Entry   Cameras Calibration, posted by Kevin Dupraz at Optical room about lasers and optics | detectors and electronics NF_Refl_calib_hex2mm__2021-06-23__16-59-26.bmpNF_Trans_calib_hex2mm__2021-06-23__17-06-49.bmpNF_Inj_calib_DIODE__2021-06-24__10-56-00.bmpNF_Inj_calib_DIODE__2021-06-24__10-42-00.bmpNF_Trans__2021-06-25__16-38-56.bmp
After installing cameras the actual calibration are :

NF_Refl: acA1920-40gm
    pixel size (real): 5.86um
Entry   Cavity lock and references, posted by Kevin Dupraz at Optical room about lasers and optics | detectors and electronics NF_Trans_2021-09-06__17-49-00.bmp
Cavity is lock and optimized with the NKT. Input power is maximized (10mW).

Cavity mode is in attachement. The reference on camera with BeamProfiler Matlab code is the following:

FF_refl: [2.0883    1.1606]
Entry   Reflection Line Polarization Measuremet, posted by Manar Amer at Optical room about lasers and optics Reflection_line.jpg
With the help of Ronic the cavity was locked in preparation to measure the polarization of the reflection line when the cavity is locked (measurement
when it is not locked was done before)

the purpose is to compare the two measurements (locked  vs not locked)
    Reply   Cameras Calibration, posted by Kevin Dupraz at Optical room about lasers and optics | detectors and electronics grab2021-12-06_10-51-34_325_microscope.bmpgrab2021-12-06_10-52-11_108_microscope_calib_05mm_ruler.bmp
 




Kevin
Dupraz wrote:



After installing cameras the actual calibration are :
    Reply   Cameras Calibration, posted by Kevin Dupraz at Optical room about lasers and optics | detectors and electronics Plot_of_NF_Trans_calib_hex2mm__2021-06-23__17-06-49.pngPlot_of_NF_Refl_calib_hex2mm__2021-06-23__16-59-26.png
After check a mistake has been found on the magnification. This seems to be the good calibration (feel free to cross check). The projection of the 2mm
hex is attached

Entry   Manip 10 12 2021 (optimization injection Burst), posted by Kevin Dupraz at Optical room about lasers and optics NF_Refl__2021-12-10__14-25-08.pngtest__2021-12-10__14-25-08_1920_25gm.png
Manip 10/12/2021
Après recherche de la résonnance principale
    - MCS miroirs plan à 5.110250mm (en derive vers les negatifs)
Entry   Manip 14 12 2021 injection optimization (and calibration), posted by Kevin Dupraz at Optical room about lasers and optics Screenshot_2021-12-14_2_125459.png
Matin :

moteurs :
M1, M2 (plans): 5.110500m
    Reply   Manip 14 12 2021 injection optimization (and calibration), posted by Kevin Dupraz at Optical room about lasers and optics Screenshot_2021-12-14_3_182645.png
Après-midi :
    * Optimisation de l'alignement en injection pour avoir le plus de puissance en transmission
    * Optimisation (vérification) de la polarisation pour avoir le plus de couplage.
Entry   Manip 16 12 2021 montée en puissance, posted by Kevin Dupraz at Optical room about lasers and optics Screenshot_2021-12-16_0_115846.png
Matin: 
* Redémarrage Tangor 4W
    - couplage 15%
    Reply   Manip 16 12 2021 montée en puissance, posted by Kevin Dupraz at Optical room about lasers and optics Screenshot_2021-12-16_1_121553.png
 




Kevin
Dupraz wrote:



Matin: 
Entry   Changements et ajout d'optiques, posted by Blanc at Optical room about lasers and optics 
Optiques ajoutées :

- splitter 90/10 au début de la voie de réflexion, pour éjecter le flux et éviter que les lambdas ne chauffent.

- deuxième polariser pour avoir une meilleur extinction.
Entry   Manip 17 12 2021 reprise réf nouvelle injection pour polarisation, posted by Kevin Dupraz at Optical room about lasers and optics Screenshot_2021-12-17_0_164132.pngScreenshot_2021-12-17_1_170202.png
Tangor : 4.2W
* Réalignement de l'injection suite aux modifications :
    - Changement de l'ordre des waveplate maintenant : lambda/4, puis lambda/2
Entry   Changement cadence et duty cycle + ref cam, posted by Blanc at Optical room about lasers and optics Ref_NF_inj_18_05_2022.pngRef_FF_inj_18_05_2022.pngScreenshot_2022-05-18_2_165654.pngScreenshot_2022-05-18_5_173440.pngScreenshot_2022-05-18_6_174947.png
- Prise des nouvelles ref caméras : NF{709;371} , FF{389;280} [Image 1 & 2]

- Placement d'une photodiode rapide en transmission pour voir le train d'impulsions du Tangor à travers la cavité.

- Lock du Tangor en quasi-continue, P_moy = 2W, 100 kHz.
Entry   Problèmes Cavitée, posted by Blanc at Optical room about lasers and optics Transmission.PNGNF_inj.PNGFF_inj.PNG
Précédentes ref pour lesquelles la manip fonctionnait :

Caméras :

709 ; 371 NF
Entry   Phase noise measurement of the Amplitude MIKAN, posted by Ronic Chiche at Optical room about lasers and optics | detectors and electronics 
Installation of the MIKAN.
powermeter (with OD1) just after the oscillator shows 440mW for 4A of the MIKAN pump current.

    Reply   Phase noise measurement of the Amplitude MIKAN, posted by Ronic Chiche at Optical room about lasers and optics | detectors and electronics 
Coupling into a 50-50% fiber coupler using the Thorlabs XYZ table NanoMax TS.
We reached 117mW after the 50% arm which means we coupled about 230mW (the coupling better than 50%).




Ronic
Entry   Mirrors Mottors Connections , posted by Manar Amer at Other about mechanics 20210804_Motor_connections_with_mirrors_inside_cavity_box.jpg20210804_Mottor_cables_behind_cavity_box.jpg
* The connections to control the motors of the mirrors are connected in the order stated :

                 M1       
-      M2       -       M3     -   
    Reply   Mirrors Mottors Connections , posted by Manar Amer at Other about mechanics | cabling 
Added the cabling to the categories 




Manar
Amer wrote:



* The connections to control the motors of the mirrors are connected
Entry   Setting up the Motors, posted by Manar Amer at Other about mechanics | cabling | software setting_up_the_motors.docx
A document attached that describes the procedure needed to:

- connect the motors

- configure the Ethernet connection
Entry   Alignment using Helium Neon Laser , posted by Manar Amer at Other about lasers and optics 20210202__Helium_Neon_laser_parameters.jpg
05 April  2021 :  A rough alignment of the cavity was done using the Helium Neon Laser.