HOME Prise de donnees MML Controle commande Simulations Notes Hardware Ligne X Laser Interaction Diagnostiques Synchronisation
Controle commande TANGO IHM CC ELLIOOS
  Notes concernant les Controle commande  Not logged in ThomX    logo
icon1.gif   Hayg; [Lima] Camera Balser , entered from 134.158.91.29 
    icon1.gif   Hayg; RE: [Lima] Camera Balser , entered from 134.158.91.29 LiveView_ScoutCamera.png
Message ID: 23     Entry time: Fri Mar 22 18:44:04 2019     In reply to: 22
 Author: Hayg 
 Status: Under Process 
 Type: Info 
 Category: Tango 
 Important:
 Subject: RE: [Lima] Camera Balser , entered from 134.158.91.29 
 Icon: icon1.gif 

Nouveaux tests avec la caméra Scout : scA640-70gm

Contexte : 

Update des librairies tango

      ▶  $ conda list | grep -i lima
lima-camera-basler             1.8.0          py27ha8d69ae_0    esrf-bcu
lima-camera-basler-tango  1.8.0                         0       esrf-bcu
lima-core                             1.8.0            py27ha8d69ae_0    esrf-bcu
lima-tango-server               1.8.0                         0    esrf-bcu
On est passé de release candidate à la version 1.8

Test de sauvegarde d'image

Script Tango qui sauve 100 images :

sans erreur on a sauvé 100 images, avec une vitesse > 20 Hz

Fini les soucis d'images divisées en 4

Test de lecture video :

Fonctionne sans erreur ... !!

Propriétés dans Tango :

Frame transmission  delay : 10 us

Inter packet delay : 100 us

Packet size : 2000

 

 Jumbo frames 
#  ip link show | grep mtu 
# 1: lo: <LOOPBACK,UP,LOWER_UP> mtu 65536 qdisc noqueue state UNKNOWN mode DEFAULT group default qlen 1000
# 2: enxe4b97ad8e69e: <BROADCAST,MULTICAST,UP,LOWER_UP> mtu 1500 qdisc fq_codel state UP mode DEFAULT group default qlen 1000
# 3: wlp59s0: <BROADCAST,MULTICAST,UP,LOWER_UP> mtu 1500 qdisc mq state UP mode DORMANT group default qlen 1000
# modify the second one which corresponds to the camera 
# sudo ip link set enxe4b97ad8e69e mtu 9000;
 

 

 

Hayg wrote:

Update du script pour sauver les images :

2 camera en test :

  • Basler ACA640 - 100gm

  • Basler Scout SCA640 - 70gm

Quelques chiffres :

Trigger : internal

50 images demandées

  • Basler ACA640 - 100gm : 70Hz testé sans perte d'images

  • Basler Scout SCA640 : plus aléatoire

    • fonctionne pas, ou bien pertes de qq images (10 environ ou moins)

    • Image semble doublée, quadriee

CF : lien github de Lima qui reporte ces les problèmes https://github.com/esrf-bliss/Lima/issues/97

 

 

 

 

Attachment 1: LiveView_ScoutCamera.png  442 kB  | Hide | Hide all
LiveView_ScoutCamera.png