HOME SBOX THOMX MINICAV Utilities
SBOX orders SBOX installation SBOX commissioning SBOX control command
  Status of commissioning, report also here plots are reports., Page 8 of 13  Not logged in ELOG logo
ID Date Author Status Type Category Location Title
  112   Tue Jul 12 10:19:30 2022 Manar AmerFixedreportlasers and optics | detectors and electronicsOptical roomFSR change & Finesse Measurements

Yesterday evening the cavity was Vacuum pumped up to pressure of 5.5*10^-2 and locked 

changed FSR to be 216.662 MHz and alignment a little and measured the Finesse

in Vacuum we have average Finesse = 30341.6265

FWHM (KHz) = 7.0592
Finesse = 30692.1961

FWHM (KHz) = 7.2186
Finesse = 30014.556

FWHM (KHz) = 7.1051
Finesse = 30493.7635

FWHM (KHz) = 7.1079
Finesse = 30481.9812

FWHM (KHz) = 7.1413
Finesse = 30339.2695

FWHM (KHz) = 7.1624
Finesse = 30249.776

FWHM (KHz) = 7.0239
Finesse = 30846.2719

FWHM (KHz) = 7.2614
Finesse = 29837.6477

FWHM (KHz) = 7.1935
Finesse = 30119.1768

 

Manar Amer wrote:

The FSR of the 2 mirror (plan-spherical) Cavity was adjusted from 210 MHz to reach 216.643 MHz

it was done by having two reference irises, one at the injection point and one at the reflection 

then changing the position of injection plan mirror to slightly closer distance and monitoring the reflection on the oscilloscope to be max.

The cavity modes were still seen, and we had to only improve the injection alignment after.

Me and Ronic locked in air and measured the Finesse, which was bigger by ~ 20%

average Finesse = 30208.53614

FWHM (KHz) = 7.0179
Finesse = 30869.9522

FWHM (KHz) = 7.1257
Finesse = 30403.005

FWHM (KHz) = 7.1287
Finesse = 30390.4014

FWHM (KHz) = 7.2884
Finesse = 29724.5531

FWHM (KHz) = 7.3055
Finesse = 29654.769

 

 

Manar Amer wrote:

Update for Finesse measurement, The cavity was put under vacuum ~ 1.1*10^-1 mbar

and the alignment and coupling improved.

FSR = 210.1 MHz

Average Finesse = 25686.46222

 FWHM (KHz) = 8.2387
Finesse = 25501.5659

FWHM (KHz) = 8.2028
Finesse = 25613.2858

FWHM (KHz) = 8.0978
Finesse = 25945.3289

FWHM (KHz) = 8.1744
Finesse = 25702.3142

FWHM (KHz) = 8.1847
Finesse = 25669.8163

Concluded from Ronic's calculations, this could be the maximum finesse we might be able to obtain with this setup

with Gain ~ 8000

On Monday we adjust the frequency to match 2160.66 MHz and lock the Pulsed,

at the same time start we start with the CELIA amplifier.

 

 

 

Manar Amer wrote:

The cavity was realigned using irises instead of pinholes, gave a better alignment.

The inside of the box, the spherical and the injection mirror were cleaned and placed back inside the box.

we see beating of fundamental mode, previously at the transmission point we placed a wedge to split the beam which resulted in an elliptical mode

we removed it and placed a very thin beam splitter, the beam is circular now.

The cavity was locked in air at a coupling of ~ 60-70 %

Finesse and line width measured five readings with a Finesse average 25095.08884  of a Gain ~ 8000

FWHM (KHz) = 8.2928
Finesse = 25323.0544

FWHM (KHz) = 7.9202
Finesse = 26514.4395

FWHM (KHz) = 8.5834
Finesse = 24465.8636

FWHM (KHz) = 8.4571
Finesse = 24831.2419

FWHM (KHz) = 8.6275
Finesse = 24340.8448

Theoretical and expected Finesse for the 2 mirror setup with the losses is calculated by Ronic for comparison between four and 2 mirror setup.

 

 

Manar Amer wrote:

The SBox cavity setup was changed to have only 2 mirrors M1 plane and M2 spherical, both from ThomX

Distance between the mirror ~ 72 cm , increased from 70 cm to take into account the thickness of the ThomX mirrors

Two lenses (300 mm @ 50 cm , 200 @ 104 cm) were placed to have the beam radius ~ 0.55 mm.

