  196   Wed Mar 15 21:13:07 2023 Manar AmerFixedreportlasers and optics | detectors and electronicsThomX iglooAmplifier Beam Profile over time This morning we took data of the beam profile
at 4 different points to see the evolution
of
 10x 
  195   Tue Mar 14 16:54:48 2023 Manar AmerFixedreportlasers and optics | detectors and electronicsThomX iglooAmplifier Beam Profile over time This morning, with Ronic, we opened the
compressor box to observe if there is any
misalignment of the mounts or the CVBG's
 comparision_Between_Beam_profile_over_time.jpg 
  194   Mon Mar 13 16:10:24 2023 Ronic ChicheFixedreportlasers and optics | softwareThomX iglooAlphanov amplifier beam profile measurementWe took reading of the beam profile of
the amplifier without a lens @ 30 cm from
the 2nd wedge for amplification percentages
 Beam_wavefront_30cm_no_Lens.jpg 
  193   Mon Mar 13 14:02:59 2023 Ronic ChicheFixedinfolasers and opticsThomX iglooCW laser installation and cleaningThis morning with Ronic , in preparation
for the mirror installation, we removed additional
mechanics and cables from the table. 
  
  192   Wed Mar 8 15:11:49 2023 Ronic ChicheFixedinfolasers and opticsThomX iglooCW laser installation and cleaningThis morning with Manar, in order to prepare
the installation of the news FP cavity mirrors,

- we installed the Koheras below
 20230308_121713.jpg 
  191   Wed Mar 8 14:49:12 2023 Ronic ChicheFixedreportlasers and optics | softwareThomX iglooAlphanov amplifier beam profile measurementthis morning with Manar,
we checked the Alphanov amplifier
issue : "The laser has been stopped
 20230308_124313.jpg 
  190   Mon Feb 27 18:23:13 2023 Ronic ChicheFixedreportlasers and optics | softwareThomX iglooAlphanov amplifier beam profile measurementThis morning with Victor we took a reading
of the beam profile using a 400 mm lens
at 70 % and 100 % amplification.
 WhatsApp_Image_2023-02-27_at_6.07.00_PM.jpeg 
  189   Thu Feb 23 10:15:30 2023 Ronic ChicheFixedreportlasers and optics | softwareThomX iglooAlphanov amplifier beam profile measurement
Yesterday,
Guillaume Machinet and Vincent Clet from
Alphanov came to fixe some issues with the
  
  188   Wed Feb 22 16:11:32 2023 Ronic ChicheFixedreportlasers and optics | softwareThomX iglooAlphanov amplifier beam profile measurementOn Monday 20th of February, Daniele and
Viktor took some beam profile measurement
data @ 50% of power amplification.
 Analyse_DonnĂ©es_20_fevrier.pptx 
  187   Tue Feb 14 12:53:09 2023 Ronic ChicheFixedreportlasers and optics | softwareThomX iglooAlphanov amplifier beam profile measurementThis morning with Daniele, we did a second
measurement of the beam profile using a different
lens of focal 400 mm.
  
  186   Fri Feb 10 09:54:46 2023 Ronic ChicheFixedreportlasers and optics | softwareThomX iglooAlphanov amplifier beam profile measurementThis morning,

- with the "Alphanov"
software => Central panel => "system"
  
  185   Wed Feb 8 14:09:50 2023 Ronic ChicheFixedreportlasers and optics | softwareThomX iglooAlphanov amplifier beam profile measurementWith Daniele after, we took a readings
of the beam with a lens of 750 mm.

From the observed data, we can
 20230208_error_watchdog.jpeg 
  184   Wed Feb 8 11:25:50 2023 Ronic ChicheFixedreportlasers and optics | softwareThomX iglooAlphanov amplifier beam profile measurementwe have had this new issue (watchdog alarm
switching off the amplifier) for several
days.
 scope.jpg 
  183   Mon Jan 30 17:40:10 2023 Manar AmerFixedreportlasers and optics | detectors and electronicsThomX iglooAlphanov amplifier beam profile Data This Morning with Victor the amplifier
was turned on @ 20% amplification (output
~ 9 W ) for approximately 3 hours and no
  
  182   Mon Jan 30 17:28:33 2023 Ronic ChicheFixedreportlasers and optics | softwareThomX iglooAlphanov amplifier beam profile measurementthe laser amplifier worked for some days
but it ended to a "watchdog" alarm
which switches off the amplifier... investigation
  
  181   Mon Jan 30 17:28:00 2023 Ronic ChicheFixedreportlasers and optics | softwareThomX iglooAlphanov amplifier beam profile measurementRonic on Friday operated the Amplifier
from the Alphanov software, switched it off
then turned on the LAL software,
  
  180   Wed Jan 11 17:38:13 2023 Ronic ChicheFixedinfolasers and opticsThomX iglooAlphanov amplifier beam profile measurementthis morning with Manar,

- we fixed the chiller problem
(+/-2°C fluctuation around the set temperature
  
  179   Wed Jan 11 17:31:00 2023 Ronic ChicheFixedinfolasers and optics | softwareThomX iglooAlphanov amplifier beam profile measurementToday with
Ronic 

For
 error_signal_Temperature_case_amplifier.jpg 
  178   Mon Jan 9 14:03:16 2023 Manar AmerFixedreportlasers and optics | detectors and electronicsThomX iglooAlphanov amplifier beam profile Data Adding the schematic of the measurements.,

the 4°
Beam Deviation, AR Coating ( B coating)
 8x 
  177   Mon Jan 9 12:11:14 2023 Ronic ChicheFixedinfolasers and opticsThomX iglooAlphanov amplifier beam profile measurementthis morning with Manar,

- we checked first the chiller
problem.
  