  54   Tue Sep 3 18:04:03 2019 Ronic ChicheFixedreportlasers and opticsOptical roomFinesse measurement after mirror cleaning using the spin coater

Following finesse measurements (start @10700): 
Clean P1 with SC (water) + isoprop (wipe) finesse @15000
Clean S2 with SC (water) + isoprop (wipe) finesse @13000
Clean P4 with SC (water) + isoprop (wipe) finesse @11000

We have to take into account that "cleaning" the mirror change the alignment axis in the cavity so the finesse can change because of that. It would explain the erratic nature of these results which seem to not show any effective cleaning.

Ronic Chiche wrote:

today I cleaned S3 with isopropanol and then with pure water on spin coater.
Finesse is only 10 700.(coupling 50% ? => to be confirmed...)

maybe we have to redo the procedure with a different way of cleanig (pushing less on the mirror ... ?)

Ronic Chiche wrote:

Last Finesse measurement in air : ~ 17 000 the 20th of may (see logbook)

from 28th to 30th of august, we cleanned, with Loïc, mirrors S3 then P1 successively using only ultra pure water with the spin coater.

we measured the finesse after each cleaning by modulation technique and in both cases, the finesse did not change !

maybe the mirror surfaces have been covered by some hydrocarbure during the summer ?
or by an other contaminant which is difficult to clean with only water ?
we decided to clean the mirrors first with an isopropanol solution and then with ultra pure water and the spin coater.

 

 

  53   Tue Sep 3 17:51:04 2019 Ronic ChicheFixedissuemechanics | lasers and optics | vacuumOptical roomVessel cleanliness

looking in the vessel of the FP cavity using a UV lamp shows a lot of dust... more dust than outside the cavity !!!

here is picture where one can see a cleaned zone using a finger (with gloves).

question : where does come from this dust ?

2 possibilities :

- large and small belows which could trap the dust and release it during the pumping process

- ionic pumps could release some clusters when they are working.

we plan to clean as effectively as possible the inner surfaces, to close bellows, make woking ion pumps to check if it released again some dust.

  52   Fri Aug 30 17:52:41 2019 Ronic ChicheFixedissuemechanics | detectors and electronicsOptical roomMotor issue

Today, Axis 9 (Z-axis of P1) has blocked the software which drives all the motors.
impossible to bypass the problem.

I had to :

- switch off the motor driver crate n°2 (motors 9-16)

- disconnect all motor cables (9-16)

- reconnect only cables 15 and 16.(same internal board of crate n°2)

- restart the software to test that, now, the software launches properly

and I did again the procedure with connecting all motor cables

  51   Fri Aug 30 17:46:15 2019 Ronic ChicheFixedreportlasers and opticsOptical roomFinesse measurement after mirror cleaning using the spin coater

today I cleaned S3 with isopropanol and then with pure water on spin coater.
Finesse is only 10 700.(coupling 50% ? => to be confirmed...)

maybe we have to redo the procedure with a different way of cleanig (pushing less on the mirror ... ?)

Ronic Chiche wrote:

Last Finesse measurement in air : ~ 17 000 the 20th of may (see logbook)

from 28th to 30th of august, we cleanned, with Loïc, mirrors S3 then P1 successively using only ultra pure water with the spin coater.

we measured the finesse after each cleaning by modulation technique and in both cases, the finesse did not change !

maybe the mirror surfaces have been covered by some hydrocarbure during the summer ?
or by an other contaminant which is difficult to clean with only water ?
we decided to clean the mirrors first with an isopropanol solution and then with ultra pure water and the spin coater.

 

  50   Fri Aug 30 09:44:56 2019 Ronic ChicheFixedreportlasers and opticsOptical roomFinesse measurement after mirror cleaning using the spin coater

Last Finesse measurement in air : ~ 17 000 the 20th of may (see logbook)

28th and 29th of august, we cleanned, with Loïc, mirrors S3 then P1 successively using only ultra pure water with the spin coater.

we measured the finesse after each cleaning by modulation technique and in both cases, F ~ 12 500 !!!

maybe the mirror surfaces have been covered by some hydrocarbure during the summer ?
or by an other contaminant which is difficult to clean with only water ?
we decided to clean the mirrors first with an isopropanol solution and then with ultra pure water and the spin coater.

