  146   Fri Nov 15 11:48:52 2024 Entered by Nicolas Delerue from 134.158.195.144 on Fri Nov 15 11:48:36 2024Reference SolutionInfoBPMs BPM transmission GUI A GUI to measure the transmission through
the BPMs has been developped. It is in the
git panneaux at Diags/BPM (a git pull is
 Screenshot_from_2024-11-15_12-03-43.png 
  133   Thu Jul 6 12:10:36 2023 Entered by Nicolas Delerue from 134.158.195.143 on Thu Jul 6 12:10:21 2023New SolutionInfoBLM Timing of the fiber oscilloscope The scope reading the ring fibers is currently
ER/DG/RAC.06-OSC.01.  It is on channel
18 of the Diag delay generator. This channel
  
  14   Fri Jan 31 16:20:01 2020 Entered by Nicolas Delerue from 134.158.195.143 on Fri Jan 31 16:19:47 2020FixedFixBPMs T2TrigCount d'un libera Un script pour connaitre le T2TrigCount
d'un libera

panneaux/Diags/BPM/get_BPM_trigcount.py
  
  2   Fri Dec 20 15:41:18 2019 Entered by Nicolas Delerue from 134.158.195.143 on Fri Dec 20 15:41:03 2019FixedFixCameras SST TL/SST.03-CCD and TL/SST.02-CCD started TL/SST.03-CCD and TL/SST.02-CCD have been
started.

TL/SST.03-CCD is in the correct
  
  119   Tue Jan 17 16:26:51 2023 Entered by Nicolas Delerue from 134.158.195.142 on Tue Jan 17 16:26:36 2023New SolutionInfoCameras SST Test vitesse des moteurs Verification de la longueur de deplacement
en fonction de la vitesse des moteurs:

Vitesse = 3:
  
  130   Tue Feb 21 17:59:43 2023 Entered by Nicolas Delerue from 134.158.195.142 on Tue Feb 21 17:59:26 2023New SolutionFixOther Outil pour verifier les derive de la phase RF par rapport a celle du laser J'ai ajouter dans PlateformeIHM un
outil "lock phase" qui permet de
recaler la phase RF canon vs laser.
 Screenshot_from_2023-02-21_17-57-13.png 
  124   Thu Feb 16 10:13:59 2023 Entered by Nicolas Delerue from 134.158.195.142 on Thu Feb 16 10:13:43 2023Under ProcessInfoCameras SST Test temps reponse IHM SST Suite a un commentaire lors de la reunion
technique de ce matin, j'ai fait une
verification en salle de controle du temps
  
  125   Thu Feb 16 10:26:55 2023 Entered by Nicolas Delerue from 134.158.195.142 on Thu Feb 16 10:13:43 2023FixedInfoCameras SST RE: Test temps reponse IHM SST Apres quelques modifications, le temps
de reponse sur client 1 de l' IHM SST
est inferieur a 4 secondes.
  
  54   Mon Jan 3 17:22:04 2022 Entered by Nicolas Delerue from 134.158.195.142 on Mon Jan 3 17:21:51 2022FixedFixOther Redemarrage Diags J'ai redemarre:

- Les Liberas 1 et 2

- Le chassis des cameras
  
  Draft   Mon Feb 5 14:44:30 2024 Entered by Nicolas Delerue from 134.158.195.142 on Mon Feb 5 14:44:15 2024 FixCameras SST En cas d'extinction des SST Si l 'on ne parvient pas a changer
le temps d'exposition des CCD dans l
'ancienne interface, sur PlateformeIHM
  
  144   Fri Sep 6 13:43:13 2024 Entered by Nicolas Delerue from 134.158.195.142 on Fri Sep 6 13:42:59 2024 ProblemCameras SST Etude declencemenent Basler ACE J'ai regarde le probleme des camera
basler ACE avec Pylon.

Il semeblerait que la camera ait
  
  73   Thu Mar 24 17:13:11 2022 Entered by ND/VC/SC from 134.158.195.144 on Thu Mar 24 17:12:56 2022Under ProcessInfoMaterial Mesures de charges Mesures de charges:

16h41 moyenne sur 50 tirs

16h57 moyenne sur 1000 tirs
 Screenshot_from_2022-03-24_16-41-08.pngScreenshot_from_2022-03-24_16-57-54.pngScreenshot_from_2022-03-24_17-03-39.pngScreenshot_from_2022-03-24_17-11-19.pngScreenshot_from_2022-03-24_17-16-17-avg10.png 
  18   Mon Oct 5 14:06:24 2020 Entered by Moutardier from 91.168.92.148 on Mon Oct 5 14:05:55 2020Under ProcessFixBPMs retour du libera 2 et échange des Libera 2 et 3 Le libera 2 est revenu de maintenance.

Il a était mis à
la place du Libéra 3 puis qu'il
  
  19   Thu Oct 22 11:00:31 2020 Entered by Moutardier from 193.55.29.171 on Thu Oct 22 11:00:17 2020Under ProcessFixBPMs échange des Libera 2 et 3 Les étiquettes des libéra
2 et 3 ont été interchanger
de façon a avoir les bonne informations
  
  22   Wed Jan 20 16:15:49 2021 Entered by Moutardier from 134.158.195.142 on Wed Jan 20 16:15:31 2021 InfoCalibrations Démarrage ds MeanStd Methode pour changer un attribut dans MeanStd 
et redémarrer le ds :

Sur jive :
  
  23   Thu Jan 28 14:17:08 2021 Entered by Moutardier from 134.158.195.142 on Thu Jan 28 14:16:54 2021FixedInfoBLM Tout les Libéra de nouveau fonctionnels L'échange entre le Libéra
2 et 3 est terminé.
Chacun est de nouveau controlable
  
  59   Tue Jan 25 09:04:06 2022 Entered by Moutardier Alexandre from 91.168.92.148 on Tue Jan 25 09:03:43 2022 InfoBPMs A faire sur le BPM du linac La conection du BPM Li/DG/BPM.01 doit être
refaite.

Une double permutation (une au
  
  63   Fri Feb 4 10:24:42 2022 Entered by Moutardier Alexandre from 91.168.92.148 on Tue Jan 25 09:03:43 2022FixedFixBPMs RE: Faire sur le BPM du linac Changement de cablage :

Fait, modification transparente
pour les mesure.
  
  79   Tue Apr 12 10:16:35 2022 Entered by Moutardier Alexandre from 91.168.92.148 on Tue Apr 12 10:16:22 2022 InfoBPMs Verification calibration BPM Verification de la calibration des BPM
:

Les valeurs de l'ATKPannel
  
  85   Mon Jul 25 15:14:52 2022 Entered by Moutardier Alexandre from 91.168.92.148 on Mon Jul 25 15:14:39 2022 ProblemCameras SST Alarme moteur Les moteur des station SST passe en "alarm"
par moment pendant quelque second.

Il s'agit d'une alarme
  