  Status of commissioning, report also here plots are reports., Page 12 of 13  Not logged in ELOG logo
    Reply   high-power experiments of 2-mirror cavity, posted by Xinyi Lu at Optical room about lasers and optics 6x
Here's a summary of our experiment last week:

The initial telescope position: 920 mm (f=+250mm) and 1148 mm (f=-150mm) from the amplifier output.

Mon May 6: We moved the concave lens 0.5mm closer to the cavity.
    Reply   D-shape mirrors positionning status, posted by Ronic Chiche at Optical room about mechanics | lasers and optics | detectors and electronics 
summary:

- the motors used to move the D-shape are the Newport Picomotors 8303-V
the sensitivity is roughly 30nm/step
Entry   CELIA 100W laser amplifier repair , posted by Ronic Chiche at Optical room about lasers and optics 
this morning with Alice, we sent the Menhir 160MHz injected in a fiber (with 6mW at the end of a long fiber) into the laser amplifier, to look for leakage
or damage in the first stages of the amplifier (the amplifier is totally off).

first of all, we checked for light scattering around the laser crate with a sensitive optical card => nothing
    Reply   CELIA 100W laser amplifier repair , posted by Ronic Chiche at Optical room about lasers and optics IMG20241002144351.jpgIMG20241002144936.jpg20241002_161657.jpgClipboard_10-02-2024_01.jpg
this afternoon, we saw some electric cables badly connected to their power supply.
we fixed it by soldering them together and screwing the result to the power supply.
(see 1st image)
Entry   Menhir 160MHz laser, posted by Alice Renaux at Optical room about lasers and optics spectre_avant_cvbg.xlsxspectre_apres_cvbg.xlsxspectre_avant_cvbg.pngspectre_apres_cvbg.png
The Menhir 160MHz has been put back in place on the CELIA amplifier setup. Its output power is measured to be 150-160mW with an attenuator as expected.
Its spectrum is available in "spectre_avant_cvbg.xlsx" and "spectre_avant_cvbg.png".

The pulses are stretched by means of a CVBG. Their spectrum is available in "spectre_apres_cvbg.xlsx" and "spectre_apres_cvbg.png".
Entry   Menhir 216MHz laser, posted by Alice Renaux at Optical room about lasers and optics spectre.xlsxspectre.png
The Menhir 216MHz laser has been put back in place on the cavity table. Its output power is measured to be 160mW with an attenuator. Its spectrum is
available in "spectre.xlsx" and "spectre.png". The main wavelength is a bit shorter (1028.75nm) and the spectrum a bit narrower (4.73nm)
than expected.
Entry   optical table cleaning, posted by Ronic Chiche at Optical room about utilities Oxydation.jpg20241203_113105.jpg20241203_125359.jpg
this morning with Alice and Daniele, we removed all the optics elements and equipments from the SBOX optical table and started to clean it.
the dust meter count 0 on all particle sizes after the cleaning.

we observed a small part of the SBOX which seems to be oxydised (see picture).
    Reply   optical table cleaning, posted by Ronic Chiche at Optical room about utilities 
this afternoon, we checked the dust meter which is at 0 for 1 and 5µm dust particles but ~ 2000-3000 for 0.3µm particles.

we opened the two top panels to let the air flow clean the inside of the vessel.

we observed some other minor oxydised regions (than the one taken in picture) on the external parts of the inox panels but at first sight, nothing
    Reply   CELIA 100W laser amplifier repair , posted by Ronic Chiche at Optical room about lasers and optics Sans_titre.png
friday morning, we add a zoom call with Jerome Lhermite about the amplifier repair.
he approximately confirmed the amplifier scheme from the Loic thesis.
he suggested to:
    Reply   optical table cleaning, posted by Ronic Chiche at Optical room about utilities 
now, it seems the table is clean enough (dust meter counts 0 particles) to install a 2 mirrors FP-cavity.




