HOME SBOX THOMX MINICAV Utilities
MINICAV orders MINICAV installation MINICAV commissioning MINICAV control command
  Status of commissioning, report also here plots are reports.  Not logged in ELOG logo
Entry   Mesure de finesse, posted by Blanc at Optical room about lasers and optics 
    Reply   Mesure de finesse, posted by Blanc at Optical room about lasers and optics Mesure_de_finesse.png
Message ID: 62     Entry time: Fri Jan 20 18:01:22 2023     Reply to this: 63
 Author: Blanc 
 Status: Fixed 
 Type: info 
 Category: lasers and optics 
 Location: Optical room 
 Title: Mesure de finesse 

Fréquence centrale : 879.48 MHz

FWHM mesurée à la main sans sweep : 120 / 130 KHz

=> Finesse = FSR/FWHM = ~7000

Apporter un analyseur de spectre pour régler le sweep du Marconi pour une mesure de finesse plus précise.

ELOG V3.1.4-395e101