  Status of commissioning, report also here plots are reports., Page 17 of 24  Not logged in ELOG logo
    Reply   !!! strange FP cavity behavior => impossible to lock !!!, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics | detectors and electronics 20230920_131634.jpg20230920_131559.jpg
This morning,

- I installed the fast feedback loop.
now, the error signal goes to the Laselock AND
    Reply   !!! strange FP cavity behavior => impossible to lock !!!, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics | detectors and electronics 
this afternoon, in adjusting the CEP and improving the parameters of the lock, I was able to get ~ 47kW stable inside the cavity, always for 30% of laser
amplification ratio.

it confirms that the Finesse, and then the gain of FP-cavity, has suddenly increased thanks to some "mirrors surface cleaning"
    Reply   Fast feedback loop between laser and FP cavity, posted by Ronic Chiche at ThomX igloo about lasers and optics | detectors and electronics 
I copy the post about Fast Feedback loop : https://elog.lal.in2p3.fr/FPC/THOMX+commissioning/265

This morning,

- I installed the fast feedback loop.
    Reply   lock oscillateur 33MHz - synthé 500MHz, posted by Ronic Chiche at Optical room about detectors and electronics 
after a sudden FP-cavity Finesse increase, we had to install the Fast Feedback Loop in order to continue to lock the cavity.

this afternoon, I was able to lock both the FP-cavity and the Laser on the 500.25MHz oscillator reference.

I had to change the CEP to reduce the effective Finesse and help to get a better stabililty => P ~ 25kW inside the FP-cavity for 30% of laser
Entry   !!! strange amplifier beam pointing fluctuations !!!, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics 
this morning, I locked the cavity to ~25kW without any problem.

but this afternoon, the reflected power exhibited low frequency (~1Hz) fluctuations of about 10% without any lock.
the reflected PhD is a DET10 which has a small surface.
Entry   Laser FP operation, posted by Daniele Nutarelli at Optical room about lasers and optics IMG_20230922_105006_resized_20230922_111612040.jpg
This morning FP Laser was operating well locked at 30 KWatts in stable vay.

There are some fluctuations due tu pointing instabilities probably dues to temperature fluctuations in de bunker.

In the attacced picture are reported the lock parameters and signals.
Entry   Tests of pointing stability of the laser CFP, posted by Daniele Nutarelli at ThomX igloo about lasers and optics IMG_20231003_155009_resized_20231009_103610100.jpgIMG_20231003_155458_resized_20231009_103610449.jpg
Last monday with Victor we have cheked the stability of de pointing of the laser FP.

The climatisation was operating since 3 days in satble way. The laser pointing was very stable before to inejct in to the cavity (picture 1)
and also for the reflexion from the cavity (picture 2).
    Reply   Strecher and fiber injection alignment, posted by Ronic Chiche at ThomX igloo about lasers and optics 
Today, the power measured at the input of the amplifier (PD_IN on the LAL amplifier software) is 2.5mW instead of >3mW generally measured

The power coming from the NKT/Onefive Origami oscillator is still >37mW (measured directly with the powermeter at the laser output without
OD2).
    Reply   !!! strange amplifier beam pointing fluctuations !!!, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics 
we confirmed the effect of the bunker temperature on the laser amplifier "beam pointing" fluctuations.
once the temperature is getting back to stable values, it doesn't happend again.

we bought a temperature data logguer to monitor them in the future: https://www.picotech.com/data-logger/tc-08/thermocouple-data-logger
Entry   some changes in the feedback scheme, posted by Ronic Chiche at ThomX igloo about lasers and optics | detectors and electronics Capture.PNG20231108_154711.jpg
from several weeks, the maximum power stored in the FP-cavity was ~ 5kW.

today with Daniele, we finished to investigate the problem, and now the power inside the FP-cavity is back
to ~50kW for 30% of laser amplifier ratio (~16W).
Entry   Ring frequency / Laser "day by day" locking, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics | detectors and electronics 
today, the equivalent Smaract position corresponding to the 500.25MHz ring frequency is +156µm on MCS-1/ch0 (closed loop mode)
Entry   Laser / FP-cavity "day by day" locking, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics | detectors and electronics 
today, we locked the FP cavity at ~50kW with 30% laser amplifier ratio (16W) during almost all the day (from 10am to 4pm)
CEP optimized for MCS-1/Ch2 = -244µm at the end of the day.

once one finds the proper CEP value to reach to correct loops gain, the cavity lock and power are very stable:
    Reply   Ring frequency / Laser "day by day" locking, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics | detectors and electronics untitled.jpgCapture1.PNG
temperature stable around 20.8°C

today, the equivalent Smaract position corresponding to the 500.25MHz ring frequency is +158.4µm on MCS-1/ch0 (closed loop mode)

it is very difficult to maintain both loops in the same time as soon as it is needed to move one motor (CFP or Laser) because of the one element
Entry   M4 motor using icepap controller and jive/Atkpanel IHM, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics 
Kevin moved the M4 mirror controller from the ISP controller to some ICEPAP controller.

the IHM to access this ICEPAP controller is accessible by launching 'jive' from any account ('operateur.thomx' for example).

once in the jive window, one has to select the 'device' tab, then select the OC=>OP=>OCH.02-MOT.03 device.
    Reply   Laser / FP-cavity "day by day" locking, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics | detectors and electronics 
as the temperature is back to a normal value ~ 20.8°C, the injected power to the amplifier (PD_IN in the Alphanov software) is back to 3.15mW without
doing any alignment.

this morning the lock was around 48.5kW with 30% laser amplifier ratio (16W) after CEP/alignment tuning.
    Reply   Laser / FP-cavity "day by day" locking, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics | detectors and electronics courbe_temperature_jeudi9-mardi14.png
temperature since we moved to temperature probe (jump at the beginning of the plot) below the ring on a metallic base.




Ronic
Chiche wrote:



as the temperature is back to a normal value ~ 20.8°C, the injected
    Reply   Ring frequency / Laser "day by day" locking, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics | detectors and electronics 
temperature stable around 21.7°C

today, the equivalent Smaract position corresponding to the 500.25MHz ring frequency is +147µm on MCS-1/ch0 (closed loop mode)




Ronic
    Reply   M4 motor using icepap controller and jive/Atkpanel IHM, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics 
one problem is the FP-cavity/laser lock loss when one moves a FP-cavity motor.

with acceleration = 0.01 units/s² and velocity = 50mm/s (here, the 'mm' unit seems strange as it is very fast), it's enough
to make a move fast enough for small displacements (10 steps for example to center PZT position) and it seems that the cavity stays locked (only the FP-cavity/laser
Entry   bunker temperature curve, posted by Ronic Chiche at ThomX igloo about detectors and electronics 2023-11-10_to_2023-11-15_-_temperature_curve.png
 
    Reply   Laser / FP-cavity "day by day" locking, posted by Ronic Chiche at ThomX igloo about mechanics | lasers and optics | detectors and electronics 
temperature @ 22.3°C

today, all the PC applications were closed except the web browser.
I had to restart all of them, then try to relock.