Ronic
Chiche wrote:



this afternoon, we checked the dust meter which is at 0 for 1 and
Entry   Alignment of a 2-mirror FP cavity, posted by Ronic Chiche at Optical room about mechanics | lasers and optics 
today with Alice,

- we measured the height of the SBOX windows center : 140mm from the optical table.
- we set the laser fiber colimator exactly at this height.
Entry   2-mirror cavity alignment, posted by Alice Renaux at Optical room about mechanics | lasers and optics 2-mirror_cavity_alignment.zip
At the beginning of the week, I installed a half-wave plate, a beamsplitter cube and a quarter-wave plate, as well as the end-of-cavity mirror, which
I unsuccessfully tried to align.

I removed the end-of-cavity mirror and started the laser injection over again with the two motorized mirrors placed in front of the vacuum chamber
Entry   2-mirror cavity alignment, posted by Alice Renaux at Optical room about lasers and optics IMG20250527160111.jpg
Yesterday and today, I replaced the OEWaves CW laser with the NKT CW laser. Its screen does not display anything, so it has to be operated through the
GraphiK software.

I then re-aligned the cavity with a new adjustment adaptation tool between the mirror mounts and the irises.
    Reply   2-mirror cavity alignment, posted by Alice Renaux at Optical room about lasers and optics Sans_titre.jpg
The NKT laser PZT sensitivity is ~0.09pm/V of wavelength variation, which is equivalent to ~ 25MHz/V
So, the full range of the scan is roughly 250MHz (more than a FSR) for 10V.

It seems impossible to get such large resonances unless the Finesse is very low => let's try to change M1 by a spare GammaFactory plan
Entry   2-mirror cavity full setup, posted by Alice Renaux at Optical room about lasers and optics 
Today with Ronic, we changed the FP cavity's mirrors to Layertek (Gamma Factory) mirrors :

- M1 : n°161186 --> fused silica, unknown absorption, unknown diffusion, T1=2500ppm transmission planar mirror ;
- M2 : n°161182 --> fused silica, unknown absorption, unknown diffusion, T2=10ppm transmission, 5m ROC mirror.
Entry   Damaged mirrors test, posted by Alice Renaux at Optical room about lasers and optics 
Today with Dorian we tested three mirrors with which we previously haven't been able to obtain any resonance or optical beating :

- C16111/11 : The mirror looks normal under a microscope, apart from a few inclusions and maybe a small scratch towards the edge. We tested it
with a 161186 M1 mirror and we weren't able to obtain any optical beating on the cavity's mirrors or resonance.
Entry   first attempt to lock, posted by Ronic Chiche at Optical room about mechanics | lasers and optics | detectors and electronics 20250623_123816.jpg20250623_123833.jpg20250623_123908.jpg
With Alice, we installed the EOWaves oscillator to be locked on the FP-cavity (T1 ~ 2500ppm, T2 ~ 10ppm => F ~ 2500)
FSR = 216MHz => LW = 86kHz.

We installed the PDH box, and we got some error signal, but the shape of the transmission signal and error signal is a bit strange...
Entry   2-mirror cavity locking, posted by Alice Renaux at Optical room about lasers and optics Figure_1.pngFigure_2.png
In June, we encountered some problems regarding the transmission and error signals (see images here),
looking as if the laser was switched off before the cavity was filled.

 
Entry   Successful lock (finally !), posted by Alice Renaux at Optical room about lasers and optics IMG_20250903_113758.jpg
Today, with Ronic, we managed to get a successful lock with the NKT laser by setting up a new PDH box from scratch (photodiode + amplifer + mixer).

Me measured the injected power (5.4mW) and the transmitted power (33µW) after a wedge (92% transmission) and a 7ppm transmission mirror,
so the intracavity power was 5.1W. We have a 950 enhancement factor for a 3100 finesse cavity, so a nominal enhancement factor of 2500. The lock was very
Entry   Installation of the avalanche photodiode, posted by Ronic Chiche at Optical room about lasers and optics | detectors and electronics 
After installing the 2nd EOM, we had some trouble to be able to lock again.
One possible reason was the very low signal level in transmission, which is important to trigger the locking system (and stop it).

See the Alice post for details, but we were able to measure only once the Finesse of the cavity at around 2600.