icon3.gif   Entered by Hayg Guler from 134.158.91.29 on Fri Jan 10 18:11:59 2020; LED output of LI/DG/SST.01-CCD.01 is now working , entered from 134.158.91.29 AtkPanel__5_5____LI_DG_SST_01-CCD_01_Basler_et_AtkPanel__5_5____li_liveviewer_sst_01-ccd_01.png
--> Use user_output_line1 to swich it ON/OFF

 

 
Entry   Iryna Chaikovska ; Restarting all the Libera. Problem with Libera 2, entered from 134.158.91.111 
On 09/01/2020 I have restarted all the Libera in order to make it accesible in the CR. Meanwhile, the DS of the Libera 2 remained to be in the unknown
state. It is green in astor, however, in Jive it says that the DS is in the unknown state. All the atribute fields are filled with crosses.

The GLPI ticket has been openned to fix this issue. 
Entry   Entered by Moutardier Alexandre from 134.158.76.123 on Wed Jan 8 09:55:02 2020; Allumage des caméras 
Les caméras de ThomX ont été rallumées hier soir.
Entry   ND+AM+SW; Tests camera ACE 
Nous avons testé les caméras ACE:

* ccdsst2-li l’adresse MAC est correcte par rapport à  
    Reply   ND+AM+SW; RE: Tests camera ACE 
Alexandre a remarqué que:

- Il ne devrait pas y avoir de caméra ccdsst2-li mais ccdsst1-el

- l'adresse MAC assignéee à ccdsst1-el est incorrecte
Entry   AM+SW+ND; Tests LED camera 
Nous avons voulu tester l'allumage des LEDs sur LI/DG/SST.01-CCD01 cependant il était impossible de redémarrer le serveur dans Astor.

 
Entry   Entered by Nicolas Delerue from 134.158.88.175 on Fri Dec 20 16:10:43 2019; All cameras shutdown for the holiday season 
All cameras have been shutdown for the holiday season.
Entry   Entered by Nicolas Delerue from 134.158.195.143 on Fri Dec 20 15:41:03 2019; TL/SST.03-CCD and TL/SST.02-CCD started 
TL/SST.03-CCD and TL/SST.02-CCD have been started.

TL/SST.03-CCD is in the correct configuration (but without light)

TL/SST.02-CCD is next to the SST station, not in the correct configuation.
Entry   Entered by Nicolas Delerue from 134.158.195.145 on Thu Dec 19 15:22:54 2019; Lens installed for TL/SST03 
The lens has been installed on the CCD at TL/SST03