    Reply   Alphanov amplifier issue, posted by Ronic Chiche at ThomX igloo about lasers and optics 
The laser amplifier controller has been sent yesterday and received today by Alphannov.

they should do the assessment quickly...




Ronic
    Reply   Alphanov amplifier issue, posted by Ronic Chiche at ThomX igloo about lasers and optics Sécurité_et_CC_ampli_Alphanov.pdf
Today, I did a Teams meeting with Guillaume.
He didn't see any problem with the controller.
normally, the controller will be shipped to the lab tomorrow.
    Reply   Alphanov amplifier issue, posted by Ronic Chiche at ThomX igloo about lasers and optics 
Yesterday, we received the controller back from Alphanov
and I tested it immediately in the optical room.
if the software is properly talking to the controller, everything works normally.
Entry   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics 
today with Manar, we brought everything to make the profile measurement.

the chiller was in error because of the water level : we filled it in.
! warning ! there is no alarm signal, only a message on the screen.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics Tutoriel_PD_PULS.pdf2023_Alphanov_Beam_Profile_setup.jpg
yesterday morning with Manar,

1) we fixed the threshold level of the laser input signal which makes the measurement of the repetition rate.
(if the rep rate is not measured properly, the amplifier safety interlock stops immediately the amplification).
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics 
this morning with Manar,

- we checked first the chiller problem.
the fluctations of +/-2° around 25°C are still there.
Entry   Alphanov amplifier beam profile Data , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 8x
Adding the schematic of the measurements.,

the 4° Beam Deviation, AR Coating ( B coating)

The 3 data points taken of the beam profile are fitted in Gaussian Beam with taking the beam waist at the output of the compressor box.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software error_signal_Temperature_case_amplifier.jpg
Today with Ronic 

For the Temperature fluctuation :


we checked the valves of the tubes bumping water into the chiller from ThomX main water supply, they were
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics 
this morning with Manar,

- we fixed the chiller problem (+/-2°C fluctuation around the set temperature value = 25°C).
we opened and closed several times the water valve of the primary circuit to remove air bubles => no effect on the temperature variations.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 
Ronic on Friday operated the Amplifier from the Alphanov software, switched it off then turned on the LAL software,

and it worked for 40 min without issues or error appearing,

 
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 
the laser amplifier worked for some days but it ended to a "watchdog" alarm which switches off the amplifier... investigation is ongoing




Ronic
Chiche wrote:



Ronic on Friday operated the Amplifier from the Alphanov software,
    Reply   Alphanov amplifier beam profile Data , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 
This Morning with Victor the amplifier was turned on @ 20% amplification (output ~ 9 W ) for approximately 3 hours and no error appeared.

I took data of the beam profile at different points without a lens after the 2 wedges 

@ 0% (only pump beam), output ~ 300 mW (taken from previous data) , output after 2 wedges 2.2 uW  
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software scope.jpg
we have had this new issue (watchdog alarm switching off the amplifier) for several days.

Alphanov mentionned it could come from the input oscillator stability (modelock loss for example).

Today,
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 20230208_error_watchdog.jpeg
With Daniele after, we took a readings of the beam with a lens of 750 mm.

From the observed data, we can state that the beam is not completely Gaussian M2 > 1, due to the variation of the ellipticity of the beam
along the focused path 
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 
This morning,

- with the "Alphanov" software => Central panel => "system" button => Watchdog "Period Max" was at 100ns
and has been set to 200ns.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 
This morning with Daniele, we did a second measurement of the beam profile using a different lens of focal 400 mm.

at a power amplification of 20%

the amplifier stayed on for around one hour and a half with no errors. 
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software Analyse_Données_20_fevrier.pptx
On Monday 20th of February, Daniele and Viktor took some beam profile measurement data @ 50% of power amplification.

data and analysis can be found in the attached file.




Ronic
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 

Yesterday, Guillaume Machinet and Vincent Clet from Alphanov came to fixe some issues with the laser amplifier:

- 1) they fixed the "rep. rate" measurement issue.
previously, the voltage range in which it was possible to detect the correct rep rate (33MHz) was very narrow ~ 50mV.
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software WhatsApp_Image_2023-02-27_at_6.07.00_PM.jpeg
This morning with Victor we took a reading of the beam profile using a 400 mm lens at 70 % and 100 % amplification.

Will process the data and add after

Note : after an hour and half of amplifier on at 70% we increased the power to 100%, and after ~ 10 minutes the amplifier turned off and showed
    Reply   Alphanov amplifier beam profile measurement, posted by Ronic Chiche at ThomX igloo about lasers and optics | software 20230308_124313.jpg
this morning with Manar,
we checked the Alphanov amplifier issue : "The laser has been stopped because its power is too far from the nominal power".

the photodiode voltage measuring the output power has drifted again and then, the estimated output power was wrong, triggering an alarm issue.