The cavity was locked with a coupling of 60 %, for Finesse measurement the sweep was taken over 100 KHz of 2 seconds.

FSR ~ 210.00 MHz, line width ~ 8.56 KHz, Finesse ~ 24 500 .

 

 

 

 

 

 

Attachment 1: Screenshot_2022-07-11_1_204143.png
Screenshot_2022-07-11_1_204143.png
Attachment 2: Screenshot_2022-07-11_0_203646.png
Screenshot_2022-07-11_0_203646.png
  111   Tue Jul 12 10:07:05 2022 Manar AmerFixedreportlasers and optics | detectors and electronicsOptical roomFSR change & Finesse Measurements

The FSR of the 2 mirror (plan-spherical) Cavity was adjusted from 210 MHz to reach 216.643 MHz

it was done by having two reference irises, one at the injection point and one at the reflection 

then changing the position of injection plan mirror to slightly closer distance and monitoring the reflection on the oscilloscope to be max.

The cavity modes were still seen, and we had to only improve the injection alignment after.

Me and Ronic locked in air and measured the Finesse, which was bigger by ~ 20%

average Finesse = 30208.53614

FWHM (KHz) = 7.0179
Finesse = 30869.9522

FWHM (KHz) = 7.1257
Finesse = 30403.005

FWHM (KHz) = 7.1287
Finesse = 30390.4014

FWHM (KHz) = 7.2884
Finesse = 29724.5531

FWHM (KHz) = 7.3055
Finesse = 29654.769

 

 

Manar Amer wrote:

Update for Finesse measurement, The cavity was put under vacuum ~ 1.1*10^-1 mbar

and the alignment and coupling improved.

FSR = 210.1 MHz

Average Finesse = 25686.46222

 FWHM (KHz) = 8.2387
Finesse = 25501.5659

FWHM (KHz) = 8.2028
Finesse = 25613.2858

FWHM (KHz) = 8.0978
Finesse = 25945.3289

FWHM (KHz) = 8.1744
Finesse = 25702.3142

FWHM (KHz) = 8.1847
Finesse = 25669.8163

Concluded from Ronic's calculations, this could be the maximum finesse we might be able to obtain with this setup

with Gain ~ 8000

On Monday we adjust the frequency to match 2160.66 MHz and lock the Pulsed,

at the same time start we start with the CELIA amplifier.

 

 

 

Manar Amer wrote:

The cavity was realigned using irises instead of pinholes, gave a better alignment.

The inside of the box, the spherical and the injection mirror were cleaned and placed back inside the box.

we see beating of fundamental mode, previously at the transmission point we placed a wedge to split the beam which resulted in an elliptical mode

we removed it and placed a very thin beam splitter, the beam is circular now.

The cavity was locked in air at a coupling of ~ 60-70 %

Finesse and line width measured five readings with a Finesse average 25095.08884  of a Gain ~ 8000

FWHM (KHz) = 8.2928
Finesse = 25323.0544

FWHM (KHz) = 7.9202
Finesse = 26514.4395

FWHM (KHz) = 8.5834
Finesse = 24465.8636

FWHM (KHz) = 8.4571
Finesse = 24831.2419

FWHM (KHz) = 8.6275
Finesse = 24340.8448

Theoretical and expected Finesse for the 2 mirror setup with the losses is calculated by Ronic for comparison between four and 2 mirror setup.

 

 

Manar Amer wrote:

The SBox cavity setup was changed to have only 2 mirrors M1 plane and M2 spherical, both from ThomX

Distance between the mirror ~ 72 cm , increased from 70 cm to take into account the thickness of the ThomX mirrors

Two lenses (300 mm @ 50 cm , 200 @ 104 cm) were placed to have the beam radius ~ 0.55 mm.

The cavity was locked with a coupling of 60 %, for Finesse measurement the sweep was taken over 100 KHz of 2 seconds.

FSR ~ 210.00 MHz, line width ~ 8.56 KHz, Finesse ~ 24 500 .

 

 

 

 

 

Attachment 1: 00mode.jpg
00mode.jpg
Attachment 2: 00mode_diameter.jpg
00mode_diameter.jpg
Attachment 3: 00mode_diameter_fit.jpg
00mode_diameter_fit.jpg
  110   Fri Jul 8 19:37:24 2022 Manar AmerFixedreportlasers and optics | detectors and electronicsOptical room2 Mirror Setup @ 216.6 MHz

Update for Finesse measurement, The cavity was put under vacuum ~ 1.1*10^-1 mbar

and the alignment and coupling improved.