  49   Mon Jul 29 17:15:35 2019 Loïc AmoudryFixedinfomechanics | lasers and opticsOptical roomSpin coater air supply removed and front panel changed

Spin coater firstly needed vacuum and air supply. The air supply being useless for our experiments, it has been disabled.

The process is the following and is reversible:

1)            Remove the hub from the top of the spin coater, (see picture 1), 4 screws hold the hub in place, once these screws are removed the hub will lift off the motor shaft.
2)            Remove the 2 screws holding the stainless steel top decks in place (see picture 2), please be careful when removing the top deck as the lid switch wire will be attached.
Once these have been removed, you can access the inside of the spin coater.
You must add a bypass wire to the timing block (see picture Purge Bypass 1), the wire must be connected to the A1 and 28 ports on the timing block and remove the red wires from connexion. This will now cause the spin coater to bypass the purge check switch and allow the machine to boot up without air/nitrogen being supplied. To revert this change, simply remove the bypass wire you have installed and plug the 2 red wires back to A1 and 28.

The "enter" button of the front panel had some trouble and has been changed.

  48   Thu Jul 4 18:30:39 2019 Loïc AmoudryFixedreportlasers and opticsOptical room2nd stage output power validated

Onefive output power : 28mW
Power before fiber injection : 7.8mW
After EOM and 99% coupling fiber : 1.8mW
After EOM and 1% coupling fiber : 14 uW (~1%), ~7.5V @ 1MOhm on DET36
Frep : 33.326 MHz
Output 1st stage + collimator & dichroic mirror: 36uW
Output 2nd stage @6A (~260mV @50Ohm on DET36): 300mW --> after 20-30min

Output 2nd stage is the same value as Onefive 133.33MHz

  47   Thu Jul 4 18:28:55 2019 Loïc AmoudryFixedissuelasers and opticsOptical roomOnefive 33MHz unmodelocked

Oscillator randomly unmodelocked on 3rd july afternoon (during oscillator power measurement, maybe a reflection from the powermeter power filter ?)

Modelocked this afternoon by putting Onefive motor and phase at reference on the PTC software. 

  46   Thu Jul 4 10:17:46 2019 Ronic ChicheFixedinfodetectors and electronicsOptical roomNew design for PDH boxes

A new design has been studied for the 5 PDH boxes to better match the low (33MHz) repetition rate of the ThomX laser.
the 5 PDH boxes have been modified and tested (see the attached report).
A mixer has been integrated to the PDH box for a more simple use.

the PDH N°1 has been lend to Huan for her experiment in China.

  43   Tue Jun 18 20:44:34 2019 Huan WangFixedinfolasers and opticsOptical roomtelescope for Koheras

telescope of Koheras is removed, information record here:

- distance between collimator edge and -50 lens mount edge is 14.7cm,

- distance between another edge of -50 lens mount and +250 lens mount edge is 14.3cm,

- distance between another edge of +250 lens mount and edge of bread board is 25cm.

  42   Wed Jun 12 17:49:20 2019 Loïc AmoudryFixedinfolasers and opticsOptical roomPolarization

Polariztion measured with Input vertically/horizontally polarized (~100% DOLP)

- Transmission (P4) locked on cavity horizontal polarization is ~50% linearly polarized and ~50% unpolarized ?

- Transmission (P4) locked on cavity vertical polarization is ~50% linearly polarized and ~50% unpolarized ?

  41   Wed Jun 12 12:07:11 2019 Loïc AmoudryFixedreportmechanics | lasers and opticsOtherAmplifier mission Bordeaux

M² mesuré à 10W.
M²x = 1.17
M²y = 1.14

Loïc Amoudry wrote:

Voyage Bordeaux au CELIA avec Jérôme LHERMITE (Loïc et Titoutan)

1er jour (34°) :

Mesure du spectre et de la puissance des 4 diodes de pompe 3ème étage jusqu’à ~60W (12A).
Mécanique et câblage ampli.

2ème jour (30°) :

Soudure des fibres des diodes de pompe 3ème étage avec les fibres du recirculateur.
Câblage entre les 2 boitiers et fixation des fibres dans le boitier du 3ème étage.

3ème jour (26°) :

Connexion refroidissement à eau du recirculateur (chiller).
Soudure fibre CFBG avant 1er étage (temporaire pour faire des mesure).
Mise en place des éléments 1er et 2ème étage sur le support de l’ampli.
Montée en puissance 35W.

