  elog des diags., Page 7 of 8  Not logged in ThomX    logo
New entries since:Thu Jan 1 01:00:00 1970
ID Date Author Status Type Category Important Subject Icon
  34   Thu Oct 14 16:23:33 2021 Entered by Iryna Chaikovska from 134.158.89.93 on Thu Oct 14 16:23:19 2021Under ProcessInfoBPMs Commissioning of the Libera ER/DG/RAC.05-LIB-01 (BPM LI/DG/BPM.01-LIB.01) 

13/10/2021

Due to the problem of finding the signal with the first BPM in the linac  LI/DG/BPM.01-LIB.01, it was decided to do the tests with the signal generator (500 MHz sine wave cw, ~100 mV).Eventually, we found that the cables (or tango names in the control room) of that BPM in the linac LI/DG/BPM.01-LIB.01 and the BPM in the EL EL/DG/BPM.01-LIB.01 were swapped.

We change the cabling the tha rack between these two BPMs, now in the control room, the tango names correspond to the nomenclature for ER/DG/RAC.05-LIB-01. What about the calibration constants in this case? To be discussed !!!!

Once the cabling was fixed, we saw well the sin signal on the libera (see the screenshot attached).

We also check the ER/DG/RAC.05-LIB-02 with the generator since there there is one TL BPM TL/DG/BPM.01-LIB.02 we will need as well for this phase of the commissioning. We found also that two BPM modules are swapped with respect to tango names used in the control room. Here it is the summary:

- in the rack5  (sticker TL/01, serial number S/N 11) corresponds well to TL/DG/BPM.01-LIB.02 in the control room =>>>> OK 

- in the rack5  (sticker TL/02, serial number S/N 4) corresponds to TL/DG/BPM.01-LIB.03 in the control room  =>>>> to be corrected  !!!!!!!

- in the rack5  (sticker TL/03, serial number S/N 18) corresponds to TL/DG/BPM.01-LIB.02 in the control room  =>>>> to be corrected  !!!!!!!

- in the rack5  (sticker TL/04, serial number S/N 16) corresponds well to TL/DG/BPM.01-LIB.04 in the control room =>>>> OK 

 

Now, we are ready to restart again with the beam (we are waiting when there will be a beam at ThomX). 

=>  need to play with the trigger delay but hopefully we will see the signal on the libera this time.

The similar tests with the generator should be done for the rest of the liberas (ER/DG/RAC.05-LIB-03, ER/DG/RAC.05-LIB-04, ER/DG/RAC.05-LIB-05) in order to eliminate such problem in the future for rest of the BPMs.

Recommendation from N. Hubert: look at the TbT signals as well during the commissioning.

Attachment 1: Screenshot_from_2021-10-13_15-56-26.png
Screenshot_from_2021-10-13_15-56-26.png
  33   Mon Oct 11 15:18:22 2021 Entered by Alexandre Moutardier from 193.55.29.169 on Mon Oct 11 15:18:09 2021FixedFixBPMs Cabage des BPM et WaveCatcher 

Câblages des BPM sur les Libéra non encore connecté en utilisant la clé dynanométrique.
Verification du câblage des Wavecatcher.

État actuel :
- BPM tous câblé correctement
- WC tous câblé correctement

  32   Mon Oct 11 15:13:16 2021 Entered by Alexandre Moutardier from 193.55.29.169 on Mon Oct 11 15:13:03 2021Not FixedProblemBPMs Libéra 5 en erreur  

Sur le devant du libéra 5 la LED E rouge est allumé.

Il me semble que le problème a déjà était rencountré et qu'il est aquitable par connection ssh sur le Libéra et utilisation d'une commande specifique

XXXX

Nicolas:

La solution est que quelqu'un avec le mot de passe tape: libera-bmc --sel --clear

J'ai fait la demande dans le ticket 12984.

 

 

 

  31   Fri Jul 23 12:24:20 2021 Entered by Nicolas Delerue from 80.119.21.73 on Fri Jul 23 12:24:08 2021Under ProcessQuestionCalibrations IHM Charge 

- Coef (sans unites) est censé être le coefficient de conversion de la charge. Veux-tu que je retire son affichage? Est-ce que tu as demandé à Hayg des modifications sur le calcul de la charge? Il n'est pas normal que le coef ne s'affiche plus.
- Aucun valeur n'est mise à zéro au moment du lancement de l'IHM et je ne suis pas favorable à le faire car si une personne lance l'IHM sur un deuxième terminal (ou si l'IHM plante), alors tout sera remis à zéro.
- Quel facteur de conversion pour passer de Dark (nV*s) à la valeur en pC?
- J'ai crée la colonne pour le Sdev, mais le DS de calcul ne renvoi pas la sdev... à mettre en ticket.
- Pilotage de la profondeur moyenne: c'est compliqué, à mettre en ticket.

