  Status of commissioning, report also here plots are reports., Page 2 of 5  Not logged in ELOG logo
ID Date Author Status Type Categorydown Location Title Text Attachments
  26   Mon Oct 4 15:24:14 2021 Aurélien MartensFixedreportlasers and optics | detectors and electronicsOptical roompiezo actuation rangeA 10V applied on the laser piezo was found
to induce a frequecny change of about 5kHz,
compatible with expectations from the piezo
  
  33   Thu Oct 7 17:53:53 2021 BlancFixedinfolasers and optics | detectors and electronicsOptical roomFiness and coupling measurementWe did the measurement of the Finess and
coupling with the NKT, with 2 EOM, without
AOM (so the lock signal is noisy), on the
 scope_10.pngscope_12.png 
  34   Fri Oct 8 16:02:19 2021 BlancFixedinfolasers and optics | detectors and electronicsOptical roomSetup and measurementsTo summerize the results of this week :

We made the setup in attachement
5 and align it to have the good mode in the
 acquisition_lock_GHz.pngSignal_d'erreur_PDH_sur_GHz.pngRésonances_dûes_au_pic_à_30kHz_dans_le_phase_noise.pngClose_loop_vs_open_loop_GHz.pngminicav_setup_Lock_GHz.PNG 
  37   Wed Oct 20 14:54:00 2021 Ronic ChicheFixedinfolasers and optics | detectors and electronicsOptical roomPhase noise measurement of the Amplitude MIKANMIKAN phase noise and RIN measurements:




 MIKAN_optical_phase_noise_comparison_with_NKT_CW_oscillator_&_RIN_@54MHz.png 
  44   Mon Dec 6 11:22:08 2021 Kevin DuprazUnder Processinfolasers and optics | detectors and electronicsOptical roomCameras Calibration 




 grab2021-12-06_10-51-34_325_microscope.bmpgrab2021-12-06_10-52-11_108_microscope_calib_05mm_ruler.bmp 
  45   Tue Dec 7 11:48:33 2021 Kevin DuprazUnder Processinfolasers and optics | detectors and electronicsOptical roomCameras CalibrationAfter check a mistake has been found on
the magnification. This seems to be the good
calibration (feel free to cross check). The
 Plot_of_NF_Trans_calib_hex2mm__2021-06-23__17-06-49.pngPlot_of_NF_Refl_calib_hex2mm__2021-06-23__16-59-26.png 
  53   Mon Mar 28 17:10:33 2022 BlancFixedinfolasers and optics | detectors and electronicsOptical roominput and output power of the AOM 50MHzInput power of the AOM :

150mW

output power of the AOM :
  
  54   Tue May 17 17:28:31 2022 BlancFixedissuelasers and optics | detectors and electronicsOptical roomDifférence de fréquence Tangor/Lockline- On a observé que le fondamental
du Tangor et celui de la Lockline n'étaient
pas situés à la même
 Freq_Tangor_Lockline_120_MHz.PNGFreq_Tangor_Lockline_80_MHz.PNGLock_Tangor_2W.PNG 
  6   Mon Apr 19 18:47:28 2021 Manar AmerFixedreportlasers and opticsOtherAlignment using Helium Neon Laser 05 April  2021 :  A rough alignment
of the cavity was done using the Helium Neon
Laser.
 20210202__Helium_Neon_laser_parameters.jpg 
  8   Tue Apr 27 13:57:15 2021 Manar AmerFixedreportlasers and opticsOtherAlignment using Koheras Following the Helium Neon Alignment + change
in the distance between the mirrors to be;
M3-M4 = 90.5 mm , M1-M2 = 80.2 mm  -→
 6x 
  9   Thu Apr 29 06:14:09 2021 Manar AmerFixedreportlasers and opticsOtherMode observed and cavity Lock ** Following up from yesterday an observation
about the TM00 mode : it has been seen with
similar dimensions and its position is the
 20212804_Fundamental_mode_in_cavity_2D_-_Position.png20212804_Cavity_Coupling_with_Koheras.jpg20212804_Temperature_Curve.jpg 
  12   Fri May 7 15:40:10 2021 Ronic ChicheFixedinfolasers and opticsOptical roomTheoretical Finesse, Gain and CouplingFor spherical mirrors M3 and M4 (batch
C117I054) the reflectance (R) is around 99.9987%
(if T+R=1 => T=13ppm)
  
  14   Wed Jun 30 20:29:17 2021 Manar AmerFixedreportlasers and opticsOtherMode observed and cavity Lock New configuration is made. The new mode
observed in NF_Trans is the following :

 NF_Trans__2021-06-25__16-38-56.bmp 
  17   Tue Sep 7 14:27:07 2021 BlancFixedinfolasers and opticsOptical roomPolarisation Results of the NKTI made many tests about polarization and
we can see some points thanks to the following
figures :
 Ellipsicité_NKT.PNGPolar_horizontale_fonction_pussance_NKT.PNGPolar_verticale_fonction_pussance_NKT.PNGPolarisation_Ag_VS_Diélec.PNG 
  19   Wed Sep 8 10:50:38 2021 BlancFixedinfolasers and opticsOptical roomPolarisation Results of the NKTTo have the polarisation, we must have
1 value because the laser in entry of the
system has 1 polar. Due to that, the graphe
 Tableau_polar.PNG 
  20   Thu Sep 9 16:34:37 2021 BlancFixedinfolasers and opticsOptical roomRing light in the cavityWe put a diaphragm in the input beam but
it doesn't work to suppress the ring
light only but it attenuates the whole transmitted
  
  21   Thu Sep 16 18:23:13 2021 BlancFixedinfolasers and opticsOptical roomPolarization Results in transmissionAfter data processing, I finally find the
polarization in transmissin of the FPC. As
you can see below, the polarisation is not
 Polarisation_transmission_cavité.png 
  22   Fri Sep 24 09:55:35 2021 Manar AmerUnder Processreportlasers and opticsOptical roomReflection Line Polarization MeasuremetWith the help of Ronic the cavity was locked
in preparation to measure the polarization
of the reflection line when the cavity is
 Reflection_line.jpg 
  32   Tue Oct 5 18:24:53 2021 Ronic ChicheFixedinfolasers and opticsOptical roomLocking with fast analog loop on AOMin addition to the standard locking scheme
with the GHz laser PZT,

we added an AOM after the PDH modulation
  
  36   Wed Oct 20 12:36:15 2021 Ronic ChicheFixedreportlasers and opticsOptical roomAmplitude GHz oscillator phase noise with PZT effect3 phase noise measurements made on the
Amplitude GHz oscillator with different PZT
configurations :
 Amplitude_GHz_phase_noise.pngGHz_oscillator_phase_noise_without_front-end-1.png 