Entered by Marie Jacquet from 134.158.89.45 on Wed May 3 20:10:26 2023 wrote:

Pour illustrer la fin de l'entree elog 745 d'aujourd'hui  :

(==> Saute de position du faisceau, cela n'est pas corrélé à une perte de puissance ou de changement de phase (phénomène impossible à reproduire avec l'atténuation section et/ou les phases canon et section) ==> E-log Marie pour visu saute)

ci-joint le resume du traitement des donnees du 24 et 26 avril ( = donnees stabilite)

On a aussi testé en bougeant la position laser cela n'a pas le même effet (mouvement sur LI/BMP.01)

On a mis "OFF" les steerer du linac mais cela ne semble pas avoir eu d'effet nottable.