FSR = 210.1 MHz

Average Finesse = 25686.46222

 FWHM (KHz) = 8.2387
Finesse = 25501.5659

FWHM (KHz) = 8.2028
Finesse = 25613.2858

FWHM (KHz) = 8.0978
Finesse = 25945.3289

FWHM (KHz) = 8.1744
Finesse = 25702.3142

FWHM (KHz) = 8.1847
Finesse = 25669.8163

Concluded from Ronic's calculations, this could be the maximum finesse we might be able to obtain with this setup

with Gain ~ 8000

On Monday we adjust the frequency to match 2160.66 MHz and lock the Pulsed,

at the same time start we start with the CELIA amplifier.

 

 

 

Manar Amer wrote:

The cavity was realigned using irises instead of pinholes, gave a better alignment.

The inside of the box, the spherical and the injection mirror were cleaned and placed back inside the box.

we see beating of fundamental mode, previously at the transmission point we placed a wedge to split the beam which resulted in an elliptical mode

we removed it and placed a very thin beam splitter, the beam is circular now.

The cavity was locked in air at a coupling of ~ 60-70 %

Finesse and line width measured five readings with a Finesse average 25095.08884  of a Gain ~ 8000

FWHM (KHz) = 8.2928
Finesse = 25323.0544

FWHM (KHz) = 7.9202
Finesse = 26514.4395

FWHM (KHz) = 8.5834
Finesse = 24465.8636

FWHM (KHz) = 8.4571
Finesse = 24831.2419

FWHM (KHz) = 8.6275
Finesse = 24340.8448

Theoretical and expected Finesse for the 2 mirror setup with the losses is calculated by Ronic for comparison between four and 2 mirror setup.

 

 

Manar Amer wrote:

The SBox cavity setup was changed to have only 2 mirrors M1 plane and M2 spherical, both from ThomX

Distance between the mirror ~ 72 cm , increased from 70 cm to take into account the thickness of the ThomX mirrors

Two lenses (300 mm @ 50 cm , 200 @ 104 cm) were placed to have the beam radius ~ 0.55 mm.

The cavity was locked with a coupling of 60 %, for Finesse measurement the sweep was taken over 100 KHz of 2 seconds.

FSR ~ 210.00 MHz, line width ~ 8.56 KHz, Finesse ~ 24 500 .

 

 

 

 

Attachment 1: Screenshot_2022-07-08_2_182405.png
Screenshot_2022-07-08_2_182405.png
  109   Fri Jul 8 17:48:48 2022 Manar AmerFixedreportlasers and optics | detectors and electronicsOptical room2 Mirror Setup @ 216.6 MHz

The cavity was realigned using irises instead of pinholes, gave a better alignment.

The inside of the box, the spherical and the injection mirror were cleaned and placed back inside the box.

we see beating of fundamental mode, previously at the transmission point we placed a wedge to split the beam which resulted in an elliptical mode

we removed it and placed a very thin beam splitter, the beam is circular now.

The cavity was locked in air at a coupling of ~ 60-70 %

Finesse and line width measured five readings with a Finesse average 25095.08884  of a Gain ~ 8000

FWHM (KHz) = 8.2928
Finesse = 25323.0544

FWHM (KHz) = 7.9202
Finesse = 26514.4395

FWHM (KHz) = 8.5834
Finesse = 24465.8636

FWHM (KHz) = 8.4571
Finesse = 24831.2419

FWHM (KHz) = 8.6275
Finesse = 24340.8448

Theoretical and expected Finesse for the 2 mirror setup with the losses is calculated by Ronic for comparison between four and 2 mirror setup.

 

 

Manar Amer wrote:

The SBox cavity setup was changed to have only 2 mirrors M1 plane and M2 spherical, both from ThomX

Distance between the mirror ~ 72 cm , increased from 70 cm to take into account the thickness of the ThomX mirrors

Two lenses (300 mm @ 50 cm , 200 @ 104 cm) were placed to have the beam radius ~ 0.55 mm.

The cavity was locked with a coupling of 60 %, for Finesse measurement the sweep was taken over 100 KHz of 2 seconds.

