| ID |
Date |
Author |
Status |
Type |
Category |
Location |
Title |
|
272
|
Tue Nov 7 14:09:25 2023 |
Ronic Chiche | Fixed | info | lasers and optics | ThomX igloo | Strecher and fiber injection alignment |
Today, the power measured at the input of the amplifier (PD_IN on the LAL amplifier software) is 2.5mW instead of >3mW generally measured
The power coming from the NKT/Onefive Origami oscillator is still >37mW (measured directly with the powermeter at the laser output without OD2).
thus, the problem should come from:
- the strecher/fiber alignment.
- or maybe from a wavelength shift of the oscillator
| Ronic Chiche wrote: |
|
Monday 7/11 morning, the power in the fiber was still 6.2 mW
| Ronic Chiche wrote: |
|
this morning : still 6.2 mW at the output of the fiber.
the powermeter has been removed and the fiber connected to amplifier input fiber (no EOM connected in between).
| Ronic Chiche wrote: |
|
This morning with Manar, we continued the strecher CVBG alignment and fiber injection procedure :
- we checked the power at the output of the laser and confirmed the measurement after the Isolator : ~ 36 mW
- we tried to improve the power after the strecher CVBG by rotating the quarter-wave plate but it seems we were already at the maximum : ~ 10.5 mW
- we aligned the 2 fiber injection mirrors + schafter-kirchoff mount z-axis : we saw 5 mW at the output of the fiber but it is very difficult to keep the power after swcrewing the 2 z-axis fixing screws of the schafter-kirchoff mount.
we decided to replace the 750mm focusing lens installed by a 1000mm lens.
- we aligned again the 2 fiber injection mirrors + schafter-kirchoff mount z-axis : now, we have ~ 6.2 mW stable at the output of the fiber after screwing the 2 z-axis fixing screws of the schafter-kirchoff mount.
the procedure for aligning the schafter-kirchoff mount is :
- unscrew the z-axis and tilt fixing screws (5 screws in total).
- improve the injection with the z-axis knob and the tilt screws
- tighten very softly all the fixing srews once the optimization is finished => you will lose a part of the alignment but not completely (~ 1 mW level)
- redo a part of the alignment with the 2 aligning mirrors => you should find back the values after optimization.
we let the power-meter at the output of the fiber to check in the next days if the injected power in the fiber changes or not....
it is important to not "kick" the fiber injection box or put anything on it ! as the schafter-kirchoff mount adjustment is soooo touchy...
| Ronic Chiche wrote: |
|
Manar + Victor, today in the morning we aligned the CVBG of the stretcher.
The double path using the mirror mounts, reached up to 5.4 mW just before the injection into the fibers
then we rotated the quarter wave plate and the power increased by a factor of ~ 2 to 10.3 mW
we started with the injection into the fiber by alignment of the 2 mirror mounts.
1) fiber not connected and have everything pass through the mount
2) connect fiber loosely and increase the power in fiber bit by bit until fiber fixed fully
reached power up to 438 uW
alignment not finished using the mounts, there is still also schafter-kirchoff mount.
| Ronic Chiche wrote: |
|
I used the 33MHz spectrum measurement : https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/153
fitted by P=P0*sech²((f-f0)/df)
f0 = c/1030m
df = (c/1030nm^2) * 2.5nm
and the strecher CVBG measured data (in attached file with reference D24-02)
to estimate the expected power to be coupled into the fiber.
the corresponding plot shows 3 curves:
- black : the 33MHz laser spectrum "manually fittted" with the sech² function (mentionned at the begining of this post) to match the measured spectrum from the elog.
- blue : the strecher CVBG reflectivity curve from the Excel measured datasheet in attachement.
- red : the corresponding output power after a double path into a CVBG (the reflectivity is applied twice).
with this simulation, one can estimate the power after CVBG to be 12.5 mW for 35 mW of input power.
or 9.6 mW after CVBG for 27 mW of input power which is exactly what has been measured in a previous post by Loic : https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/70
| Ronic Chiche wrote: |
|
This morning with Manar, we measured the power at different points:
- direct measurement at the output of the 33MHz laser : 35 mW
inside the injection box :
- after the Isolator + focalization lens : 35 mW
- just at the input of the strecher CVBG : 35 mW
- atfer optimizing the alignment of the double path CVBG, in between the 2 fiber injection coupling mirrors : 5,7 mW
(the power is measured after going through the quarter-waveplate and PBS)
I found an old post from Loic claiming that with 27mW input power, we got 9.6 mW after the PBS instead of 5,7 mW !
https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/70
we have to check if the laser wavelength shifted, if the alignment could be improved, if the quarter-waveplate has the right angle,....
then, we tried to couple this 5,7mW inside the fiber using the schafter-kirchoff mount (SKM) but it is a nightmare.
changing the focus and the internal fiber angle is very sensitive, not always predictable and rarely reproductible...
