87
Wed Apr 26 16:39:59 2023
Blanc
Fixed
info
lasers and optics
Optical room
Couplage and CEP
Pour vérifier la cohérence
des mesures précédentes, j'ai
fait la même analyse avec l'intensité
88
Fri Apr 28 15:28:06 2023
Blanc
Fixed
info
lasers and optics
Optical room
Verification CEP
On a voulu véifier que le changement
de couplage induit par la CEP était
une conséquence du changement de CEP
89
Fri May 26 18:08:27 2023
Blanc
Fixed
info
lasers and optics
Optical room
Modulation EOM
EOM dans la lockline:
Image 1 (sans 2è étage):
on voit la modulation en réflexion
24
Mon Oct 4 14:57:04 2021
Aurélien Martens
Fixed
info
detectors and electronics | software
Optical room
ip adress of smaract driver
the ip adress of the smaract driver is
now 10.0.53.10:5000
please think about putting such
28
Tue Oct 5 11:35:16 2021
Aurélien Martens
Fixed
report
detectors and electronics | software
Optical room
new position of smaract motors
a cavity length measurement gave 880.9Mhz.
The m1/M2 have been put to +5.1000mm both
M3/M4 srill at -2.9mm
29
Tue Oct 5 12:18:34 2021
Aurélien Martens
Fixed
report
detectors and electronics | software
Optical room
new position of smaract motors
to find the fundamental mode we moved the
cavity mirrors to new positions:
+5.104mm for both planar mirrors
30
Tue Oct 5 12:23:51 2021
Aurélien Martens
Fixed
report
detectors and electronics | software
Optical room
new position of smaract motors
the M1 is not at 5.1045mm to put the fundamental
on the scanning range of the laser piezo
M2 still at 5.104mm
31
Tue Oct 5 16:23:04 2021
Aurélien Martens
Fixed
report
detectors and electronics | software
Optical room
new position of smaract motors
after 3hours (1pm --> 4pm) the cavity
moved by 12um in total
new positions: +5.1065mm on M1
27
Mon Oct 4 15:38:43 2021
Aurélien Martens
Fixed
info
detectors and electronics
Optical room
phone is noise
doing the sensitivty measurement of the
piezo we observed satellite peaks around
800 to 900 Mhz. Removing the phones from