FSR ~ 210.00 MHz, line width ~ 8.56 KHz, Finesse ~ 24 500 .

 

 

 

Attachment 1: 00mode.jpg
00mode.jpg
Attachment 2: 00mode_diameter.jpg
00mode_diameter.jpg
Attachment 3: Screenshot_2022-07-08_0_164115.png
Screenshot_2022-07-08_0_164115.png
Attachment 4: Slide1.JPG
Slide1.JPG
Attachment 5: Slide2.JPG
Slide2.JPG
  108   Fri Jul 8 17:10:44 2022 Manar AmerFixedreportlasers and opticsOptical roomImages of ThomX mirrors

Mirror images in the previous log showing the metal deposit of the coating of the spherical mirrors

(not sent by mail due to large size of images)

Manar Amer wrote:

Under the microscope,

the spherical mirrors show the deposit of metal dots on the reflecting surface and not on the back of the mirrors

The mirror with fewer spots near the center was used in the setup.

 

 

  107   Fri Jul 8 17:09:26 2022 Manar AmerFixedreportlasers and opticsOptical roomImages of ThomX mirrors

Under the microscope,

the spherical mirrors show the deposit of metal dots on the reflecting surface and not on the back of the mirrors

The mirror with fewer spots near the center was used in the setup.

 

Attachment 1: ThomX_Spherical_Front_coating.jpg
ThomX_Spherical_Front_coating.jpg
Attachment 2: ThomX_Spherical_2_Front_coating.jpg
ThomX_Spherical_2_Front_coating.jpg
Attachment 3: ThomX_Spherical_back.jpg
ThomX_Spherical_back.jpg
  106   Wed Jul 6 19:57:17 2022 Manar AmerFixedreportlasers and optics | detectors and electronicsOptical room2 Mirror Setup @ 216.6 MHz

The SBox cavity setup was changed to have only 2 mirrors M1 plane and M2 spherical, both from ThomX

Distance between the mirror ~ 72 cm , increased from 70 cm to take into account the thickness of the ThomX mirrors

Two lenses (300 mm @ 50 cm , 200 @ 104 cm) were placed to have the beam radius ~ 0.55 mm.

The cavity was locked with a coupling of 60 %, for Finesse measurement the sweep was taken over 100 KHz of 2 seconds.

FSR ~ 210.00 MHz, line width ~ 8.56 KHz, Finesse ~ 24 500 .

 

 

Attachment 1: 20220706_Setup.jpg
20220706_Setup.jpg
Attachment 2: Screenshot_2022-07-06_1_181116.png
Screenshot_2022-07-06_1_181116.png
Attachment 3: Finesse_Fit.jpg
Finesse_Fit.jpg
Attachment 4: 00mode.jpg
00mode.jpg
Attachment 5: 00mode_diameter.jpg
00mode_diameter.jpg
  105   Wed Jul 6 18:39:38 2022 Manar AmerFixedreportlasers and optics | vacuumOptical roomSBox Finesse at 216.6625 MHz

 

Manar Amer wrote:

Max Finesse obtained @ 4 mirror setup was ~ 19 300 

With M1 from ThomX , M2/M3/M4 from SBox

Manar Amer wrote:

The Sbox currently has the injection from ThomX installed T ~ 120 ppm , M2 planar from Sbox spare and M3 / M4 spherical mirrors from Sbox installed.

The last finesse measurement that was done gave a Finesse of ~ 19300 (After cleaning M2 using ethanol + aceton + pure water)

on Friday the whole optical table was cleaned again and the inside of the box was cleaned (even with a lot of dust inside the box the Finesse was 19300 )

all the mirrors were cleaned after using the three steps, M3 and M4 had clear visible dust on them.

D-shaped mirrors were installed again between M1 and M2

The alignment was redone and the beating observed ,  There is still a need for alignment improvement.

the cavity is under vacuum ~ 4.8*10^-2 mbar on Friday evening (~ 19h00) , The cavity is yet to be locked and to measure its Finesse

A PowerPoint showing the last finesse measurement and the error signal of the menhir laser lock

 

 

 

 

  104   Wed Jul 6 18:39:25 2022 Manar AmerFixedreportlasers and optics | vacuumOptical roomSBox Finesse at 216.6625 MHz

Max Finesse obtained @ 4 mirror setup was ~ 19 300 

With M1 from ThomX , M2/M3/M4 from SBox

Manar Amer wrote:

The Sbox currently has the injection from ThomX installed T ~ 120 ppm , M2 planar from Sbox spare and M3 / M4 spherical mirrors from Sbox installed.