I have to ask Guillaume how he used this mount...
the best power we saw in the fiber is 1mW but after screwing the fixing screws of the SKM, we lose almost all the alignment and we have something around 100-300µW...
| Ronic Chiche wrote: |
|
In preparation for tomorrow morning, I did a test with a 45° mount to have a power pickup inside the injection box.
I used the 133MHz laser.
output power measured with the powermeter : 45mW
output power measured with the powermeter + OD2 : 2mW
output power measured with the 45° mount with BB1-E03 mirror with the powermeter + OD2 : 2mW
as expected the BB1-E03 mirrors have a very good reflectance for AOI=45°
the specs give >99% @ 1030nm for both S and P polarizations.
| Ronic Chiche wrote: |
|
This morning with Viktor, we started the alignment of the CVBG and fiber.
we did a better alignment of the 2-pass CVBG.
we are able to the see a spot after PBS (after the 2 CVBG pass) which means the alignment is OK even if it can be improved.
then, we started the fiber injection alignment with the 2 last mirrors (7 & 8 on Alphanov documentation).
we saw that if we unswcrew the fiber to play on the focal position, we are able to improve a lot the power in the fiber.
2-3µW with the fiber screwed => 500µW with the fiber unscrewed.
it means the beam is not enough focused.
I will ask Guillaume Machinet his advice when injecting the 33MHz... do we need to replace the long focal lens just after the Isolator ?
and with which value approximatively ?
we played also with the Schafter+Kirchhoff mount of the fiber colimator (see attached documentation).
we loosen the 2 screws around the eccentric key which adjust the focal position + play on this eccentic key + tighten the 2 screws again.
now, we reached 330µW with the fiber properly screwed.
we have to check the available power before the fiber injection but we have very few place to place the powermeter.
maybe with a small mirror ?
| Ronic Chiche wrote: |
|
Today, after laser was modelocked, we checked the power inside the fiber coming from the Alphanov Strecher box.
it was almost zero, at the nW level.
after doing some very rough alignment, we clearly saw some power (with the powermeter) correlated with the position of mirrors, when turning the alignment knobs.
this is a first step but the output power in the fiber is still very low, about 20nW !
an additionnal 2mm L-shaped hex key is needed to do some walking alignment on the mounts.
|
|
|
|
|
|
|
|
|
|
|
273
|
Wed Nov 8 17:51:30 2023 |
Ronic Chiche | Fixed | issue | mechanics | lasers and optics | ThomX igloo | !!! strange amplifier beam pointing fluctuations !!! |
we confirmed the effect of the bunker temperature on the laser amplifier "beam pointing" fluctuations.
once the temperature is getting back to stable values, it doesn't happend again.
we bought a temperature data logguer to monitor them in the future: https://www.picotech.com/data-logger/tc-08/thermocouple-data-logger
this post close this thread.
| Ronic Chiche wrote: |
|
this morning, I locked the cavity to ~25kW without any problem.
but this afternoon, the reflected power exhibited low frequency (~1Hz) fluctuations of about 10% without any lock.
the reflected PhD is a DET10 which has a small surface.
we checked the OneFive oscillator power which is perfectly stable.
we changed the DET10 PhD for a DET100 PhD with ~1cm surface : we don't see any power fluctuation => the amplifier power seems stable.
we put back the DET10 PhD : we see these fluctuations of about 10% => it could be some pointing effect !!!
when one locks the FP-cavity, we clearly see exactly the same power fluctuations at the Transmission PhD but complementary => the sum is constant.
so, it seems clear that the beam coupling to the cavity is fluctuating due to some pointing fluctuation of the incoming beam.
a reason of these fluctuations could be the thermal jump done today because of the air cooling system of the bunker :
the temperature jumped from 25°C yesterday (and maybe still this morning ?) to 18°C this afternoon !!!
thus, some mechanical parts (the compressor CVBG ?) could be moving and then could produce these pointing fluctuations...
|
|
|
274
|
Wed Nov 8 18:52:44 2023 |
Ronic Chiche | Fixed | info | lasers and optics | detectors and electronics | ThomX igloo | some changes in the feedback scheme |
from several weeks, the maximum power stored in the FP-cavity was ~ 5kW.
today with Daniele, we finished to investigate the problem, and now the power inside the FP-cavity is back to ~50kW for 30% of laser amplifier ratio (~16W).
we optimized the signal received by the PDH photodiode by installing a large DET100 to collect more light.
if one installs a small photodiode (DET10) in the middle of the beam, the carrier signal when a FP-cavity crosses a resonance is larger because the photodiode "sees" only the part of the beam which is geometrically coupled to the cavity in its small active area, but :
1- once we will improve the geometrical coupling, the part of the incoming beam coupled to the FP-cavity will increase.
