    Reply   2-mirror cavity alignment, posted by Alice Renaux at Optical room about lasers and optics Sans_titre.jpg
The NKT laser PZT sensitivity is ~0.09pm/V of wavelength variation, which is equivalent to ~ 25MHz/V
So, the full range of the scan is roughly 250MHz (more than a FSR) for 10V.

It seems impossible to get such large resonances unless the Finesse is very low => let's try to change M1 by a spare GammaFactory plan
Entry   2 mirror cavity high power experiments, posted by Xinyi Lu at Optical room about lasers and optics record_20240425.pngScreenshot_2024-04-25_4_155354-155kw.pngScreenshot_2024-04-25_1_154630-155kw.png60kW_highordermode2.jpg
Today, Ronic and I recorded some intracavity power and cavity mode size as shown in Fig. 1.

Coupling was calculated using the locking curve of this overcoupled cavity. Pr/Pi = 1-Cgeo*Cimp, Cimp = 1-|1-2T1/RTL|^2

We can see that the effective gain, coupling, and mode size decrease with increasing power. And the beam is constantly moving.
Entry   2 Mirror Setup @ 216.6 MHz, posted by Manar Amer at Optical room about lasers and optics | detectors and electronics 20220706_Setup.jpgScreenshot_2022-07-06_1_181116.pngFinesse_Fit.jpg00mode.jpg00mode_diameter.jpg
The SBox cavity setup was changed to have only 2 mirrors M1 plane and M2 spherical, both from ThomX

Distance between the mirror ~ 72 cm , increased from 70 cm to take into account the thickness of the ThomX mirrors

Two lenses (300 mm @ 50 cm , 200 @ 104 cm) were placed to have the beam radius ~ 0.55 mm.
    Reply   2 Mirror Setup @ 216.6 MHz, posted by Manar Amer at Optical room about lasers and optics | detectors and electronics 00mode.jpg00mode_diameter.jpgScreenshot_2022-07-08_0_164115.pngSlide1.JPGSlide2.JPG
The cavity was realigned using irises instead of pinholes, gave a better alignment.

The inside of the box, the spherical and the injection mirror were cleaned and placed back inside the box.

we see beating of fundamental mode, previously at the transmission point we placed a wedge to split the beam which resulted in an elliptical
    Reply   2 Mirror Setup @ 216.6 MHz, posted by Manar Amer at Optical room about lasers and optics | detectors and electronics Screenshot_2022-07-08_2_182405.png
Update for Finesse measurement, The cavity was put under vacuum ~ 1.1*10^-1 mbar

and the alignment and coupling improved.

FSR = 210.1 MHz
    Reply   2 Mirror Setup @ 216.6 MHz, posted by Manar Amer at Optical room about lasers and optics manip2miroirs.pptx2mirror_plan_shperical_beam_size.png
Adding information about the 2 mirror cavity setup (plan - spherical) that is currently installed.

From Aurélien at the start of the manipulation.

@ 0 is where the injection mirror is located
Entry   100W CELIA laser amplifier "how's to" document, posted by Ronic Chiche at Optical room about lasers and optics | software How_to_start_the_Amplifier_v1.pdf
1st version of the document.
if some information is not correct or missing, give any comment by replying to this post.
Entry   100W CELIA laser amplifier "Power vs Pump current" curve, posted by Ronic Chiche at Optical room about lasers and optics | detectors and electronics CELIA_100W_amplifier_Power_vs_Current.png
We measured again the 100W CELIA laser amplifier with a pump current until 8A.

as the first current pump of the amplifier has a Peltier issue, we don't exceeded 6A on this stage and we compensated with the 3 other stages.

7A average current is obtained with 6A / 7.3A / 7.3A / 7.4A
Entry   "ThomX" CELIA Amplifier commissionning, posted by Ronic Chiche at Optical room about lasers and optics 
The "ThomX" CELIA amplifier is installed on the optical table.
I added all the rubber pieces available between the 2 racks to isolate as best as possible the top rack, which embeds some fans, from the bottom
rack from which the laser beam is going out to the FP cavity.
    Reply   "ThomX" CELIA Amplifier commissionning, posted by Ronic Chiche at Optical room about lasers and optics Capture.PNGCapture2.PNG
Beam size at the amplifier output:

to make this measurement, I switched ON only 1st and 2nd stage.

the dichroïc mirror seems to work the best in reflection at normal AOI !!!
    Reply   "ThomX" CELIA Amplifier commissionning, posted by Ronic Chiche at Optical room about lasers and optics 
this afternoon, with Fabian, we did some measurements on the 3rd stage.

finally we used the Dichroïc mirror at AOI=45°C as first mirror and a Thorlabs BB1-E03 as 2nd mirror to send the
beam to the powermeter.