The last finesse measurement that was done gave a Finesse of ~ 19300 (After cleaning M2 using ethanol + aceton + pure water)

on Friday the whole optical table was cleaned again and the inside of the box was cleaned (even with a lot of dust inside the box the Finesse was 19300 )

all the mirrors were cleaned after using the three steps, M3 and M4 had clear visible dust on them.

D-shaped mirrors were installed again between M1 and M2

The alignment was redone and the beating observed ,  There is still a need for alignment improvement.

the cavity is under vacuum ~ 4.8*10^-2 mbar on Friday evening (~ 19h00) , The cavity is yet to be locked and to measure its Finesse

A PowerPoint showing the last finesse measurement and the error signal of the menhir laser lock

 

 

 

  103   Wed Jul 6 18:37:02 2022 Manar AmerFixedreportlasers and opticsOptical roomDivergence of the CW beam after collimator

Closing the series of the telescope for the beam @ 133.33 MHz

 

 

Manar Amer wrote:

Divergence of the CW beam out of the collimator measured after 1 meter to be 

  • 2 \Thetax = 0.7425  mrad  --> \Thetax = 0.37125  mrad
  • 2 \Thetay = 0.7115  mrad  --> \Thetay = 0.3575    mrad

two lenses were placed to decrease the diameter of the injected beam

 

  102   Wed Jul 6 18:35:38 2022 Manar AmerFixedreportlasers and opticsOptical roomFundamental Mode TEM00

Closing series of 00 mode for FSR @ 133.33 MHz

 

Manar Amer wrote:

Images of the Transmitted beam at M2 mirror at different days. 

The first one was from last week, the second from yesterday.

There is misalignment (drift due to temperature) from the position where the beam was initially. ( reference taken between centers of mode and direct injected beam) 

it is most clear in the vertical position.

Manar Amer wrote:

Adding images from the transmission after the spherical mirrors M3 and M4 window for a resonating mode inside the cavity.

There appears to be some misalignment from the center of the mirror ???

 

Manar Amer wrote:

The current setup for the SBox has a beam viewer and 2 photodiodes placed at reflection and transmission.

The CW laser piezoelectric crystal was driven using the lase lock system to observe the 00 mode.

Mode Observed and has no reflection (from what was previously seen)

higher order modes are resonating with 00 mode , example 11 mode (shown in image attached)

the coupling is yet to be seen and measured.

 

 

 

  101   Wed Jul 6 18:34:12 2022 Manar AmerFixedreportlasers and optics | detectors and electronics | softwareOptical roomCavity Lock

Closing the series for the setup

The cavity was only locked @ 133.33 MHz but the Finesse was not measured

The FSR was changed to match the new laser MENHIR 216.6625 MHz 

Manar Amer wrote:

Cavity locked today under vacuum and no airflow, it is stable over a long time

the alignment was improved and an increase of the transmission by 20 % from its original value

unfortunately,  data from the oscilloscope was saved, but it was not in the correct format (image) to show.

Attached ; the mode shape, the new motor positions. 

 

Manar Amer wrote:

Current setup at SBox for locking the cavity

After spacial matching, the coupling increased to ~ 15 - 20 % 

The locking was done for the fundamental mode, tried to improve the coupling with the motors, but the lock is very unstable to be investigated. (Note: the cavity in under normal pressure)

Manar Amer wrote:

Start with the cavity lock images attached, taken from last week, at start we have a coupling ~ 5%  and started improving on the alignment .

The image of 00 modes attached corresponds to the coupling seen on the oscilloscope.

We have a higher coupling with the 11 mode due to the size of the injected beam in comparison to mode

lenses were added to improve it.

 

 

 

  100   Mon Jun 27 08:52:46 2022 Manar AmerFixedreportlasers and optics | vacuumOptical roomSBox Finesse at 216.6625 MHz

The Sbox currently has the injection from ThomX installed T ~ 120 ppm , M2 planar from Sbox spare and M3 / M4 spherical mirrors from Sbox installed.

