned and improved on the fundamental mode.
  • On Tuesday with Daniele and Kevin
    •  I added a low pass filter between 1 - 1.9 MHz on the reflected signal, to reject the oscillator signal (33.33 MHz)
    • removed the external resistors on the signal, only the internal resistance of the oscilloscope used ( transmission : 1 M ohm  , Reflection : 50 ohm  )
    • We connected the motors to control the CEP and adjusted on them until we reached ~ 5 -10% coupling
  • Images:
    • 1st : showing the size of the beam just before injecting into the cavity @ 10%
    • 2nd: signal with only improved alignment (not CEP adjustment yet), no low pass filter added yet
    • 3rd : zoom on the signals (reflection and error) while adjusting on the CEP
    • 4th :attempting to lock the cavity , after getting the max coupling using the CEP motors

Note : the voltage ranges are not the same between the images

Manar Amer wrote:

Yesterday,

  • we installed the second high power reflecting mirrors at the output of the amplifier.
  • We aligned and installed a telescope that has adjustable distance between the 2 lenses of -100 mm, and + 200 mm, with approximate diameter at the injection window of  ~ 4.5 mm.
  • Amplifier was turned on only at the first stage only (output power ~ 300 mW) aligned and injected into the cavity.
  • Measured the repetition frequency of the One Five oscillators, Frep = 33.326239 MHz
  • Changed the FSR of the cavity to match it.
  • Using the Piezo scan on the oscillator, we observed 00 Mode, Fundamental. the shape is the same 
  • We see transmission on the diode, but there was zero coupling observed (yet to be investigated experimentally )
    • a reason could be the CEP of the oscillator ,
    • bad matching of the beam waist position,
    • noise from the amplifier (as we were operating it at very low power)
    • additional alignment needed,

 

 

 

 

 

 

 

 

 

Entry   Onefive Origami 33MHz specs, posted by Loïc Amoudry at Optical room about lasers and optics 6x

Power during ~20h, power after turn on, spectrum image and excel datas, polarisation and pulse length.

Powermeter measurments were done with a /10 filter.. 3 mW means 30 mW

Important: Onefive does not automatically locks after being turned on.

Entry   End of vacuum test and cavity ready for cleaning, posted by Loïc Amoudry at Optical room about vacuum IMG_20181221_1110152.jpgIMG_20181221_1119082.jpgIMG_20181221_1113122.jpgIMG_20181221_1124492.jpg

Vacuum tests finished on 19 december 2018.

Spent last two days to discuss the importance of removing aluminium and baking wires before to insert mirrors for cleanliness issues. Finally we took pictures to support our demand.

Cavity should be ready to clean for the new term 2019

Le 19/12/2018 à 14:56, Bruno Mercier a écrit :

Bonjour Daniele, Loïc,

Nous sommes à votre disposition pour effectuer une entrée d'azote sur
la cavité.

Bruno
Entry   Cleaning outside IP cavity, posted by Loïc Amoudry at Optical room about lasers and optics | utilities Avant_nettoyage.jpgApres_une_semaine_nettoyage_lingettes_et_brosseadents.jpgSans_UV_traces_scotch_nettoyees_apres_baking.jpg

From 7th january to 11th january.

 

Entry   End of cleaning, plastic fixed on cavity, posted by Loïc Amoudry at Optical room about utilities covered_by_plastic.pngcovered_by_plastic2.png

08/02/19 - End of the cleaning process.

Entry   Mirrors checked with binocular microscope, posted by Loïc Amoudry at Optical room about lasers and optics 41_THOMX_SphericalZoom1x.jpg49_THOMX_M1ReflectiveZoom4x.jpg

29/01/19 - ThomX mirrors have also been checked with binocular microscope. They shown the same spots as S-BOX mirrors. 

1st image shows reflective face of a spherical mirror.

2nd image shows reflective face of M1 (sapphire).

Entry   Mirrors sent to LMA, posted by Loïc Amoudry at Other about lasers and optics 

30/01/19 - As LMA people came, we discussed about the spots on ThomX mirrors which should be clean. As mentionned in "SBOX commissioning", these spots are normal.

An other question appeared during the conversation about the storage time and location of these mirrors. They have been stored for more than two years in plastic boxes. LMA people explained that some company do not let anything more than 6 months in these boxes for pollution issues.

We merged into the same conclusion, clean the mirrors in case of pollution. Finally, they took the mirrors and we will go to LMA to recover them between 11/02 and 14/02. 

Entry   Basic alignment & alignment motors check, posted by Loïc Amoudry at Optical room about mechanics | lasers and optics 

14/02/19: The 4 motors but Mby control works. Mby motor home does not bring it to its "middle" position but to the end of its range. Issue circumvented using the higher rights software.

The beam is now going through the cavity windows (from P1 to S2).