2- one need to work with a diffuser in front of this photodiode to precisely adapt the feedback loops gain : in that case the photodiode is sensitive to the whole input beam, whatever his active area size.
so, we decide to put a DET100 (which is given for 35ns rise time / 10MHz BW when connected on 50ohms).
see the scheme in attached file.
and a picture of the desktop with all the lock parameters :
the quality of the lock, seen on the reflected power signal is very good !
and the stability is only limited by the necessity to act on the laser Smaract motors to let the PZT in its working range.
dark blue : transmitted signal
green : PZT
pink : error signal
light blue : reflected signal
conclusion : it is not clear that the cavity Finesse have significantly increased during the last weeks, as we are roughly at the same level than before (47kW).
but as we precisely adapted the signal levels in the feedback scheme (PDH S/N ratio and Laselock parameters), the result is a more stable lock.
|
| Attachment 1: Capture.PNG
|
 |
| Attachment 2: 20231108_154711.jpg
|
 |
|
275
|
Thu Nov 9 16:40:16 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | detectors and electronics | ThomX igloo | Ring frequency / Laser "day by day" locking |
today, the equivalent Smaract position corresponding to the 500.25MHz ring frequency is +156µm on MCS-1/ch0 (closed loop mode) |
|
276
|
Thu Nov 9 17:02:26 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | detectors and electronics | ThomX igloo | Laser / FP-cavity "day by day" locking |
today, we locked the FP cavity at ~50kW with 30% laser amplifier ratio (16W) during almost all the day (from 10am to 4pm)
CEP optimized for MCS-1/Ch2 = -244µm at the end of the day.
once one finds the proper CEP value to reach to correct loops gain, the cavity lock and power are very stable:
one looses the lock only when one needs to use the Smaract motors to follow the long temperature drifts.
several elements of the ThomX machine have been powered ON progressively during the lock without any lock perturbation excepting for a very short time when switching ON the RF cavity (to be confirmed) and when one tried to inject electrons into the ring (loss of the electrons after few turns only)... but it's not very clear. the lock is still stable but some time one sees a lock loss without "reason"... could it be the electron loss or some bad compensation of the noise due to feedback, it's hard to say.
at the end of the day, I had to realign the FP-cavity injection and change the CEP more often than in the morning,
and surprisingly, the intra-cavity power drops a little bit at the begining of the lock (~50kW) and after some tenth of seconds (=> ~47-48kW).
it is not so much but it is very repeatable at each try.
I tried to optimize the CEP, the injection alignment, the PID parameters => it helps but at the end, I still have this slow power drop of few kW over tenth of seconds that I didn't see at the begining of the day... to be investigated.
|
|
277
|
Fri Nov 10 09:43:26 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | detectors and electronics | ThomX igloo | Ring frequency / Laser "day by day" locking |
temperature stable around 20.8°C
today, the equivalent Smaract position corresponding to the 500.25MHz ring frequency is +158.4µm on MCS-1/ch0 (closed loop mode)
it is very difficult to maintain both loops in the same time as soon as it is needed to move one motor (CFP or Laser) because of the one element (RF reference or Laser or CFP) is drifting in frequency.
to try to understand why these 3 elements seems to drift so fast one from each other, we only measure the beating frequency between the RF reference and the free running laser (without lock of the CFP)... and we see a drift around several Hz by second of the beating signal => who is guilty ? Laser or RF synthesizer ?
one can compare their respective phase noise to have an idea of their relative phase/frequency stability :
the OneFive phase noise gives +40dBc/Hz @10Hz offset in optical frequency which is 300000 (110dB) more than at 1GHz => 40 - 110 = -70dBc/Hz @10Hz offset @ 1GHz
to be compared to the SMA100A which gives -85dBc/Hz @10Hz offset @ 1GHz
conclusion : the RF reference should be more stable in long term and the drifts we see should come from the laser...
| Ronic Chiche wrote: |
|
today, the equivalent Smaract position corresponding to the 500.25MHz ring frequency is +156µm on MCS-1/ch0 (closed loop mode)
|
|
| Attachment 1: untitled.jpg
|
 |
| Attachment 2: Capture1.PNG
|
 |
|
278
|
Fri Nov 10 09:51:22 2023 |
Ronic Chiche | Fixed | report | mechanics | lasers and optics | ThomX igloo | M4 motor using icepap controller and jive/Atkpanel IHM |
Kevin moved the M4 mirror controller from the ISP controller to some ICEPAP controller.
the IHM to access this ICEPAP controller is accessible by launching 'jive' from any account ('operateur.thomx' for example).
once in the jive window, one has to select the 'device' tab, then select the OC=>OP=>OCH.02-MOT.03 device.
an AtkPanel is launched in which one can change the step values which are direclty the motor steps. |
|
279
|
Fri Nov 10 12:37:42 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | detectors and electronics | ThomX igloo | Laser / FP-cavity "day by day" locking |
as the temperature is back to a normal value ~ 20.8°C, the injected power to the amplifier (PD_IN in the Alphanov software) is back to 3.15mW without doing any alignment.
this morning the lock was around 48.5kW with 30% laser amplifier ratio (16W) after CEP/alignment tuning.
| Ronic Chiche wrote: |
|
today, we locked the FP cavity at ~50kW with 30% laser amplifier ratio (16W) during almost all the day (from 10am to 4pm)
CEP optimized for MCS-1/Ch2 = -244µm at the end of the day.