The last finesse measurement that was done gave a Finesse of ~ 19300 (After cleaning M2 using ethanol + aceton + pure water)

on Friday the whole optical table was cleaned again and the inside of the box was cleaned (even with a lot of dust inside the box the Finesse was 19300 )

all the mirrors were cleaned after using the three steps, M3 and M4 had clear visible dust on them.

D-shaped mirrors were installed again between M1 and M2

The alignment was redone and the beating observed ,  There is still a need for alignment improvement.

the cavity is under vacuum ~ 4.8*10^-2 mbar on Friday evening (~ 19h00) , The cavity is yet to be locked and to measure its Finesse

A PowerPoint showing the last finesse measurement and the error signal of the menhir laser lock

 

 

Attachment 1: 20220624_SBox_MENHIR.pptx
  99   Tue Jun 21 23:52:28 2022 Manar AmerFixedreportmechanics | lasers and optics | detectors and electronicsOptical roomMenhir Pulsed Laser

The SBox Length (FSR), was changed to match the repetition rate of the Menhir pulsed laser Frep = 216.6625 MHz

It was changed by removing the 2 D-shaped mirrors placed between M1 and M2, distance between M1-M2 ~= 169 mm 

it was still needed to be adjusted with M1 motor to reach the correct FSR.

The pulsed laser is placed on the able with injection into a fiber

The Finesse of the cavity has progressed from ~2500  to ~ 18 000

 

** Note: PowerPoint attached shows the progress over the last two weeks on the laser and 

Attachment 1: MENHIR_injection_into_fiber.jpg
MENHIR_injection_into_fiber.jpg
Attachment 2: SBox_M.jpg
SBox_M.jpg
Attachment 3: 20220620_SBox_MENHIR.pptx
  98   Thu Jun 9 19:59:42 2022 Manar AmerFixedreportlasers and optics | detectors and electronics | softwareOptical roomCavity Lock

Cavity locked today under vacuum and no airflow, it is stable over a long time

the alignment was improved and an increase of the transmission by 20 % from its original value

unfortunately,  data from the oscilloscope was saved, but it was not in the correct format (image) to show.

Attached ; the mode shape, the new motor positions. 

 

Manar Amer wrote:

Current setup at SBox for locking the cavity

After spacial matching, the coupling increased to ~ 15 - 20 % 

The locking was done for the fundamental mode, tried to improve the coupling with the motors, but the lock is very unstable to be investigated. (Note: the cavity in under normal pressure)

Manar Amer wrote:

Start with the cavity lock images attached, taken from last week, at start we have a coupling ~ 5%  and started improving on the alignment .

The image of 00 modes attached corresponds to the coupling seen on the oscilloscope.

We have a higher coupling with the 11 mode due to the size of the injected beam in comparison to mode

lenses were added to improve it.

 

 

Attachment 1: 00_Mode_Locked_Under_vacuum.bmp
Attachment 2: Ma1_Ma2_Horizontal_end-of-day.jpg
Ma1_Ma2_Horizontal_end-of-day.jpg
Attachment 3: Ma1_Ma2_Vertical_end-of-day.jpg
Ma1_Ma2_Vertical_end-of-day.jpg
  97   Wed Jun 8 11:31:47 2022 Manar AmerFixedreportlasers and optics | detectors and electronics | softwareOptical roomCavity Lock

Current setup at SBox for locking the cavity

After spacial matching, the coupling increased to ~ 15 - 20 % 

The locking was done for the fundamental mode, tried to improve the coupling with the motors, but the lock is very unstable to be investigated. (Note: the cavity in under normal pressure)

Manar Amer wrote:

Start with the cavity lock images attached, taken from last week, at start we have a coupling ~ 5%  and started improving on the alignment .

The image of 00 modes attached corresponds to the coupling seen on the oscilloscope.

We have a higher coupling with the 11 mode due to the size of the injected beam in comparison to mode

lenses were added to improve it.

