ID Date Author Status Type Categorydown Location Title Text Attachments
  38   Wed Oct 20 15:27:51 2021 Ronic ChicheFixedreportlasers and opticsOptical roomAmplitude GHz oscillator phase noise with PZT effectThorlabs PZT datasheet.
Reference: PC4QR

 PC4QR-SpecSheet-1.pdf 
  46   Fri Dec 10 16:57:56 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 10 12 2021 (optimization injection Burst)Manip 10/12/2021
Après recherche de la résonnance
principale
 NF_Refl__2021-12-10__14-25-08.pngtest__2021-12-10__14-25-08_1920_25gm.png 
  47   Tue Dec 14 12:59:07 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 14 12 2021 injection optimization (and calibration)Matin :

moteurs :
M1, M2 (plans): 5.110500m
 Screenshot_2021-12-14_2_125459.png 
  48   Thu Dec 16 10:40:55 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 14 12 2021 injection optimization (and calibration)Après-midi :
    * Optimisation
de l'alignement en injection pour avoir
 Screenshot_2021-12-14_3_182645.png 
  49   Thu Dec 16 12:12:10 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 16 12 2021 montée en puissanceMatin: 
* Redémarrage Tangor 4W
    - couplage 15%
 Screenshot_2021-12-16_0_115846.png 
  50   Thu Dec 16 12:13:52 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 16 12 2021 montée en puissance 




 Screenshot_2021-12-16_1_121553.png 
  51   Thu Dec 16 18:16:12 2021 BlancUnder Processinfolasers and opticsOptical roomChangements et ajout d'optiquesOptiques ajoutées :

- splitter 90/10 au début
de la voie de réflexion, pour éjecter
  
  52   Fri Dec 17 17:03:34 2021 Kevin DuprazUnder Processreportlasers and opticsOptical roomManip 17 12 2021 reprise réf nouvelle injection pour polarisationTangor : 4.2W
* Réalignement de l'injection
suite aux modifications :
 Screenshot_2021-12-17_0_164132.pngScreenshot_2021-12-17_1_170202.png 
  55   Thu May 19 10:04:05 2022 BlancUnder Processreportlasers and opticsOptical roomChangement cadence et duty cycle + ref cam- Prise des nouvelles ref caméras
: NF{709;371} , FF{389;280} [Image 1 &
2]
 Ref_NF_inj_18_05_2022.pngRef_FF_inj_18_05_2022.pngScreenshot_2022-05-18_2_165654.pngScreenshot_2022-05-18_5_173440.pngScreenshot_2022-05-18_6_174947.png 
  56   Tue Jun 14 11:16:37 2022 BlancUnder Processissuelasers and opticsOptical roomProblèmes CavitéePrécédentes ref pour lesquelles
la manip fonctionnait :

Caméras :
 Transmission.PNGNF_inj.PNGFF_inj.PNG 
  57   Tue Jan 10 18:04:20 2023 BlancFixedinfolasers and opticsOptical roomDonnées pour la baisse en cadencePosition Cam :

692;363 NF_inj
520;193 FF_inj
 7x 
  58   Wed Jan 18 17:52:33 2023 BlancFixedinfolasers and opticsOptical roomTests et mesures locklineMesures de puissance fibres lockline :

AVANT POLISSAGE

injection premier étage
  
  59   Thu Jan 19 14:56:27 2023 BlancFixedinfolasers and opticsOptical roomTests et mesures lockline 




 AOM_250_drivé_à_160_MHz_sortie_fibre_bleue.pngschéma_mesures_lockline_18_01_2023.PNG 
  60   Thu Jan 19 15:05:58 2023 BlancFixedinfolasers and opticsOptical roomSuite Mesures de puissance locklineChangement d'AOM pour passer sur l'AOM
200 MHz.

Le schéma suivant récapitule
 schéma_mesures_lockline_19_01_2023.PNG20230119_124318.jpg 
  61   Thu Jan 19 18:20:02 2023 BlancFixedinfolasers and opticsOptical roomSuite Mesures de puissance lockline 




  
  62   Fri Jan 20 18:01:22 2023 BlancFixedinfolasers and opticsOptical roomMesure de finesseFréquence centrale : 879.48 MHz

FWHM mesurée à la
main sans sweep : 120 / 130 KHz
  
  63   Wed Jan 25 12:12:36 2023 BlancFixedinfolasers and opticsOptical roomMesure de finesse 




 Mesure_de_finesse.png 
  64   Wed Jan 25 12:18:25 2023 BlancFixedinfolasers and opticsOptical roomPosition des miroirs et gain photodiodeNouvelle position des miroirs :

5.198865 ; 5.199305 ; -2.9 ; -2.9

Photodiode de transmission, gain
  
  65   Thu Jan 26 22:49:46 2023 BlancFixedinfolasers and opticsOptical roomNouvelles positions moteurs et calibrations cameraPositions moteurs :

5.198865 ; 5.198596 ; -2.9 ; -2.9

Calibration caméras :
 Lock_lockline.pngSignal_d'erreur_PDH.png 
  66   Mon Jan 30 14:39:52 2023 BlancFixedinfolasers and opticsOptical roomMesures de PolarisationEn récupérant la transmission
du miroir M3 sphérique sortant de
la cavité, on a pu faire une mesure
 Paramètres_de_Stokes.pngDOP_DOLP_DOCP.pngPhase_Ellipticité_Êta.pngEllipse_polarisation_cavité.png 