  Status of commissioning, report also here plots are reports., Page 5 of 5  Not logged in ELOG logo
ID Date Author Status Type Categorydown Location Title Text Attachments
  87   Wed Apr 26 16:39:59 2023 BlancFixedinfolasers and opticsOptical roomCouplage and CEPPour vérifier la cohérence
des mesures précédentes, j'ai
fait la même analyse avec l'intensité
 Fit_Gain_en_fonction_de_la_CEP_Preamp.pngFit_Gain_en_fonction_de_la_CEP_Preamp&Etage1Lockline.png 
  88   Fri Apr 28 15:28:06 2023 BlancFixedinfolasers and opticsOptical roomVerification CEP On a voulu véifier que le changement
de couplage induit par la CEP était
une conséquence du changement de CEP
  
  89   Fri May 26 18:08:27 2023 BlancFixedinfolasers and opticsOptical roomModulation EOMEOM dans la lockline:

Image 1 (sans 2è étage):
on voit la modulation en réflexion
 osc_preamp_AOM100_etage1_EOM.pngosc_preamp_AOM100_etage1_EOM_etage2.pngosc_preamp_AOM100_EOMbreadboard_matlab.pngosc_preamp_AOM100_EOMbreadboard.png 
  24   Mon Oct 4 14:57:04 2021 Aurélien MartensFixedinfodetectors and electronics | softwareOptical roomip adress of smaract driverthe ip adress of the smaract driver is
now 10.0.53.10:5000

please think about putting such
  
  28   Tue Oct 5 11:35:16 2021 Aurélien MartensFixedreportdetectors and electronics | softwareOptical roomnew position of smaract motorsa cavity length measurement gave 880.9Mhz.
The m1/M2 have been put to +5.1000mm both
M3/M4 srill at -2.9mm
  
  29   Tue Oct 5 12:18:34 2021 Aurélien MartensFixedreportdetectors and electronics | softwareOptical roomnew position of smaract motorsto find the fundamental mode we moved the
cavity mirrors to new positions:

+5.104mm for both planar mirrors
  
  30   Tue Oct 5 12:23:51 2021 Aurélien MartensFixedreportdetectors and electronics | softwareOptical roomnew position of smaract motorsthe M1 is not at 5.1045mm to put the fundamental
on the scanning range of the laser piezo

M2 still at 5.104mm
  
  31   Tue Oct 5 16:23:04 2021 Aurélien MartensFixedreportdetectors and electronics | softwareOptical roomnew position of smaract motorsafter 3hours (1pm --> 4pm) the cavity
moved by 12um in total

new positions: +5.1065mm on M1
  
  27   Mon Oct 4 15:38:43 2021 Aurélien MartensFixedinfodetectors and electronicsOptical roomphone is noisedoing the sensitivty measurement of the
piezo we observed satellite peaks around
800 to 900 Mhz. Removing the phones from
  