  24   Mon Oct 4 14:57:04 2021 Aurélien MartensFixedinfodetectors and electronics | softwareOptical roomip adress of smaract driverthe ip adress of the smaract driver is
now 10.0.53.10:5000

please think about putting such
  
  25   Mon Oct 4 15:10:17 2021 Aurélien MartensFixedreportmechanics | detectors and electronics | softwareOptical roomnew position of smaract motorsfollowing the measuremetn at 879.9 MHZ
of the laser theoptical cavity length has
been adjusted to that value from 876MHz setting
  
  26   Mon Oct 4 15:24:14 2021 Aurélien MartensFixedreportlasers and optics | detectors and electronicsOptical roompiezo actuation rangeA 10V applied on the laser piezo was found
to induce a frequecny change of about 5kHz,
compatible with expectations from the piezo
  
  27   Mon Oct 4 15:38:43 2021 Aurélien MartensFixedinfodetectors and electronicsOptical roomphone is noisedoing the sensitivty measurement of the
piezo we observed satellite peaks around
800 to 900 Mhz. Removing the phones from
  
  28   Tue Oct 5 11:35:16 2021 Aurélien MartensFixedreportdetectors and electronics | softwareOptical roomnew position of smaract motorsa cavity length measurement gave 880.9Mhz.
The m1/M2 have been put to +5.1000mm both
M3/M4 srill at -2.9mm
  
  29   Tue Oct 5 12:18:34 2021 Aurélien MartensFixedreportdetectors and electronics | softwareOptical roomnew position of smaract motorsto find the fundamental mode we moved the
cavity mirrors to new positions:

+5.104mm for both planar mirrors
  
  30   Tue Oct 5 12:23:51 2021 Aurélien MartensFixedreportdetectors and electronics | softwareOptical roomnew position of smaract motorsthe M1 is not at 5.1045mm to put the fundamental
on the scanning range of the laser piezo

M2 still at 5.104mm
  
  31   Tue Oct 5 16:23:04 2021 Aurélien MartensFixedreportdetectors and electronics | softwareOptical roomnew position of smaract motorsafter 3hours (1pm --> 4pm) the cavity
moved by 12um in total

new positions: +5.1065mm on M1
  
  43   Thu Dec 2 16:12:17 2021 Aurélien MartensFixedreportmechanics | softwareOptical roomnew smaract positionsM1 and M2 at 5.1085mm and M3 and M4 at
-2.9mm
  