 

  30   Fri Jul 23 09:27:24 2021 Entered by vincent.chaumat from 134.158.195.144 on Fri Jul 23 09:27:09 2021 InfoCalibrations vérification fonctionnement IHM charge avec parametres d'étalonnage 

Vérification du fonctionnement de l'IHM_charge.py avec les variables d'étalonnage

Les 5 voies de mesures fonctionnent correctement.

 

Remarques sur l'IHM:

Une colone "Coef (sans unité)" est affichée sans données => peux t'on l'a retirer?

Par défaut au lancement de l'IHM la valeur Dak(nV*s) est renseigné pour la ligne LI/DG/ICT.01  => peut elle etre à mise à "0" par défaut comme les autres lignes?

Peut 'on ajouter un colonne " Sdev" en pC après la colonne Calc (pC)  calculée avec la meme routine que RAW(nV*s)   => Calc(pC)

La colonne Dark (nV*s) peut elle etre exprimée en pC?

Peut'on rajouter un parametre dans l'IHM pour piloter la profondeur du moyennage  effectué dans le DS mean_std?

Le nom de la colonne Calc(pC) peut il etre changé en Q_Beam(pC)

 

 

  29   Wed Jul 21 15:01:02 2021 Entered by vincent.chaumat from 134.158.195.144 on Wed Jul 21 15:00:38 2021 InfoCalibrations DS wavecatcher 

Test reglage de la largeur de la fenetre d'acquisition.

les parametres sont

FMP_RefCellForCharge  => début de la fenetre (numero du sample)

FMP_ChargeLength => largeur de la fenetre (nombre entier = nB samples/16  ie 100ns = 20 à 3.2GS/s)

 

pour changer les valeurs: (dans le DS wavecatcher)

1- stop acquisition

2- choisir les valeurs des parametres

3- set firmwave measurement parameters

4- get firmware measurement parameters (pour vérifier les valeurs)

5- start acquisition

 

mesure

Pour une fenetre de 100ns le bruit de charge des wavecatchers (quelque soit la voie) est de 0.03nVs  Sdev => 0.6pC

pour la fenetre entiere (1000 samples) le bruit de charge des wavecatchers (quelque soit la voie) est de 0.08nVs  Sdev => 1.6pC

 

 

Remarque

dans le DS ER/CA/RAC0.5-ELR.01-WAC.02

la valeur AcqTime est a 1.94e+227   !!!! dans les autres DS WAC1 et 3 la valeur est 0.

au bout de 2 minutes de fonctionnement le DS charge ne rafraichi plus les valeurs "RAW" et EventID de charge

d'abord affecte le wavecatcher 2 puis le 1 et 3

 

  28   Tue Jul 20 14:38:24 2021 Entered by Nicolas Delerue from 134.158.195.145 on Tue Jul 20 14:38:00 2021FixedFixMaterial Changements config WAC 

Sur WAC1:

Polling

ch0: 10=> 50

ch0_charge: 20 => 50

 

  27   Mon Jul 12 10:27:31 2021 Entered by vincent.chaumat from 134.158.195.144 on Mon Jul 12 10:27:17 2021Under ProcessProblemOther   

test IHM diag mesure charge:

Géné pulse trigger en externe (DEG3-CH15) sur les 3 Wavecatchers   (50mV 50ns)

[LI-DG-ICT01 WAC1-CH0]

[TL-DG-FC01  WAC2-CH0]

[EL-DG-FC01   WAC3-CH0]

les 3 DS Wavecatcher fonctionnent 

 

IHMcharge 2021-07-12 10-39-31.png 

=> colonne RAW seule la ligne  EL/DG/FC/01 se rafraichie. (à 5Hz)

=> les cases coef ligne 1 et 4 grisées (pB DS calc)

 

sur le DS de reboot

=> adressage calc/ict/charge_li-ict.01_wac.01-ch0  ==== charge/ict.01

=> adressage calc/ict/charge_tl-ict.01_wac.02-ch3 ====charge/ict.01

 

 

Les DS calc ne sont pas rebbootables... voir attachment2.

les DSmean std non plus

 

 

 

 

 

 

 

 

 

 

Attachment 1: IHMcharge_2021-07-12_10-39-31.png
IHMcharge_2021-07-12_10-39-31.png
Attachment 2: calc-EL-FC-01-WAC3-ch0.png
calc-EL-FC-01-WAC3-ch0.png
  26   Fri Jul 9 11:11:04 2021 Entered by vincent.chaumat from 134.158.195.144 on Fri Jul 9 11:10:45 2021Under ProcessProblemOther   

Test IHM ICT

Injection d'un pulse 100mV-100ns en entrée des voies des wavcatchers.

