HOME SBOX THOMX MINICAV Utilities
THOMX orders THOMX installation THOMX commissioning THOMX control command
  Status of commissioning, report also here plots are reports., Page 16 of 24  Not logged in ELOG logo
Entry   Spherical mirrors at closest position. FSR, finesse and beam size., posted by Loïc Amoudry at Optical room about lasers and optics data_finesse.isffit_finesse_17000.pngbeamsize.PNGfit.PNG

The spherical mirrors are at the closest position to get a FSR ~=33.3435 MHz. Planar mirrors are at the largest position.

The finesse has been measured about 17 000. 

Beam diameter (2*w) is ~= 4.6 mm x and y.

Entry   Amplifier output power , posted by Loïc Amoudry at Optical room about lasers and optics dichroic.PNGcourbe.PNGpower.xlsx

The ouput power of the ThomX amplifier has been measured as a function of the 3rd stage pumps current (0A is 1st and 2nd stage turned on).

We measured the total power, the total power without the "donuts" pump signal dumped by an iris and the transmission of a dichroic mirror (for ref see image).

    Reply   Amplifier output power , posted by Loïc Amoudry at Optical room about lasers and optics courbedichro.PNG

 

Loïc Amoudry wrote:

The ouput power of the ThomX amplifier has been measured as a function of the 3rd stage pumps current (0A is 1st and 2nd stage turned on).

We measured the total power, the total power without the "donuts" pump signal dumped by an iris and the transmission of a dichroic mirror (for ref see image).

 

    Reply   Onefive Origami 33MHz specs, posted by Loïc Amoudry at Optical room about lasers and optics beam_size.xlsxBeam_size_Onefive_33.rar

Beam Size Onefive added

Loïc Amoudry wrote:

Power during ~20h, power after turn on, spectrum image and excel datas, polarisation and pulse length.

Powermeter measurments were done with a /10 filter.. 3 mW means 30 mW

Important: Onefive does not automatically locks after being turned on.

 

Entry   Onefive output power, posted by Loïc Amoudry at Optical room about lasers and optics 

The pump voltage has been increased few days ago when looking for a modlock of the laser.

The 33MHz Onefive output power is now 3.58mW with OD1 filter ==> ~36mW.

There are 2µW @ 1% coupling fiber with OD1 filter ==> ~20µW ==> 2mW @99% coupling fiber (2mW to the amplifier).

Entry   CVBG reference and Eric Cormier calculation, posted by Loïc Amoudry at Optical room about lasers and optics compresseurs.jpgEtireur.jpgcalcul_Eric.jpg

The Stretcher is the "D 24-02-II", it must be used in double path before the amplifier injection.

There are two compressors in single path, "D 25-14-I" and "D 25-14-II".

Eric calculations are attached below.

    Reply   Onefive output power, posted by Loïc Amoudry at Optical room about lasers and optics 

Onefive output power measured today. 3.52 mW with OD1 filter ==> ~35mW.

Loïc Amoudry wrote:

The pump voltage has been increased few days ago when looking for a modlock of the laser.

The 33MHz Onefive output power is now 3.58mW with OD1 filter ==> ~36mW.

There are 2µW @ 1% coupling fiber with OD1 filter ==> ~20µW ==> 2mW @99% coupling fiber (2mW to the amplifier).

 

Entry   Beam pulse duration, posted by Ronic Chiche at Optical room about lasers and optics interferences.avipulse_duration.png

We installed the 2 CVBG for compression after the amplifier.
We used an interferometric technique with a delay line and combining the two paths in a CCD to measure interferences... see interferences.avi video file
One can notice some misalignement at the end of the video.

After supressing the global shape of the superposed pulses, one measures the amplitude of the remaining fringes (peak-peak or standard deviation) each 250µm of the delay line (500µm of round-trip). one gets the interferences pulse shape with a FWHM of 6ps...
It seems that an 'after pulse' is visible in the interferences.

An other meausurement using a 2 photons photodiode will be used to confirm this measurement.

 

    Reply   Beam pulse duration, posted by Ronic Chiche at Optical room about lasers and optics pulse_width.png

Yesterday, Loic installed a telescope before CVBG's to reduce the spot size on CVBG's and reduced the incident angle 0.5-1° on them (on the datasheet the specified incident angle is 2.8°).