once one finds the proper CEP value to reach to correct loops gain, the cavity lock and power are very stable:
one looses the lock only when one needs to use the Smaract motors to follow the long temperature drifts.
several elements of the ThomX machine have been powered ON progressively during the lock without any lock perturbation excepting for a very short time when switching ON the RF cavity (to be confirmed) and when one tried to inject electrons into the ring (loss of the electrons after few turns only)... but it's not very clear. the lock is still stable but some time one sees a lock loss without "reason"... could it be the electron loss or some bad compensation of the noise due to feedback, it's hard to say.
at the end of the day, I had to realign the FP-cavity injection and change the CEP more often than in the morning,
and surprisingly, the intra-cavity power drops a little bit at the begining of the lock (~50kW) and after some tenth of seconds (=> ~47-48kW).
it is not so much but it is very repeatable at each try.
I tried to optimize the CEP, the injection alignment, the PID parameters => it helps but at the end, I still have this slow power drop of few kW over tenth of seconds that I didn't see at the begining of the day... to be investigated.
|
|
|
280
|
Tue Nov 14 10:35:16 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | detectors and electronics | ThomX igloo | Laser / FP-cavity "day by day" locking |
temperature since we moved to temperature probe (jump at the beginning of the plot) below the ring on a metallic base.
| Ronic Chiche wrote: |
|
as the temperature is back to a normal value ~ 20.8°C, the injected power to the amplifier (PD_IN in the Alphanov software) is back to 3.15mW without doing any alignment.
this morning the lock was around 48.5kW with 30% laser amplifier ratio (16W) after CEP/alignment tuning.
| Ronic Chiche wrote: |
|
today, we locked the FP cavity at ~50kW with 30% laser amplifier ratio (16W) during almost all the day (from 10am to 4pm)
CEP optimized for MCS-1/Ch2 = -244µm at the end of the day.
once one finds the proper CEP value to reach to correct loops gain, the cavity lock and power are very stable:
one looses the lock only when one needs to use the Smaract motors to follow the long temperature drifts.
several elements of the ThomX machine have been powered ON progressively during the lock without any lock perturbation excepting for a very short time when switching ON the RF cavity (to be confirmed) and when one tried to inject electrons into the ring (loss of the electrons after few turns only)... but it's not very clear. the lock is still stable but some time one sees a lock loss without "reason"... could it be the electron loss or some bad compensation of the noise due to feedback, it's hard to say.
at the end of the day, I had to realign the FP-cavity injection and change the CEP more often than in the morning,
and surprisingly, the intra-cavity power drops a little bit at the begining of the lock (~50kW) and after some tenth of seconds (=> ~47-48kW).
it is not so much but it is very repeatable at each try.
I tried to optimize the CEP, the injection alignment, the PID parameters => it helps but at the end, I still have this slow power drop of few kW over tenth of seconds that I didn't see at the begining of the day... to be investigated.
|
|
|
| Attachment 1: courbe_temperature_jeudi9-mardi14.png
|
 |
|
281
|
Tue Nov 14 10:51:57 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | detectors and electronics | ThomX igloo | Ring frequency / Laser "day by day" locking |
temperature stable around 21.7°C
today, the equivalent Smaract position corresponding to the 500.25MHz ring frequency is +147µm on MCS-1/ch0 (closed loop mode)
| Ronic Chiche wrote: |
|
temperature stable around 20.8°C
today, the equivalent Smaract position corresponding to the 500.25MHz ring frequency is +158.4µm on MCS-1/ch0 (closed loop mode)
it is very difficult to maintain both loops in the same time as soon as it is needed to move one motor (CFP or Laser) because of the one element (RF reference or Laser or CFP) is drifting in frequency.
to try to understand why these 3 elements seems to drift so fast one from each other, we only measure the beating frequency between the RF reference and the free running laser (without lock of the CFP)... and we see a drift around several Hz by second of the beating signal => who is guilty ? Laser or RF synthesizer ?
one can compare their respective phase noise to have an idea of their relative phase/frequency stability :
the OneFive phase noise gives +40dBc/Hz @10Hz offset in optical frequency which is 300000 (110dB) more than at 1GHz => 40 - 110 = -70dBc/Hz @10Hz offset @ 1GHz
to be compared to the SMA100A which gives -85dBc/Hz @10Hz offset @ 1GHz
conclusion : the RF reference should be more stable in long term and the drifts we see should come from the laser...
| Ronic Chiche wrote: |
|
today, the equivalent Smaract position corresponding to the 500.25MHz ring frequency is +156µm on MCS-1/ch0 (closed loop mode)
|
|
|
|
282
|
Tue Nov 14 11:33:45 2023 |
Ronic Chiche | Fixed | report | mechanics | lasers and optics | ThomX igloo | M4 motor using icepap controller and jive/Atkpanel IHM |
one problem is the FP-cavity/laser lock loss when one moves a FP-cavity motor.
with acceleration = 0.01 units/s² and velocity = 50mm/s (here, the 'mm' unit seems strange as it is very fast), it's enough to make a move fast enough for small displacements (10 steps for example to center PZT position) and it seems that the cavity stays locked (only the FP-cavity/laser is locked).
