Message ID: 233     Entry time: Fri Mar 21 18:27:36 2025
 Author: Alice Renaux 
 Status: Under Process 
 Type: report 
 Category: mechanics | lasers and optics 
 Location: Optical room 
 Title: 2-mirror cavity alignment 

At the beginning of the week, I installed a half-wave plate, a beamsplitter cube and a quarter-wave plate, as well as the end-of-cavity mirror, which I unsuccessfully tried to align.

I removed the end-of-cavity mirror and started the laser injection over again with the two motorized mirrors placed in front of the vacuum chamber (but left Ronic's setup as it was).

Today, I aligned the laser by putting an iris on the cavity mirrors' mounts. I tested the repeatability of such a setup, both by removing and putting back the same iris and by using a different iris. The result is in the attached pictures. Is it good enough ?

The mirrors' motors positions are :

- 1 : 3.349670 mm ;

- 3 : 2.060280 mm ;

- 4 : 3.484560 mm ;

- 5 : 3.269710 mm.

Attachment 1: 2-mirror_cavity_alignment.zip  5.042 MB