We used an interferometric technique with a delay line and combining the two paths in a CCD to measure interferences.

After supressing the global shape of the superposed pulses, one measures the amplitude of the remaining fringes (standard deviation) each 50µm of the delay line (100µm of round-trip). one gets the interferences pulse shape with a FWHM of 2 ps... (see curve)
we still see an 'after pulse'.

if the pulse is 1ps long and 100kW is stored in the cavity, it means 3GW peak for the whole beam... it is comparable with the damage threshold of the mirror !!!

 

 

 

Ronic Chiche wrote:

We installed the 2 CVBG for compression after the amplifier.
We used an interferometric technique with a delay line and combining the two paths in a CCD to measure interferences... see interferences.avi video file
One can notice some misalignement at the end of the video.

After supressing the global shape of the superposed pulses, one measures the amplitude of the remaining fringes (peak-peak or standard deviation) each 250µm of the delay line (500µm of round-trip). one gets the interferences pulse shape with a FWHM of 6ps...
It seems that an 'after pulse' is visible in the interferences.

An other meausurement using a 2 photons photodiode will be used to confirm this measurement.

 

 

    Reply   Beam pulse duration, posted by Ronic Chiche at Optical room about lasers and optics interferences_pulse_duration.png

Yesterday, Loïc put the CVBG's back to their specified angles.

We used an interferometric technique with a delay line and combining the two paths in a CCD to measure interferences.

After supressing the global shape of the superposed pulses, one measures the amplitude of the remaining fringes (standard deviation) each 50µm of the delay line (100µm of round-trip). one gets the interferences pulse shape with a FWHM of 2.5 ps... (see curve)

from Fabian calculation, at 100kW, with w=2mm, the fluence on mirror should be around 0.05J/cm^2.

from this article (https://www.sciencedirect.com/science/article/pii/S0030402618313275), the damage threshold for SiO2/Ta2O5 multilayers should be around 4.8J/cm^2 @ 1030nm

 

Ronic Chiche wrote:

Yesterday, Loic installed a telescope before CVBG's to reduce the spot size on CVBG's and reduced the incident angle 0.5-1° on them (on the datasheet the specified incident angle is 2.8°).

We used an interferometric technique with a delay line and combining the two paths in a CCD to measure interferences.

After supressing the global shape of the superposed pulses, one measures the amplitude of the remaining fringes (standard deviation) each 50µm of the delay line (100µm of round-trip). one gets the interferences pulse shape with a FWHM of 2 ps... (see curve)
we still see an 'after pulse'.

if the pulse is 1ps long and 100kW is stored in the cavity, it means 3GW peak for the whole beam... it is comparable with the damage threshold of the mirror !!!

 

 

 

Ronic Chiche wrote:

We installed the 2 CVBG for compression after the amplifier.
We used an interferometric technique with a delay line and combining the two paths in a CCD to measure interferences... see interferences.avi video file
One can notice some misalignement at the end of the video.

After supressing the global shape of the superposed pulses, one measures the amplitude of the remaining fringes (peak-peak or standard deviation) each 250µm of the delay line (500µm of round-trip). one gets the interferences pulse shape with a FWHM of 6ps...
It seems that an 'after pulse' is visible in the interferences.

An other meausurement using a 2 photons photodiode will be used to confirm this measurement.

 

 

 

Entry   Telescope for compressors and ThomX, posted by Loïc Amoudry at Optical room about lasers and optics 

CVBGs telescope is done with w ~= 1mm. 

ThomX telescope is done with w ~= 2.3mm in x and y.

Newt step is to lock with these short pulses (measured ~2.5ps with FWHM standard deviation interference method).

Entry   ThomX Locked with CVBGs, posted by Loïc Amoudry at Optical room about lasers and optics tek00001.pngtek0002CH1.isftek0002CH2.isftek0002CH3.isftek0002CH4.isf

ThomX has been locked with a new telescope while using compression CVBGs.
Coupling is ~50-55% and lock is stable.

Power up to ~50kW should follow up soon (stop when HOM are observed). We'll not go over ~100kW to not risk any breakdown due to the short pulse length (~2.5 ps).