4ème jour :

Adaptation du mode de sortie en contrôlant la position de la fibre avant le collimateur.
Mesure de M².
 

 

  40   Tue Jun 11 18:03:10 2019 Loïc AmoudryFixedreportmechanics | lasers and opticsOtherAmplifier mission Bordeaux

Voyage Bordeaux au CELIA avec Jérôme LHERMITE (Loïc et Titoutan)

1er jour (34°) :

Mesure du spectre et de la puissance des 4 diodes de pompe 3ème étage jusqu’à ~60W (12A).
Mécanique et câblage ampli.

2ème jour (30°) :

Soudure des fibres des diodes de pompe 3ème étage avec les fibres du recirculateur.
Câblage entre les 2 boitiers et fixation des fibres dans le boitier du 3ème étage.

3ème jour (26°) :

Connexion refroidissement à eau du recirculateur (chiller).
Soudure fibre CFBG avant 1er étage (temporaire pour faire des mesure).
Mise en place des éléments 1er et 2ème étage sur le support de l’ampli.
Montée en puissance 35W.

4ème jour :

Adaptation du mode de sortie en contrôlant la position de la fibre avant le collimateur.
Mesure de M².
 

  39   Thu May 23 17:30:03 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomTransmission of Sapphire mirror

See image.

  38   Thu May 23 10:10:56 2019 Ronic ChicheFixedreportlasers and opticsOptical roomFinesse measurement by decay time technique

5 Files attached:

- Note on the technique and results for the present cavity on 2019, May 21th : F ~ 18 000

- 2Tektronix scope raw data

- 2 matlab files to fit the scope raw data and extract the Finesse

  37   Tue May 21 17:49:30 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomFinesse after cleaning (new suprasil mirror)

15/05/19-20/05/19:

Cleaning of the mirrors:

Initial finesse 14000 @air
S2 with isopropanol, finesse 17000 @air
P4 with isopropanol, finesse 17000 @air
S3 with isopropanol, finesse 17000 @air
S3 with first contact, finesse 17000 @air
S2 with first contact, finesse 17000 @air

  36   Tue May 21 17:44:46 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomNew Suprasil mirror from LMA

14/05/19:

New Suprasil mirror received. It will replace the S-BOX suprasil mirror on ThomX which was already replacing the sapphire mirror.

First finesse measurement: 14000

Features sent by Christophe Michel:

"- Transmission mesurée (1064 nm)= 172 ppm ; Transmission estimée (1031 nm)= 140 ppm
- Réflexion (1064 nm)= 1025 ppm; doit être inférieure à 1031 nm
L'AR n'est pas bon alors que l'on obtenu 170 ppm pour le run de calage!
J'espère que vous pourrez néanmoins l'utiliser.
- Diffusion moyenne sur diamètre 12 mm: 2.6 ppm"

 

  35   Thu Apr 18 16:17:53 2019 Loïc AmoudryFixedissuelasers and opticsOptical room33MHz Onefive power overtime decrease

Power of the 33MHz Onfive decreases over time. Have to pay attention to that.
Note that air dryers have never been stopped.

10th april 2018: ~40mW (first picture, 10dB filter)
25th september 2018: ~36mW (second picture, 10dB filter)
16th april 2019: ~29mW (third picture, no filter)
 

  34   Thu Apr 18 16:07:23 2019 Loïc AmoudryFixedinfolasers and opticsOptical roomInjection of 33MHz Onefive through fiber

Output power of the Onefive, and power after injection and different components have been measured.

Coupling ~= 90%
Each of our two EOM roughly have the same loses ~3dB

  33   Wed Apr 10 13:11:46 2019 Loïc AmoudryFixedreportlasers and opticsOptical roomMirror's cleaning and finesse

Summary of what have been cleaned the last week:

P1 with isopropanol, finesse 11000
P4 with isopropanol, finesse 13000
S2 with isopropanol, finesse 13000
S3 with isopropanol, finesse 14000
S2 with isopropanol, finesse 14000
P4 with isopropanol, finesse 12500
P4 with isopropanol, finesse 12500

Finesse decreased with last steps which probably means our cleaning technique is at its maximum. We will probably try a cavity axis change with the mirror's motors to increase the finesse  to the maximum like that.
Next step is to clean P1 with acetone + pur ethanol, then install the Onefive.