HOME Prise de donnees MML Controle commande Simulations Notes Hardware Ligne X Laser Interaction Diagnostiques Synchronisation
Diagnostiques
  elog des diags.  Not logged in ThomX    logo
Message ID: 105     Entry time: Thu Nov 3 11:41:38 2022
 Author: ND 
 Status: New Solution 
 Type: Info 
 Category: BPMs 
 Important:
 Subject: Update on scan charge phase 
 Icon:  

An update has been done on the script launched by the scan charge phase IHM.

Now if the first scan is 0-360 degrees, it launches a second scan on the area where the max charge has been found during the first scan.

At the end of the second scan the phase is set to max charge -50.

The plots indicate the max charge value computed and max charge - 50 degree (max energy).

Un git pull a ete fait.

 