7h50 : mise en chauffe en remote (frigo/chillers, modulateur, laser, cyclage)
09h30 HV is ON
10h23 Energy = 61.5 MeV, Laser Phase 39 dg., Sect. Phase 225 dg., Sect. Att. 0.952 dB, I_Gun -78.1953 pC, I_Linac -122.7687 pC, I_TL -102.6104 pC RF_Ampl. 148.7161 kV, RF_Phase -45.5 dg.
*** Mesure Monochromateur 61.5MeV ***
* angle initiale mono : Ry_mono(0) = -0.1° (ajouté aux valeurs d'angles pour la position moteur)
* Position fentes initiale [-19;-16]
Point On Axis :
- Angle Mono 1.5° --> position RSample : 3.5°
- Angle Mono 1.6° --> position RSample : 3.7°
- Angle Mono 1.7° --> position RSample : 3.9°
- Angle Mono 1.8° --> position RSample : 4.1°
H +30mm (position hexapod TX = -12):
- Angle Mono 1.8° --> position RSample : 3.7°
- Angle Mono 2° --> position RSample : 4.1°
H +40mm (position hexapod TX = -23):
- Angle Mono 1.8° --> position RSample : 3.6°
- Angle Mono 2° --> position RSample : 4°
H -30mm (position hexapod TX = 53):
- Angle Mono 1.5° --> position RSample : 3.9°
- Angle Mono 1.6° --> position RSample : 4.1°
- Angle Mono 1.8° --> position RSample : 4.5°
H -40mm (position hexapod TX = 64):
- Angle Mono 1.5° --> position RSample : 4°
- Angle Mono 1.6° --> position RSample : 4.2°
- Angle Mono 1.8° --> position RSample : 4.6°
V +30mm (position hexapod TX = 21, TY : 30):
- Angle Mono 1.5° --> position RSample : 3.5°
- Angle Mono 1.6° --> position RSample : 3.7°
- Angle Mono 1.8 --> position RSample : 4.1°
V -30mm (position hexapod TX = 21, TY : -30):
- Angle Mono 1.5° --> position RSample : 3.5°
- Angle Mono 1.6° --> position RSample : 3.7°
- Angle Mono 1.8 --> position RSample : 4.1°
*** Images pilatus (thershold 30keV) ==> position On Axis, avec Rsample = 2°
1.8° :
452 (expo 2s), 453 (expo 5s), 454 (expo 10s), 455 (expo 20s)
1.5° :
456 (expo 20s), 457 (expo 20s), 458 (expo 10s)
2° :
459 (expo 20s), 460 (expo 20s), 461 (expo 30s), 465 (expo 20s)
*** Re-reglage streak camera (car completement dereglee)
longueur paquet : 40 ps fwhm (idem injection et apres injection) |