       Reply   ND; RE: RE: mauvaise calibration TL_SST1 calibration_SST_20230220.pptx2023-02-20_14_24_42_li_dg_sst_01-ccd_01_RAW_16bits.tiff2023-02-20_13_38_55_tl_dg_sst_02-ccd_01_RAW_16bits.tiff2023-02-15_04_24_18_image_tl_dg_sst_01_pos_-41.0_expo_0.1.png2023-02-13_04_48_22_image_tl_dg_sst_01_pos_-41.0_expo_0.1.png
Message ID: 127     Entry time: Mon Feb 20 10:50:45 2023     Reply to this: 128
 Author: KD 
 Status: New Solution 
 Type: Info 
 Category: Cameras SST 
 Important:
 Subject: mauvaise calibration TL_SST1 
 Icon:  

Bonjour,

après les shifts de la semaine dernière, il semblerait que la calibration de la caméra TL/SST1 soit erronée. Les valeurs diffèrent beaucoup des autres stations diag. Après une calibration rapide on trouve bien une différence :

on trouve v = 28.7um/pixel, h = 28.6 um/pixel

au lieu de v = 54.7 um/pixel, h = 52.8 um/pixel

La station TL/SST2 semble avoir le même problème. Il doit y avoir un bug dans la calibration.

Bonne journée,