=> to be checked when both FP-cavity/laser and FP-cavity/RF-reference feedback loops are running.
| Ronic Chiche wrote: |
|
Kevin moved the M4 mirror controller from the ISP controller to some ICEPAP controller.
the IHM to access this ICEPAP controller is accessible by launching 'jive' from any account ('operateur.thomx' for example).
once in the jive window, one has to select the 'device' tab, then select the OC=>OP=>OCH.02-MOT.03 device.
an AtkPanel is launched in which one can change the step values which are direclty the motor steps.
|
|
|
283
|
Wed Nov 15 09:36:35 2023 |
Ronic Chiche | Fixed | report | detectors and electronics | ThomX igloo | bunker temperature curve |
|
| Attachment 1: 2023-11-10_to_2023-11-15_-_temperature_curve.png
|
 |
|
284
|
Thu Nov 16 17:48:29 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | detectors and electronics | ThomX igloo | Laser / FP-cavity "day by day" locking |
temperature @ 22.3°C
today, all the PC applications were closed except the web browser.
I had to restart all of them, then try to relock.
(maybe I let the terminal window open with all the apps and someone tried to unlog by removing them ?)
the cavity height was pretty misaligned.
after a rough alignment, the power inside the cavity was back at 49kW for 30% amplifier ratio.
the CEP motor needs to be adjusted a lot during the cavity heating process
| Ronic Chiche wrote: |
|
temperature since we moved to temperature probe (jump at the beginning of the plot) below the ring on a metallic base.
| Ronic Chiche wrote: |
|
as the temperature is back to a normal value ~ 20.8°C, the injected power to the amplifier (PD_IN in the Alphanov software) is back to 3.15mW without doing any alignment.
this morning the lock was around 48.5kW with 30% laser amplifier ratio (16W) after CEP/alignment tuning.
| Ronic Chiche wrote: |
|
today, we locked the FP cavity at ~50kW with 30% laser amplifier ratio (16W) during almost all the day (from 10am to 4pm)
CEP optimized for MCS-1/Ch2 = -244µm at the end of the day.
once one finds the proper CEP value to reach to correct loops gain, the cavity lock and power are very stable:
one looses the lock only when one needs to use the Smaract motors to follow the long temperature drifts.
several elements of the ThomX machine have been powered ON progressively during the lock without any lock perturbation excepting for a very short time when switching ON the RF cavity (to be confirmed) and when one tried to inject electrons into the ring (loss of the electrons after few turns only)... but it's not very clear. the lock is still stable but some time one sees a lock loss without "reason"... could it be the electron loss or some bad compensation of the noise due to feedback, it's hard to say.
at the end of the day, I had to realign the FP-cavity injection and change the CEP more often than in the morning,
and surprisingly, the intra-cavity power drops a little bit at the begining of the lock (~50kW) and after some tenth of seconds (=> ~47-48kW).
it is not so much but it is very repeatable at each try.
I tried to optimize the CEP, the injection alignment, the PID parameters => it helps but at the end, I still have this slow power drop of few kW over tenth of seconds that I didn't see at the begining of the day... to be investigated.
|
|
|
|
|
285
|
Fri Nov 17 11:44:35 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | detectors and electronics | ThomX igloo | Laser / FP-cavity "day by day" locking |
temperature @ 21.9°C
48.5kW inside CFP for 30% amplifier ratio
Frep Laser motor = 123.4µm
CEP motor = -67.8µm
below the image of the relative voltage range between CFP PZT (pink) and laser PZT (green).
the CFP PZT is driven by 10x the voltage range showed by the scope (0-100V HV output and 0-10V monitor on the Laselock)
one can see on the scope that the relative range of both PZT voltage range is roughly the same which means that the real CFP PZT sensitivity is 10x smaller than the Laser PZT sensitivity.
as the CFP PZT is driven by 0-100V and the Laser PZT is driven by 0-10V voltages, they have approximately the same range which is ~200nm (calibrated with the Laser motor in closed loop mode).
in a previous email, the CFP PZT should be given with a sensitivty of about 4nm/V => 400nm/100V of length range => 800nm/100V of CFP roundtrip range => 80e-9/100V of relative sensitivity => 40Hz of frequency peak-peak range for 500MHz carrier frequency.
there is a discrepancy between measurements (~10Hz peak-peak range for 500MHz carrier) and expected value !
| Ronic Chiche wrote: |
|
temperature @ 22.3°C
today, all the PC applications were closed except the web browser.
I had to restart all of them, then try to relock.