Trigger sur la voie d'injection avec un retard de 100ns (pour simplification pas de réglage de timing)

 

le wavecatcher 2 ne fonctionne pas correctement :(voir fichier joint)

 

pour l'IHM de charge:

aucun affichage des valeurs "RAW" 

les valeur moyenne et std sont calculées et affichées

pas de calcul de la charge en pC

les boutons reset average et set dark current fonctionnent

Attachment 1: Wavecatcher2
  25   Tue Jun 22 10:12:22 2021 Entered by Moutardier Alexandre from 193.55.29.169 on Tue Jun 22 10:11:55 2021FixedInfoCameras SST Vérification switch/moteur station SST et fente 

La vérification du fonctionnement des moteurs ci-dessous ainsi que des butés associées ont été faites récemment. Aucun problème à déplorer.

Station SST :
- LI/DG/SST.01-MOT.01
- TL/DG/SST.01-MOT.01
- TL/DG/SST.02-MOT.01
- TL/DG/SST.03-MOT.01
- EL/DG/SST.01-MOT.01

Fente :
- TL/DG/SRH.01-MoG.01-MOT.01
- TL/DG/SRH.01-MoG.02-MOT.01

  24   Wed Jun 9 14:50:45 2021 Entered by Nicolas Delerue from 80.119.21.73 on Wed Jun 9 14:50:31 2021FixedInfoCameras SST Résolution LI/DG/SST1 

Cf présentation ci-joint.

Conclusions:

Le grossissement au centre de l’image est de 40um par pixel verticalement et 30um par pixel horizontalement.
La resolution varie de  50um en bas de l’image à 70um en haut de l’image.
Le grossissement peut être changé sur demande (accès à la caméra nécessaire) mais un grossissement plus grand ne permettra pas de voir tout l’écran.

Attachment 1: Resolution_LI_DG_SST_1.pptx
  23   Thu Jan 28 14:17:08 2021 Entered by Moutardier from 134.158.195.142 on Thu Jan 28 14:16:54 2021FixedInfoBLM Tout les Libéra de nouveau fonctionnels 

L'échange entre le Libéra 2 et 3 est terminé.
Chacun est de nouveau controlable avec Jive, ainsi tout les Libéra sont fonctionnels.

Dans l'ordre de haut en bas dans la baie 5 nous avont les Libéra : 3 4 5 2 1
Les étiquètes on toutes était modifié aussi pour être les bonnes.

Lorsque j'ai rallumé le Libéra 3, il a fallu que j'ouvre toute les interfaces BPM sur jive pour faire un "Init" car ils étaient tous en alarme (main pas pour le Libéra 2).

Tous les Libéra ont était rééteint apès les tests.

  22   Wed Jan 20 16:15:49 2021 Entered by Moutardier from 134.158.195.142 on Wed Jan 20 16:15:31 2021 InfoCalibrations Démarrage ds MeanStd 

Methode pour changer un attribut dans MeanStd  et redémarrer le ds :

Sur jive :
dans  MeanStd/test/MeanStd/'tmp/DG/MeanStd'/paramètre, modifier "ManagedAttributeList" pour avoir les valeurs voulues.

Sur Astor :
"Kill" le serveur "MeanStd" dans 'Salle de controle'/client2/'Not Controlled' (ATTENTION, les ds Not controlled ne sont pas affiché par défault, il faut double cliquer sur Not Controlled)

Sur un terminal depuis le pc client2 tapper la commande :
/data/shared/MeanStd/MeanStd test &

Sur jive :
Ouvrir l'ATKPanel : MeanStd/test/MeanStd/'tmp/DG/MeanStd'
Lancer une commande start.

  21   Tue Dec 15 16:59:26 2020 Entered by Nicolas Delerue from 80.119.21.73 on Tue Dec 15 16:59:10 2020Not FixedProblemCameras SST Trigger défaillant TL/DG/SST.02-CCD-01 

La caméra TL/DG/SST.02-CCD-01 semble ne pas pouvoir se déclencher sur signal externe. Pour l'instant elle reste en déclenchement interne.

Nous la changerons avant le démarrage de la ligne de transfert.