Fabian and Ronic discussed LIDT (laser induced damage threshold) for Ta2O5 at 2.5 ps is => 1J/cm²
We have w > 2 mm ( => surface ~= 0.126 cm²) . At 100 kW it means 800 kW/cm². At 33.33 MHz it means 24 mJ/cm². With 2.5 ps, peak power is 10 GW/cm² (1.2 GW peak).

 

 

Entry   Powerup with 2.5ps pulses, posted by Loïc Amoudry at Optical room about lasers and optics report.PNGtek00004.png4A.PNG

A power up has been performed on ThomX until 65kW intracavity power. We didn't see any modes but stored power was really unstable since ~60kW (see image tek00004.png).

Measurement report is shown in the table below. Gain seems to decrease against intracavity power.

As a conclusion, pulse length compression does not seems to bring any trouble in the PDH loop. But at relatively high intracavity power, power start to be really unstable even if we did not see any mode.
We can notice that we tried to align, change CEP but it had no important impact on the quality of the lock. See further experiments with Ronic and D-shape.

    Reply   Beam pulse duration, posted by Ronic Chiche at Optical room about lasers and optics Interferences_pulse_width.png

Last Thursday (20th of February), Loïc and Titouan realigned the stretcher CVBG to its nominal angle and they used an interferometric technique with a delay line and combining the two paths in a CCD to measure interferences.

After supressing the global shape of the superposed pulses, one measures the amplitude of the remaining fringes (standard deviation) each 50µm of the delay line (100µm of round-trip).
one gets the interferences pulse shape with a FWHM of 2.3 ps... (see curve)

Ronic Chiche wrote:

Yesterday, Loïc put the CVBG's back to their specified angles.

We used an interferometric technique with a delay line and combining the two paths in a CCD to measure interferences.

After supressing the global shape of the superposed pulses, one measures the amplitude of the remaining fringes (standard deviation) each 50µm of the delay line (100µm of round-trip). one gets the interferences pulse shape with a FWHM of 2.5 ps... (see curve)

from Fabian calculation, at 100kW, with w=2mm, the fluence on mirror should be around 0.05J/cm^2.

from this article (https://www.sciencedirect.com/science/article/pii/S0030402618313275), the damage threshold for SiO2/Ta2O5 multilayers should be around 4.8J/cm^2 @ 1030nm

 

Ronic Chiche wrote:

Yesterday, Loic installed a telescope before CVBG's to reduce the spot size on CVBG's and reduced the incident angle 0.5-1° on them (on the datasheet the specified incident angle is 2.8°).

We used an interferometric technique with a delay line and combining the two paths in a CCD to measure interferences.

After supressing the global shape of the superposed pulses, one measures the amplitude of the remaining fringes (standard deviation) each 50µm of the delay line (100µm of round-trip). one gets the interferences pulse shape with a FWHM of 2 ps... (see curve)
we still see an 'after pulse'.

if the pulse is 1ps long and 100kW is stored in the cavity, it means 3GW peak for the whole beam... it is comparable with the damage threshold of the mirror !!!

 

 

 

Ronic Chiche wrote:

We installed the 2 CVBG for compression after the amplifier.
We used an interferometric technique with a delay line and combining the two paths in a CCD to measure interferences... see interferences.avi video file
One can notice some misalignement at the end of the video.

After supressing the global shape of the superposed pulses, one measures the amplitude of the remaining fringes (peak-peak or standard deviation) each 250µm of the delay line (500µm of round-trip). one gets the interferences pulse shape with a FWHM of 6ps...
It seems that an 'after pulse' is visible in the interferences.

An other meausurement using a 2 photons photodiode will be used to confirm this measurement.

 

 

 

 

Entry   Optical spectrum analyzer resolution, posted by Ronic Chiche at Optical room about lasers and optics Capture.PNG

The Koheras @0.5mW is directly connected with fibers FC/APC -> FC/PC to the Avantes optical spectrum analyzer.

The FWHM wavelength, measured with the Avantes software, is 0.126nm

    Reply   Optical spectrum analyzer resolution, posted by Ronic Chiche at Optical room about lasers and optics Capture2.PNG

Same measurement with Koheras @100mW and using fiber coupling lenses between Koheras and Avantes OSA.

The FWHM wavelength, measured with the Avantes software, is 0.116nm

Ronic Chiche wrote:

The Koheras @0.5mW is directly connected with fibers FC/APC -> FC/PC to the Avantes optical spectrum analyzer.