(maybe I let the terminal window open with all the apps and someone tried to unlog by removing them ?)
the cavity height was pretty misaligned.
after a rough alignment, the power inside the cavity was back at 49kW for 30% amplifier ratio.
the CEP motor needs to be adjusted a lot during the cavity heating process
| Ronic Chiche wrote: |
|
temperature since we moved to temperature probe (jump at the beginning of the plot) below the ring on a metallic base.
| Ronic Chiche wrote: |
|
as the temperature is back to a normal value ~ 20.8°C, the injected power to the amplifier (PD_IN in the Alphanov software) is back to 3.15mW without doing any alignment.
this morning the lock was around 48.5kW with 30% laser amplifier ratio (16W) after CEP/alignment tuning.
| Ronic Chiche wrote: |
|
today, we locked the FP cavity at ~50kW with 30% laser amplifier ratio (16W) during almost all the day (from 10am to 4pm)
CEP optimized for MCS-1/Ch2 = -244µm at the end of the day.
once one finds the proper CEP value to reach to correct loops gain, the cavity lock and power are very stable:
one looses the lock only when one needs to use the Smaract motors to follow the long temperature drifts.
several elements of the ThomX machine have been powered ON progressively during the lock without any lock perturbation excepting for a very short time when switching ON the RF cavity (to be confirmed) and when one tried to inject electrons into the ring (loss of the electrons after few turns only)... but it's not very clear. the lock is still stable but some time one sees a lock loss without "reason"... could it be the electron loss or some bad compensation of the noise due to feedback, it's hard to say.
at the end of the day, I had to realign the FP-cavity injection and change the CEP more often than in the morning,
and surprisingly, the intra-cavity power drops a little bit at the begining of the lock (~50kW) and after some tenth of seconds (=> ~47-48kW).
it is not so much but it is very repeatable at each try.
I tried to optimize the CEP, the injection alignment, the PID parameters => it helps but at the end, I still have this slow power drop of few kW over tenth of seconds that I didn't see at the begining of the day... to be investigated.
|
|
|
|
|
| Attachment 1: Screenshot_2023-11-17_11-37-50.png
|
 |
|
287
|
Thu Nov 23 12:25:07 2023 |
Ronic Chiche | Fixed | issue | mechanics | lasers and optics | detectors and electronics | ThomX igloo | Temperature control of the FP-cavity length |
as the CFP motors displacements induce a cavity unlock, we try to change the cavity length by changing the temperature of elment of the cavity.
yesterday, we tried to put a "heating cable" borrowed to the vacuum group to change the temperature of one bellows between the FP-cavity and the electron ring.
we chose a below because it is flexible and should not apply a too strong force on the cavity vessels.
we heated the cable at ~30°C but we didn't see a clear effect on the FP-cavity frequency measurement.
then, we put a heating cable around the X-ray output flange of the FP-cavity vessel and we saw a clear effect : a relatively fast (at a "second" level) frequency change.
the problem is the heating system is not remotely driven.
the cable is R=55ohms impedance and can reach 450°C for 1kW dissipated.
so today, we will try to use 2 remotely controlled Siglent SPD3303X power supplies.
they can reach V=120V DC => P=V²/R=260W => we could reach more than 100°C |
|
288
|
Thu Nov 23 12:26:26 2023 |
Ronic Chiche | Fixed | issue | lasers and optics | ThomX igloo | Temperature control of the Laser length with the inside Peltier |
|
bunker temperature @ 22.5°C
today, Sebastien Pitrel improved its python software to manually control the laser peltier.
we were able to make to some test. unfortunately, doing only one step change the 500MHz relative frequency by ~400Hz (see the attached plot from 480Hz to 100Hz which is the laser harmonic @500MHz compared to the RF reference @500MHz), which is equivalent to a round trip length variation of 8µm !!!
the present PZT's ranges are equivalent to 400nm of the round trip length. the laser PZT range could be extended by a factor 10 if one drives it using 100V instead of 10V actually, but it would not be enough to be able to use the Peltier ! :-(
if the Peltier changes the inox baseplate of the laser, the relative length change is 1e-5 /K which is equivalent to 100µm/K of the round trip length.
it means the internal temperature steps, done by the peltier, are around 0.08K.
maybe we could try to have smaller steps by using an external thermal setup ?
|
|
289
|
Fri Nov 24 12:56:39 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | ThomX igloo | Laser cavity temperature/pressure sensitivity |
laser cavity :
when one decreases the laser motor position, the laser repetition rate increases (laser cavity length decreases).
=> +/- laser motor step => +/- laser cavity roundtrip length => -/+ laser repetition rate => -/+ laser harmonic @500MHz
=> +/- 100nm => +/- 200nm => -/+ 0.7Hz @33.33MHz => -/+ 10Hz @500MHz
here is the natural variation of the laser cavity frequency beating with RF @500MHz over 1h (~1.6s / iteration)
one can see some oscillations equivalent to ~1µm of roundtrip length with ~10 minutes period and maybe a slower drift or oscillation with ~2µm of roundtrip range over the hour.