 

Attachment 1: 20220607_SBox_Setup.png
20220607_SBox_Setup.png
Attachment 2: Ma1_Ma2_Horizontal_end-of-day.jpg
Ma1_Ma2_Horizontal_end-of-day.jpg
Attachment 3: Ma1_Ma2_Vertical_end-of-day.jpg
Ma1_Ma2_Vertical_end-of-day.jpg
  96   Wed Jun 8 11:24:35 2022 Manar AmerFixedreportlasers and opticsOptical roomTelescope - Spacial Matching

Divergence of the CW beam out of the collimator measured after 1 meter to be 

  • 2 \Thetax = 0.7425  mrad  --> \Thetax = 0.37125  mrad
  • 2 \Thetay = 0.7115  mrad  --> \Thetay = 0.3575    mrad

two lenses were placed to decrease the diameter of the injected beam

Attachment 1: SBox_Spacial_Matching_Telescope.png
SBox_Spacial_Matching_Telescope.png
  95   Wed Jun 8 11:10:17 2022 Manar AmerFixedreportlasers and optics | detectors and electronics | softwareOptical roomCavity Lock

Start with the cavity lock images attached, taken from last week, at start we have a coupling ~ 5%  and started improving on the alignment .

The image of 00 modes attached corresponds to the coupling seen on the oscilloscope.

We have a higher coupling with the 11 mode due to the size of the injected beam in comparison to mode

lenses were added to improve it.

Attachment 1: Screenshot_2022-05-30_0_125247.png
Screenshot_2022-05-30_0_125247.png
Attachment 2: 20220530_00_Mode.bmp
Attachment 3: 11.bmp
  94   Wed Jun 8 10:58:15 2022 Manar AmerFixedreportlasers and opticsOptical roomFundamental Mode TEM00

Images of the Transmitted beam at M2 mirror at different days. 

The first one was from last week, the second from yesterday.

There is misalignment (drift due to temperature) from the position where the beam was initially. ( reference taken between centers of mode and direct injected beam) 

it is most clear in the vertical position.

Manar Amer wrote:

Adding images from the transmission after the spherical mirrors M3 and M4 window for a resonating mode inside the cavity.

There appears to be some misalignment from the center of the mirror ???

 

Manar Amer wrote:

The current setup for the SBox has a beam viewer and 2 photodiodes placed at reflection and transmission.

The CW laser piezoelectric crystal was driven using the lase lock system to observe the 00 mode.

Mode Observed and has no reflection (from what was previously seen)

higher order modes are resonating with 00 mode , example 11 mode (shown in image attached)

the coupling is yet to be seen and measured.

 

 

Attachment 1: 20220530_00_Mode.bmp
Attachment 2: 00_Mode_Beam_Transmitted_Position_drift_due_to_temperature.bmp
  93   Wed Jun 8 10:47:45 2022 Manar AmerFixedreportlasers and opticsOptical roomSBox Alignment

Closing Series for Alignment.

Beating observed, beam is in the locking stages

 

Manar Amer wrote:

The external and internal alignment was redone.

External Alignment : the collimator position moved so in order to have a good start I placed 2 metallic mirrors and did a rough retrace of it's line through the cavity box.

It helped as I did not have to do a lot of adjustment on Ma1 and Ma2 position (new position attached)

 

Internal Alignment: I replaced the 2 pinholes inside the cavity box with another set where the holes are more symmetric, it shows in the alignment.

Images attached of the beam diffraction pattern after each mirror pinhole.

The injection mirror was placed at the end of the alignment.

Alignment Done !!!!!!!!!!!!!

 

Manar Amer wrote:

M2 Sbox changed to spare. Alignment of M2 was redone. 

With the help of Ronic , We observe fundamental mode of the cavity with higher order modes. A frequency scan on the piezo was done to find the beating frequency , the shape is elliptical with no clear deformation.

The beam seems to be hitting something, which causes a reflection seen on the beam profiler.

We started a slow scan to improvement of the outside alignment.

At the end of the day, I can only get the TM01 and TM10 modes,

alignment to be continued tomorrow + the scan offset

Note: observation on TM01 and TM10 modes, we clearly see that the beam position is different from the mode center position !!

 

Manar Amer wrote:

Yesterday, to find the source of why the mode is deformed, I tried to rotate the mirrors to see if there is a change.

I started with M2 and rotated it by ~ 90 degrees, there is a change in the shape of the beam.(image attached)

Will see if changing M2 to the spare M2 from SBox can fix the issue.

At the end of the day, alignment was lost.

will be redone today while changing M2 mirror.

Manar Amer wrote:

Today; replaced the M1 SBox mirror with M1 from gamma factory , following the previous alignment.

Removed all mirrors and cleaned them one by one.

Started with no M1, placed M2 did its alignment through M3 pinhole, then placed M3 aligned it through M4 pinhole, then placed M4 aligned it through M1 pinhole.

