HOME SBOX THOMX MINICAV Utilities
THOMX orders THOMX installation THOMX commissioning THOMX control command
  Status of commissioning, report also here plots are reports., Page 18 of 24  Not logged in ELOG logo
Entry   FP cavity alignment , posted by Manar Amer at ThomX igloo about lasers and optics 2021_12_02_M2_Transmission.jpg2021_12_02_M3_transmission_90deg.PNG
Note the direction of injection is     M1 - M2 - M3 - M4

                                                      
P1 - S2 - S3 - P4
    Reply   FP cavity alignment , posted by Manar Amer at ThomX igloo about lasers and optics 
The alignment is on hold until next week 17th - 18th  Jan

an Alignment attempt will be done when the ring part close to the FP cavity is opened.




Manar
    Reply   FP cavity alignment , posted by Manar Amer at ThomX igloo about lasers and optics Interior_of_cavity_P1_S3_windowOpen.jpgInterior_of_cavity_P1_S3_windowOpen_irrisOnP1.jpgCavityAlignment_cavityOpen_S2_Output.jpgCavityAlignment_cavityOpen_S3_Output.jpeg
***** Continuation of the alignment ---- FP Cavity Open --- ****

The cavity was put under outer pressure and was opened for the alignment

The alignment was done using CW koheras infrared laser and  the inside mirrors irises
    Reply   FP cavity alignment , posted by Manar Amer at ThomX igloo about lasers and optics 
Yesterday, we did the alignment again using Iris and beam profilers to obtain a more precise result.
at the end of the alignment procedure, we successfully obtained the beating modes.

the geometrical alignment seems good (weak odd modes) but we could need to put a telescope on the Koheras line as the beam size seems quite different
    Reply   FP cavity alignment , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics ThomX_FP_setup_01-02-2022.pngtek0002.png
Installed and inputs: 


CW laser ; power = 101 mW
PDH ;

at output

Low pass filter 50 ohm  DC-1.9 MHz


at
input

freq = 8.4 MHz , Ampl = 1.2 V  , phase = 160 /
voltage =   ~ 6 V




EOM
    Reply   FP cavity alignment , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 
This afternoon, I did some alignment of the injection mirrors with a fast scan on the LaseLock to get a regular transmission photodiode signal.
the coupling increased to 10-15%

I got a first lock of the cavity only with the PZT.
    Reply   FP cavity alignment , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 
This morning with Manar, we installed the AOM+RF amplifier and the associated fast feedback loop.
Now the locking with the Koheras is good with a coupling of 20%.
Tomorrow is dedicated to the measurement of the Finesse.... we will have to add the 2nd EOM.
    Reply   FP cavity alignment , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics tek0003.pngtek0006.pngtek0007.png
Adding Oscilloscope images


delocked + locked intervals showing the 20% coupling
lock when the housing panels are closed + airflow on .
lock
when the housing panels are closed + air flow at lowest setting


an additional factor to the PZT voltage sensitivity
    Reply   FP cavity alignment , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 
Ending alignment series !!




Manar
Amer wrote:



Adding Oscilloscope images
    Reply   FP cavity alignment , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 
As it can be seen on the first plot, even with a good locking (good reduction of "high" frequencies noise: we had better locking than on the
picture) we still have very low frequency (~ 1Hz) fluctuations

these fluctuations prevent having a good measurement of the Finesse and they need to be understood.
Entry   Cavity Lock and Finess measurment , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 6x
This morning, me and Ronic managed to obtain the FSR and image of the finesse which is to be analyzed.

Adding:


Beam Image at P4 propagation, no lens is added.
Image of the Finesse on Oscilloscope


more information to
    Reply   FP cavity alignment , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 
The fluctuation problem has been solved.
It was simply the new scheme to inject 2 RF frequencies in a single EOM.
It maybe produces some standing waves in the EOM RF input and creates some phase noise.
    Reply   Cavity Lock and Finess measurment , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics 


to measure the Finesse, instead of having 2x EOM for the PDH and for the modulation sweep, I simply used 2x generators coupled with DC-blocks
to a T connector (SMA) screwed directly on the EOM input.
    Reply   Cavity Lock and Finess measurment , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics Finesse_Fit.png


After changing the way of injecting modulation for PDH and modulation for FSR scanning (we split the modulations to 2 different EOMs), the locking
is very stable and we can measure the Finesse.
    Reply   Cavity Lock and Finess measurment , posted by Manar Amer at ThomX igloo about lasers and optics | detectors and electronics apercu_03.jpgBras_manipulateur.jpg
one thing which could reduce the Finesse is the L-shape metal piece if it is slightly inserted in the FP-cavity mode path.
understanding where this L-shape effectively is is not easy... some pictures are attached.




Manar
Entry   Cavity beam Image , posted by Manar Amer at ThomX igloo about lasers and optics Capture.PNGCapture2.PNG
Updated beam image after change on the mode axis.

Taken after P4.
Entry   Current Installed Setup, posted by Manar Amer at ThomX igloo about mechanics | lasers and optics | detectors and electronics 18-02-2022_ThomX_FPCavity_Finesse_Setup.jpg
Attached is the current setup in ThomX
Entry   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 8x
Today: Manar, Ronic and Aurlien started the study of the beam profile of the alphanov amplifier at high power.

The setup shown in the image shows how the power is reduced by using Anti-reflective mirrors,

pick up 1 : Anti-reflective coating on both sides , pick up 2 : Anti-reflective coating on one side and High
    Reply   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 
 




Manar
Amer wrote:



Today: Manar, Ronic and Aurlien started the study of the beam profile
    Reply   Alphanov amplifier Beam Profile, posted by Manar Amer at ThomX igloo about lasers and optics 6x
Adding the fitted beam in 1D in Horizontal and vertical axis.

for 10%   to  60%




Manar
ELOG V3.1.4-395e101