I mention that I moved the laser "PZT" motor before taking the data : could it be the reason of the 10-20min oscillations ?
during the same time, here is a probe temperature curve (the probe in stuck on the end flange, close to X-hutch, of the FP-cavity, inside the housing... not close to the laser position).
the temperature variation range is ~2.5/100 °C which induces on inox (relative length thermal effect : 17e-6 /K) a length variation of 4µm of roundtrip (10m) which could be compatible to the laser cavity length variation measured. |
| Attachment 1: Screenshot_2023-11-24_11-44-53.png
|
 |
| Attachment 2: probe_temp.png
|
 |
|
290
|
Fri Nov 24 18:16:37 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | ThomX igloo | Laser cavity temperature/pressure sensitivity |
Now, I placed a temperature probe stuck on the laser housing itself.
fig. 1
one can compare the temperature measured at the surface of the laser housing and the beating frequency with the 500MHz reference oscillator
one sees a possible very long term correlation but there is no correlation at the minute level when we see the frequency oscillation after t=2000s.
the laser housing temperature seems not to induce directly a frequency variation.
fig 2 / 3
we applied 15W on the heating wire rolled around the FP-cavity flange.
in red, we see the temperature increasing on the probe rolled around the wire, reaching almost 30°C.
we heat the inside of the housing (airflow stopped) during more than 30 minutes
in green, we don't see any variation (even if one makes a zoom) of the temperature of the probe stuck on the laser housing.
in same time, on fig 3, one can see the frequency drift.
there is no correlation between the oscillations and the temperature.
CONCLUSION :
the laser frequency fluctuations does not seem to come from the outside temperature.
| Ronic Chiche wrote: |
|
laser cavity :
when one decreases the laser motor position, the laser repetition rate increases (laser cavity length decreases).
=> +/- laser motor step => +/- laser cavity roundtrip length => -/+ laser repetition rate => -/+ laser harmonic @500MHz
=> +/- 100nm => +/- 200nm => -/+ 0.7Hz @33.33MHz => -/+ 10Hz @500MHz
here is the natural variation of the laser cavity frequency beating with RF @500MHz over 1h (~1.6s / iteration)
one can see some oscillations equivalent to ~1µm of roundtrip length with ~10 minutes period and maybe a slower drift or oscillation with ~2µm of roundtrip range over the hour.
I mention that I moved the laser "PZT" motor before taking the data : could it be the reason of the 10-20min oscillations ?
during the same time, here is a probe temperature curve (the probe in stuck on the end flange, close to X-hutch, of the FP-cavity, inside the housing... not close to the laser position).
the temperature variation range is ~2.5/100 °C which induces on inox (relative length thermal effect : 17e-6 /K) a length variation of 4µm of roundtrip (10m) which could be compatible to the laser cavity length variation measured.
|
|
| Attachment 1: temp_freq.png
|
 |
| Attachment 2: Screenshot_2023-11-24_16-47-37.png
|
 |
| Attachment 3: Screenshot_2023-11-24_16-48-02.png
|
 |
|
291
|
Fri Nov 24 18:47:25 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | ThomX igloo | Laser cavity temperature/pressure sensitivity |
the last thing we did with Daniele, is to start/stop the airflow on top of the housing (closed) to see a possible pressure effect on the laser frequency drift.
fig 1 : in green, the temperature measured with the probe stuck on the laser housing.
one can clearly see the 2 "start - wait ~10mins - stop" we did at 16h50 then at 17h30.
the air temperature blowed by the airflow is cooler than the housing temperature and we see the effect on the probe.
fig 2 : this is the laser frequency drift during the 1st airflow start/stop
the airflow has been turned on at 100 iterations and stopped at 500 iterations (~5 mins)
we don't see any correlation
fig 3 : this is again the laser frequency drift during the 2nd airflow start/stop
the airflow has been turned on at 1850 iterations and stopped at 2550 iterations (~10 mins)
we don't see any correlation
CONCLUSION : neither external temperature change or pressure variations can explain the 10-20min period oscillations observed on the laser frequency variations.
it can be either the laser temperature regulation or the RF reference oscillator temperature regulation (due to the oven of the quartz)
| Ronic Chiche wrote: |
|
Now, I placed a temperature probe stuck on the laser housing itself.
fig. 1
one can compare the temperature measured at the surface of the laser housing and the beating frequency with the 500MHz reference oscillator
one sees a possible very long term correlation but there is no correlation at the minute level when we see the frequency oscillation after t=2000s.
the laser housing temperature seems not to induce directly a frequency variation.
fig 2 / 3
we applied 15W on the heating wire rolled around the FP-cavity flange.
in red, we see the temperature increasing on the probe rolled around the wire, reaching almost 30°C.
we heat the inside of the housing (airflow stopped) during more than 30 minutes
in green, we don't see any variation (even if one makes a zoom) of the temperature of the probe stuck on the laser housing.
in same time, on fig 3, one can see the frequency drift.
there is no correlation between the oscillations and the temperature.
CONCLUSION :
the laser frequency fluctuations does not seem to come from the outside temperature.
| Ronic Chiche wrote: |
|
laser cavity :
when one decreases the laser motor position, the laser repetition rate increases (laser cavity length decreases).