Placed M1 Gamma factory at the end

Note: mirror is new was not used before, took image on microscope there was no dust in the center, but there was little on the edges (cleaned it using spin-coater)

beating of fundamental mode observed (at M2 transmission) immediately with very higher order modes, slight adjustment on M4 suppressed them.

The beam is not circular and still has some deformation, M1 was changed, so it could be a damaged point at one of the other mirrors (to be investigated tomorrow)!!

No significant damage seen on the other mirrors.

 

 

Manar Amer wrote:

Last week In FP cavity we had all SBox mirrors including the damaged M1 mirror.

The beating was observed with just a slight adjustment on M4 alignment.

In addition, it was also observed while driving the piezo of the CW laser.

In the attached image of the beam at transmission point at M2 mirror, the shape of the beam is deformed.

one of the factors could be related to the damaged spot on M1 mirror. The deformation position can also be seen in the higher order modes.

after changing the injection mirror to Gamma factory M1, this deformation should disappear.

Manar Amer wrote:

Finished the alignment of the SBox with M1 of SBox placed, to be changed to gamma factory mirror later

Observed what seems to be transmission through M2 mirror, to be confirmed. image attached shows it

a PowerPoint attached showing the details of today's work.

 

 

Manar Amer wrote:
  • Adjust the placement of the Irises  (done at the start of the day, adjusted at the end when M1 mirror was placed)
  • Improvement on the Ma1 and Ma2 (The Alignment Mirrors)
    • Improves the alignment a bit. There is an observation where there is no symmetry in the image, concluded to be because of the pinholes that are usually used for the alignment the circle cut is not symmetric (took images on phone of pinholes)
    • This explains why, for the pinhole used on M1, we see a little more symmetry than the pinhole used on M2. 
    •  
    • Placed M1 mirror from Sbox (with damage spot in the middle), after cleaning it using the spincoater, only used to adjust the injection into the mirror where it will give Horizontal displacement due to the thickness of the mirror.
    • Last position of the motors attached
  • Installed the 2 dielectric mirrors for the reflection line
  • Did a rough alignment for the reflection line to have it exit through the reflection window.

Note about mirrors : They were cleaned a year a go and placed in the boxes. When opened the box and seen under microscope they were very dirty with dust particle 

Specially from the back, I believe the boxes have dust in them (to be verified under the UV light /  Microscope  !!!!!!!! )

 

Manar Amer wrote:

Remember to update the position of the reference Irises tomorrow !!!!

Manar Amer wrote:

After pointers and Advice from Aurélien,

improvement of the beam alignment through M1 and M2 irises was done.

To improve the image on the beam profiler, I cleaned the alignment mirrors and the windows using pure ethanol and optical lens cleaning tissue,

a significant improvement after cleaning the inside of M2 Window. Images have less dust diffraction rings.

Attached new images of the beam:

  • New Ma1 and Ma2 motor position
  • After M1 pinhole optimized (saturated)
  • After M2 and M1 pinhole optimizing (saturated and not)
Manar Amer wrote:

SBox alignment, using CW laser Koheras power 1 mW

  • Finished M1 and M2 alignment using the diffraction patterns seen on the beam profiler,  the beam fit is not very good ?? (image taken)
  • Positions of Ma1 and Ma2 motors are registered (image taken)
  • Placed Fixed 2 irises at M1 window and M2 window (Image taken while both are closed)
  • Distance between the collimator and beam profiler ~ 2 m
  • Distance between collimator and M1 diaphragm   ~ 1 m
  • Distance between collimator and M2 diaphragm ~ 1.6 m 

 

Manar Amer wrote:

Starting with the alignment of the KBox using CW Koheras.

The size of the CW output from the collimator at 1 meter from is ~ 2.4 mm, to be adjusted with lenses to match the cavity !!! (attached image)

For the 2 alignment mirrors, I am only able to control 3 axes for now (motor issue for the fourth to be fixed !!) , Horizontal and vertical for Ma1 and Horizontal for Ma2(closes to injection window).

Started with only irises placed on both the M1 and M2. Placed a beam profiler at the window behind M2 to observe an output beam.

Motors position for the 3 axes and beam shape and position after M2 attached.

 

 

 

 

 

 

 

 

 

 

 

ELOG V3.1.4-395e101