  20   Tue Dec 8 12:59:13 2020 Entered by Nicolas Delerue from 80.119.21.73 on Tue Dec 8 12:58:57 2020Under ProcessFixBPMs Libera arrêtés 

Une intervention aura lieu sur les baies de l'IGLEX cet après-midi.
J'ai donc arrêté les liberas.

Nicolas

  19   Thu Oct 22 11:00:31 2020 Entered by Moutardier from 193.55.29.171 on Thu Oct 22 11:00:17 2020Under ProcessFixBPMs échange des Libera 2 et 3 

Les étiquettes des libéra 2 et 3 ont été interchanger de façon a avoir les bonne informations inscrite dessus.

Ainsi nous avons en partant du haut de la baie 5 :
Libéra 3 : 2/3 Ri-C1 (BPM 1 à 4)
Libéra 4 : 1/3 Ri-C1 (BPM 5 à 6) + 1/3 Ri-C1 (BPM 1 à 2)
Libéra 5 : 2/3 Ri-C2 (BPM 3 à 6)
Libéra 2 : TL (BPM 1 à 4)
Libéra 1 : EL + Li

Il reste a verrifier la connection au réseau du Libéra 3 et a faire les branchements des sorties des BPM sur les Libéra 2 et 3.

  18   Mon Oct 5 14:06:24 2020 Entered by Moutardier from 91.168.92.148 on Mon Oct 5 14:05:55 2020Under ProcessFixBPMs retour du libera 2 et échange des Libera 2 et 3 

Le libera 2 est revenu de maintenance.

Il a était mis à la place du Libéra 3 puis qu'il ont été echangé (attention les etiquettes sur ces 2 Libéra n'ont pas encore été échangées).

Le cc doit encore modifier l'addresse IP et le nom de l'ex Libéra 2 afin qu'il soit utilisable depuis jive.

  17   Thu Jul 23 17:21:59 2020 Entered by Nicolas Delerue from 193.55.29.171 on Thu Jul 23 17:21:46 2020Under ProcessProblemBPMs Le libera 2 a été renvoyé en maintenance. 

Le libera 2 a été renvoyé en maintenance. Il est donc indisponible jusqu'à son retour.

  16   Fri Jul 10 08:41:52 2020 Entered by Moutardier Alexandre from 91.168.92.148 on Fri Jul 10 08:41:36 2020Under ProcessInfoCameras SST Caméra SST 

Modification interface IHM_SST pour permetre de sauver des images lorsque l'on ouvre l'interface avec le panneau interfaces. A tester.

Les éclairages LED :
- li/dg/sst.01-ccd.01 : Installé, tester, fonctionnelle
- tl/dg/sst.01-ccd.01 : Installé, tester, fonctionnelle
- tl/dg/sst.02-ccd.01 : Installé, tester, fonctionnelle
- tl/dg/sst.03-ccd.01 : Installé, tester, fonctionnelle
- el/dg/sst.01-ccd.01 : Installé, NON tester, LED fonctionnelle (Impossible de contrôler les LED depuis le cc car c'est (avec la spare) une caméra ACE et non Scout)

2 LED en plus : 1 jeu fonctionnelle mais pas installable sur ThomX (au sous sol du bât 208), 1 en réparation chez Thomas Chabaud

La caméra "spare" est au sous sol du bat 208.

  15   Tue Jun 23 09:42:22 2020 Entered by Moutardier Alexandre from 193.55.29.171 on Tue Jun 23 09:41:14 2020 InfoCameras SST Bilan des caméra diagnostic au 23/06/2020 

Ouverture interface IHM_SST avec le panneau_interface OK, allume correctement les serveurs des 5 caméras installé sur ThomX (li/dg/sst.01-ccd.01, tl/dg/sst.01-ccd.01, tl/dg/sst.02-ccd.01, tl/dg/sst.03-ccd.01, el/dg/sst.01-ccd.01)
Caméra "spare" non testé car pas sur ThomX (au sous sol du 208 pour faire les tests sur les objectifs).

Interface IHM_SST ouvert avec le panneau_interface ne permet pas de saugarder des images.
Interface IHM_SST ouvert en ligne de commande si. (A voir avec Hayg)

Les éclairages LED :
- li/dg/sst.01-ccd.01 : OK
- tl/dg/sst.01-ccd.01 : OK
- tl/dg/sst.02-ccd.01 : Pas de Led d'installée, fonctionnait avant le confinement
- tl/dg/sst.03-ccd.01 : OK
- el/dg/sst.01-ccd.01 : Impossible de contrôler les LED car c'est (avec la spare) une caméra ACE et non Scout
- spare                      : Il existe un système d’éclairage en rab mais il n'est pas montable sur les montures des caméras mais qui est utilisable sur le banc de test des caméras