=> +/- laser motor step => +/- laser cavity roundtrip length => -/+ laser repetition rate => -/+ laser harmonic @500MHz
=> +/- 100nm => +/- 200nm => -/+ 0.7Hz @33.33MHz => -/+ 10Hz @500MHz
here is the natural variation of the laser cavity frequency beating with RF @500MHz over 1h (~1.6s / iteration)
one can see some oscillations equivalent to ~1µm of roundtrip length with ~10 minutes period and maybe a slower drift or oscillation with ~2µm of roundtrip range over the hour.
I mention that I moved the laser "PZT" motor before taking the data : could it be the reason of the 10-20min oscillations ?
during the same time, here is a probe temperature curve (the probe in stuck on the end flange, close to X-hutch, of the FP-cavity, inside the housing... not close to the laser position).
the temperature variation range is ~2.5/100 °C which induces on inox (relative length thermal effect : 17e-6 /K) a length variation of 4µm of roundtrip (10m) which could be compatible to the laser cavity length variation measured.
|
|
|
| Attachment 1: Screenshot_2023-11-24_18-24-43.png
|
 |
| Attachment 2: Screenshot_2023-11-24_17-18-09.png
|
 |
| Attachment 3: Screenshot_2023-11-24_18-37-35.png
|
 |
|
292
|
Tue Nov 28 10:24:47 2023 |
Ronic Chiche | Fixed | info | mechanics | lasers and optics | ThomX igloo | Laser cavity temperature/pressure sensitivity |
today, I swapped the 500MHz RF reference oscillator for a Siglent 500MHz DDS oscillator.
see the attached plot : the beating frequency with the 500MHz laser harmonics produces the same behavior as before.
so, the oscillations should come from the laser temperature regulation.
| Ronic Chiche wrote: |
|
the last thing we did with Daniele, is to start/stop the airflow on top of the housing (closed) to see a possible pressure effect on the laser frequency drift.
fig 1 : in green, the temperature measured with the probe stuck on the laser housing.
one can clearly see the 2 "start - wait ~10mins - stop" we did at 16h50 then at 17h30.
the air temperature blowed by the airflow is cooler than the housing temperature and we see the effect on the probe.
fig 2 : this is the laser frequency drift during the 1st airflow start/stop
the airflow has been turned on at 100 iterations and stopped at 500 iterations (~5 mins)
we don't see any correlation
fig 3 : this is again the laser frequency drift during the 2nd airflow start/stop
the airflow has been turned on at 1850 iterations and stopped at 2550 iterations (~10 mins)
we don't see any correlation
CONCLUSION : neither external temperature change or pressure variations can explain the 10-20min period oscillations observed on the laser frequency variations.
it can be either the laser temperature regulation or the RF reference oscillator temperature regulation (due to the oven of the quartz)
| Ronic Chiche wrote: |
|
Now, I placed a temperature probe stuck on the laser housing itself.
fig. 1
one can compare the temperature measured at the surface of the laser housing and the beating frequency with the 500MHz reference oscillator
one sees a possible very long term correlation but there is no correlation at the minute level when we see the frequency oscillation after t=2000s.
the laser housing temperature seems not to induce directly a frequency variation.
fig 2 / 3
we applied 15W on the heating wire rolled around the FP-cavity flange.
in red, we see the temperature increasing on the probe rolled around the wire, reaching almost 30°C.
we heat the inside of the housing (airflow stopped) during more than 30 minutes
in green, we don't see any variation (even if one makes a zoom) of the temperature of the probe stuck on the laser housing.
in same time, on fig 3, one can see the frequency drift.
there is no correlation between the oscillations and the temperature.
CONCLUSION :
the laser frequency fluctuations does not seem to come from the outside temperature.
| Ronic Chiche wrote: |
|
laser cavity :
when one decreases the laser motor position, the laser repetition rate increases (laser cavity length decreases).
=> +/- laser motor step => +/- laser cavity roundtrip length => -/+ laser repetition rate => -/+ laser harmonic @500MHz
=> +/- 100nm => +/- 200nm => -/+ 0.7Hz @33.33MHz => -/+ 10Hz @500MHz
here is the natural variation of the laser cavity frequency beating with RF @500MHz over 1h (~1.6s / iteration)
one can see some oscillations equivalent to ~1µm of roundtrip length with ~10 minutes period and maybe a slower drift or oscillation with ~2µm of roundtrip range over the hour.
I mention that I moved the laser "PZT" motor before taking the data : could it be the reason of the 10-20min oscillations ?
during the same time, here is a probe temperature curve (the probe in stuck on the end flange, close to X-hutch, of the FP-cavity, inside the housing... not close to the laser position).
the temperature variation range is ~2.5/100 °C which induces on inox (relative length thermal effect : 17e-6 /K) a length variation of 4µm of roundtrip (10m) which could be compatible to the laser cavity length variation measured.
|
|
|
|
| Attachment 1: Screenshot_2023-11-28_10-21-33.png
|
 |