The FWHM wavelength, measured with the Avantes software, is 0.126nm

 

Entry   Optical spectrum before and after CVBG, posted by Ronic Chiche at Optical room about lasers and optics OSA_after_amplifier_CFBG_and_before_CVBG.PNGOSA_after_amplifier_CFBG_2_CVBG.PNG

Measurement of the optical spectrum with Avantes OSA before (1st picture) and after CVBG (2nd picture) with 2nd stage on the laser amplifier.
The 1st CVBG stretches the beam horizontally due to the default incident angle and the fact that different wavelength are reflected in the CVBG with different depth.
as a result, the optical spectrum is varying along the transversal axis of the beam.
The 2nd CVBG is injected with the opposiste incident angle and should compensate the stretch effect to get back a circular beam.
spatially, the beam is quite circular but one can still see a dependance between position in the beam and optical spectrum.

Then, it is difficult to show the "right" optical spectrum after CVBG (one could use a diffuser for that) but it is clear that the spectral width is quite the same (~ 2nm due to the CFBG at the input of the amplifier) before and after CVBG.

Entry   Amplifier laser diodes issue, posted by Ronic Chiche at Optical room about lasers and optics Problem_TEC.PNGparam_diode_1.csvparam_diode_2.csv

This is a long time issue for the ThomX amplifier : on the 4 pumping diodes available, the 1st diode has a higher temperature than the others around 40°C (see the picture).
the problem comes from the TEC which is not activated (see diodes parameter files in execel format). the related software windows are shaded.

I phoned to Jerome to ask him if one can securely activate the TEC, and he answered "yes".
but once the TEC is activated by loading parameters and modifying the line of the TEC activation, the temperature does'nt fall down as it seems the TEC does'nt work properly.
maybe it is not connected properly ? to be checked...

the result is, as the diode temperature is too high, an alarm is fired and the diode is deactivated... impossible to increase the current.

one has to deactivate the TEC and shut down electrically the amplifier to go back to the starting point.

Conclusion :
- the TEC of the diode 1 does'nt work properly.
- one can't activate it otherwise an alarm is fired and the diode is disabled.
- one should check the TEC connections in the amplifier

Entry   Pointing stability with CVBG, posted by Ronic Chiche at Optical room about lasers and optics Pointing_stability_without_airflow_and_no_one_walking_around.pngPointing_stability_without_airflow_vs_amplifier_current.png

Yesterday, with Titouan, we made some measurements on pointing stability of the laser beam after amplifier and CVBG.
the amplifier beam go through a first telescope to be small enough and colimated before going to CVBG's, then go to CVBG's and then is sent through the FP-cavity telescope to the FP-cavity itself.
the total length is about 6m to mirror M1.
the surface of the mirror M1 is imaged with several wedges to a Basler CCD.
(x is for veritcal position and y is for horizontal position)

* the 1st plot shows the pointing stability at low power of amplifier without airflow and walking around during about 5 minutes extracted from the Basler CCD video.
when walking around or with airflow the pointing stabily is much worse.

* the 2nd plot shows the pointing stability vs amplifier current.
it is comparable (a litle bit worse) to the pointing stability at low power.
one clearly see the beam expanding in vertical direction and also in horizontal direction.
the effect could come from the CVBG telescope lenses which are standard lenses and not high power lenses.
the pointing effect could come also from the same effect if the lens is not perfectly centered.

we took also some picture of the CVBG at different power with the Thermal camera but we need to get the data from camera (old software not compatible with Windows 10)

Entry   Finesse measurement after confinement, posted by Loïc Amoudry at Optical room about lasers and optics sweep_parameters.PNGtek00007.pngtek0009CH1.isftek0010CH1.isftek0011CH1.isf

After 3 months without human presence, finesse has been measured 3 times with average optimization: 
- 16150
- 16067
- 16172
Last measurement was on 6th of november 2019, finesse was 17000. So it didn't change or just slightly.

We used the last Onefive telescope (used with CVBGs). Coupling reached ~30% (see pink curve on image attached).
Cavity vacuum is ~1.2e-9 mbar and pneumatic valves were still openned after these 3 months.

ELOG V3.1.4-395e101